510(k) Data Aggregation

    K Number
    K213641

    Validate with FDA (Live)

    Date Cleared
    2022-09-15

    (301 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K160597

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MONDRIAN ALIF Cage with Supplementary Fixation Plate System (MONDRIAN ALIF Cage) is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). It is to be used in patients who have had at least six months of non-operative treatment. This device is intended to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous and/or corticocancellous bone.

    The MONDRIAN ALIF Cage with Supplementary Fixation Plate System is intended for use with either two or four titanium alloy screws which are provided with the system. When used with screws and implants with lordotic angles ≤20°, the MONDRIAN ALIF Cage with Supplementary Fixation Plate System is a stand-alone interbody fusion device. If the physician chooses to use fewer than the number of screws compatible with the plate, then a supplemental spinal fixation system that is cleared for use in the lumbosacral spine must be used. Hyperlordotic cage offerings (>20°) require the use of a supplemental fixation system (e.g. facet screws or posterior fixation).

    Device Description

    The MONDRIAN ALIF Cage with Supplemental Fixation Plate System are anterior lumbar interbody fusion devices used to provide structural stability in skeletally mature individuals. The system is comprised of spacers, plates, screws and instruments to facilitate the installation of the implants. The spacers are manufactured from PEEK per ASTM F2026 and/or Titanium Alloy per ASTM F136, plates and screws are manufactured from Titanium Alloy per ASTM F136. The instruments are manufactured from Stainless Steel per ASTM F899. Implants and instruments will be provided in non-sterile configuration and will require steam sterilization prior to use.

    AI/ML Overview

    The document describes the MONDRIAN ALIF Cage with Supplementary Fixation Plate System, an intervertebral body fusion device.

    Acceptance Criteria and Reported Device Performance:

    The document states that the rationale for substantial equivalence is based on the following general characteristics being similar to predicate devices:

    Acceptance Criteria (based on similar characteristics to predicate devices K160597 and K192863)Reported Device Performance (Summary)
    Intended UseSimilar to K160597 and K192863.
    Indications for UseSimilar to K160597 and K192863.
    MaterialsMONDRIAN ALIF Cage with Supplemental Fixation Plate implants manufactured from PEEK per ASTM F2026 and/or Titanium Alloy per ASTM F136. Instruments manufactured from Stainless Steel per ASTM F899. These are commonly used materials in orthopedic implants and instruments and similar to materials used in K160597 and K192863.
    Design FeaturesSimilar to those in currently marketed devices cleared in K160597 and K192863. Design differences have not identified any issues that would impact the safety and effectiveness of the device.
    SterilizationDevice offered in non-sterile configuration, requiring steam sterilization by the user prior to use, which is similar to the devices cleared in K160597 and K192863.
    Non-Clinical TestsThe device underwent: - Axial Dynamic Compression (per ASTM F2077) - Screw Push-Out Testing - Plate-Cage Disassembly Testing Results indicate that the device will perform within the intended uses and no new issues of safety and effectiveness have been raised.
    Clinical TestsNo specific clinical tests are mentioned as being performed for this submission ("o" is listed under clinical tests with no further details). The document relies on substantial equivalence to predicate devices, implying that clinical performance is considered similar based on the similarities in the device's characteristics and the non-clinical test results.

    Study to Prove Device Meets Acceptance Criteria:

    The document describes a non-clinical study to demonstrate substantial equivalence to previously cleared predicate devices (K192863 and K160597).

    1. A table of acceptance criteria and the reported device performance: This is provided above.

    2. Sample size used for the test set and the data provenance:
      The document does not explicitly state the sample sizes for the non-clinical tests (Axial Dynamic Compression, Screw Push-Out, Plate-Cage Disassembly Testing). It also does not mention data provenance as these are laboratory/bench tests, not patient data studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. These were non-clinical, mechanical tests performed according to ASTM standards. Expert consensus would not be used to establish "ground truth" for these types of tests; rather, the results would be compared against predefined criteria from the standard or predicate device performance.

    4. Adjudication method for the test set: Not applicable. This was a non-clinical, mechanical testing study, not a study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an implantable medical device (ALIF cage), not an AI-assisted diagnostic or imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithmic device. The non-clinical tests assess the mechanical performance of the physical device.

    7. The type of ground truth used: For the non-clinical tests, the "ground truth" or reference for acceptance would be the performance criteria (e.g., strength, durability) specified by relevant ASTM standards (e.g., ASTM F2077) and/or comparison to the mechanical performance of the predicate device(s).

    8. The sample size for the training set: Not applicable. This is a non-clinical performance study of a physical medical device, not a machine learning model.

    9. How the ground truth for the training set was established: Not applicable, as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170119

    Validate with FDA (Live)

    Date Cleared
    2017-09-29

    (259 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K161230, K160958, K160597, K153782, K112036

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Calix® Lumbar Spinal Implant System is intended for spinal fusion procedures at one or two contiguous levels (L2 – S1 inclusive) in skeletally mature patients with degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the lumbosacral spine. DDD patients may also have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved level(s). These implants are to be packed with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft, and implanted via an anterior, posterior, and/or transforaminal approach. Patients should receive at least six (6) months of non-operative treatment prior to treatment with a lumbosacral intervertebral fusion device.

    This device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).

    Device Description

    The X-spine Calix® Lumbar Spinal Implant System is an intervertebral fusion device for use with bone graft, designed for surgical placement between adjacent vertebrae of the lumbar spine as an adjunct to fusion.

    The X-spine Calix® Lumbar Spinal Implant System is a generally hollow box or oval shaped device manufactured from Invibio PEEK-Optima™ LT1 per ASTM F2026. The device contains an array of holes located throughout its geometry as well as teeth on the superior and inferior surfaces to help prevent implant dislodgement or expulsion once placed in its desired location. The implants are available with or without a medical grade commercially pure titanium (CP Ti) plasma coating (per ASTM F1580) on the superior and inferior surfaces of the device.

    The hollow center of the implant allows the device to be packed with bone graft: autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft,

    The single use implants are supplied in several width, height and lordosis options to accommodate variations in patient anatomy and differing pathologies. The devices contain radiographic markers made from tantalum per ASTM F560 to aid in final placement verification. Plasma coated implants are provided clean and sterile. Nonplasma coated implants and instruments are provided non-sterile. All implants are intended for single use only, and should not be reused under any circumstances.

    The system does not contain software/firmware.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device (Calix® Lumbar Spinal Implant System), which is a regulatory submission demonstrating substantial equivalence to a legally marketed predicate device. This type of submission primarily relies on non-clinical performance data (mechanical and material testing) to demonstrate that the new device performs as well as or better than the predicate, especially when modifications are made.

    The document does not describe an AI/ML-driven device or diagnostic tool that would require a study involving human readers, ground truth establishment through expert consensus (radiologists), or MRMC/standalone performance metrics typical of AI models. Therefore, many of the requested elements (sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable to this type of device submission.

    The acceptance criteria and proof of performance for this particular device (a spinal implant) are based on mechanical and material testing as per established ASTM standards and FDA recommendations for intervertebral body fusion devices.

    Here's a breakdown of the relevant information from the provided text, addressing the points where applicable and noting when information is not present or relevant:


    Acceptance Criteria and Device Performance (for a spinal implant)

    The acceptance criteria for this medical device are primarily based on demonstrating mechanical equivalence and safety to predicate devices, adhering to recognized standards. The "performance" here refers to the physical and functional integrity of the implant, not diagnostic accuracy.

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria AreaSpecific Tests / RequirementsReported Device Performance
    Mechanical Performance- ASTM F2077: Test Methods for Intervertebral Body Fusion Devices (Dynamic Compression Shear, Static Compression, Static Compression-Shear)The tests demonstrate that the addition of the medical grade commercially pure titanium plasma coating does not present a new worst case, and does not impact the performance of the implants.
    - Expulsion testing as recommended by FDAPerformance is equivalent to or better than the predicate device.
    Subsidence- ASTM F2267: Test Methods for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression(Not explicitly stated in detail for the new modifications, but implied to be acceptable based on the predicate's performance and no new worst case.)
    Material Characterization- ASTM F1877: Standard Practice for Characterization of Particles (for titanium plasma coating)The characterization confirms the material properties and safety of the new coating.
    Biocompatibility(Implied by use of PEEK-Optima™ LT1 per ASTM F2026 and CP Ti per ASTM F1580, which are well-established biocompatible materials for implants.)(Not explicitly detailed in the performance section, but these are FDA-recognized materials for implants.)
    Compatibility with Bone Grafts- Expansion of Indications for Use to include allograft comprised of cancellous and/or corticocancellous bone graft.The expansion "does not impact the mechanical performance" and "does not introduce new risks."

    Detailed Information as per Request (with applicability notes):

    1. A table of acceptance criteria and the reported device performance

      • See table above. The acceptance criteria are largely adherence to specified ASTM standards and demonstrating that the modifications (titanium coating, allograft use) do not negatively impact performance compared to the predicate.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • N/A. For this type of device (spinal implant), "sample size" refers to the number of physical devices or material samples tested in a lab, not a population of patients or images. The document does not specify the exact number of implants tested for each mechanical test, which is common for 510(k) summaries where the methodology typically follows established engineering test standards.
      • Data Provenance: The data originates from the manufacturer's (X-spine Systems, Inc.) internal testing, conducted in their facilities or contracted labs. This is not clinical data, so terms like "retrospective" or "prospective" are not applicable.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • N/A. "Ground truth" in the context of this spinal implant device is established through physical mechanical testing and material characterization according to validated engineering standards (ASTM). It does not involve expert clinical assessment or image interpretation for establishing a "ground truth" diagnosis.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • N/A. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in human-reader studies (e.g., in diagnostic imaging). This regulatory submission is for a physical implant, not a diagnostic AI device.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • N/A. An MRMC study is designed to evaluate the performance of diagnostic devices, particularly AI-assisted ones, by comparing human reader performance with and without AI assistance. This submission is for an orthopedic implant and does not involve AI or human readers for diagnosis.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • N/A. This refers to the standalone performance of an AI algorithm. The device is a physical implant, not an algorithm.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

      • The "ground truth" for this device's performance is based on engineering standards and established physical properties. For example, the acceptable load-bearing capacity, expulsion resistance, and material characteristics determined by industry-standard mechanical testing are the "ground truth" for its functional performance.
      • For the Indications for Use, the "ground truth" for degenerative disc disease (DDD) is defined clinically: "back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies." This is a clinical definition for patient selection, not a ground truth for testing the device itself.
    8. The sample size for the training set

      • N/A. There is no "training set" as this is not an AI/ML device.
    9. How the ground truth for the training set was established

      • N/A. No training set exists for this device.

    Summary of Acceptance and Performance:

    The Calix® Lumbar Spinal Implant System's acceptance and demonstrated performance hinge on its ability to meet or exceed the performance characteristics of its predicate device, the Calix® Lumbar Spinal Implant System (K131350), and other reference devices. The key elements of its "proof" are:

    • Mechanical Testing: Demonstrating compliance with ASTM F2077 (for static/dynamic compression/shear) and ASTM F2267 (for subsidence) and FDA-recommended expulsion testing. The critical finding is that the new titanium plasma coating "does not present a new worst case, and does not impact the performance of the implants."
    • Material Characterization: Showing the properties of the new CP Ti plasma coating via ASTM F1877.
    • Indications for Use Expansion: Justifying that the addition of allograft use "does not impact the mechanical performance" and "does not introduce new risks," citing that allograft has been cleared for many other spinal implants.
    • Substantial Equivalence: Ultimately, the study (mechanical and material testing) served to prove that the modified device is substantially equivalent to legally marketed predicate devices in terms of safety and effectiveness, despite the added coating and expanded graft options.