Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221178

    Validate with FDA (Live)

    Date Cleared
    2022-11-17

    (206 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K162602, K142518, K100988, K151686

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Insulin Pen Needle(Model CT) : Disposable Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.

    Disposable Insulin Pen Needle(Model ST) : Disposable Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. Additionally, it has sharps injury protection feature that it can reduce the occurrence of accidental needlesticks from the patient end of the needle due to the effect of safe-lock guard.

    Device Description

    Disposable Insulin Pen Needle, could be divided into Model CT (normal type) and model ST(safety type) according to whether there is sharps protection features. Model CT consists of needle hub, needle container, needle tube, needle shield and seal (dialyzing paper). Model ST consists of needle hub, needle container, protective cap, needle tube, spring, slider, protective jacket and seal (dialyzing paper). The device is sterilized by EO gas and for single use. The shelf life is 5 years.

    Needle tube is fixed in the center of the needle hub with UV glue, the needle tip is protected with needle shield or protective jacket before use, medical dialyzing paper is covered and sealed needle container to maintain sterile of the cartridge end of needle can be inserted into the rubber of insulin pen, meanwhile the needle hub is connected to insulin pen with screw thread to provide sterile fluid path for injection of insulin during use. When Disposable Insulin Pen Needle is connected to insulin pen operates on the principles of common piston syringes. The patient end and the cartridge end of the needle tube are lubricated with silicone oil for ease of injection and rubber penetration.

    Model ST has sharps injury protection feature that it can reduce the occurrence of accidental needlesticks from the patient end of the needle due to the effect of safe-lock guard.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a Disposable Insulin Pen Needle. It focuses on demonstrating substantial equivalence to predicate devices through design comparisons and a summary of performance testing. The document does not describe a study that involves acceptance criteria for device performance based on AI or diagnostic accuracy. Instead, it details the physical and functional performance of the device based on established medical device standards.

    Therefore, I cannot provide information about:

    • A table of acceptance criteria and reported device performance related to diagnostic accuracy or AI.
    • Sample size used for the test set and data provenance for an AI model.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for a test set.
    • MRMC comparative effectiveness study or human reader improvement with AI.
    • Standalone algorithm performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data).
    • Sample size for the training set.
    • How ground truth for the training set was established.

    However, I can extract information regarding the device's general performance testing and criteria based on the provided document:

    The device's performance is evaluated against recognized international standards for medical devices and sterility.

    Here's a summary of the performance testing and criteria mentioned, which serves as the "acceptance criteria" in the context of this medical device submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance Criteria (Standard / Requirement)Reported Device Performance
    Sterilization- EO gas sterilization with Sterility Assurance Level (SAL) of 10⁻⁶, validated per ISO 11135:2014. - Maximum EO residual < 0.8mg. - Maximum ECH residual < 1.8mg, according to ISO 10993-7:2008.- Device is sterilized by EO gas; SAL is 10⁻⁶. - Process validated per ISO 11135:2014. - EO residual < 0.8mg. - ECH residual < 1.8mg.
    Product Performance (General)- Meet criteria specified in ISO 11608-2:2012 (Needle-based injection systems). - Meet criteria specified in ISO 9626:2016 (Stainless steel needle tubing). - Meet criteria specified in ISO 7864:2016 (Sterile hypodermic needles).- Meets criteria specified in ISO 11608-2, ISO 9626, and ISO 7864 and other applicable standards.
    Functional Compatibility- Compatible with commonly used insulin pens in the U.S. market.- Tests completed; insulin pens declared in label are compatible with the needle. (Specific compatible pens listed: NovoPen Echo, Humapen Luxura, Humapen Luxura HD, HumanPen Ergo II)
    Transport- Meet requirements of ASTM D4169-16 (Standard Practice for Performance Testing of Shipping Containers and Systems).- Simulated transportation test completed per ASTM D4169-16. Test result: passed (for Model CT and Model ST).
    Shelf Life- Maintain performance (ISO 11608-2), chemical properties, bacterial endotoxin, and leakage (ASTM F1929-2015) after accelerated aging.- Performance, chemical properties, bacterial endotoxin, and leakage tested before accelerated aging, after 4 years, and 5 years of accelerated aging. Determined shelf life: 5 years. (Accelerated aging test by ASTM F88/F88M-2015).
    Biocompatibility- Meet requirements of ISO 10993 series standards. - Specific tests: Cytotoxicity, skin sensitization, intracutaneous reactivity, acute systemic toxicity, hemocompatibility, subchronic toxicity, material-mediated pyrogenicity. - Particulate matter for injections per USP <788>.- Meets requirements of ISO 10993 series standards. - All listed specific tests performed. - Particulate matter for injections performed per USP <788>.
    Sharps Protection Features- For Model ST: Meet requirements of safe-lock guard activation and safety overriding/unlocking force after activation (triggering performance) as per ISO 23908:2011 (Sharps injury protection).- Model ST meets the requirements of safe-lock guard activation and safety overriding/unlocking force after activation per ISO 23908:2011.
    Clinical Simulated Use- For Model ST: Demonstrate sharps injury prevention feature through clinical simulated use testing, according to ISO 23908:2011 and FDA Guidance for Industry and FDA Staff-Medical Devices with Sharps Injury Prevention Features.- Clinical simulated use testing completed for Model ST. Test results show good sharps injury prevention feature.
    Hub/Needle Bond Strength- Conform with ISO 11608-2.- Conform with ISO 11608-2.
    Method of Attachment- Screw structure, conform with ISO 11608-2.- Screw structure, conform with ISO 11608-2.
    Needle Tip Configuration- Tri-bevel edge needle.- Tri-bevel edge needle.
    Endotoxin Limit- 20 EU per device.- 20 EU per device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" in the context of diagnostic performance or AI. Instead, it refers to samples of the device being manufactured and subjected to various physical, chemical, and biological tests according to the standards listed above. The provenance of these test samples would be the manufacturer's production line (Ningbo Medsun Medical Co., Ltd., China). The tests are prospective in the sense that they are performed on the manufactured device models to demonstrate compliance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable, as this is not a diagnostic device involving expert interpretation or AI. The "ground truth" here is compliance with established international and national medical device standards, which are defined by committees of experts in the field.

    4. Adjudication Method for the Test Set

    Not applicable. The criteria are objective measurements and validations against standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not a diagnostic device or AI-driven system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is a physical medical device.

    7. The type of ground truth used

    The "ground truth" is defined by the technical specifications and performance requirements outlined in the cited international and national standards (e.g., ISO 11135, ISO 11608-2, ISO 9626, ISO 7864, ISO 10993 series, ASTM D4169-16, ASTM F1929-2015, USP <788>, ISO 23908:2011). These standards represent established scientific and engineering principles for medical device safety and efficacy.

    8. The Sample Size for the Training Set

    Not applicable, as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1