Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K180949

    Validate with FDA (Live)

    Device Name
    Steribite
    Manufacturer
    Date Cleared
    2018-11-08

    (211 days)

    Product Code
    Regulation Number
    882.4840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K140927

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steribite® Rongeur is a manually operated instrument indicated for cutting bone during surgery involving the skull or spinal column.

    Device Description

    Steribite® is a system of disposable manual surgical Kerrison rongeurs. The instruments are offered in a 40° up configuration having shaft lengths of 8 and 11 inches and bite sizes 1mm to 5mm to accommodate variations in surgical need. The devices are sold sterile and single use only.

    AI/ML Overview

    The document describes a 510(k) premarket notification for the "Steribite®" manual rongeur, which is a device used in surgery involving the skull or spinal column. The submission aims to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Mechanical Testing
    Cut PerformanceFull profile cut in simulated use (based on Steris IMS test standard).Demonstrated full profile cut in simulated use.
    Cut ForceNot explicitly stated as a pass/fail criterion, but assessed against a reference device.Assessed by applying a controlled load (10-35 pounds) in 5-pound step-wise fashion to cut a 2mm thick bone analog compared to the reference device. (Comparison results are not detailed as specific pass/fail values but implied to be acceptable for substantial equivalence).
    Handle StrengthForce to cause failure was at least twice the required cutting force.Applied a continuous load to failure; demonstrated that the force to cause failure was at least twice the required cutting force.
    Sterilization
    Sterilization ValidationIn accordance with ISO 11137-1:2006 and ISO 11137-2006 for radiation sterilization.Validation conducted in accordance with specified ISO standards.
    Bacterial Endotoxin TestingIn accordance with AAMI ST72:2011/(R)2016.Testing conducted in accordance with specified AAMI standard.
    Packaging
    Packaging Performance & StabilityIn accordance with ANSI/AAMI/ISO 11607-1:2006/IR)2010.Testing conducted in accordance with specified ANSI/AAMI/ISO standard.
    Biocompatibility
    CytotoxicityNot explicitly stated, but implies meeting safety standards of ISO 10993-5 Third Edition 2009-06-01.Assessed for cytotoxicity.
    SensitizationNot explicitly stated, but implies meeting safety standards of ISO 10993-10 Third Edition 1020-08-01.Assessed for sensitization.
    Irritation/Intracutaneous ReactivityNot explicitly stated, but implies meeting safety standards of ISO 10993-10 Third Edition 1020-08-01.Assessed for irritation/intracutaneous reactivity.
    Acute Systemic ToxicityNot explicitly stated, but implies meeting safety standards for medical devices.Assessed for acute systemic toxicity.
    Material-Mediated PyrogenicityNot explicitly stated, but implies meeting safety standards for medical devices.Assessed for material-mediated pyrogenicity.
    HemolysisNot explicitly stated, but implies meeting safety standards for medical devices.Assessed for hemolysis.
    Residual Manufacturing Material CleanlinessBased on ASTM F2459.Applied ASTM F2459 to demonstrate residual manufacturing material cleanliness.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set: The document mentions "the subject device models, 1-5 mm bite sizes and 8-11 in handles," but does not explicitly state the number of samples (individual devices) used for each mechanical test.
    • Data Provenance: The data is generated from laboratory testing of the device itself. There is no mention of country of origin for data or whether it's retrospective or prospective, as it pertains to device performance testing rather than clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This section is not applicable to this submission. The "Steribite®" is a manual surgical instrument, and its performance and safety validation relies on non-clinical mechanical, sterilization, packaging, and biocompatibility testing against established industry standards and comparison to a predicate device. There is no mention of "ground truth" established by experts in the context of clinical outcomes or diagnostic accuracy, as would be relevant for software or diagnostic devices.

    4. Adjudication Method for the Test Set

    This section is not applicable for the reasons stated above. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving human interpretation or clinical endpoints, which are not part of this 510(k) submission for a manual surgical instrument.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not performed. This type of study is relevant for AI/radiology devices where human readers' performance is augmented by AI. The Steribite® is a manual surgical rongeur, not an AI or diagnostic device that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone study was not performed. This device is a manual surgical instrument and does not incorporate an algorithm or AI component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This section is not applicable in the traditional sense. The "ground truth" for the non-clinical tests conducted is defined by established engineering principles, industry standards (e.g., ISO, AAMI, ASTM), and the expected performance characteristics of a manual rongeur compared to its predicate. For example, for "cut performance," the ground truth is a "full profile cut in simulated use" as per a specific test standard.

    8. The sample size for the training set

    Not Applicable. This device is a mechanical surgical instrument and does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not Applicable. As there is no training set, there is no ground truth established for one.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162921

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-03-13

    (146 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K140927

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Navigated Instrument System is indicated for use during the preparation and placement of Orthofix screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Navigated Instruments are specifically designed for use with the BrainLab Vector Vision system and the Medronic StealthStation System which are indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a skull, a long bone, or vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks for the anatomy.

    Device Description

    The Navigated Instrument System is comprised of manual surgical instruments for use with the BrainLab VectorVision System and the Medtronic StealthStation Navigation System to assist surgeons in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures for preparation and placement of pedicle screw system implants. This surgical imaging technology provides surgeons visualization for complex and MIS procedures and confirms the accuracy of advanced surgical procedures. Use of these navigation systems provides the surgeon access to real-time, multi-plane 3D images (and 2D images) providing confirmation of hardware placement. Use of the Navigated Instrument System is limited to Taps ranging in sizes of 4.5mm to 8.5mm and bone screws ranging from 4.5mm to 8.5mm with length ranging from 25mm to 55mm.

    The Navigated Instrument System Medtronic compatible instruments are comprised of Bone Awl, Bone Taps, and Bone Probes only.

    The Navigated Instrument System BrainLab compatible instruments are comprised of Bone Awl. Bone Taps, Bone Probes and a variety of Screw Drivers.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the "Navigated Instrument System." This document aims to demonstrate substantial equivalence to predicate devices, not necessarily to prove a device meets specific performance acceptance criteria through clinical studies in the typical sense (e.g., diagnostic accuracy on a test set).

    Here's an attempt to answer your questions based on the available information, noting that much of it is not explicitly stated in this type of regulatory document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria or detailed performance results in the format of a table. Instead, it makes a general statement about substantial equivalence based on "compatibility, accuracy, function and performance."

    Acceptance Criteria (Implied)Reported Device Performance (Implied)
    Compatibility with BrainLab VectorVision and Medtronic StealthStation systems.Demonstrated compatibility with both systems.
    Accuracy (specific numerical target not provided)."Accuracy" demonstrated through engineering analysis.
    Functionality in spinal surgery (preparation and placement of Orthofix screws)."Function" demonstrated through engineering analysis.
    Performance during spinal surgery (assisting in locating anatomical structures in open or MIS procedures)."Performance" demonstrated through engineering analysis.
    Substantial equivalence to predicate devices (K153442 and K070106).Concluded to be substantially equivalent based on engineering analysis and non-clinical tests.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Engineering analysis and performance data" and "non-clinical test." This implies in-vitro or bench testing rather than studies involving patient data. Therefore:

    • Sample size for the test set: Not specified, but likely refers to a set of instruments/fixtures used for bench testing, not human subjects or patient data.
    • Data provenance: Not explicitly stated, but given it's a 510(k) submission from Orthofix Inc. in Lewisville, Texas, the testing was likely conducted in the US or by an affiliated entity. The data would be non-clinical/bench testing data, not clinical in retrospect or prospect.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable or not provided in the context of this 510(k) summary. "Ground truth" in this context would typically refer to clinical data or expert consensus on diagnoses, which is not what was used given the "engineering analysis and performance data" for a surgical instrument system. The ground truth for bench testing would be defined by metrology standards or physical measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable or not provided. This usually refers to how disagreements among expert readers are resolved in diagnostic image interpretation studies. Since this device relates to surgical instrumentation and its performance was assessed via engineering analysis, such a method would not be used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is a surgical instrument system, not an AI-powered diagnostic device. Therefore, an MRMC study assessing human reader improvement with AI assistance is not relevant and was not performed.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    Not applicable. This device is an instrument system, not an algorithm. Its function is to assist a human surgeon. Standalone performance as an algorithm would not be relevant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the "engineering analysis and performance data" would be based on physical measurements and engineering specifications. For example, the accuracy of instrument tip localization might be verified against known physical coordinates or measurements of a phantom. It is not clinical ground truth like pathology reports or patient outcomes.

    8. The sample size for the training set

    The concept of a "training set" is primarily relevant for machine learning or AI models. This device is a traditional mechanical/navigated instrument system. Therefore, there is no training set in the AI sense. The "training" for the system would involve calibration procedures and design verification.

    9. How the ground truth for the training set was established

    As there is no training set in the AI context, this question is not applicable. The design and performance were validated against engineering principles and predicate device characteristics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1