510(k) Data Aggregation

    K Number
    K170233

    Validate with FDA (Live)

    Date Cleared
    2017-10-13

    (261 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K142575, K130908, K071423

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PVC Urinary Catheter is launched for clean intermittent catheterization (CIC) treatment. It is intended for use in the drainage of fluid from the urinary tract. It is indicated for use in male, female, and pediatric patients 2-12 years old.

    The Silicone Urinary Catheter is intended to be placed in the urethra, to drain urine into a collection device. Two-way catheters are intended for urological bladder drainage only. Three-way catheters are intended for urological bladder drainage and bladder irrigation. It is indicated for use in male, female, and pediatric patients 2-12 years old.

    The Latex Urinary Catheter is intended to be placed in the bladder, through the urethra, to drain urine into a collection device. Neleton catheters are launched for clean internittent catheterization (CIC). Two-way catheters are intended for urological bladder drainage only. Three-way catheters are intended for urological bladder irrigation. It is indicated for use in male, female, and pediatric patients 2-12 years old.

    Device Description

    Disposable urinary catheter is a kind of disposable product which is used for draining the bladder, for temporary or short term. Based on the materials, the disposable urinary catheter includes PVC Urinary Catheter, Silicone Urinary Catheter and Latex Urinary Catheter.

    PVC Urinary Catheter only has nelaton type structure, which is made of PVC and is intended for draining the bladder for short term. PVC Urinary Catheter consists of drainage funnel, drainage lumen, eyelets and tip. It is supplied in French size ranging from 8 to 20F. The catheters come in sizes from 12Fr-20Fr for adults and 8Fr-10Fr for pediatrics 2 -12 years old.

    Silicone Urinary Catheter is made of pure silicone. It is intended to be placed in the bladder, through the urethra, to drain urine into a collection device and has two-way type and three-way type. The two-way type consists of funnel, drainage lumen, drainage eyes, balloon, inflating connector and valve. The three-way type consists of Funnel, Drainage lumen, Injection connector, injection lumen, drainage eyes, balloon, inflating connector and valve. Both of them are supplied in French size ranging from 8 to 24F. The catheters come in sizes from 12Fr-24Fr for adults and 8Fr-10Fr for pediatrics 2 -12 years old.

    Latex Foley Catheter is made of nature latex, which has nelaton type, two-way type and three-type.

    The nelaton Latex Urinary Catheter is intended for draining the bladder for short term and consists of drainage funnel, drainage lumen, eyelets and tip. The nelaton catheters come in sizes from 12Fr-20Fr for adults and 8Fr-10Fr for pediatrics 2-12 years old.

    The two-way type and three-way type are intended to be placed in the bladder, through the urethra, to drain urine into a collection device. The two-way type consists of funnel, drainage lumen, drainage eyes, balloon, inflating connector and valve. The three-way type consists of Funnel, Drainage lumen, Injection connector, injection lumen, drainage eye. Both of them are supplied in French size ranging from 8 to 24F. The catheters come in sizes from 12Fr-24Fr for adults and 8Fr-10Fr for pediatrics 2 -12 years old.

    AI/ML Overview

    This document describes a 510(k) premarket notification for Disposable Urinary Catheters (PVC, Silicone, and Latex). It focuses on demonstrating substantial equivalence to predicate devices, rather than a de novo evaluation of clinical effectiveness with an AI component. Therefore, much of the requested information regarding AI-specific studies (e.g., sample sizes for test sets, number of experts for ground truth, MRMC studies, standalone algorithm performance, training set details) is not applicable or present in this document.

    Here's an analysis of the provided information, addressing the applicable points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally based on meeting specified performance standards. The reported device performance is that the device "Passed" all tests against these requirements.

    Test CategorySpecific Test / Acceptance CriteriaReported Device Performance
    PVC Urinary Catheter (based on EN 1616:1997(R2007))
    Surface finishMeet the requirement of 4.2 of EN 1616:1997(R2007).Passed
    DimensionsMeet the requirement of 4.3 of EN 1616:1997(R2007).Passed
    StrengthMeet the requirement of 4.4 of EN 1616:1997(R2007).Passed
    Connector securityMeet the requirement of 4.5 of EN 1616:1997(R2007).Passed
    Flow ratesMeet the requirement of 4.8 of EN 1616:1997(R2007).Passed
    Biocompatibility (based on ISO 10993 series)
    CytotoxicityISO 10093-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityPassed
    SensitizationISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPassed
    IrritationISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationPassed
    Systemic toxicityISO 10993-11 Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic ToxicityPassed
    SterilitySAL ≤ 10^-6^Passed
    Silicone Urinary Catheter (based on ASTM F623-99(2013))
    Flow rate through drainage lumenMeet the requirement of 4.1 of ASTM F623-99(2013).Passed
    Balloon IntegrityMeet the requirement of 4.2 of ASTM F623-99(2013).Passed
    Inflated Balloon Response to TractionMeet the requirement of 4.3 of ASTM F623-99(2013).Passed
    Balloon Volume MaintenanceMeet the requirement of 4.4 of ASTM F623-99(2013).Passed
    Balloon Size and Shaft SizeMeet the requirement of 4.5 of ASTM F623-99(2013).Passed
    Deflation Reliability (Failure to Deflate)Meet the requirement of 4.6 of ASTM F623-99(2013).Passed
    Biocompatibility (same ISO standards as PVC)CytotoxicityPassed
    SensitizationPassed
    IrritationPassed
    Systemic toxicityPassed
    Implantation (Biological Evaluation Of Medical Devices - Part 6: Tests For Local Effects After Implantation)Passed
    SterilitySAL shall be less than 10^-6^ .Passed
    Latex Urinary Catheter (based on EN 1616:1997(R2007) and ASTM F623-99(2013))
    Surface finishMeet the requirement of 4.2 of EN 1616:1997(R2007).Passed
    DimensionsMeet the requirement of 4.3 of EN 1616:1997(R2007).Passed
    StrengthMeet the requirement of 4.4 of EN 1616:1997(R2007).Passed
    Connector securityMeet the requirement of 4.5 of EN 1616:1997(R2007).Passed
    Flow ratesMeet the requirement of 4.8 of EN 1616:1997(R2007) and 4.1 of ASTM F623-99(2013).Passed
    Balloon IntegrityMeet the requirement of 4.2 of ASTM F623-99(2013).Passed
    Inflated Balloon Response to TractionMeet the requirement of 4.3 of ASTM F623-99(2013).Passed
    Balloon Volume MaintenanceMeet the requirement of 4.4 of ASTM F623-99(2013).Passed
    Balloon Size and Shaft SizeMeet the requirement of 4.5 of ASTM F623-99(2013).Passed
    Deflation Reliability (Failure to Deflate)Meet the requirement of 4.6 of ASTM F623-99(2013).Passed
    Biocompatibility (same ISO standards as above)CytotoxicityPassed
    SensitizationPassed
    IrritationPassed
    Systemic toxicityPassed
    ImplantationPassed
    SterilitySAL ≤ 10^-6^.Passed

    2. Sample size used for the test set and the data provenance

    This document describes non-clinical "bench testing" and "biocompatibility tests" performed in accordance with standards like EN 1616, ASTM F623, and ISO 10993. These types of tests do not typically involve patient-specific data or test sets in the context of AI. The sample sizes for these bench tests are not specified in this document, nor is the data provenance in terms of country of origin or retrospective/prospective status as it would be for a clinical study. The device manufacturer is Zhanjiang Star Enterprise Co., Ltd based in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-clinical, bench testing submission. No "ground truth" established by experts in a clinical context (e.g., radiologists) is mentioned or required for this type of submission.

    4. Adjudication method for the test set

    Not applicable. There are no clinical test sets or expert evaluations requiring adjudication in this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a urinary catheter, not an AI-powered diagnostic or assistive technology. No MRMC studies were conducted or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device (urinary catheter) and does not involve an algorithm or AI.

    7. The type of ground truth used

    For the performance tests, the "ground truth" is defined by the technical specifications and requirements outlined in the referenced international standards (e.g., EN 1616, ASTM F623). For biocompatibility, the "ground truth" is adherence to the criteria set forth in the ISO 10993 series of standards. This is not clinical "ground truth" such as pathology or outcomes data.

    8. The sample size for the training set

    Not applicable. This device does not involve machine learning or AI, and therefore, no training set.

    9. How the ground truth for the training set was established

    Not applicable. No training set exists for this device.