Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K232177

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-07-10

    (352 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K151604, K233037

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Lightin is a needle for injection laryngoplasty with an optical fiber. Lightin Generator is an AC-powered light source including Light Guide which is fiber optic cable that transmits light at the tip of fiber optics embedded within Lightin. Lightin System is used for injection laryngoplasty to address glottic incompetence to indicate the injection site with visible light.

    Device Description

    Lightin System consists of Lightin and Lightin Generator. Lightin is a light-guided needle used injection laryngoplasty and Light generator delivers the light energy to Lightin by generating LED light from an AC-powered light source. Lighin consists of injection needle, Y-connector and SMA connector/cable. The light transmitted from the Lightin Generator is emitted from the tip of fiber optics embedded within Lightin. Lightin system used for injection laryngoplasty procedures to indicate the injection site with visible light. This system does not contain or deliver drugs.

    AI/ML Overview

    The provided document is a 510(k) summary for the Solmedix Lightin System, which is a medical device for injection laryngoplasty. This summary primarily focuses on establishing substantial equivalence to predicate devices rather than directly presenting acceptance criteria and detailed study results in a format typical for an AI/ML device.

    Here's an analysis based on the information provided, specifically addressing the points requested, while noting that many are not applicable or not explicitly detailed for this type of device submission:

    1. Table of acceptance criteria and the reported device performance

    For a Class I device seeking 510(k) clearance, the "acceptance criteria" are generally compliance with recognized standards and demonstration that the device performs as intended and is as safe and effective as a predicate device. This document lists the standards the device conforms to and describes various tests performed. It does not provide specific quantitative performance metrics with associated acceptance thresholds in the way an AI/ML device submission often would.

    However, based on the non-clinical performance testing sections, we can infer some criteria and reported performance:

    Criterion TypeAcceptance Criteria (Inferred from testing)Reported Device Performance
    Lightin (Needle)
    BiocompatibilityConformity to ISO 10993 series for cytotoxicity, skin sensitization, irritation, systemic toxicity, and pyrogenicity.Lightin was "accessed for conformity" and presumably passed all listed ISO 10993 tests (ISO 10993-5, -10, -23, -11) and USP <151> Pyrogen Test.
    Small-bore ConnectorsConformity to ISO 80369-7 and ISO 80369-20 for intravascular/hypodermic connectors and common test methods.Lightin was "accessed for conformity" and presumably passed ISO 80369-7 and ISO 80369-20.
    Hypodermic NeedlesConformity to ISO 7864 for sterile hypodermic needles.Lightin was "accessed for conformity" and presumably passed ISO 7864.
    Stainless Steel Needle TubingConformity to ISO 9626 for stainless steel needle tubing.Lightin was "accessed for conformity" and presumably passed ISO 9626.
    Sterilization (Ethylene Oxide)Conformity to ISO 11135 for EO sterilization development, validation, and routine control.Lightin was "accessed for conformity" and presumably passed ISO 11135.
    Packaging IntegrityConformity to ASTM F1980 (Accelerated Aging), ASTM F1929 (Dye Penetration), and ASTM F88/F88M (Seal Strength).Lightin was "accessed for conformity" and presumably passed ASTM F1980, ASTM F1929, and ASTM F88/F88M.
    SterilityConformity to USP <71> Sterility Test.Lightin was "accessed for conformity" and presumably passed USP <71>.
    Light OutputInternal test standard for light output.Lightin was "accessed for conformity" and presumably passed the internal light output test. (No specific quantitative value provided).
    Thermal MeasurementInternal test standard for thermal measurement.Lightin was "accessed for conformity" and presumably passed the internal thermal measurement test. (No specific quantitative value provided).
    Performance (Laryngoplasty)Successfully directs injection material to target area within the vocal fold.The Lightin System "successfully directs the laryngoplastic injection material to the target area within the vocal fold." (Qualitative statement).
    Tissue Heating ConcernsAddresses tissue heating concerns by tip illumination.Tissue heating concerns by tip illumination were addressed. (Qualitative statement, likely related to the "Thermal measurement test").
    Lightin Generator
    Electrical Safety & EMCConformity to IEC 60601-1 (general safety) and IEC 60601-1-2 (EMC).Lightin Generator was "accessed for conformity" and presumably passed IEC 60601-1 and IEC 60601-1-2 (including IEC TR 60601-4-2 guidance).
    Photobiological SafetyConformity to IEC 62471 for photobiological safety of lamps and lamp systems.Lightin Generator was "accessed for conformity" and presumably passed IEC 62471.
    Light OutputInternal test standard for light output.Lightin Generator was "accessed for conformity" and presumably passed the internal light output test. (No specific quantitative value provided beyond the max output ≤ 8mW in the comparison table).

    2. Sample size used for the test set and the data provenance

    The document does not specify sample sizes for any of the tests conducted or the data provenance (e.g., country of origin, retrospective/prospective). This level of detail is typically not required or provided in a 510(k) summary for a Class I manual surgical instrument unless specific performance claims necessitate it. The tests listed are primarily engineering, electrical, and biocompatibility tests on device components or finished products, not clinical studies with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable. The device is a surgical instrument, not an AI/ML diagnostic or prognostic tool that relies on expert interpretation of data to establish ground truth for training or testing. The "ground truth" for this device relates to its physical performance, safety, and functionality (e.g., biocompatibility, sterility, light output), which are assessed through standardized laboratory and technical tests.

    4. Adjudication method for the test set

    This is not applicable for the reasons stated in point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable, as the device is not an AI/ML product designed to assist human readers/interpreters.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable, as the device is not an AI/ML algorithm. It is a physical medical instrument.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established through adherence to recognized international and national standards (e.g., ISO, IEC, ASTM, USP) for materials, electrical safety, biocompatibility, sterilization, and basic functionality. For instance, successfully passing a cytotoxicity test per ISO 10993-5 serves as the "ground truth" for non-cytotoxicity. The statement that the system "successfully directs the laryngoplastic injection material to the target area" is a functional performance claim, likely verified through internal testing or benchtop models.

    8. The sample size for the training set

    This is not applicable, as there is no mention of a training set for an AI/ML model for this device.

    9. How the ground truth for the training set was established

    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182646

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-05-01

    (219 days)

    Product Code
    Regulation Number
    886.4690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K121187, K151604

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VitreQ Disposable Laser Probes, 90° Directional Laser Probes and 90° Directional Illuminated Laser Probes are intended for use in vitreoretinal surgery to perform endo-ocular laser photocoagulation treatments at operating wavelengths of 500nm to 900nm.

    The 90° Directional Illuminated Laser Probes, Light Fibers and Chandelier are for illumination during ophthalmic surgery and should only be used with the light wavelength range of 425nm to 700mm.

    Device Description

    The VitreQ Disposable Laser Probe is an ophthalmic instrument, which is used in posterior segment eye surgery and is appropriate for photocoagulation. The device can be connected to an ophthalmic laser generator, which is not a part of this application.

    The VitreQ Disposable Laser Probe is introduced into the posterior segment of the eye through an incision or an ophthalmic entry port, after the vitreous is removed during vitreoretinal surgery.

    The Disposable Laser Probe guides the laser energy to the intended surgical site to provide photocoagulation treatment. The laser enerqy intensity or power output is not controlled, altered or significantly reduced by the disposable laser probe. The laser spot size on the target tissue can be varied by altering the distance between the tissue and the probe tip.

    To reach the target tissue in the periphery section of the 90° Directional Laser Probe tip can be directed due to bending the metal tip. The angle of the tip in the most directed position is about 90°. An angled laser probe reduces the need of a strong indentation of the eye and is intended to prevent accidently touching the crystal lens.

    The VitreQ Disposable Laser Probes, 90° Directional Laser Probes and (90°) Directional Illuminated Laser Probes are constructed with an optical laser connector, a glass fiber optic covered by a protective sheath and one handle for surgeon manipulation, metal tubing is extending from the handle which penetrates the surgical site. The device is intended for singleuse only.

    The VitreQ Disposable Laser Probes can only be used with a medical laser at an operating wavelength range of 500nm to 900nm.

    For the VitreQ Disposable Laser Probes, (90°) Directional Illuminated Laser Probe the functionality is combined with illumination. To achieve the illumination of the surgical site, the probe is constructed with an additional plastic fiber, which has a connector to attach the fiber to an ophthalmic light source, which is not a part of this application.

    The VitreQ Disposable Light Fibers are constructed with a handle, the plastic illumination fiber, and the connector to attach the fiber to an ophthalmic light source. To match the connection to the ophthalmic light source, there are several reusable light source adaptors available.

    The VitreQ (90°) Directional Illuminated Laser Probes, Light Fibers and Chandelier can only be used with the light wavelength range of 425nm to 700nm.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification from Vitreq B.V. for their Disposable Laser Probes, Light Fibers, and Chandelier. It seeks to demonstrate substantial equivalence to previously cleared predicate devices.

    Here's an analysis of the provided information regarding acceptance criteria and performance studies:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly present a formal "acceptance criteria" table with corresponding numerical performance results in the format requested. Instead, it describes performance characteristics and compares them to the predicate device, emphasizing substantial equivalence.

    Based on the "Summary of technological characteristics" and "Differences in performance" sections, we can infer some performance metrics and their comparison:

    CharacteristicAcceptance Criteria (Implied from Predicate/SE Discussion)Reported Device Performance (Subject Device)SE Discussion (Comparison)
    Laser Performance
    Laser Wavelength500nm to 900nm500nm to 900nmNot applicable (identical)
    Max Threshold of Laser Fiber3 Watt3 WattNot applicable (identical)
    Transmission of Treatment Laser99.6% (Predicate)99.8%Substantial Equivalent. The difference between predicate and subject device are negligible.
    Laser Power Efficiency0.996 (Predicate)0.998Substantial Equivalent. The difference between predicate and subject device are negligible.
    Laser Output Power$500 \pm 20%$ mW (Predicate)$500 \pm 15%$ mWSubstantial Equivalent. The Vitreq device has more strict tolerances than the predicate device.
    Laser Size (NA value)between 0.1 and 0.2 (Predicate)between 0.1 and 0.2Not applicable. Bench testing is performed (see Tab 18).
    Illumination Performance
    Light Wavelength425nm to 700nm425nm to 700nmNot applicable (identical)
    Transmission of Illumination Light (Illuminated Laser Probes)Unknown (Predicate) / Less than light fibers/chandeliers12.5%Identical, transmission of illuminated laser probes is less than light fibers/chandeliers.
    Transmission of Illumination Light (Light Fibers/Chandelier)39% (Predicate)39%Identical.
    Light TypeXenon or Mercury (Predicate)LED, Xenon or MercurySubstantial Equivalent. Vitreq also tested the LED, art technology. The values Xenon and LED are substantial equivalent. Bench testing is performed see Tab 18.
    Safety/Biocompatibility/Sterilization
    Sterilant ResidualsBelow allowable limits (implied)Below allowable limitsSuccessful adoption to validated Ethylene Oxide method.
    Bacterial EndotoxinsBelow allowable limits (implied)Below allowable limitsSuccessful adoption to validated Ethylene Oxide method.
    Shelf Life5 years after sterilization (implied via study)5 years after sterilizationShelf life study performed to establish expiration date.
    BiocompatibilityISO 10993-1 and FDA quidance compliant (implied)Evaluated and tested according to ISO 10993-1 and related FDA guidance.Claimed compliant.
    Light Hazard ProtectionAccording to ISO 15004-2 (Predicate) / ANSI Z80.36 (Subject)According to ANSI Z80.36-2016Information to the user is provided according to standard ANSI Z80.36-2016.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not provide explicit sample sizes for any specific test sets for performance. It mentions "Performance Testing Summary" indicating that the devices "have been tested to meet the product requirements and requirements from (safety) standards," covering "performance testing and human factors testing." However, details on the methodology, sample sizes, and specific data provenance (country, retrospective/prospective) are not provided in this public summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not available in the provided document. The document describes a medical device (laser probes and light fibers) for surgical use, not an AI/diagnostic device that typically involves expert ground truth establishment for test sets.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable or not available in the provided context. Adjudication methods are typically relevant for diagnostic studies where expert disagreement on ground truth occurs. This document focuses on the physical and functional performance of a surgical instrument.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable as the device described is a physical surgical tool (laser probes and light fibers), not an AI-assisted diagnostic or decision support system that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable as the device is a physical surgical tool, not an algorithm or software-only device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the performance testing mentioned (e.g., laser transmission, power output, light transmission, biocompatibility), the "ground truth" would be established by well-defined engineering specifications, international standards (e.g., ISO, ANSI), and validated testing methodologies against which the device's physical and functional properties are measured. It does not involve expert consensus, pathology, or outcomes data in the way a diagnostic AI device would.

    8. The sample size for the training set:

    This information is not applicable. The device is a physical medical instrument, not an AI model that requires a training set.

    9. How the ground truth for the training set was established:

    This information is not applicable as the device is a physical medical instrument, not an AI model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1