Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K201057

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-14

    (633 days)

    Product Code
    Regulation Number
    870.4100
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K142673,K143083,K180453

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bio-Medicus Life Support catheters and introducers are single-lumen drainage or reinfusion peripheral-access catheters to be used in ECMO or ECLS with an extracorporeal circuit intended for use in adult and pediatric patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.

    Device Description

    The Bio-Medicus Life Support™ catheter is a single-lumen catheter used to drain or reinfuse blood. The introducer facilitates proper insertion and placement of the appropriately sized catheter over a guidewire within the vessel. These devices are intended to perfuse vessels or organs in conjunction with extracorporeal support, including Extracorporeal Membrane Oxygenation (ECMO) and Extracorporeal Life Support (ECLS). Catheter models with tip lengths of 18 cm (7.09 in), 50 cm (19.7 in), or 55 cm (21.7 in), with and without additional side holes may be used as either drainage or reinfusion catheters.

    These devices are sterile, nonpyrogenic, disposable, intended for single use only. Bench studies were performed after device preconditioning including exposure (21 days) to simulated in vivo use conditions to demonstrate safety and reliability. Do not store the product above 25°C (77°F).

    AI/ML Overview

    The provided document is a 510(k) summary for the Medtronic Bio-Medicus Life Support Catheter and Introducer. It describes a medical device, specifically a catheter and introducer for use in ECMO/ECLS, and assesses its substantial equivalence to a predicate device.

    Crucially, this document is for a medical device (catheter and introducer), not an AI/ML device or a software device. Therefore, the standard acceptance criteria and study designs typically associated with AI/ML devices (e.g., standalone performance, MRMC studies, expert ground truth for image analysis, etc.) are not applicable to this submission.

    The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance tests and regulatory requirements for a physical medical device. They are focused on mechanical properties, biocompatibility, sterility, shelf-life, and in vivo performance in an animal model, rather than AI/ML model performance metrics like accuracy, sensitivity, or specificity.

    Therefore, I cannot populate the requested table and answer the questions in the context of an AI/ML device. The document does not contain information on:

    • A table of acceptance criteria and reported device performance related to AI/ML metrics.
    • Sample size for a "test set" or "training set" in the AI/ML sense.
    • Number of experts for ground truth or adjudication methods for an AI/ML model.
    • MRMC studies or effect sizes for human readers with/without AI assistance.
    • Standalone algorithm performance.
    • Type of ground truth (expert consensus, pathology, outcomes data) for an AI/ML context.
    • How ground truth for a training set was established for an AI/ML model.

    What the document does provide relevant to device acceptance is:

    • Acceptance Criteria (Special Controls for the device type): These are outlined under "Comparison to Predicate (Special Controls)" on page 5 and include:
      • Technological Characteristics (geometry, design, compatibility)
      • Biocompatibility (ISO 10993-1:2009 compliance)
      • Sterility and Shelf-life (demonstrates sterility, integrity, durability, reliability)
      • Non-clinical Performance (bench, mechanical integrity, durability, reliability testing)
      • In vivo Evaluation (performance over 7-day duration of use)
      • Labeling (detailed summary of non-clinical and in vivo evaluations, instructions, compatibility, maintenance)
    • Study Proving Device Meets Criteria (Summary of Performance Data): This is detailed on pages 5-6 under "Summary of Performance Data" and includes:
      • Pre-clinical bench testing: Used to verify performance characteristics.
        • 21-day Simulated use durability testing (Tensile strength after life conditioning, Pressure test after life conditioning)
        • Blood trauma testing
        • Kink testing
        • Birds Mouth testing
        • ID occlusion due to suture ring
        • Suture Collar Slide Force
        • Introducer removal and insertion forces
      • In vivo animal testing: 7-day GLP in vivo study.

    In conclusion, the requested information pertains to AI/ML device validation, which is not the subject of this 510(k) submission. This document describes the testing and regulatory compliance for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201100

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-15

    (447 days)

    Product Code
    Regulation Number
    870.4210
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K142673,K143083,K180453

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Bio-Medicus Life Support catheters are peripheral-access catheters used to perfuse vessels or organs in conjunction with extracorporeal cardiopulmonary bypass (CPB) procedures. The catheter introducer is intended to facilitate proper insertion and placement of the appropriately-sized catheter within the vessel. Catheter models with tip lengths of 18 cm (7.09 in), 50 cm (19.7 in), or 55 cm (21.7 in) and without additional side holes may be used as either drainage or reinfusion catheters. This product is intended for use in adult and pediatric patients for up to 6 hours.

    Device Description

    The Bio-Medicus Life Support catheter is a single-lumen catheter used to drain or reinfuse blood. These devices are sterile, nonpyrogenic, disposable, and intended for single use only. Do not store the product above 25°C (77°F).

    AI/ML Overview

    This document is a 510(k) summary for the Bio-Medicus Life Support Catheter and Introducer. It is a premarket notification for a medical device seeking clearance from the U.S. FDA, not a study report that details acceptance criteria and device performance in the way a clinical trial or a deep learning algorithm validation would.

    Therefore, the requested information elements related to AI models, ground truth establishment, expert adjudication, sample sizes for training/test sets, MRMC studies, or standalone algorithm performance are not applicable (N/A) to this device or document. This document pertains to a physical medical device (catheter and introducer) and its pre-clinical testing, not an AI/ML algorithm.

    However, I can extract the information that is available about the "device performance" in the context of this 510(k) submission, interpreted as the pre-clinical bench and in-vivo animal testing conducted.

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document summarises that "Pre-clinical bench testing was used to verify the performance characteristics of this device. In vivo animal testing was also completed to establish substantial equivalence with the predicate device." It also lists the types of performance tests conducted.

    However, the document does not provide specific acceptance criteria values or detailed quantitative results for each test. It only states that the data is "sufficient to demonstrate that the Medtronic Bio-Medicus Life Support Catheter and Introducer is substantially equivalent to the legally marketed predicate devices."

    Test CategoryAcceptance Criteria (Not explicitly stated in detail)Reported Device Performance (Summary)
    Catheter testingTo demonstrate performance and safety profileData considered sufficient for substantial equivalence
    Catheter and Introducer TestingTo demonstrate performance and safety profileData considered sufficient for substantial equivalence
    Introducer TestingTo demonstrate performance and safety profileData considered sufficient for substantial equivalence
    Blood trauma testingTo ensure minimal blood damage during useData considered sufficient for substantial equivalence
    Packaging TestingTo ensure sterility and product integrityData considered sufficient for substantial equivalence
    Sterilization TestingTo confirm effective sterilizationData considered sufficient for substantial equivalence
    Catheter and Introducer Life TestingTo confirm durability and functional lifeData considered sufficient for substantial equivalence
    Biocompatibility TestingTo ensure material compatibility with biological systemsData considered sufficient for substantial equivalence
    Pressure Drop TestingTo evaluate flow characteristics through the deviceData considered sufficient for substantial equivalence
    In-vivo animal testingTo establish substantial equivalence with predicatesDemonstrated substantial equivalence with predicate devices

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not specified in this summary. The tests listed are "pre-clinical bench testing" and "in vivo animal testing," which implies physical testing samples rather than human clinical data sets. The exact number of catheters/introducers tested or animals used is not detailed.
    • Data Provenance: Not specified in this summary. These are pre-clinical tests, so "country of origin of data" is not typically a relevant factor like it would be for a clinical trial with human subjects. The tests would have been conducted in a lab environment.
    • Retrospective or Prospective: These would be considered prospective tests, as they are specifically conducted to evaluate the performance of the new device for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • N/A. This information is not relevant for a physical device's pre-clinical performance testing. There is no "ground truth" to be established by human experts in the context of catheter performance testing, such as flow rates, tensile strength, or blood trauma. These are typically measured by laboratory equipment and standardised protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • N/A. Not applicable for pre-clinical bench or animal testing of a physical medical device. Adjudication methods are typically used in clinical trials or for subjective assessments by human readers/experts, which is not the case here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • N/A. This is a clearance for a physical medical device (catheter and introducer), not an AI/ML diagnostic or assistive tool. Therefore, no MRMC study, human readers, or AI assistance is involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • N/A. Not applicable for a physical medical device. This is relevant for AI/ML algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • N/A. For pre-clinical bench testing, the "ground truth" would be objective measurements obtained through calibrated equipment and validated test methods (e.g., specific flow rates under defined pressures, measured tensile strength, observed material characteristics under stress, analysis of blood components for trauma). For animal testing, it would involve direct physiological measurements or observation of biological responses, with controls. It's not "ground truth" in the sense of expert diagnosis or pathology.

    8. The sample size for the training set:

    • N/A. There is no "training set" in the context of a physical medical device. This term applies to machine learning algorithms.

    9. How the ground truth for the training set was established:

    • N/A. As there is no training set for a physical device, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1