Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243227

    Validate with FDA (Live)

    Device Name
    B-Scan
    Date Cleared
    2025-07-11

    (276 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K070943, K123349, K152573

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B-Scan module is used for imaging the internal structure of the eye, including the opaque media and posterior pathology, for the purpose of diagnosing pathological or traumatic conditions in the eye. The probe is intended to be used on both adult and pediatric patients that require imaging of the eye.

    Device Description

    The B-Scan device is designed as an ultrasound B-Scan, which uses pulsed echo ultrasound to image the structure of the eye. It utilizes an eye-contact probe to generate and receive the ultrasound pulse signals and provides a graphic display of returning pulse echoes to indicate the various structures.

    All of the critical functions of the B-Scan are calculated in the same manner as in the predicate device, B-Scan Plus. The software algorithms for clinically critical functions remain the same as in the predicate device. However, the user interface and the workflow of B-Scan have enhancement to support cybersecurity implementation.

    Both the subject device and predicate device are compatible with the Connect Software (K070943, K123349) and 4Sight (K152573). The software improvements in Connect were focused on enhancing features that optimize integration with the personal computer's processing, data storage, display, and printing capabilities.

    The energy source for the B-Scan is USB power as in the predicate device, B-Scan Plus.

    The software utilized on both the Connect and 4Sight platforms is fundamentally identical in core clinical functions, including image scanning, rendering, IOL calculations, data storage, and report formatting. Additionally, the key features within the B-Scan module do not differ significantly between the two systems, nor do they make an impact on the established clinical workflow.

    Track 1 is being followed for this 510(k) submission.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the B-Scan Ultrasonic Imaging System, based on the provided FDA 510(k) clearance letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary primarily focuses on demonstrating substantial equivalence to a predicate device rather than explicitly stating pre-defined "acceptance criteria" for a novel device performance. However, we can infer acceptance criteria from the comparison table (Table 2) and the performance data section, where the subject device's performance is either "Same" or "SE" (Substantially Equivalent) to the predicate, or explicitly lists performance metrics.

    For quantitative metrics, the "acceptance criteria" for the subject device can be interpreted as performing comparably or within acceptable limits relative to the predicate device, or meeting specific new specifications.

    ParameterAcceptance Criteria (Inferred/Stated)Reported Device Performance (B-Scan)Remarks/Proof
    Clinical Accuracy (Line)±3% (for 60mm Depth Setting)12MHz: 60mm Depth Setting Clinical Accuracy: ±3%Bench Test: Physical Accuracy and Range Test
    Accuracy Range (Line)60 mm (for 60mm Depth Setting)12MHz: 60mm Depth Setting Accuracy Range: 60 mmBench Test: Physical Accuracy and Range Test
    Clinical Accuracy (Area)±15% (for 60mm Depth Setting)12MHz: 60mm Depth Setting Clinical Accuracy: ±15%Bench Test: Physical Accuracy and Range Test
    Accuracy Range (Area)60 mm (for 60mm Depth Setting)12MHz: 60mm Depth Setting Accuracy Range: 60 mmBench Test: Physical Accuracy and Range Test
    Image Preview TimeDisplayed within 2 seconds of pressing probe buttonImage preview is displayed within 2 seconds of pressing probe button (average time recorded across 5 probes).Bench Test: The specific bench test for this is not named explicitly, but is mentioned within the comparison table as part of the performance metrics.
    Cleaning & DisinfectionValidation per procedures in Instruction for UseReprocessing validation test was conducted on the proposed device.Reprocessing Validation Test
    BiocompatibilityMeet requirements of ISO 10993 series (Cytotoxicity, Ocular Irritation, Skin Sensitization)The proposed device has been tested and met the requirements according to the ISO 10993 series standard for Biocompatibility.Biocompatibility Testing
    Software V&VDocumentation provided as recommended by FDA guidanceSoftware verification and validation testing were conducted, and documentation was provided.Software Verification and Validation Testing
    Cybersecurity ComplianceDocumentation provided in accordance with FDA guidanceCybersecurity compliance was implemented, and documentation was provided.Cybersecurity Implementation and Documentation
    Electrical Safety & EMCMeet requirements of IEC 60601-1, IEC 60601-1-2The proposed device has been tested and met the requirements.Thermal, mechanical, and electrical safety and electromagnetic compatibility testing
    Mechanical/Thermal DurabilityMeet design specifications (implied)Accelerated Thermal Cycling Test; Plastic Component UV Assessment; Transit Performance Test; USB cable Pull Test; B-Scan Integrated Life Testing Verification ReportBench Tests (listed)
    Button ActuationMeet design specifications (implied)Button actuator Validation TestBench Test: Button actuator Validation Test

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a sample size for patients or images used in any clinical or test set. The performance data mostly refers to bench testing and validation against standards.

    • The only mention of a "sample" related to performance is "average time recorded across 5 probes" for the image preview time. This refers to hardware units, not patient data.
    • Data Provenance: Not applicable, as no external data set or clinical study on patients is described beyond bench tests. The focus is on device specifications and in-house validation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. The submission does not describe a study involving expert readers establishing ground truth for a test set of images for diagnostic performance evaluation. The "Performance" section outlines bench tests and compliance with recognized standards.

    4. Adjudication Method for the Test Set

    Not applicable. Since there's no mention of a clinical test set requiring human expert review to establish ground truth, there is no adjudication method described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not describe a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor does it quantify improvement in human readers with AI assistance. The device is purely an imaging system, and there is no mention of AI assistance for image interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-loop Performance)

    The device described is an "B-Scan Ultrasonic Imaging System," which is a diagnostic tool that produces images for human interpretation. It does not contain an AI algorithm for standalone diagnostic performance. Its performance is related to image acquisition parameters, accuracy of measurements, and adherence to safety/design standards.

    7. Type of Ground Truth Used

    For the quantitative performance claims (e.g., Clinical Accuracy Line/Area), the ground truth was established by physical measurements against known standards or calibrated references during bench testing ("Physical Accuracy and Range Test"). For other aspects, the ground truth is adherence to technical specifications, safety standards, and validated manufacturing/reprocessing procedures.

    8. Sample Size for the Training Set

    Not applicable. This device is an ultrasound imaging system, not an AI-driven diagnostic algorithm that requires a "training set" of data in the typical machine learning sense. The software aspects mentioned are primarily for device control, image rendering, data storage, and report formatting, enhanced for cybersecurity and usability – not for learning from data to perform a diagnostic task.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set mentioned for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1