Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K191052

    Validate with FDA (Live)

    Date Cleared
    2020-01-09

    (265 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K150833

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hyalo GYN® Vaginal Moisturizing Suppositories are personal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. Hyalo GYN® Vaginal Moisturizing Suppositories are not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

    Device Description

    Hyalo GYN® Vaginal Moisturizing Suppositories are non-sterile, glyceride-based personal lubricants. The subject lubricant is intended to moisturize and lubricate the vaginal epithelium to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. The Hyalo GYN® Vaginal Moisturizing Suppositories are provided in peel-to-open pre-cut strips, in an outer cardboard box with a package insert.

    AI/ML Overview

    The provided text describes the regulatory clearance of a medical device, Hyalo GYN® Vaginal Moisturizing Suppositories, and presents a 510(k) summary. This type of document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific clinical performance criteria in the way an AI algorithm might. Therefore, the information requested about acceptance criteria and a study to prove meeting those criteria, in the context of an AI device, is not fully applicable or directly available in this document.

    However, I can extract the closest analogous information regarding the device's specifications and supporting performance data provided in the 510(k) summary. It details technical requirements and testing conducted to ensure safety and effectiveness, which are regulatory "acceptance criteria" for this type of product.

    Here's an interpretation based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    For a medical device like Hyalo GYN® Vaginal Moisturizing Suppositories, the "acceptance criteria" are the defined specifications that the product must meet. The "reported device performance" indicates that the device did meet these specifications during testing.

    ParameterAcceptance Criteria (Specification)Reported Device Performance
    AppearanceWhite to ivory white suppositories in suitable plastic pre-formed shells, tightly sealed & easy to openMet specification (implied by 510(k) clearance and conclusion stating device is safe and effective)
    OdorWax-like smellMet specification (implied by 510(k) clearance and conclusion stating device is safe and effective)
    Antimicrobial EffectivenessMeets requirements per the standard. (USP <51>)Met specification
    Uniformity of dosage unitMeets requirements for Content Uniformity (Ph. Eur. <2.9.40>, current ed.)Met specification
    Hyaff11p50 assay0.18-0.22 % w/wMet specification
    Disintegration timeNo more than (NMT) 60 minutes (Ph. Eur. <2.9.2>, current ed.)Met specification
    Osmolality≤ 1200 mOsm/kg (Ph. Eur. <2.2.35>, current ed.)Met specification
    Consistency0.1-5.0 mmMet specification
    Peroxide value (Ip)≤ 10 mEqO2 / kg (Ph. Eur. <2.5.5>, current ed., method A)Met specification
    Microbiological quality (TAMC)≤100 cfu/g (Ph. Eur. <2.6.12>, current ed., Ph. Eur. <2.6.13>, current ed.)Met specification
    Microbiological quality (TYMC)≤10 cfu/g (Ph. Eur. <2.6.13>, current ed.)Met specification
    Microbiological quality (pathogens)Pseudomonas aeruginosa: Absent; Staphylococcus aureus: Absent; Candida albicans: AbsentMet specification
    BiocompatibilityDemonstrated to be biocompatible per ISO 10993-1, 10993-5, 10993-10, 10993-11. (Specific criteria for each test not detailed in summary but implied)Demonstrated to be biocompatible (Results of testing "demonstrated that the subject lubricants are biocompatible.")
    Condom CompatibilityNot compatible with natural rubber latex, polyisoprene, and polyurethane condoms (this is a stated characteristic, not a performance target to "meet" as much as a determination)Demonstrated that the device is not compatible with natural rubber latex, polyisoprene, and polyurethane condoms based on ASTM D7661-10. This is reflected in labeling
    Shelf-lifeMaintain all specifications for 24 monthsMet all specifications at baseline and following aging for 24-month shelf life.

    2. Sample size used for the test set and the data provenance

    The document does not provide specific sample sizes for each of the performance (stability, biocompatibility, microbiology, etc.) or material compatibility tests. The tests refer to established standards (USP <51>, Ph. Eur., ASTM D7661-10, ISO 10993 standards). The data provenance is laboratory testing performed by the manufacturer, Fidia Farmaceutici S.p.A., to support device specifications and substantial equivalence. It is not patient or clinical data, thus concepts like country of origin for patient data or retrospective/prospective are not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this type of device and submission. This is not an AI/diagnostic device that requires expert ground truth for image or data interpretation. The "ground truth" for this device's performance is established by standardized laboratory test methods and their defined pass/fail criteria.

    4. Adjudication method for the test set

    This question is not applicable. There's no "adjudication" in the sense of reconciling multiple expert opinions for data labeling for an AI algorithm. The tests are scientific and follow established protocols with objective outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The device is a personal lubricant, not a diagnostic imaging or AI-assisted device. Therefore, MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical product, not an algorithm.

    7. The type of ground truth used

    For the laboratory performance tests and material compatibility, the "ground truth" is defined by the validated and established scientific methods and standards themselves (e.g., USP <51> for antimicrobial effectiveness, Ph. Eur. for disintegration time and osmolality, ISO 10993 for biocompatibility, ASTM D7661-10 for condom compatibility). The criteria within these standards define what constitutes an acceptable result.

    8. The sample size for the training set

    This question is not applicable. This is a physical medical device, not an AI or machine learning model that requires a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1