Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K230563

    Validate with FDA (Live)

    Date Cleared
    2023-09-27

    (210 days)

    Product Code
    Regulation Number
    888.3510
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVOLUTION® Hinge Knee System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

    1. noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis:

    2. inflammatory degenerative joint disease including rheumatoid arthritis;

    3. correction of functional deformity:

    4. revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

    The EVOLUTION® Hinge Knee System implants are for cemented use only.

    Device Description

    The EVOLUTION® Hinge Knee System is a hinged, constrained cemented total knee system being introduced as a line extension to MicroPort Orthopedics Inc.'s EVOLUTION® Revision knee portfolio. The system is composed of:

    • a femoral component offered in 7 sizes (sizes 2-8) with femoral yoke fixation screw for ● attachment to the yoke assembly
    • a hinge yoke assembly offered in 9 sizes based on femur size (size 2-3, 4-6, and 7-8) and tibial insert thickness (small, medium, large)
    • a yoke extension stop offered in 3 options (0°, 5°, 10°)
    • a tibial insert offered in 8 sizes (sizes 1-8) and in 5 thickness options (12, 14, 17, 20, 24mm)
    • a tibial base offered in two variants:
      • modular (8 sizes, 1-8) with tibia locking screw for attachment to stem extensions o and stem extension adapters
      • monolithic (3 sizes, 1-3) with fixed small keel o
    • . Offset adapters (2 sizes, 4mm x 25mm and 8mm x 25mm) and valgus stem adapters (2 sizes, 1° x 25mm and 2° x 25mm)
    • Universal femoral augments
      • distal augments offered in 7 sizes (sizes B-H) in 4, 8, 12, 16, and 20mm o thicknesses
      • posterior femoral augments in 7 sizes (sizes B-H) in 4, 8, and 12mm thicknesses, O
    • Medial and lateral tibial augments offered in 8 sizes (sizes 1-8) in 5, 10, and 15mm thicknesses

    Components are manufactured from cobalt chrome alloy, titanium alloy, UHMWPE and crosslinked UHMWPE, and PEEK conforming to ASTM F75, F1537 - Alloy 1, F136, F648, F2026, respectively.

    The system includes new and existing instrumentation from K140735, K162026, and K171389 to facilitate device implantation.

    The subject femoral offset adapters and femoral valgus adapters are not for use with the EVOLUTION® Hinge tibial bases. The subject system is compatible with previously cleared stem extensions, stem extension adapters, and modular keels cleared in K162026 and patellae from K953439 and K122218. Furthermore, the subject universal femoral augments introduced in this 510(k) are backwards compatible with the EVOLUTION® Femurs cleared in K142550, K171389, and K182125.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about the acceptance criteria or a study proving that a device meets acceptance criteria. The document is an FDA 510(k) premarket notification for a knee system, primarily discussing its substantial equivalence to previously marketed devices based on non-clinical testing, design features, and materials.

    It explicitly states:

    • "Clinical data were not provided for the subject device." (Page 7, Section E. Clinical Testing)

    Therefore, I cannot provide the requested information, which includes:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size used for the test set and data provenance.
    3. Number of experts used and their qualifications for ground truth.
    4. Adjudication method for the test set.
    5. Results of a multi-reader multi-case (MRMC) comparative effectiveness study.
    6. Results of a standalone (algorithm only) performance study.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    This document focuses on mechanical and material testing, and substantial equivalence, not a clinical study involving an AI or diagnostic device with acceptance criteria for performance metrics like sensitivity or specificity.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213817

    Validate with FDA (Live)

    Date Cleared
    2022-01-05

    (29 days)

    Product Code
    Regulation Number
    888.3565
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MicroPort knee instruments are accessory devices and are intended to be used to assist in the implantation of MicroPort Total Knee Systems in their cleared Indications for Use as provided below:

    MicroPort Total Knee Systems are indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

    1. noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis:

    2. inflammatory degenerative joint disease including rheumatoid arthritis;

    3. correction of functional deformity;

    4. revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

    Non-porous MicroPort total knee replacement implants are for cemented use only.

    Porous coated MicroPort total knee replacement implants, including ADVANCE® BIOFOAM® Tibial System and EVOLUTION® BIOFOAM® Tibial System implants, are for use without bone cement.

    MicroPort's PROPHECY® Pre-Operative Navigation Alignment Guides are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in quiding the marking of bone before cutting. The PROPHECY® Pre-Operative Navigation Alignment Guides are intended for use with MicroPort's ADVANCE® and EVOLUTION® Total Knee Systems and their cleared indications for use, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. The PROPHECY® Pre-Operative Navigation Alignment Guides are intended for single use only.

    Device Description

    The device modification consists of an alteration to the sterilization instructions for FDAcleared MicroPort Orthopedics' (MPO) non-sterile knee orthopedic joint replacement instruments. The subject instruments are part of MicroPort Orthopedics' 510(k)-cleared knee product lines and are required to facilitate total knee arthroplasty procedures. The modification will allow the option to sterilize the subject instruments using an FDA-cleared containment device. The subject devices will be placed in an FDA-cleared containment device, which will be double-wrapped in an FDA-cleared CSR wrap or similar type nonwoven, medical grade wrapping material, and then steam sterilized.

    The modified sterilization process of the subject instruments was successfully challenged and validated through worst-case load configurations using an FDA-cleared containment device. The intended use and sterilization parameters, such as cycle, temperature, and exposure time, remain identical to the predicate devices.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided document:

    Acceptance Criteria and Device Performance Study for MPO Knee Instruments Reprocessing Change (K213817)

    This submission (K213817) focuses on a modification to the sterilization instructions for existing, FDA-cleared MicroPort Orthopedics' non-sterile knee instruments. The device itself (the instruments) remains unchanged in its design or intended use. The study specifically validates the modified sterilization process.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this submission is for a sterilization process change rather than a new diagnostic or AI-driven device, the acceptance criteria are related to the effectiveness of the sterilization and the integrity of the instruments after the new process.

    Acceptance CriteriaReported Device Performance
    Sterilization Efficacy (per AAMI ST77:2013 and AAMI ST79:2017): Achieves sufficient device sterility when instruments are placed in an FDA-cleared containment device (single-level and double-level trays) and steam sterilized.Met: "The modified sterilization process of the subject instruments was successfully challenged and validated through worst-case load configurations using an FDA-cleared containment device." "MicroPort non-sterile orthopedic joint replacement instruments are able to withstand the reported sterilization cycles and achieve sufficient device sterility."
    Instrument Integrity after Sterilization: Instruments maintain their functionality and structural integrity after undergoing the modified sterilization process.Met: Implied by the "Technological Characteristics Comparison" stating that instruments successfully withstand sterilization cycles, and the absence of any reported failures in nonclinical testing.
    Vibration Resistance (per ISTA 2A): Packaging and instruments withstand vibration during transport after sterilization.Met: "Vibration Test of MicroPort orthopedic joint replacement instruments using FDA-cleared containment device per ISTA 2A."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a numerical "sample size" in terms of how many individual instruments were tested or how many sterilization cycles were run. However, the testing was conducted under "worst-case load configurations," implying a robust validation approach.

    • Sample Size: Not explicitly stated as a numerical count of individual instruments or cycles. Testing was performed under "worst-case load configurations" and for "MicroPort orthopedic joint replacement instruments."
    • Data Provenance: The studies were non-clinical validation tests performed by MicroPort Orthopedics Inc. There is no mention of country of origin for data or whether it was retrospective or prospective, as it pertains to laboratory validation of a manufacturing process change.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable to this submission. The device is a set of surgical instruments, and the study is a non-clinical validation of a sterilization process. There is no "ground truth" to be established by human experts in the way that would apply to diagnostic imaging or AI devices. The conformity to sterilization standards (AAMI ST77:2013 and AAMI ST79:2017) and vibration testing (ISTA 2A) serves as the "ground truth" for the performance of the modified process.

    4. Adjudication Method for the Test Set

    This section is not applicable. Since there are no human experts establishing a ground truth for a diagnostic outcome, there is no adjudication method. The test results are objective measurements against established engineering and sterilization standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No. A MRMC comparative effectiveness study is designed for evaluating human performance, often with and without AI assistance, typically in diagnostic tasks. This submission is for surgical instruments and a sterilization process change, so such a study is not relevant or required.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This question typically applies to AI algorithms. The device being discussed is physical surgical instrumentation, and the study is evaluating a sterilization process, not an algorithm's performance.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission is established by adherence to recognized industry standards and validated testing protocols for sterilization and material integrity:

    • Sterility: Demonstrated by successful validation according to AAMI ST77:2013 and AAMI ST79:2017, likely involving biological indicator testing and sterility assurance level (SAL).
    • Physical Integrity/Durability: Demonstrated by successful vibration testing according to ISTA 2A.

    8. The Sample Size for the Training Set

    This section is not applicable. There is no "training set" in the context of validating a sterilization process for physical surgical instruments. Training sets are relevant for machine learning algorithms.

    9. How the Ground Truth for the Training Set was Established

    This section is not applicable, as there is no training set for this type of submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171389

    Validate with FDA (Live)

    Date Cleared
    2017-08-08

    (89 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    Reference Devices :

    K990030 ADVANCE® Revision (CCK System), K142550 EVOLUTION® MP Revision Femur, K063731 ADVANCE® STATURE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EVOLUTION® Revision CCK System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

    1. Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis
    2. Inflammatory degenerative joint disease, including rheumatoid arthritis;
    3. Correction of functional deformity
    4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.
      The EVOLUTION® Medial-Pivot Total Knee System Nonporous implants are for cemented use only.
    Device Description

    EVOLUTION® Revision CCK (Constrained Condylar Knee) is being introduced to supplement MicroPort Orthopedics' knee product lines to provide options in the case of revision or complex primary total knee replacements. The design features are summarized below:

    • Components manufactured from cobalt chrome alloy, titanium alloy, and UHMWPE conforming to ASTM F75, F136, and F648, respectively
    • . Available in 8 standard sizes, left and right
    • Available in 2 plus sizes, left and right
    • . System includes femoral component, augments, stem offset adapters, tibial insert and stem cap
    AI/ML Overview

    The provided text describes the MicroPort Orthopedics EVOLUTION® Revision CCK System, a knee joint prosthesis. However, it does not contain a study that establishes acceptance criteria and then proves the device meets those criteria in the typical format of a diagnostic or AI/software device.

    Instead, this 510(k) summary focuses on demonstrating "substantial equivalence" to predicate devices based on design features, materials, and non-clinical performance testing. The information provided is primarily related to mechanical and material performance rather than diagnostic accuracy or human performance with or without AI.

    Therefore, I cannot fulfill all parts of your request as posed because the document is not an "acceptance criteria study" in the sense you are asking for. It is a regulatory submission demonstrating substantial equivalence.

    Based on the provided text, I can extract information relevant to non-clinical performance testing:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" for each test in a pass/fail format alongside numerical performance values. Instead, it describes comparative performance against a predicate device. The general "acceptance criteria" can be inferred as performing "as well as" or "greater than" the predicate device for various mechanical properties to demonstrate substantial equivalence.

    Acceptance Criterion (Inferred)Reported Device Performance
    Bacterial Endotoxin Limit (< 20 EU/device)Less than the USP endotoxin limit of 20 EU/device.
    Patellofemoral Contact Area (Perform as well as predicate)Performed as well as the predicate K990030 implants.
    Tibiofemoral Contact Area (Perform as well as predicate)Possessed greater maximum contact area than the predicate K990030 implants.
    Tibial Post Shear Fatigue (Perform as well as predicate)Demonstrated greater fatigue resistance than the predicate K990030.
    Taper Disassembly (Withstand typical in vivo forces)Withstand typical in vivo disassociation forces as reported in peer-reviewed literature.
    Taper Fatigue (Achieve values greater than predicate)Achieved taper disassembly values greater than those of the predicate K162026 for Femoral constructs.

    2. Sample size used for the test set and the data provenance: Not applicable. This document describes non-clinical (laboratory) testing, not a test set derived from patient data. The "samples" would be the manufactured device components themselves, tested in a lab setting. The provenance is internal laboratory testing by MicroPort Orthopedics Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is not a study requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a physical knee prosthesis, not an AI/software device for interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The "truth" for these non-clinical tests would be the measurement results against established engineering and material standards or comparisons to predicate devices' performance.

    8. The sample size for the training set: Not applicable. This device is not an AI algorithm.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1