510(k) Data Aggregation

    K Number
    K192684

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-01-22

    (118 days)

    Product Code
    Regulation Number
    876.5550
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K150870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroPearl® Microspheres are intended for the embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH).

    Device Description

    The HydroPearl TM Microspheres are a pre-formed, compressible, precisely calibrated, spherical embolic agent consisting of a biocompatible hydrogel. The HydroPearl TM Microspheres are offered in a variety of diameters ranging from 75-1100µm and are provided in a sterile syringe pre-filled with microspheres in phosphate buffered saline. The pre- filled syringe is packaged in a sealed sterile dispenser tray. The HydroPearl TM Microspheres are delivered to the treatment site through a delivery catheter.

    AI/ML Overview

    The provided document describes the acceptance criteria and study for the HydroPearl Microspheres, specifically for the expanded indication of prostatic artery embolization (PAE) for symptomatic benign prostatic hyperplasia (BPH).

    Here's an organized breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria with specific numerical thresholds (e.g., "IPSS score reduction must be X%"). Instead, it reports the observed clinical outcomes and concludes that these results support safe and effective use. The "acceptance criteria" here are implied through the positive clinical outcomes and the comparison to established understanding of PAE effectiveness.

    Acceptance Criteria (Implied)Reported Device Performance (HydroPearl Microspheres)
    Primary Endpoint: Reduction in IPSS score during follow-upMean baseline IPSS: 20.7 (range; 4-30) Mean IPSS at latest follow-up (8.6 months): 9.1 (range; 2-18) (p<0.05)
    Secondary Endpoint: Improvement in Quality of Life (QoL)Mean baseline QoL score: 3.7 (range; 1-6) Mean QoL score at follow-up: 1.2 (range; 0-3) (p<0.05)
    Technical Success: Bilateral prostatic artery embolizationAchieved in all 17 patients (100%)
    Clinical Success (based on symptom improvement/medication reduction)Achieved in 16/17 patients (94%) One patient considered a "clinical failure" (had LUTS well controlled on medication, wanted to cease medication, which was achieved, but symptoms mildly worsened from IPSS 4 to 10). Half of patients on BPH medications at baseline no longer required them after PAE; two additional patients reduced daily medications. Patient relying on CIC: Catheterized 5 times/day before PAE, 0 or 1 time/day after PAE.
    Safety: Low rate of adverse eventsOne major complication: hospitalization for acute prostatitis (n=1), resolved without residual effects. Minor complications: hematuria (n=1), hematospermia (n=1), considered expected for PAE and likely due to mild non-target embolization.
    Equivalence to Predicate DeviceComparative mechanical testing showed equivalent performance to Embosphere Microspheres. Pre-clinical animal testing showed similar performance to Embosphere Microspheres with favorable embolization results.

    2. Sample size used for the test set and the data provenance

    • Sample Size (Test Set): 17 patients
    • Data Provenance: Retrospective analysis from a single US center.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts for establishing ground truth for the clinical data. The data appears to be derived from standard clinical assessments (IPSS, QoL scores, observation of complications) and procedural records (technical success of embolization), rather than an explicit "ground truth" established by independent expert panel review.

    4. Adjudication method for the test set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the clinical outcomes in the retrospective study. Clinical outcomes were reported based on collected patient data.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study focuses on the device's clinical performance in patients, not on comparing AI assistance to human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the HydroPearl Microspheres are a medical device (embolization agent), not an AI algorithm. The study assesses the performance of the physical device when used by clinicians.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the clinical study's effectiveness was established through patient-reported outcomes data and objective clinical assessments:

    • International Prostate Symptom Score (IPSS) and Quality of Life (QoL) scores (patient-reported).
    • Observation of BPH medication usage reduction/cessation.
    • Objective assessment of technical success (bilateral PAE).
    • Adverse event reporting.

    8. The sample size for the training set

    The document does not describe a "training set" in the context of an AI algorithm. For the clinical performance anaysis, it refers to a retrospective analysis of existing data.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI algorithm study requiring a training set with established ground truth in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171946

    Validate with FDA (Live)

    Device Name
    Gel-Bead
    Date Cleared
    2017-10-25

    (118 days)

    Product Code
    Regulation Number
    870.3300
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K150870

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gel-Bead embolization spheres are intended for use in embolization of hypervascular tumors.

    Device Description

    The Gel-Bead embolization spheres (Gel-Bead) consists of biodegradable gelatin spheres prefilled in a 20 ml syringe. The syringe contains 2 ml of spheres suspended in 4 ml of saline. Gel-Bead is offered in four nominal size ranges: 100-300 um, 300-500 um, 500-700 um and 700-1000 um. The spheres are intended to be used with a delivery catheter with an inner diameter that is adequate for sphere delivery (not included). The finished product is sterilized by Gamma irradiation and is intended for single use only.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Gel-Bead device, based on the requested information:

    This device is a medical device (embolization spheres), not an AI/ML device. Therefore, many of the requested fields pertinent to AI/ML device studies (like those related to AI effectiveness, ground truth for training sets, etc.) are not applicable in this context and will be marked as such.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance CharacteristicReported Device Performance (Met/Not Met)
    Sphere suspension in contrastMet acceptance criteria
    Pepsin digestionMet acceptance criteria
    Supplied volumeMet acceptance criteria
    DeliverabilityMet acceptance criteria
    Glutaraldehyde residualsMet acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for each of the bench tests (sphere suspension, pepsin digestion, supplied volume, deliverability, glutaraldehyde residuals). It also does not explicitly mention the data provenance in terms of country of origin or whether it was retrospective or prospective, as these are bench tests rather than clinical studies with patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. These are bench tests measuring physical and chemical properties of the device, not diagnostic interpretations requiring expert ground truth in the context of AI/ML.

    4. Adjudication Method for the Test Set

    Not applicable, as ground truth establishment by experts is not relevant for these bench tests. The acceptance criteria were likely objective measurements defined by the manufacturer based on regulatory guidelines and validated methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is not an AI/ML device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    Not applicable. This is not an AI/ML device.

    7. The Type of Ground Truth Used

    The "ground truth" for these bench tests would be the established scientific and engineering principles, regulatory standards, and internal specifications for the physical and chemical properties of the embolization spheres. For example, for "supplied volume," the ground truth would be the expected nominal volume (2 ml) and acceptable deviation from that volume. For "glutaraldehyde residuals," the ground truth would be the acceptable safe limit for residuals.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device.


    Summary of the Study Proving Device Meets Acceptance Criteria:

    The device, "Gel-Bead" embolization spheres (2 ml fill), underwent a series of bench tests to demonstrate substantial equivalence to its predicate device (Gel-Bead, K133237). The studies focused on evaluating the device's functionality and technological characteristics.

    The specific tests conducted were:

    • Sphere suspension in contrast: This likely assesses how well the spheres remain suspended in a contrast agent, which is crucial for their navigability and visualization during clinical use.
    • Pepsin digestion: This test evaluates the biodegradability of the gelatin spheres, confirming they break down as intended over time after embolization.
    • Supplied volume: This verifies that the actual volume of spheres provided in the syringe meets the specified amount (2 ml).
    • Deliverability: This assesses the ability of the spheres to be successfully delivered through compatible catheters without clumping or clogging, which is critical for clinical efficacy.
    • Glutaraldehyde residuals: This measures any residual glutaraldehyde (a cross-linking agent sometimes used in gelatin products), ensuring it is below safe limits.

    The document states that "The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues." This indicates that for each of these measured characteristics, the device performed within the predefined acceptable ranges or thresholds. The exact numerical acceptance criteria are not provided in this document, but the overall conclusion is that they were successfully met.