510(k) Data Aggregation

    K Number
    K180559

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-10-29

    (242 days)

    Product Code
    Regulation Number
    866.3309
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K151906

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HSV 1&2 ELITe MGB® Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of Herpes Simplex Virus 1 and 2 (HSV-2) DNA in cutaneous or mucocutaneous lesion swab specimens from patients with signs and symptoms of HSV-2 infection. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 infections.

    The HSV 1&2 ELITe MGB Assay is not FDA cleared for use with cerebrospinal fluid (CSF) specimens. The assay is not intended to be used for prenatal screening or for screening blood or blood products.

    Device Description

    The HSV 1&2 ELITe MGB Assay is a qualitative in vitro diagnostic Real-Time PCR Assay for the direct detection of Herpes Simplex Virus (HSV) DNA (glycoprotein D gene for HSV-1 and glycoprotein G gene for HSV-2) in symptomatic male and female patients using DNA purified from swab specimens collected from individuals with cutaneous or mucocutaneous herpetic lesions.

    The HSV 1&2 ELITe MGB Assay system is comprised of three major processes: (1) automated preparation of unprocessed sample to extract nucleic acids from primary swab specimens using the ELITe InGenius SP 200 Extraction Cartridge, (2) PCR amplification of target DNA sequences using HSV-1 and HSV-2 specific primers, and (3) real-time detection of fluorescent-labeled HSV-1 and HSV-2 specific oligonucleotide detection probes.

    An Internal Control (IC), containing unrelated randomized DNA sequence, is added to all samples prior to extraction and monitors the integrity of the reagents, equipment function, and the presence of inhibitors in the samples. A positive signal in the Internal Control channel in the absence of HSV DNA indicates that the PCR has not been inhibited.

    The amplification reagents, Positive Control and Internal Control are packaged as part of the HSV 1&2 ELITe MGB Assay.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the HSV 1&2 ELITe MGB Assay, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the HSV 1&2 ELITe MGB Assay are presented in terms of Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) against a composite reference method.

    MetricAcceptance Criteria (Implied by achieved performance and regulatory clearance)Reported Device Performance (95% CI) - Cutaneous Lesion SamplesReported Device Performance (95% CI) - Mucocutaneous Lesion Samples
    HSV-1 PPAHigh (e.g., typically >90-95%)98.7% (93.2-99.8%)99.2% (95.7-99.9%)
    HSV-1 NPAHigh (e.g., typically >90-95%)98.5% (96.9-99.3%)97.6% (95.8-98.6%)
    HSV-2 PPAHigh (e.g., typically >90-95%)96.2% (91.3-98.3%)97.6% (94.0-99.1%)
    HSV-2 NPAHigh (e.g., typically >90-95%)98.6% (96.9-99.3%)98.2% (96.6-99.1%)

    Study Information

    1. Sample Size Used for the Test Set and Data Provenance:

      • Test Set Sample Size: 1,174 prospectively collected archived swab samples were initially collected. After exclusions:
        • 1,171 samples were analyzed for HSV-1.
        • 1,170 samples were analyzed for HSV-2.
        • A separate contrived oral panel study involved 75 individual negative cheek swab samples spiked with HSV-2, plus 10 HSV-1 positive samples and 15 HSV-1/HSV-2 negative oral samples.
      • Data Provenance: The samples were "left-over prospectively collected archived swab samples" from symptomatic patients. The country of origin is not explicitly stated, but the context of an FDA submission generally implies US-based data.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • The document does not explicitly state the number of experts or their qualifications for establishing the ground truth. The ground truth was established using a composite reference method.
    3. Adjudication Method for the Test Set:

      • The adjudication method for the test set was a "composite reference method" defined as: "an FDA cleared assay and a validated HSV 1&2 PCR followed by bi-directional sequencing of gel electrophoresis-positive samples."
      • "A positive result by the composite reference method is defined as a positive by the FDA cleared PCR or the validated sequencing. Two negative results are needed to confirm a negative." This implies a type of adjudicted consensus, where the "truth" is determined by agreement or hierarchical resolution between different methods, rather than human experts per se.
    4. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

      • No. This document describes the analytical and clinical performance of an in vitro diagnostic (IVD) assay, which is an algorithm-only (standalone) test. It does not involve human readers in the primary diagnostic decision-making process that would warrant an MRMC study comparing human performance with and without AI assistance.
    5. If a Standalone Performance Study Was Done:

      • Yes. The entire clinical performance section (Section 11.b) is a standalone performance study where the HSV 1&2 ELITe MGB Assay's performance is compared directly against a composite reference method. The assay operates without human intervention in the interpretation of results.
    6. The Type of Ground Truth Used:

      • Composite Reference Method: This ground truth combined an FDA-cleared PCR assay and a validated HSV 1&2 PCR followed by bi-directional sequencing. This type of ground truth integrates highly sensitive and specific molecular methods, reflecting a robust laboratory-based gold standard.
    7. The Sample Size for the Training Set:

      • The document does not explicitly mention a "training set" in the context of an AI/ML algorithm. This device is a real-time PCR assay, which is a chemical and enzymatic reaction rather than a machine learning model that would typically require a separate training set. The various analytical performance studies (Limit of Detection, Analytical Reactivity, Specificity, etc.) serve to characterize the assay's performance across different conditions, which is analogous to how a system is "trained" or refined during development, but not in the sense of a distinct machine learning training set. The "assay cut-off" was established using a separate set of 141 clinical samples, which could be considered a form of calibration/training for the threshold of positivity.
    8. How the Ground Truth for the Training Set Was Established:

      • As noted above, a traditional "training set" for an AI model is not applicable here. For the establishment of the assay cut-off (using 141 clinical samples), the ground truth was established using the same "composite reference method" (FDA-cleared real-time PCR assay combined with PCR amplification and bidirectional sequencing). For other analytical studies (like LoD, inclusivity, cross-reactivity), quantitated viral strains or characterized organisms were used as the reference.