Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171984

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-11

    (103 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K150662

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Clickfine and Penfine Classic Pen Needle are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.

    Device Description

    The subject devices are sterile, non-pyrogenic, single use pen needles designed to be used with commercially available pen-injectors. The back end of the needle punctures the rubber injection port of the drug reservoir in the pen. When the injection is needed, the needle is inserted into the chosen site and the pen-injector delivers the fluid through the needle. The materials of construction, (siliconized stainless-steel needle encased in a polypropylene hub with a polypropylene protective cap) as well as the sterile packaging (a polypropylene outer protective container and the peel tab) which provides the sterile barrier and the manufacturing (which includes radiation sterilization) are unchanged from the currently marketed devices. They subject devices will be marketed in the same sizes as the predicate devices - 32 and 31 gauges and 4, 4.5, 5, 6, 8, 10, and 12 mm lengths.

    AI/ML Overview

    This FDA 510(k) summary focuses on the substantial equivalence of a medical device (pen needle) to a predicate device, rather than the performance of an AI/ML algorithm. Therefore, many of the requested categories related to AI/ML studies are not applicable.

    Here's an analysis based on the provided text, addressing the applicable points:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device ModificationVerification ActivityAcceptance CriteriaResults of Verification
    Larger inner diameter of needle on 31 Gauge and 32 Gauge needlesTesting for resistance to breakage was conducted acc. ISO 9626:2016When tested acc. Annex C of ISO 9626:2016, the tubing shall not show visible breakage when examined by normal or corrected vision after testing.All samples tested met the acceptance criteria for resistance to breakage acc. ISO 9626:2016.
    Larger inner diameter of needle on 31 Gauge and 32 Gauge needlesTesting for stiffness was conducted acc. ISO 9626:2016The tubing shall show a maximum defined deflection when tested in accordance with annex B of ISO 9626:2016: max. deflection 0.22 mm for 32 G and max. 0.21 mm for 31 G.All samples tested met the acceptance criteria for stiffness acc. ISO 9626:2016.

    2. Sample size used for the test set and the data provenance

    The document does not specify a numerical sample size for the tests. It states "All samples tested" for both breakage and stiffness, implying that a sufficient number was tested to meet the requirements of the specified ISO standards, but the exact number isn't provided. Data provenance is not applicable here as it's a physical device performance test, not a data-driven model.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth for physical device performance is established by standardized measurement methods defined by international standards (ISO). No human expert consensus for "ground truth" in this context is mentioned.

    4. Adjudication method for the test set

    Not applicable. The tests are pass/fail based on objective measurements against ISO standards. No adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device.

    7. The type of ground truth used

    The ground truth for the device's performance (breakage and stiffness) is defined by the objective measurement criteria and methodologies outlined in the referenced ISO standards (ISO 9626:2016).

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device; there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device; there is no "training set" or corresponding ground truth establishment process.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1