510(k) Data Aggregation

    K Number
    K251148

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-25

    (164 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K143142, K182081, K233896, K050406

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JDEvolution Plus L and JDEvolution Plus LE implant system is intended for surgical placement in the upper jaw.

    JDEvolution Plus L and JDEvolution Plus LE implant system provides a means for prosthetic attachment in partially or fully edentulous spans with multiple teeth utilizing delayed, or immediate loading, or as a terminal or intermediary abutment for fixed ore removable bridgework or to retain overdentures. JDEvolution Plus L and JDEvolution Plus LE is intended for immediate function multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function.

    JDEvolution Plus L and JDEvolution Plus LE implants with lengths of 18, 20, 22, 24, 26 mm are only indicated for multiple unit restorations in splinted applications with at least two implants when placed in the maxilla.

    Device Description

    JDEvolution Plus L and LE, are tissue level implantable devices produced in commercially pure titanium, intended for surgical placement in the upper jaw to support prosthetic devices, such as artificial teeth, in order to restore chewing function.

    JDEvolution Plus L and LE are endosseous dental implants intended for surgical placement in the upper jaw. JDEvolution Plus L and LE implants includes dental implant fixtures only.

    JDEvolution Plus L and LE provide a means for prosthetic attachment partially or fully edentulous spans with multiple teeth utilizing delayed or immediate loading when good primary stability is achieved in order to restore patient chewing function.

    The subject implant bodies are compatible with abutments cleared in K143142 and K233896

    Knowledgeable clinicians are in charge for the following of a delayed instead immediate loading, evaluating the stability of the implant sites after the surgical operation.

    The device insertion occurs at the end of the surgical procedure which consists in an osteotomy of the upper jaw.

    JDEvolution Plus L and LE implants are single use devices, supplied sterile and that cannot be re-cleaned or re-sterilised.

    JDEvolution Plus L and LE Implant System is a two-piece implant made of commercially pure titanium.

    The body of the implant fixture (Endosseous Dental Implant) is surgically placed in the upper or lower jaw, while the Abutment (several type for several clinical application) is screwed into the fixture to support the prosthesis.

    The connection is done through an internal hexagon: abutments and other accessories are exclusively designed for JDEvolution Plus L and LE Implant System.

    Endosseous dental implant and abutment will represent the support, for the dentist, for building the artificial replacement tooth.

    The artificial replacement tooth component is not part of this submission.

    The main features of JDEvolution Plus L and LE Implant System are summarized in the table below:

    JDEvolution Plus:
    Materials: Titanium grade 4 (diameter 4.0)

    Design:
    General Features: Endosseous implant with connection with internal hexagon
    Shape:
    Collar: Machined gingival collar of 1,5mm length for all the lengths
    Body: Double thread with 0.6 mm lead: straight cylindrical of 4,0 mm diameter for a total length (including collar) of 20 / 22 / 24 / 26mm. Tapered apex with bone cutting flutes.
    Double thread with 1.2 mm lead: tapered body of 4,0 mm for a total length (including collar) of 18 / 20 / 22 / 24 / 26mm
    Thread: Double thread with 0.6 mm or 1.2 mm lead
    Abutment: To be connected with the Conical abutment cleared under K143142, K233896
    Ti gr.5: Straight– angled 15° - angled 17° - angled 25°- angled 30°

    Surface:
    Collar: Machined, without any surface treatment
    Body: Treated through sandblasting followed by acid etching (SLA) for the full length of the endosseous portion for all variants (the surface treatment is the same of the JDentalCare Implant System JDIcon cleared under K 182081)

    Thread:
    Body: Double thread
    outline: trapezoidal
    Thread lead: 0.6 mm (for tapered body version)
    Thread lead: 1.2 mm (for straight cylindrical body version)
    Self-Threading Capacity: Self-threading capacity in both direction

    Dimensions:
    Mean Diameter: 4.0 mm
    Total Length: 18mm (only for JD Evolution Plus LE Implant), 20mm, 22mm, 24mm and 26mm
    Implanted Length: 16.5mm (only for JD Evolution Plus LE Implant), 18.5mm, 20.5mm, 22.5mm and 24.5mm

    Connection:
    Shape: Internal hexagonal connection (Hexagon dimension of 2.425 mm)

    Packaging:
    Type of package: Blister with internal vial and protective cap for implants
    External carton box as commercial packaging
    Blister material: Tyvek
    Polyester PET MO65
    Vial material: Polystyrene Empera 116 vial closed with Eralyte PET cap

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233896

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-16

    (249 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JDentalCare Dental Implant System, family names JDEvolution Plus, JDIcon, JDIcon Plus, is intended to replace missing masticatory functional units (teeth) within the maxilla or mandible.

    .DentalCare Dental Implant System is comprised of dental implant fixtures and prosthetic devices. It provides a means for prosthetic attachment in single tooth restorations and partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading, or as a terminal or intermediary abutment for fixed or removable bridgework or to retain overdentures. Prosthetic devices provide support and retention for screw-retained or cemented restorations in mandible and maxilla. JDentalCare Dental Implant System is intended for immediate function on single tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. When a one-stage surgical approach is applied, the immediately loaded when good primary stability is achieved and the functional load is appropriate.

    Implants 20 mm when placed in the maxilla are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    JDental Care implant system JDIcon 2.75mm D Dental Implant shall only be used to replace maxillary lateral incisors and mandibular lateral and central incisors for single stage or two stage procedures. It is for immediate implantation in extraction sites or implantation in partially healed or completely healed alveolar ridge situations

    Device Description

    JDentalCare S.r.l. manufactures Dental Implant Systems made by Endosseous Dental Implants, Abutment and related accessories, in order to support dentist in performing Dental Implant Surgery.

    In order to grant Dentists with a wide range of prosthesis solution, the JDentalCare® Dental Implant System are proposed with different trade name within families lines JDEvolution® plus, JDIcon®, JDIcon® plus, characterized mainly by same mechanical and surface treatment solution and different measure of the hexagon in the connection system. All families have the internal hexagon connection.

    JDentalCare® Dental Implant System is a two-piece implant made of commercially pure titanium. The body of the implant fixture (Endosseous Dental Implant) is surgically placed in the upper or lower jaw, while the Abutment (several type for several clinical application) is screwed into the fixture to support the prosthesis.

    The connection is done through an internal hexagon: abutment and other accessories are exclusively designed for JDentalCare® Dental Implant System.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a dental implant system. It details the device's description, indications for use, comparison to predicate devices, and performance data. However, it does not contain information about an AI/ML-enabled medical device.

    Therefore, I cannot extract information related to acceptance criteria and the study proving the device meets those criteria for an AI/ML device, as the document describes a physical dental implant system and not an AI/ML product.

    The sections regarding acceptance criteria, sample sizes, expert involvement, MRMC studies, standalone performance, ground truth, and training sets are all relevant to the evaluation of AI/ML models. Since this document is about a physical medical device (dental implants), these concepts are not applicable in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160536

    Validate with FDA (Live)

    Date Cleared
    2017-01-05

    (314 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K131682, K143353, K143142, K081806, K142260, K063341

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Chaorum Implant System is intended for use in partially or fully edentulous mandibles and maxilla, in support of single of multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. Chaorum Implant System is for single stage and two stage surgical procedures. This system is intended for delayed loading.

    Device Description

    Chaorum Implant System are devices made of titanium intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices such as artificial teeth, and to restore the patients chewing function. Chaorum Implant System consists of fixtures, abutments, and screws. Its material, structure and intended use are substantially equivalent to the predicate devices. It offers three different implants in RBM treatment and SLA treatment.

    AI/ML Overview

    The provided text describes a 510(k) submission for the "Chaorum Implant System," which is an endosseous dental implant system. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving the device meets specific performance acceptance criteria for an AI/ML algorithm.

    Therefore, many of the requested categories about acceptance criteria, study design for AI/ML performance, sample sizes for AI/ML test/training sets, expert ground truth establishment, MRMC studies, or standalone algorithm performance are not applicable to this document. This submission pertains to a physical medical device (dental implants), not a software or AI/ML-driven device.

    However, I can extract information regarding the non-clinical testing performed to validate the safety and substantial equivalence of the physical device.

    Here's the relevant information that can be extracted, acknowledging that it does not fit the typical format for AI/ML algorithm evaluation:

    1. A table of Acceptance Criteria and the Reported Device Performance:

    The document does not explicitly state "acceptance criteria" in a quantitative, pass/fail manner for each test, nor does it report specific performance values against those criteria. Instead, it lists the types of non-clinical tests conducted to validate the device's safety and substantial equivalence. The conclusion states that "Based on the similarities and the test results of the validation activities, we conclude that the Chaorum Dental Implant System is substantially equivalent to the predicate device." This implies that the device met the unstated acceptance criteria for each test to support the substantial equivalence claim.

    Therefore, for AI/ML-specific criteria, this section is N/A as this is not an AI/ML device.

    For the physical device, a table summarizing the types of non-clinical tests performed is provided in the document:

    NoTest TitleTest StandardImplied Acceptance
    1Physical tests (visual, packaging, packaging seal efficacy, dimension, sterility - direct transfer method)Not explicitly listed for each, but general quality control and sterility standards apply.The device's physical attributes, packaging integrity, and sterility were confirmed to meet established standards. For sterility, the direct transfer method was used, implying successful sterilization according to protocol. Dimensions and visual inspection would have ensured conformance to design specifications.
    2RBM/SLA Surface treatment tests (roughness average, developed surface area ratio, surface characteristics (SEM), surface composition analysis (EDX))Not explicitly listed for each, but general material and surface characterization standards apply.The RBM and SLA surface treatments were characterized and confirmed to be consistent with the intended design and expected properties for dental implants, ensuring appropriate surface characteristics for osseointegration.
    3TiN coating tests (surface roughness and abrasion testing)Not explicitly listed, but material and coating standards apply.The TiN coating was evaluated for its roughness and abrasion resistance, confirming its suitability for the device's function and durability.
    4Mechanical properties test (adaptation accuracy (Implant to abutment compatibility), 35° compressive loads, torsional breaking force, removal torque force, fatigue test)Not explicitly listed, but industry standards for dental implant mechanical testing apply.The mechanical integrity and performance of the implant system (fixtures, abutments, screws) were confirmed through various tests, demonstrating appropriate strength, compatibility, and durability under simulated physiological loads to ensure device functionality and patient safety. This would include ensuring the device can withstand chewing forces without failure and maintain a stable connection between components.
    5Sterilization validation and shelf-life testsNot explicitly listed, but ISO 11137 series and ISO 11607 series are common standards.Sterilization processes (gamma irradiation) were validated to ensure the device is sterile at the point of use, and shelf-life testing confirmed that the sterility and integrity of the device are maintained over its intended storage period.
    6Biocompatibility tests (Cytotoxicity, Acute Systemic Toxicity, Intracutaneous Reactivity, Pyrogen Test, Local Lymph Node Assay (LLNA), Bone Implantation Test)ISO10993 series (specific parts listed in the document)The device materials and design were proven to be biocompatible, meaning they do not cause adverse biological reactions (e.g., cell toxicity, irritation, sensitization, systemic toxicity, fever, or undesirable tissue reactions) when in contact with human tissue or implanted in bone. The bone implantation test would demonstrate appropriate osteointegration or tissue response in a living model.

    2. Sample size used for the test set and the data provenance:

    N/A. This does not refer to an AI/ML algorithm's test set. The non-clinical tests described involve physical devices and material samples. The document does not specify the number of units or samples used for each physical or biocompatibility test, nor the provenance in the context of "data" (e.g., country of origin, retrospective/prospective). These are laboratory tests on manufactured articles.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    N/A. This is not an AI/ML device that requires human experts to establish ground truth for image interpretation or diagnosis. The "ground truth" for the non-clinical tests would be the accepted standards and specifications for material science, mechanical engineering, and biocompatibility, as evaluated by engineers and scientists in accredited labs.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    N/A. Adjudication methods like 2+1 or 3+1 are used for resolving disagreements among human readers in a diagnostic or interpretive task (often in AI/ML performance evaluation). This is not relevant to the non-clinical physical and biological testing of a dental implant.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    N/A. MRMC studies are used to evaluate diagnostic systems, often with AI components assisting human readers. This document describes a physical medical device, not a diagnostic system with an AI component.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    N/A. This is not an AI/ML algorithm; it is a physical dental implant system.

    7. The type of ground truth used:

    For the non-clinical tests, the "ground truth" is defined by:

    • Engineering specifications and design requirements: For dimensional accuracy, mechanical properties, and compatibility.
    • Material standards: ASTM F136, ASTM F67 for titanium alloys.
    • International standards: ISO10993 series for biocompatibility.
    • Sterilization validation protocols: For sterility and shelf-life.

    There is no "expert consensus" or "pathology" in the sense of clinical diagnostic ground truth for these types of tests, other than expert evaluation within the relevant engineering and scientific fields to ensure compliance with standards.

    8. The sample size for the training set:

    N/A. This is not an AI/ML algorithm. There is no concept of a "training set" for the physical device.

    9. How the ground truth for the training set was established:

    N/A. As above, no training set is involved.