Search Results
Found 2 results
510(k) Data Aggregation
(111 days)
The Tyber Medical Trauma Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair and fracture fixation of bones appropriate for the size of the devices. Screws are intended for single use only. The Tyber Medical Trauma Screws are not for spinal use.
This traditional 510(k) submission adds configurations of the previously cleared Tyber Medical Trauma Screws under K192975 and K153575. The Tyber Medical Trauma Screw is designed to allow compression and fixation between two adjacent bone segments. The screw is available with thread diameters of 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 8.0, and 9.0mm. The screw is configured to have either full thread, long thread, or short thread. The multiple thread lengths come in various options which can assist to ensure fit for various bone sizes. General trauma screws are intended for compression and fixation of bone.
This Submission also references Washers and Trays. The Trays are specific to the Tyber Medical Trauma Screws.
N/A
Ask a specific question about this device
(267 days)
Minimally invasive reconstruction of fractures and joints; Adjuvant for osteosynthesis in complex joint fractures; Multifragment joint fractures; Simple metaphyseal fractures of the wrist, ankle, elbow, and shoulder, Condylar fractures; Osteochrondriis dissecans, Ostero-Chondral Fractures, Ligament avulsion injuries, Ligament fixation, Other small fragment, cancellous bone fractures, Small joint fusion. Areas where accurate sorew placement is vital. Metatarsal and phalangeal osteotomies; Fractures of the tarsals, metatarsals and other fractures of the foot; Avulsion fractures and fractures of metatarsal V; Tarso-metatarso-phalangeal arthrodesis; Tarsal Fusions; Calcaneal and talar fractures, Subtalar arthrodesis. Ankle arthrodesis. Fractures of small joints, such as: Ankle fractures, Fractures of the fibula, malleolus, and calcaneus, Distal tibia and pilon fractures, Acetabular fractures of the pelvic ring; Fractures of the femoral head and neck, Supracondylar femoral fractures, Slipped capital femoral epiphyses, An adjunct to DHIS in basilar neck fractures, Intractures, Intracapsular fractures of the hip, Fractures of the distal femur and proximal tibia, Patellar fractures. Small fragments of the hand and wrist, Fractures of the carpals and metacarpals, Carnal and metacarpal arthrodesis. Scaphoid fracture and other fractures of the hand, Phalangeal and interphalangeal fractures of the ulna and radius. Radial head fractures. Fractures of the olecranon and distal humeral head fractures, Ligament fixation at the proximal humerus, Glenoid fractures.
The S4 Screw System is collection of cannulated, headless screws having fenestrations on the head and within the thread pitch.
This document, a 510(k) Summary for the S4 Screw System, does not describe a study that uses AI or machine learning technology. It pertains to a physical medical device (bone screws) and its mechanical properties.
Therefore, many of the requested elements regarding acceptance criteria for an AI/ML device and the details of a study proving its performance (e.g., sample size for test/training sets, expert ground truth, MRMC studies) are not applicable to the content provided.
The document focuses on non-clinical performance testing (bench tests) and comparisons to predicate devices for substantial equivalence, which is typical for Class II devices like bone screws.
Here's an analysis based on the information provided in the document, framed by what would be relevant if it were an AI/ML device, to highlight why the full set of questions cannot be answered:
Based on the provided FDA 510(k) Summary for the S4 Screw System (K191995), the following information can be extracted regarding its acceptance criteria and the study proving it meets them:
Note: This submission does not involve an AI/ML device. Therefore, the concepts of a "test set," "training set," "ground truth experts," "adjudication methods," or "MRMC studies" as they apply to software-based diagnostic or prognostic devices are not relevant or described in this document. The "study" mentioned refers to non-clinical bench testing of the physical screw system.
1. Table of Acceptance Criteria and Reported Device Performance:
For a physical device like a bone screw, "acceptance criteria" are typically defined by engineering standards to ensure its mechanical integrity and safety. "Performance" refers to the results of tests against these standards.
| Acceptance Criterion (Standard) | Reported Device Performance (Summary) |
|---|---|
| Mechanical Performance: | |
| ASTM F543 (Standard Specification and Test Methods for Metallic Medical Bone Screws) | Tested per ASTM F543. |
| - Torsional Properties | Testing performed. (Specific values/pass criteria not detailed in the summary, implied to meet standard requirements). |
| - Insertion and Removal Torque | Testing performed. (Specific values/pass criteria not detailed). |
| - Pullout Strength | Testing performed. (Specific values/pass criteria not detailed). |
| Biocompatibility/Material Composition: | |
| ASTM F136-13 (Ti-6Al-4V ELI) | Material specified to meet this standard. |
| ASTM F899-12b (Stainless Steel) | Material specified to meet this standard. |
| ASTM F1185-03 (Hydroxyapatite) | Material specified to meet this standard; information referenced in Master File. |
| Sterilization Efficacy: | |
| ISO 11137-1:2017 (Radiation Sterilization) | Sterilization validation study completed. (Implied to meet standard requirements). |
| Packaging Integrity: | |
| Packaging Performance Study | Packaging performance study completed to validate packaging performance. (Implied to meet standard requirements). |
| Risk Assessment: | |
| (FMEA for modified intended use, additional sizes, HA coating) | Analysis determined modifications resulted in an acceptable low risk category. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size (Test Set): Not explicitly stated in terms of number of screws tested for each mechanical property. For bench testing, samples are typically specific numbers of manufactured devices.
- Data Provenance: The studies are non-clinical, bench-top laboratory tests conducted by the manufacturer, Subchondral Solutions, Inc. No information regarding country of origin of data (beyond the company's US location) or retrospective/prospective nature is applicable in the context of human clinical data.
3. Number of Experts Used to Establish Ground Truth and Their Qualifications:
- Not applicable. For a physical device, "ground truth" is established through adherence to national/international engineering standards and validated test methods, not by expert interpretation of data or images.
4. Adjudication Method for the Test Set:
- Not applicable. This concept relates to resolving disagreements among human readers or annotators for ground truth in AI/ML performance studies.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, this was not done. MRMC studies are specific to evaluating the clinical performance of diagnostic or screening devices, often involving human readers and potentially AI assistance. This submission is for a physical orthopedic implant.
6. Standalone (Algorithm Only) Performance:
- Not applicable. This refers to the performance of an AI algorithm without human intervention. The S4 Screw System is a physical device, not an algorithm.
7. Type of Ground Truth Used:
- For this physical device, "ground truth" is defined by engineering specifications, material science standards, and established mechanical test methodologies (e.g., ASTM F543). It is not derived from expert consensus, pathology, or outcomes data in the way an AI/ML diagnostic would be.
8. Sample Size for the Training Set:
- Not applicable. There is no "training set" as this is not an AI/ML device.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable. No training set exists for this type of device.
In summary, the provided document details the mechanical, material, and sterilization validation of a bone screw system (S4 Screw System) through non-clinical bench testing against established industry standards for similar devices. It does not contain any information relevant to the development, validation, or performance of an AI/ML driven medical device.
Ask a specific question about this device