Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180948

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-07-20

    (100 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K153555

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TruDi™ NAV Suction Instruments are intended for use with the TruDi™ Navigation System during surgical procedures in ENT medicine and skull base surgery to provide navigation of the instruments to the targeted anatomy, and evacuation of gases, liquids, and fragments.

    TruDi™ NAV Suction Instruments are intended to be used by ENT surgeons or support staff.

    Device Description

    TruDi™ NAV Suction Instruments (0°, 70°, 90°) are electromagnetically- navigated reusable devices intended to be used in conjunction with the TruDi™ Navigation System. When used with the TruDi™ Navigation System, the TruDi™ NAV Suction Instruments provide navigation of the devices to targeted anatomy and evacuate gases, liquids, and fragments. The TruDi™ Navigation System is the trade name for the ACCLARENT® ENT Navigation System cleared in K173628.

    TruDi™ NAV Suction Instrument Set consists a 0° suction instrument, 70° suction instrument, 90° suction instrument, and a single-use sterile cable called the TruDi™ NAV Cable. The TruDi™ NAV Cable may be used to connect the suction instruments to the TruDi™ Navigation System.

    The TruDi™ NAV Suction Instruments include an EM trackable single axis sensor, which is integrated at the distal tip of the device. The TruDi™ Navigation System acquires the position and orientation of the distal tip of the device and displays it in real-time view over the patient's pre-operative CT scan to confirm access of target anatomy. Following confirmation, the physician operates the surgical instrument at the target anatomical structure.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the TruDi™ NAV Suction Instruments. It outlines the regulatory process, device description, comparison to predicate devices, and non-clinical performance data.

    However, it does not contain information related to a study proving the device meets acceptance criteria regarding clinical performance or a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, or details about the training and test sets as typically found in submissions for AI/ML-based medical devices.

    The document explicitly states: "Clinical data was not necessary for the TruDi™ NAV Suction Instruments." and "The performance data demonstrated that the device performs as intended." This indicates that substantial equivalence was primarily established through non-clinical testing and comparison to predicate devices, rather than a clinical trial involving human performance assessment with the device.

    Therefore, many of the requested details, such as sample sizes for test and training sets, expert qualifications for ground truth, adjudication methods, MRMC study results, and specifics on AI algorithm ground truth establishment, are not present in the provided text because the device is a navigation suction instrument and not an AI/ML diagnostic tool.

    I can, however, extract the acceptance criteria related to the device's physical and functional performance, along with the reported outcomes from the non-clinical tests.

    Acceptance Criteria and Reported Device Performance (Non-Clinical)

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    Dimensional SpecificationsNot explicitly stated in quantitative terms but implied to be within design specifications.Met all performance acceptance criteria including dimensional specifications.
    Suction Flow PerformanceNot explicitly stated in quantitative terms but implied to be sufficient for intended use.Met all performance acceptance criteria including suction flow performance. Simulated use testing successfully tested suction performance.
    DeflectionNot explicitly stated in quantitative terms.Met all performance acceptance criteria including deflection.
    Location Accuracy (Sensor Sensitivity)≤ 2mm RMS over the entire navigation volume (as stated under "System or Instrument Accuracy Requirements"). The predicate device also provided 1.54 mm at 95% confidence and 99% reliability in controlled environment and 1.73 mm at 5% confidence and 99% reliability in simulated surgical environment.Met all performance acceptance criteria including location accuracy (sensor sensitivity). Testing was performed to verify the navigation accuracy of the subject device when used with the TruDi™ Navigation System (K173628). Performance data demonstrated the device performs as intended. Specific quantitative results matching the ≤ 2mm RMS acceptance criteria are implied to have been met.
    Reprocessing ReliabilityCapable of thorough cleaning and sterilization via specified methods, demonstrating sterility assurance level of 10⁻⁶.Reprocessing and sterilization testing was conducted and met all acceptance criteria. Sterilization process validated per AAMI/ANSI/ISO 11135:2014, demonstrating a sterility assurance level of 10⁻⁶ via steam or STERRAD methods (overkill/half-cycle).
    Electrical FunctionalityNot explicitly stated in quantitative terms.Met all performance acceptance criteria including electrical functionality.
    EEPROM VerificationNot explicitly stated in quantitative terms.Met all performance acceptance criteria including EEPROM verification.
    BiocompatibilityBiocompatible per ISO 10993-1.Biocompatibility testing was successfully completed to determine that the TruDi™ NAV Suction Instruments are biocompatible per ISO 10993-1.
    Mechanical AspectsFunctions in accordance with design specifications.Simulated use testing successfully tested the mechanical aspects of the subject device.
    Clinical Accuracy (Simulated Use)Functions in accordance with design specifications and intended use.Simulated use testing was performed with ENT surgeons and support staff, which successfully tested the clinical accuracy of the subject device. The testing demonstrated that the subject device functions in accordance with design specifications and intended use.
    TruDi™ NAV Cable FunctionalityFunctions as intended with the suction instruments.Simulated use testing successfully tested the functionality of the TruDi™ NAV cable. Testing included sensitivity and connectivity verification.
    Packaging Shelf Life (TruDi™ NAV Cable)Meet a shelf life of two months.Established through accelerated aging via ASTM F1980-07, ASTM F88/F88M-09, and ASTM F2096-04 requirements and confirmed to meet a shelf life of two months.

    Here's an analysis of the requested information based on the provided document, highlighting what is not applicable (N/A) given the nature of the device and the submission:

    1. A table of acceptance criteria and the reported device performance: See table above.

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Test Set Sample Size: Not explicitly stated for specific non-clinical tests (e.g., number of instruments tested for accuracy, deflection, etc.). The mention of "simulated use testing with ENT surgeons and support staff" doesn't provide the number of participants or simulated cases.
      • Data Provenance: Not specified. As these are non-clinical tests, country of origin is less relevant than for clinical data. This was a premarket notification (510(k)) based on non-clinical performance and substantial equivalence.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • N/A. Ground truth in the context of AI/ML diagnostic devices (e.g., disease presence/absence) is not relevant here. For the "simulated use testing," the "experts" were "ENT surgeons and support staff," who were evaluating the device's functional and mechanical performance, not establishing a diagnostic ground truth. No specific number or qualifications beyond "ENT surgeons and support staff" are provided.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • N/A. Adjudication methods are typically used to resolve discrepancies in expert labeling a dataset for AI/ML ground truth. This is not applicable to the non-clinical and simulated use testing described.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. The document explicitly states: "Clinical data was not necessary for the TruDi™ NAV Suction Instruments." This means no MRMC study was conducted. This device is a surgical instrument, not an AI-based diagnostic tool.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • N/A. There is no standalone AI algorithm discussed for this device. The device itself is an instrument that works with a navigation system. Its "performance" (accuracy) is a standalone measurement of the physical system (instrument + navigation system), but this is not an AI algorithm's standalone performance in the typical sense.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • N/A. For the non-clinical tests, "ground truth" was established through engineering measurements, validated test methods (e.g., AAMI/ANSI/ISO for sterility, ASTM for packaging), and design specifications. For "clinical accuracy" in simulated use, it implies the device performed as expected by the ENT surgeons and support staff based on its design and intended function, not against a medical ground truth (like a biopsy result or patient outcome).
    8. The sample size for the training set:

      • N/A. This device does not involve a training set for an AI/ML algorithm.
    9. How the ground truth for the training set was established:

      • N/A. This device does not involve a training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1