510(k) Data Aggregation

    K Number
    K212790

    Validate with FDA (Live)

    Device Name
    XLase Plus
    Manufacturer
    Date Cleared
    2022-08-15

    (348 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K193367, K053628, K143723, K201111

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    XLase Plus is indicated for use as follows:

    Diode 808/760 nm 4000W ALEX PRO Handpiece: Indicated for the treatment of hair removal with static and dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
    · Treatment of Pseudofolliculitis barbae (PFB)
    · Use on all skin types (Fitzpatrick I-VI).

    SLP ND:YAG 1064 nm Handpiece: Indicated for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime. (Skin Types Fitzpatrick I-VI), photocoagulation and hemostasis of benign pigmented and benign vascular lesions, treatment of benign pigmented lesions, such as but not limited to warts, telangiectasia, leg veins and spider veins.

    Q Switch ND:YAG 1064 nm Handpiece: Indicated for removal of dark tattoos and treatment of benign pigmented lesions.

    CPL Handpiece: Indicated for use Fitzpatrick skin types I – IV, as shown in the table below. Indicated for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime. Photocoagulation of benign vascular lesions, photo thermolysis of blood vessels, treatment of benign pigmented lesions.

    ERBIUM:YAG FRACTIONAL 2940 nm Handpiece: Indicated for procedure requiring resurfacing of soft tissue with fractionated handpiece.

    Diode 2800W 810nm Handpiece: Indicated for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin. Indicated for:

    • the treatment of benign vascular and benign pigmented lesions,
    • permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Diode 1200W 810nm Handpiece:
    Indicated for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
    Indicated for the treatment of benign vascular lesions, including angiomas, hemangiomas, telangiectasia and other benign vascular lesions, and the treatment for pseudo folliculitis barbae.
    Indicated for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
    Indicated for the treatment of benign pigmented lesions and leg veins.

    ALEX 755nm Handpiece: Indicated for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime. Treatment of benign pigmented lesions. Treatment of wrinkles and the photocoagulation of dermatological benign vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). On all skin types (Fitzpatrick I- VI) including tanned skin.

    Device Description

    XLase Plus, laser for medical applications. It features two models: XLase Plus Standing (with wheels) and XLase Plus Table (tabletop).

    The two medical devices are identical, contain the same electronic boards and the same critical components, use the same software and are able to perform the same aesthetic treatments. The generic components of XLase Plus Table are a subset of those of XLase Plus Standing. The difference between the two devices is that XLase Plus Standing is built on wheels while XLase Plus Table has no wheels but can be easily placed on a table due to its low weight. The device is for prescription use only.

    The XLase Plus Standing and XLase Plus Table medical devices consists of:

    • . System Console
    • Operator control panel .
    • . Touch screen monitor
    • Power cable
    • Accessories and components .
    AI/ML Overview

    Based on the provided text, the device in question is the "XLase Plus," a medical laser and pulsed light platform. The document is a 510(k) premarket notification summary submitted to the FDA, indicating that it aims to demonstrate substantial equivalence to legally marketed predicate devices, rather than proving novel clinical effectiveness through a new clinical study with specific acceptance criteria related to disease diagnosis or treatment efficacy per se.

    Therefore, the acceptance criteria and supporting "study" (in the context of this 510(k) submission) are focused on non-clinical performance and substantial equivalence to existing devices, rather than on the direct clinical performance of an AI algorithm in a diagnostic or treatment outcome context.

    Here's an breakdown based on the provided document:

    Device: XLase Plus (Medical Laser and Pulsed Light Platform)

    Overall Goal of the Submission: To demonstrate Substantial Equivalence (SE) to legally marketed predicate devices. This means proving that the new device is as safe and effective as a legally marketed device and does not raise new questions of safety and effectiveness.

    Acceptance Criteria and Reported Device Performance (Non-Clinical Focus):

    Since this is a 510(k) for a laser/pulsed light platform, the "acceptance criteria" are not related to diagnostic accuracy (like sensitivity/specificity for an AI) but rather to the device's technical specifications and safety/performance standards compared to predicates.

    Acceptance Criteria CategorySpecific Criteria (Met by Device)Reported Device Performance / Means of Proof
    BiocompatibilityEvaluation of potential biological risks from patient contact materials.Performed Tests: Skin Sensitization (ISO 10993-10:2010), Irritation (ISO 10993-23:2021), Cytotoxicity (ISO 10993-5:2009). The document implies these tests were passed, demonstrating acceptable biocompatibility.
    Electrical SafetyCompliance with recognized electrical safety standards for medical devices.Performed Test: IEC 60601-1. The device and accessories are reported to comply with this standard.
    Electromagnetic Compatibility (EMC)Compliance with recognized EMC standards for medical devices to ensure proper functioning in an electromagnetic environment without causing unacceptable electromagnetic disturbance to other devices.Performed Test: IEC 60601-1-2. The device and accessories are reported to comply with this standard.
    Software Verification & ValidationSoftware meets its specified requirements and performs its intended functions correctly and reliably, commensurate with its "level of concern."Testing Conducted: Verification and validation testing on the software interface and firmware. Documentation provided as recommended by FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software/firmware was considered "moderate" level of concern. The implication is that V&V passed, demonstrating the software is fit for purpose.
    Performance Testing (Bench)The device emits set energy parameters within specifications as intended, across various handpieces and their respective modes (hair removal, vascular lesions, pigmented lesions, soft tissue resurfacing).Performed Tests: Verified according to Biotec Italia SRL procedures for product design and development. The document states that testing ensures the device "emits set energy parameters within specifications, as intended." This includes parameters like: - For various handpieces (Diode, SLP ND:YAG, Q Switch ND:YAG, CPL, ERBIUM:YAG, ALEX 755nm): Laser Wavelength, Fluence (J/cm2), Pulse Duration (ms/ns), Pulse Repetition Rate (Hz), Handpiece Spot Size (mm), Cooling Temperature (specific values provided for some). These parameters are compared against those of predicate/reference devices to establish substantial equivalence.
    Pulsed Light Wavelength RangesConformance to specified operational wavelength ranges for different indications and skin types (Fitzpatrick I-IV).The document provides two tables (Table 1: Predicate Device Wavelength Ranges; Table 2: CPL – Subject Device Wavelength Range) showing comparable wavelength ranges for hair reduction, benign vascular lesions, blood vessels, and benign pigmented lesions across different skin types. This comparative data implicitly shows the XLase Plus (Subject Device) operates within ranges acceptable for its intended uses, similar to the predicate.

    Since the submission is a 510(k) for a physical device (laser/pulsed light platform) and not an AI/ML medical device, many of the typical categories for AI software are not applicable. Here's how the other points are addressed:

    1. Sample sizes used for the test set and the data provenance:
      This is not applicable in the context of this 510(k) for a physical device, as it did not involve a clinical "test set" of patient data for an AI algorithm. The "test set" here refers to the actual physical device units undergoing non-clinical testing. The document doesn't specify the number of devices tested but implies standard product testing.
      Data provenance (country of origin, retrospective/prospective) is also not applicable for this type of non-clinical device testing.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      Not applicable. Ground truth, in the AI context, refers to expert annotations or pathological confirmation. For a laser device, "ground truth" is established by manufacturing tolerances and measurement standards/equipment to ensure the device performs according to its specifications (e.g., emits the correct energy, pulse duration). This is a technical validation process, not one involving expert human interpreters for diagnoses.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
      Not applicable, as there is no "test set" of patient cases requiring adjudication of interpretations.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This device is a treatment device, not a diagnostic AI assistance tool for human readers.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable. While the device has software, it's an embedded system controlling a hardware device. Its "performance" is assessed by its ability to deliver specified energy parameters, not by its ability to independently diagnose or interpret clinical images. The "standalone" performance is its electrical, mechanical, and software functionality in producing the desired laser/light output.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For the non-clinical testing of the device's performance, the "ground truth" would be engineering specifications, verified by calibrated measurement equipment and adherence to international standards (e.g., ISO, IEC). For example, a power meter would provide the "ground truth" for laser energy output, or an oscilloscope for pulse duration.

    7. The sample size for the training set:
      Not applicable. There is no AI model being trained on patient data for this device. The software is a control system for the hardware.

    8. How the ground truth for the training set was established:
      Not applicable, as there is no training set for an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182045

    Validate with FDA (Live)

    Date Cleared
    2018-12-27

    (150 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K093162, K143723, K172639

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LOTUS III Multi-Pulsed Er: YAG Laser System is intended for coagulation, vaporization, ablation or cutting of soft tissue (skin) in dermatology, plastic surgery (including aesthetic surgery), oral surgery, and ophthalmology (skin around the eyes).

    Device Description

    Laseroptek Co. Ltd.'s LOTUS III Pulsed Er:YAG Laser System is an Erbrium:YAG laser with a wavelength of 2940 nm. A set of Er: YAG lasers is controlled by an embedded processor, to be used in dermatology. The laser system uses focusing optics to deliver a pattern of thermal energy to the epidermis and dermis. This system consists of main body, color touch screen, articulated arm, hand piece and foot switch.

    AI/ML Overview

    Based on the provided text, the device in question is the "LOTUS III Multi-Pulsed Er:YAG Laser System," which is a medical laser system. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical study with detailed performance metrics against specific acceptance criteria for a new AI/imaging device.

    Therefore, the requested information regarding acceptance criteria and a study proving the device meets them (especially in the context of AI performance metrics like sensitivity, specificity, human reader improvement, ground truth establishment, training sets, etc.) is not present in this type of regulatory submission for this particular device.

    This document is for a traditional medical device (a laser system) and primarily relies on non-clinical performance testing and comparison of technical characteristics to predicate devices to establish substantial equivalence. It does not involve a study of the kind typically performed for AI-powered diagnostic devices that would have the specific performance metrics and study design details requested in your prompt.

    Here's why the requested information cannot be found in the provided text:

    • Device Type: This is a laser system (Er:YAG laser) for general and plastic surgery and dermatology. It is a physical medical device, not an AI/software device that analyzes images or data.
    • Regulatory Pathway: This is a 510(k) submission, which focuses on demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving novel effectiveness or performance against new, specific clinical acceptance criteria through a full-scale clinical trial like a PMA, or a clinical validation study for an AI algorithm.
    • Performance Data (Section 6): The "Performance Data" section explicitly states that "Non-clinical tests" were conducted, referring to compliance with mandatory performance standards for laser products (e.g., 21 CFR 1040.10 and 1040.11), electromagnetic compliance (IEC 60601-1-2), usability (IEC 60601-1-6, IEC 62366), safety of laser products (IEC 60825-1, IEC 60601-2-22), risk management (ISO 14971), and biocompatibility (ISO 10993). These are engineering and safety standards, not clinical performance metrics for a diagnostic AI.
    • Absence of Clinical Study Details: There is no mention of:
      • Sensitivity, specificity, AUC, or other diagnostic performance metrics.
      • Test sets, training sets, or data provenance (countries, retrospective/prospective).
      • Experts establishing ground truth, adjudication methods, or MRMC studies.
      • Standalone algorithm performance or human-in-the-loop improvement metrics.

    In summary, as this document pertains to a traditional laser medical device obtaining 510(k) clearance, the requested details for AI/imaging device performance criteria and studies are not applicable and thus not present in the provided text.