510(k) Data Aggregation

    K Number
    K203475

    Validate with FDA (Live)

    Date Cleared
    2021-02-05

    (72 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K153204

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LimaCorporate Kirschner wires are indicated as guide pins for insertion of other implants.

    Device Description

    The LimaCorporate Kirschner wire is a metallic smooth wire, intended as non-implantable medical device.
    This instrument is needed to guide the surgeon during orthopedic surgeries for the bone preparation in partial or total primary or revision shoulder joint replacement.
    The LimaCorporate Kirschner wire is available in single diameter (2.5mm) and various lengths (150mm, 200mm and 240mm).
    The LimaCorporate Kirschner wire is made from stainless steel 316L (AISI 316L), according to ISO 5832-1 and it does not have any coating.
    The LimaCorporate Kirschner wire is designed for single use. It is supplied sterile and it is sterilized by means of radiation.

    AI/ML Overview

    This document (K203475) describes a 510(k) premarket notification for a medical device called the "LimaCorporate Kirschner Wire." The core of a 510(k) submission is to demonstrate "substantial equivalence" to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as typically done for new, higher-risk devices.

    Therefore, the information provided does not contain the typical acceptance criteria and rigorous study design associated with demonstrating performance for an AI/ML medical device. Instead, it focuses on comparing the new device's characteristics to a predicate device.

    Here's an analysis based on the provided document, addressing your points where applicable, and noting where the information is absent due to the nature of a 510(k) for a physical, non-AI device:


    Device Name: LimaCorporate Kirschner Wire
    Device Type: Smooth or threaded metallic bone fixation fastener (Specifically, a smooth metallic wire for guiding other implants).
    Regulatory Class: Class II
    Product Code: HTY

    This document is a 510(k) submission for a physical, non-AI medical device. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are related to demonstrating substantial equivalence to a predicate device, primarily through non-clinical testing and comparison of attributes, not a clinical performance study with defined performance metrics like sensitivity, specificity, or AI model accuracy.

    1. Table of Acceptance Criteria and Reported Device Performance

    For this type of device and submission, the "acceptance criteria" are not performance metrics of an AI model, but rather successful demonstration that the device's technical characteristics are substantially equivalent to the predicate device.

    Acceptance Criteria (for Substantial Equivalence)Reported Device Performance / Evidence
    Identical Design - Kirschner wire design (smooth version).The LimaCorporate Kirschner wire has "identical design compared to the predicate device (K153204. in smooth version)."
    Similar Size Range - Comparison of available diameters and lengths.The LimaCorporate Kirschner wire has "similar size range when compared to the predicate device (K153204)." (Specifically, 2.5mm diameter; 150mm, 200mm, 240mm lengths are mentioned).
    Same Raw Material - Material composition meeting appropriate ISO standard.The LimaCorporate Kirschner wire has the "same raw material, when compared to the predicate device (K153204) and both of them meet appropriate ISO standard." (Specifically, Stainless Steel 316L (AISI 316L) according to ISO 5832-1).
    Included Indications for Use - New device's indications fall within predicate's.The "LimaCorporate Kirschner wire indication for use is included in those of the predicate device (K153204)." (Indication: "as guide pins for insertion of other implants.")
    Sterilization Validation - Proof of sterility for single-use, radiation-sterilized device."Sterilization validation" was performed. (Details of the validation study, e.g., sterility assurance level (SAL), methods, or results are not provided in this summary, but would be part of the full submission).
    Biocompatibility Evaluation - Proof that material is safe for contact with tissue."Biocompatibility evaluation" was performed. (Details, e.g., specific ISO 10993 tests performed or results, are not provided in this summary, but would be part of the full submission).
    Conformity to International Standards - Device meets relevant material standards.The LimaCorporate Kirschner wire "conforms to the international standard ISO 5838-1 (2013)."

    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable to this 510(k) submission, as it is for a physical medical device (Kirschner wire), not an AI/ML software. There is no "test set" of data in the sense of medical images or patient records used to evaluate an algorithm's performance. The "testing" refers to non-clinical bench testing (sterilization validation, biocompatibility) and direct comparison of physical attributes to the predicate.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. There is no AI/ML algorithm or "test set" requiring expert-established ground truth in this submission.

    4. Adjudication Method for the Test Set

    Not applicable, as there is no AI/ML algorithm or "test set" requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. This type of study is relevant for imaging devices or AI-assisted diagnostic tools where human interpretation is involved. This submission is for a Kirschner wire (a surgical instrument).

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    No. This pertains to AI/ML devices.

    7. The Type of Ground Truth Used

    Not applicable in the context of AI/ML. For this device, "ground truth" generally refers to verified physical properties, material composition, and functional characteristics validated through engineering testing and adherence to standards.

    8. The Sample Size for the Training Set

    Not applicable. There is no AI/ML model to train for this physical device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no AI/ML model or training set.


    Conclusion from the document:

    The provided document (K203475) details a 510(k) clearance for the LimaCorporate Kirschner Wire. The basis for clearance is demonstrating substantial equivalence to an existing predicate device (In2Bones® Kirschner Wire, K153204) rather than proving de novo safety and effectiveness through extensive clinical trials or performance studies as would be required for a novel device or AI/ML product. The manufacturer states that "Testing, therefore, is not needed to demonstrate that the subject devices are substantially equivalent to other legally marketed Kirschner wires," implying that the comparison of design, materials, size, and indications, alongside standard non-clinical tests (sterilization, biocompatibility), are sufficient evidence for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191803

    Validate with FDA (Live)

    Date Cleared
    2020-04-14

    (286 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K153204

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Orthopedic Fixation Pin is indicated for fixation of bone fractures, bone reconstruction, and as guide pins for insertion of other implants.

    Device Description

    The Orthopedic Fixation Pin is straight, partially threaded; and has trocar point on one end. The device is available in various diameters and lengths to accommodate the specific indication. The pins are made from stainless steel in accordance with ASTM F138-13.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for an Orthopedic Fixation Pin. It focuses on demonstrating substantial equivalence to a predicate device based on material, design, and intended use, primarily through non-clinical performance data (bench testing and biocompatibility).

    Crucially, this document does not describe a study that uses an AI/ML device, nor does it contain acceptance criteria or performance data for such a device. The "Orthopedic Fixation Pin" is a physical medical device, not a software or AI-driven diagnostic or assistive tool.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as the provided input does not pertain to an AI/ML device.

    Here's why the information you're asking for is absent in the document:

    • Device Type: The "Orthopedic Fixation Pin" is a physical implant used for bone fixation. It is not an AI/ML algorithm or system.
    • Performance Data: The performance data presented (static pullout, static torsion, tensile testing, biocompatibility) are standard engineering and materials tests for physical medical implants, not metrics like sensitivity, specificity, AUC, or reader performance improvement typically used for AI/ML devices.
    • Clinical Studies: The document explicitly states, "No data from human clinical studies have been included to support the substantial equivalence of the proposed device, Orthopedic Fixation Pin, as clinical studies are not required for this medical device." This confirms that no clinical performance data, let alone data relevant to AI/ML device evaluation (like MRMC studies or human-in-the-loop performance), was generated or submitted.
    • Ground Truth: The concept of "ground truth" as it applies to AI/ML (e.g., expert consensus on image interpretation, pathology reports) is not relevant to a physical pin. The "ground truth" for the pin is its adherence to material specifications and mechanical properties.

    If you have a document describing an AI/ML medical device, please provide that, and I would be able to answer your questions.