Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181367

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-28

    (97 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K153299

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tenex Health TX System with the TXP MicroTip is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of both soft and hard (e.g.: bone) tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery.

    The Tenex Health TX System with the TXP MicroTip is also indicated for use in the debridement of wounds, such as, but not limited to diabetic ulcers, in application, in the physician's judgement would require the use of an ultrasonic aspirator with sharp debridement.

    Device Description

    The Tenex Health TX System with the TXP MicroTip is an ultrasonic surgical instrument that fragments, emulsifies and aspirates soft and hard tissue. It also provides sharp debridement of soft and hard tissue in wounds such as neuropathic ulcers. The Tenex Health TX System with the TXP MicroTip is intended for use in an office-based setting, clinical environment or hospital environment. The system consists of a console, ultrasonic MicroTip, inflation cuff, foot pedal and supply kit. The console provides power and control for the user functions including irrigation, aspiration, and ultrasonic fragmentation/emulsification. It has a large, color LCD and employs a touch-screen with a graphical user interface for selection of required settings. The console provides audible tones for confirmation of selections. The console also houses the irrigation valve, the irrigation pump and the aspiration pump. The Console is non-sterile when used and does not require processing.

    The ultrasonic MicroTip connects to the console for electrical power, the transfer of irrigation fluid to the surgical site and the aspiration of emulsified tissue back to the collection bag. The MicroTip is single use, sterile packaged. A single use, disposable collection bag is integrated into the MicroTip tubing to collect the aspirated materials. The connections are made via an integrated wiring connector and tubing set. The handpiece is constructed from various biocompatible polymers and metals. The tubing is made of biomedical grade PVC. The handpiece and tubing are provided sterile, via electron beam irradiation. The handpiece and tubing are a single use disposable component of the system. Irrigation fluid is delivered from a user-provided surgical saline bag, under pressure to the surgical site by operation of an air pump residing in the console. The regulated output of the air pump pressurizes an inflation cuff that is fitted over the saline bag.

    The foot pedal is used by the user to remotely control the function of the console. It connects to the back of the console via an electrical cable. It provides on/off functionality and is rated IPX8 for protection against liquids.

    The System also includes accessories. A Travel Case is provided for shipping, handling and storage of the TX Console. A power cord is provided for connecting the Console to facility electrical power in order to operate the device. A replacement fuse set is also provided in case of the need for replacement of electrical fuses in the TX Console. The accessories are non-sterile when used and does not require processing.

    The supply kit contains sterile packaged, single use, OEM surgical supply items for use in the surgical site preparation and closure. These include a syringe and needle, scalpel, an ultrasound probe cover, gauze, closure strips and dressing. The itemized contents of the kit with their manufacturer and regulatory information are provided in Table 4 below.

    The Tenex Health TX System with the TXP MicroTip is an upgrade to the Tenex Health TX System (K153299), with the addition of the TXP MicroTip, an upgraded handpiece with indications for use for the treatment of hard tissue and the sharp debridement of wounds such as neuropathic ulcers.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Tenex Health TX System with the TXP MicroTip. It describes the device, its indications for use, and a comparison to predicate devices, along with performance data. However, the document does NOT contain information about acceptance criteria and a study proving the device meets those criteria in the context of AI/ML or human reader performance.

    The "Performance Data" section details non-clinical performance tests which include:

    • Electro-mechanical bench testing (e.g., aspiration vacuum, irrigation flow, operating frequency, needle stroke length)
    • EMC and electrical safety testing
    • Biocompatibility evaluation
    • Sterilization validation
    • Packaging Validation
    • Simulated use testing in cadaver (evaluating design, usability, surgical technique, and effectiveness in sharp debridement of soft and hard tissue in wounds)
    • Simulated use testing in bovine tissue (validating durability, cutting, fragmentation, emulsification, aspiration, and quantifying tissue removal rates)

    These tests are primarily focused on the physical and functional aspects of the surgical device, its safety, and its ability to perform its intended mechanical actions (fragmentation, emulsification, aspiration of tissue). There is no mention of an AI/ML component, diagnostic accuracy studies, or human reader performance studies (MRMC studies).

    Therefore, I cannot provide the requested information regarding:

    1. A table of acceptance criteria and reported device performance related to AI/ML or diagnostic accuracy.
    2. Sample size for a test set in the context of AI/ML, data provenance, number of experts, adjudication methods, MRMC studies, standalone AI performance, or ground truth for such studies.
    3. Training set details related to AI/ML.

    The device is a physical surgical instrument, not a diagnostic or AI-based software device. My capabilities are limited to extracting and rephrasing information explicitly present in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1