    Intended Use

    The Solex 7 Intravascular Heat Exchange Catheter connected to the Coolgard/Thermogard Thermal Regulation System is indicated for use:

    · In cardiac surgery patients to achieve and/or maintain normothermia during surgery and recovery/intensive care. (Maximum use period: 4 days)

    · To induce, maintain and reverse mild hypothermia in neurosurgery and recovery/intensive care.

    (Maximum use period: 4 days)

    · In fever reduction, as an adjunct to other antipyretic therapy, in adult patients with cerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated. (Maximum use period: 7 days)

    The Cool Line Catheter Model CL-2295A, when used with the ZOLL Thermal Regulation System, is indicated for use in fever reduction, as an adjunct to antipyretic therapy. in adult patients with cerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.

    The ZOLL ICY Intravascular Heat Exchange Catheter Model IC-3893A, connected to the ZOLL Coolgard/Thermogard Thermal Regulation System, is indicated for use:

    • In cardiac surgery adult patients to achieve and/or maintain normothermia during surgery and recovery/intensive care, and

    • · To induce, maintain and reverse mild hypothermia in neurosurgery adult patients in surgery and recovery/intensive care.
      The ZOLL Quattro Catheter Model IC-4593, connected to a ZOLL Thermal Regulation System, is indicated for use:

    • In cardiac surgery adult patients to achieve and/or maintain normothermia during surgery and recovery/intensive care, and

    • · To induce, maintain and reverse mild hypothermia in neurosurgery adult patients in surgery and recovery/intensive care.
    Device Description

    The ZOLL Intravascular Heat Exchange Catheters (Quattro® Intravascular Heat Exchange Catheter Kit, Quattro® Intravascular Heat Exchange Catheter Premium Access Kit, Cool Line® Intravascular Heat Exchange Catheter, ICY® Intravascular Heat Exchange Catheter) are sterile, single use heparin coated flexible catheters designed for placement in the femoral, jugular, or subclavian veins. The Solex 78 Intravascular Heat Exchange Catheter Premium Access Kit is a sterile, single use 9.3F flexible catheter designed for placement in the superior vena cava from an insection site in the jugular and subclayian veins. The Cool Line catheter contains two heat exchange balloons, the ICY catheter contains three heat exchange balloons, the Quattro catheter contains four heat exchange balloons and the Solex 7® catheter consists of a serpentine balloon. The ZOLL catheters are connected to a single use, disposable Start-Up Kit (SUK), the Coolgard 3000® Console and Thermogard XP® Console. The catheters connect to the console coolant well via tubing integral to the SUK. The catheter is connected to the SUK by connecting the male outflow Luer of the SUK to the female inflow Luer of the catheter and the female inflow Luer of the SUK to the male outflow Luer of the catheter. Both SUK Luers, in turn, are connected via tubing to a heat exchange coil through which saline circulates. The coil is placed in a coolant well located in the console. The controlled temperature saline is circulated through the closed-loop circuit of the SUK and catheter using the console pump, after which the saline is then returned within the SUK to the console heater and chiller coolant well via the catheter's outflow lumen. The catheters, SUK, the Coolgard 3000® Console and the Thermogard XP® Console are supplied separately. The ZOLL Intravascular Heat Exchange System is also designed for use with an off-the-shelf temperature probe, which is supplied separately and not manufactured by ZOLL.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (intravascular heat exchange catheters and associated consoles). It describes the device, its intended use, a comparison to predicate devices, and nonclinical testing that was performed. However, it explicitly states that clinical evaluations were NOT performed for this change (see section V. SUMMARY OF CLINICAL TESTS PERFORMED on page 18).

    Therefore, it is not possible to extract the requested information regarding acceptance criteria and studies that prove the device meets these criteria, as the submission focuses on demonstrating substantial equivalence through nonclinical testing and comparison to predicate devices, rather than a de novo clinical study with specific performance acceptance criteria.

    The clinical trial information presented in the document (the table with CI, ICH, PTBI, SAH patient cohorts, and mortality percentages) is not a study performed to prove the acceptance criteria of this specific device, but rather a reference to a randomized controlled trial that influenced the FDA's decision to add a "Warning - Fever Reduction" limitation to the device's labeling. This trial evaluated the safety and effectiveness of the Cool Line Intravascular Heat Exchange Catheter generally, and its results (increased mortality in certain cohorts compared to standard of care) led to the specific warning about its use for fever reduction in patients with subarachnoid hemorrhage or primary traumatic brain injury. It does not represent a study designed to meet pre-defined acceptance criteria for the device in this 510(k) submission.

    Given this, I cannot fill the table or answer the specific questions about "proving the device meets acceptance criteria" as no such study is presented in the provided document for this 510(k) submission. The submission relies on substantial equivalence to predicate devices and nonclinical testing for "MR Conditional" labeling.

    If you had a document describing a clinical study undertaken specifically to demonstrate the performance characteristics of this device against defined acceptance criteria, I would be able to provide the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153226

    Validate with FDA (Live)

    Date Cleared
    2016-03-24

    (139 days)

    Product Code
    Regulation Number
    870.5900
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K150046

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Solex 7™ Intravascular Heat Exchange Catheter connected to the CoolGard 3000®/Thermal Regulation System is indicated for use:

    · In cardiac surgery patients to achieve and or maintain normothermia during surgery and recovery/intensive care. (Maximum use period = 4 days)

    · To induce, maintain and reverse mild hypothermia in neurosurgery and recovery/intensive care. (Maximum use period = 4 days)

    · In fever reduction, as an adjunct to other antipyretic therapy, in adult patients with cerebral infarcerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated. (Maximum use period = 7 days)

    Device Description

    The Solex 7 Intravascular Heat Exchange Catheter (Solex 7 Catheter) is a sterile, single use 9.3F flexible catheter designed for placement in the Superior Vena Cava from an insertion site in the Jugular and Subclavian Veins. The Solex 7 Catheter is connected to a single use, disposable CoolGard 3000 or Thermogard XP® Start-Up Kit (SUK) and the CoolGard 30000 or Thermogard XP® Console, all of which comprise the ZOLL Intravascular Heat Exchange The Start-Up Kit (SUK) and the CoolGard 3000 or Thermogard XP Console are System. supplied separately. The ZOLL Heat Exchange System is also designed for use with an off-theshelf temperature probe. The Solex 7 Catheter is comprised of a polyurethane shaft and a serpentine shaped PET balloon at the distal end. The blood contact surfaces of the catheter incorporate a hydrophilic heparin coating.

    The catheter has five lumens, two of which when connected to the Start-Up Kit, are used to circulate sterile saline in a closed loop circuit for heat exchange with the blood in the central venous system. Warmed or chilled saline is pumped through the heat exchange lumens, inflating the diameter of the serpentine balloon that interfaces with the patient's blood to warm or cool circulating blood. The inflow/outflow lumens form a closed-loop system through which the warmed or chilled saline circulates. The warmed or chilled saline is not infused into the patient.

    Additional lumens of the Solex 7 Catheter consist of a 0.032" guidewire compatible lumen that can also be used as a primary infusion lumen, and two additional infusion lumens within the catheter shaft.

    AI/ML Overview

    The document provided describes the Solex 7™ Intravascular Heat Exchange Catheter and Start-Up Kit and its substantial equivalence to predicate devices, based on nonclinical testing. This is a 510(k) premarket notification, which demonstrates a device is at least as safe and effective as a legally marketed predicate device, rather than proving effectiveness through clinical trials required for a PMA.

    Therefore, the study described here is primarily a set of nonclinical tests to show that the modified device performs similarly to its predicate devices and meets its own design specifications.

    Here's an analysis of the acceptance criteria and the "study" (nonclinical testing) that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document provides multiple tables (Table 3, Table 4, Table 5, Table 6, Table 7) detailing various tests, their methods, and results. These tables effectively serve as the acceptance criteria and reported device performance. I will consolidate key examples below.

    Test CategoryAcceptance Criteria (Test Method Summary)Reported Device Performance (Results)
    Solex 7 Catheter Bench Performance
    Visual InspectionEnsure catheter is smooth without major pits.Catheter samples met the acceptance criterion for smooth appearance with no pits.
    Catheter Indwell Life (7 days)Verify no saline leakage during simulated use for seven days.Catheter samples met the acceptance criterion for normal function for seven days without leakage.
    Particulate TestingDetermine catheter coating integrity following simulated indwell and flow rate conditions.Solex 7 samples and Cool Line samples tested demonstrated levels of particulate below that required in USP <788>.
    Heat Exchange TestingDetermine heat exchange capability of catheter.Catheter samples met the acceptance criterion for a heat exchange power of 125 Watts minimum.
    In Vitro Thrombogenicity TestingMeasurement of thrombogenic behavior of catheter following 7 day simulated use by radioactive platelet quantification and visual inspection.Coated catheter samples showed statistically significant decreases (p < 0.05) in thrombogenicity when compared to normalized thrombogenic behavior of uncoated catheters.
    Start-Up Kit (SUK) Packaging Testing
    SUK Package Seal Peel TestEnsure seal peel data is not less than specified acceptance criteria.SUK package seals met the acceptance criterion for peel strength.
    Start-Up Kit Performance Testing
    SUK Flow TestVerify that air flows through the SUK system.SUK systems met the acceptance criterion for air flow.
    Modified SUK Indwell Life TestVerify that Custom Luers and SUK Tubing are able to function normally at specified temperatures for the labeled indwell period.Custom Luers and SUK tubing met the acceptance criterion for normal function for 7 days at the specified temperature.
    Custom Luer Catheter Bench Performance
    Guidewire PassageVerify ability to frontload, backload, and remove J-tip 0.032" guidewire from catheter.Catheters met the acceptance criterion for the J-tip 0.032" guidewire to be frontloaded, backloaded, and removed from the catheters.
    Balloon Deflation Using Deflation Syringe (20 mL Slip-Fit)Verify that balloons collapse upon aspiration at specified temperature using supplied slip-fit syringe.Catheter balloons met the acceptance criterion of collapse upon aspiration using the supplied slip fit syringe.
    Luer Standards Performance Testing
    Water/Liquid LeakageTested using test method similar to test specified in ISO594-1&2 (§4.2.1).All samples met the requirements of the test method used.
    Stress CrackingTested using test method similar to that specified in ISO594-2 (§4.7).All samples met the requirements of the test method used.
    Biocompatibility Testing
    CytotoxicityMEM Elution Test.The reactivity grade of all test article extract samples was “0”.
    Hemocompatibility (ASTM Hemolysis)ASTM Hemolysis (Extract Method) Test.The difference between the hemolytic indexes of the test article and the negative control equals 0.00 percent. This places the test article in the non-hemolytic range.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document generally refers to "samples" or "catheter samples" without specifying precise numerical sample sizes for most bench tests. However, for "Particulate Testing," it mentions "Solex 7 samples and Cool Line samples."

    • Sample Size: Not explicitly stated numerically for most tests. It refers to "catheter samples," "SUK systems," "all samples," etc.
    • Data Provenance: This is a nonclinical study conducted by ZOLL Circulation, Inc. (submitter). The tests are laboratory-based and simulated, rather than involving human subjects or real-world clinical data. Therefore, concepts like country of origin for data or retrospective/prospective don't directly apply in the same way they would for a clinical study. The tests were performed to verify performance against international standards (ISO, EN ISO, ASTM, USP).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable (N/A). The "ground truth" for these tests is defined by established engineering and medical device standards (e.g., ISO, ASTM, USP criteria for tensile strength, leakage, biocompatibility, etc.) and the device's own design specifications. Expert consensus in the traditional sense (like for medical image interpretation) is not directly involved in setting the pass/fail criteria for these bench tests. The human factors testing involved "users" but not experts establishing ground truth for objective performance metrics in the same way.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This section is N/A. Adjudication methods like 2+1 or 3+1 are used in clinical studies where subjective expert review is required (e.g., for ambiguous medical findings). For these nonclinical bench tests, results are typically objective measurements or observations against predefined pass/fail criteria outlined in the test methods.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is N/A. The document describes a medical device (intravascular heat exchange catheter and start-up kit), not an AI diagnostic or assistance system. Therefore, an MRMC comparative effectiveness study or assessment of human reader improvement with AI assistance is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This section is N/A. The device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on a combination of:

    • International Standards: ISO 10555-1, ISO 594-1, ISO 594-2, EN ISO 10555-1, USP <788>, ASTM Hemolysis standards.
    • Device Design Specifications: Specific dimensional tolerances, flow rates, heat exchange power, tensile strength, burst/leak resistance, indwell life, and biocompatibility criteria defined by the manufacturer for the device.
    • Human Factors Test Scenarios: User tasks designed to assess the ease and correctness of device usage as per the Instructions for Use (IFU).

    8. The sample size for the training set

    This section is N/A. This is not a machine learning or AI device, so there is no "training set" in the context of algorithm development.

    9. How the ground truth for the training set was established

    This section is N/A for the same reason as above.