Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K233440

    Validate with FDA (Live)

    Date Cleared
    2024-07-02

    (258 days)

    Product Code
    Regulation Number
    884.2740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K101600, K023931, K140535

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Avalon CL Fetal & Maternal (F&M) Pod & Patch is a device indicated for use by healthcare professionals in a clinical setting for non-invasive monitoring of maternal heart rate (aHR), fetal heart rate (aFHR), and uterine activity (aToco) in women who are at >36 completed weeks, in labor, with singleton pregnancy, using surface electrodes on the maternal abdomen.

    Device Description

    The Avalon CL Fetal & Maternal (F&M) Pod and the Avalon CL Fetal & Maternal (F&M) Patch is a beltless battery-powered maternal-fetal monitoring system that non-invasively measures abdominal fetal heart rate (aFHR), abdominal uterine activity (aToco), and abdominal maternal heart rate (aHR). The Avalon CL Fetal & Maternal (F&M) Patch is a single-use disposable adhesive electrode patch designed to be affixed to the maternal abdomen. The Avalon CL Fetal & Maternal (F&M) Pod is a reusable device which, when connected to the Avalon CL Fetal & Maternal (F&M) Patch, picks up electrical signals and converts it to Short Range Radio (SRR). The Avalon CL Fetal & Maternal Pod communicates the data measurement values to the Avalon CL Base Station using Short-Range Radio (SRR). The Avalon CL Base Station in turn relays the information to the connected Philips Fetal-Maternal (FM) Monitor (i.e., FM20, FM30, FM40, and FM50).

    AI/ML Overview

    The provided FDA 510(k) summary for the Philips Avalon CL Fetal & Maternal (F&M) Pod & Patch focuses heavily on demonstrating substantial equivalence to a predicate device through non-clinical testing and comparison of technical characteristics rather than a detailed clinical study report with specific acceptance criteria and performance metrics for the device's accuracy in monitoring FHR, MHR, and UA.

    Therefore, much of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of clinical performance (e.g., accuracy, sensitivity, specificity, agreement with ground truth for FHR, MHR, and UA) is not explicitly detailed in this document. The document primarily discusses non-clinical tests for safety, electrical performance, and biocompatibility.

    However, based on the information provided, here's what can be extracted and inferred:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific clinical performance acceptance criteria (e.g., accuracy ranges for FHR) and reported device performance from an effectiveness standpoint. Instead, it details non-clinical technical acceptance criteria related to safety, electrical performance, and biocompatibility, which the device met.

    Criterion CategorySpecific Criterion / TestAcceptance Criterion (Implicit)Reported Device Performance (Implicit)
    BiocompatibilityCytotoxicity (ISO 10993-5)Met acceptance criteria as defined in test requirementsMet
    Sensitization (ISO 10993-10)Met acceptance criteria as defined in test requirementsMet
    Irritation (ISO 10993-10)Met acceptance criteria as defined in test requirementsMet
    Electrical SafetyANSI AAMI ES60601-1Compliance with standard for basic safety and essential performancePassed
    EMC/WirelessIEC 60601-1-2Compliance with standard for electromagnetic disturbancesPassed
    IEEE ANSI C63.27Compliance with standard for evaluation of wireless coexistencePassed
    IEC/TR 60601-4-2Compliance with standard for electromagnetic immunityPassed
    Alarm SystemsIEC 60601-1-8Compliance with standard for alarm systemsPassed
    Battery SafetyIEC 62133-2Compliance with standard for lithium systemsPassed
    Software/FirmwareFDA Guidance complianceCompliance with "Content of Premarket Submissions for Device Software Functions"Documentation provided and reviewed
    CybersecurityFDA Guidance complianceCompliance with "Cybersecurity in Medical Devices" guidanceDocumentation provided and reviewed
    Performance BenchInspection of labeling and pouch sealingN/A (Visual inspection)Met
    Impedance/tensile strength/pull-off force/noise level/conductivity/offset voltage/defibrillation overload (new and aged patches)Met acceptance criteria as defined in test requirementsMet
    In vivo testing: integrity, detachment/reattachment, and performance (impedance, noise level, MHR, conductivity) after shower and usageMet acceptance criteria as defined in test requirementsMet
    Peel-off force of each electrode and central stickerMet acceptance criteria as defined in test requirementsMet
    MHR/FHR/UA accuracy after storage at various temperaturesMet acceptance criteria as defined in test requirementsMet
    Signal transmission continuityMet acceptance criteria as defined in test requirementsMet

    Regarding MHR/FHR/UA accuracy, the document states for "Performance Bench" that "MHR/FHR/UA accuracy after stored in room (23℃), high (32℃) and low (2-8℃) temperature" were conducted and "met the acceptance criteria as defined in the test requirements." However, the specific numerical acceptance criteria for accuracy (e.g., mean absolute difference, percentage agreement, etc.) and the reported numerical performance regarding MHR/FHR/UA accuracy are not provided in this summary. This suggests that these accuracy tests were likely bench tests under controlled conditions, not a clinical trial comparing device readings to a clinical ground truth.

    2. Sample Size for Test Set and Data Provenance

    The document does not explicitly mention a "test set" in the context of a clinical performance study with human subjects to evaluate the accuracy of FHR, MHR, and UA measurements. The in-vivo testing mentioned under "Performance Bench" refers only to "integrity, detachment/reattachment, and performance (impedance, noise level, MHR, conductivity) after shower and usage (8 hours/32 hours) for the patch (Novii Patch)." This does not sound like a large-scale clinical accuracy study.

    Therefore, based on the provided text alone:

    • Sample size for the test set: Not explicitly stated for clinical performance as commonly understood for device accuracy. The "in vivo testing" details are too limited to determine sample size or its direct relation to device accuracy claims.
    • Data provenance: Not explicitly stated. The type of testing suggests it might be internal company testing rather than an independent clinical trial.

    3. Number of Experts and Qualifications for Ground Truth

    Given the lack of a detailed clinical performance study report, there is no information provided regarding the number or qualifications of experts used to establish ground truth for a clinical test set for FHR, MHR, or UA.

    4. Adjudication Method

    Again, due to the absence of a detailed clinical performance study, there is no information provided on any adjudication method (e.g., 2+1, 3+1) for a clinical test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss human readers or AI assistance in this context. This device appears to be a monitoring system for physiological parameters, not an AI-assisted diagnostic imaging or interpretation tool.

    6. Standalone Performance

    The device itself is a "standalone" monitoring system in the sense that it performs its measurements (aHR, aFHR, aToco) via its electrodes and pod, then relays this data to a Philips Fetal-Maternal (FM) Monitor for display. The performance tests ("Performance Bench") assess the device's ability to measure these parameters. However, the exact "standalone" clinical accuracy metrics (e.g., sensitivity, specificity, accuracy vs. a gold standard) are not provided. The phrase "standalone performance" is generally associated with diagnostic algorithms, which doesn't seem to be the primary claim here.

    7. Type of Ground Truth Used

    For the non-clinical performance "MHR/FHR/UA accuracy after stored in room (23℃), high (32℃) and low (2-8℃) temperature," the type of ground truth used is not specified. It likely refers to controlled laboratory measurements against calibrated reference standards, rather than clinical ground truth like pathology, expert consensus, or outcomes data. For clinical performance data (which is not detailed), common ground truths for FHR, MHR, and UA would be internal fetal monitoring (IUPC for UA, fetal scalp electrode for FHR) or expert interpretation of existing monitoring tracings (though this isn't mentioned).

    8. Sample Size for the Training Set

    No information is provided about a "training set." This term is typically associated with machine learning or AI algorithm development. While the device uses signal processing (template matching, filtering, confidence tagging) to identify fECG and mECG complexes, the document does not describe the development or training of such algorithms or any associated data sets used for this purpose.

    9. How Ground Truth for the Training Set Was Established

    As no training set is discussed, there is no information provided on how ground truth for a training set was established.

    In summary, the provided FDA summary focuses on demonstrating substantial equivalence through technical and non-clinical performance and safety testing. It lacks detailed clinical performance data (e.g., accuracy, sensitivity, specificity) against a clinical ground truth, specific sample sizes for clinical evaluations, or information about expert consensus or adjudication methods for such clinical data, which are typically found in clinical study reports for devices claiming diagnostic or interpretative capabilities.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231964

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-12-08

    (158 days)

    Product Code
    Regulation Number
    884.2740
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K072976, K140535

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Novii+ Pod is an antepartum and intrapartum Maternal/Fetal Monitor that non-invasively measures fetal heart rate (FHR), uterine activity (UA) and maternal heart rate (MHR). The Novii+ Pod acquires the FHR tracing from abdominal surface electrodes that pick up the fetal ECG (fECG) signal. Using the same surface electrodes, the Pod also acquires the UA tracing from the uterine electromyography (EMG) signal and the MHR tracing from the maternal ECG signal (mECG). The Pod is indicated for use on women who are at 34 weeks and 0/7 days and greater with singleton pregnancies, using surface electrodes on the maternal abdomen.

    The Novii Patch is an accessory to the Novii+ Pod that connects directly to the Novii+ Pod and contains the surface electrodes that attach to the abdomen.

    The Novii+ Interface is an accessory to the Novii+ Pod which provides a means of interfacing the wireless output of the Novii+ Pod to the transducer inputs of a Maternal/Fetal Monitor. The Novii+ Interface enables signals collected by the Novii+ Pod to be printed and displayed on a Maternal/Fetal Monitor and sent to a central network, if connected.

    The Novii+ Pod Maternal/Fetal Monitor and its accessories are intended for use by healthcare professionals in a clinical setting.

    Device Description

    The Novii+ Wireless Patch System (Novii+ system) is a battery-powered maternal-fetal monitoring system that measures abdominal fetal heart rate (FHR), abdominal uterine activity (UA), and abdominal maternal heart rate (MHR). The Novii+ Wireless Patch system is designed as an ambulatory device for the monitoring of a pregnant mother. The monitor enables the abdominal electrophysiological signal to be picked up from three different positions on the maternal abdomen using the 5 electrodes on the Novii Patch. The monitor filters the abdominal signals, converts the abdominal electrophysiological data into a digital format and then processes it in real time to extract the fetal heart rate, maternal heart rate and uterine activity. The result of the processing is transmitted via a Bluetooth connection to the Novii+ Interface device which is an accessory to the Novii+ Pod.

    The Novii+ Pod (proposed device) is indicated for use on women who are at 34 weeks and 0/7 days and greater with singleton pregnancies with cephalic fetal presentation.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Novii+ Wireless Patch System, based on the provided FDA 510(k) summary:

    1. Acceptance Criteria and Reported Device Performance

    Parameter (Metric)Acceptance Criteria (Lower limit of 95% two-sided CI)Reported Device PerformanceOutcome
    FHR (PA)>80%83.45%PASS
    MHR (PA)>80%97.26%PASS
    UA (RI)>80%100%PASS
    UA (PPA)>80%84.67%PASS
    FHR Deming Slope0.958 - 1.042 (95% two-sided CI)1.02PASS
    FHR Deming Intercept-10 to 10 BPM (95% two-sided CI)-3.18 BPMPASS
    MHR Deming Slope0.958 - 1.042 (95% two-sided CI)1.01PASS
    MHR Deming Intercept-10 to 10 BPM (95% two-sided CI)-1.18 BPMPASS
    MHR RMSE (Novii vs. GS)< 2 BPM< 2 BPMPASS
    MHR RMSE% (Novii vs. GS)< 3.2 BPM< 3.2 BPMPASS
    Ratio of RMSEs (Novii+ vs. Novii for MHR)< 1.25(Implicitly passed)PASS
    MAE (Novii / Novii+ vs. GS for MHR)< 1.6 BPM< 1.6 BPMPASS
    MAE% (Novii / Novii+ vs. GS for MHR)< 1.6 BPM< 1.6 BPMPASS

    Note:

    • PA: Percent Agreement
    • RI: Relative Interpretability
    • PPA: Positive Percent Agreement
    • RMSE: Root Mean Square Error
    • MAE: Mean Absolute Error
    • GS: Gold Standard

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Study: 30 subjects
      • 14 subjects with threatened pre-term labor (non-laboring)
      • 16 subjects with pre-term labor and delivery
    • Data Provenance: The document does not explicitly state the country of origin. Given that GE HealthCare is the submitter and their address is in Wauwatosa, WI, USA, it is likely the study was conducted in the United States. The study type is prospective clinical study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical study. Instead, the ground truth was established by comparator devices:

    • FHR ground truth: Doppler Fetal Heart Rate device
    • MHR ground truth: SpO2 Maternal Heart Rate device and 12-lead ECG data (Gold Standard)
    • UA ground truth: TOCO Uterine Activity device

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method (e.g., 2+1, 3+1). The primary comparison was against established medical devices (Doppler, SpO2, TOCO, 12-lead ECG) which inherently serve as the reference/ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study's focus was on the Novii+ Wireless Patch System's equivalence to existing, established monitoring devices (Doppler, SpO2, TOCO, ECG) for capturing physiological data, not on how human readers' interpretations improve with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The study implicitly evaluated the standalone performance of the Novii+ Wireless Patch System. It directly compared the measurements generated by the Novii+ algorithms (FHR, UA, MHR) against the measurements from the comparator devices. There is no mention of human-in-the-loop performance evaluation in this specific context.

    7. The Type of Ground Truth Used

    • Fetal Heart Rate (FHR): Doppler Fetal Heart Rate device
    • Maternal Heart Rate (MHR): SpO2 Maternal Heart Rate device, and a 12-lead ECG data (referred to as "MECG Gold Standard (GS)")
    • Uterine Activity (UA): TOCO Uterine Activity device

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set. This particular document focuses on the performance evaluation of the device in a clinical study for premarket notification, not the development or training of its underlying algorithms.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established, as details about the training process are not included in this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212271

    Validate with FDA (Live)

    Device Name
    smart-sync
    Date Cleared
    2022-02-04

    (199 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K140535

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    smart-sync is intended to generate a gating signal for MR scanners (1.5T and 3T) for motion compensation by detecting the heartbeat of a patient using Doppler ultrasound. Thus, smart-sync is intended to improve the image quality by reducing movement artefacts for MR images of the whole body, where motion compensation of the heart movement is necessary.

    smart-sync is intended to be used on pregnant women, where the fetus is examined. smart-sync is also intended to be used for adult MR examinations.

    smart-sync is not intended to monitor any physiological parameters.

    Device Description

    smart-sync is an MRI-compatible Doppler ultrasound system that determines the cardiac cycle of patients and synchronizes it with the MRI imaging, smart-sync can be used for the synchronization of the heartbeat both for fetuses and adults.

    The heartbeat of the patient is detected by positioning an ultrasound sensor above the patient's heart. The ultrasound sensor is fixated with a flexible belt. The reflection of Doppler ultrasound signals is analyzed, and a trigger signal is derived for every heartbeat. smart-sync sends the Doppler ultrasound signal and the trigger signal (synchronization signal) to the MRI scanner utilizing a wireless connection.

    AI/ML Overview

    The provided document is a 510(k) summary for the device "smart-sync." It describes the device's intended use, technological characteristics, and a summary of non-clinical and clinical performance data. However, it does not provide detailed information about specific acceptance criteria or a comprehensive study report with quantitative performance metrics for the AI component of the device.

    Based on the provided text, the "smart-sync" device does not appear to be an AI/ML-driven device in the context of image analysis or diagnostic aid. It is described as a Doppler ultrasound system that generates a gating signal for MR scanners to compensate for motion by detecting a patient's heartbeat. Its primary function is to improve image quality by reducing movement artifacts, not to perform analysis or diagnosis using AI.

    Therefore, many of the requested points regarding AI/ML device testing (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not applicable to the "smart-sync" device as described in this 510(k) summary.

    Here's what can be extracted and inferred from the provided text, addressing the points where information is available or can be logically deduced:


    Device Name: smart-sync
    Product Code: LNH
    Regulation Name: Magnetic resonance diagnostic device

    Acceptance Criteria and Device Performance (as far as can be inferred for this type of device):

    The "acceptance criteria" for a motion-gating device like smart-sync would primarily revolve around its ability to reliably detect the heartbeat and provide an accurate gating signal to the MRI machine, leading to improved image quality. The document refers to "clinical performance" and "effectiveness in improving image quality," but it does not specify quantitative metrics or thresholds for these.

    Since this is a non-AI device focused on mechanical/electrical performance, traditional AI performance metrics (like sensitivity, specificity, AUC) are not directly relevant. The "performance" is about its ability to generate a correct and timely gating signal and the resulting visual improvement in MR images.

    Acceptance Criterion (Inferred for a Gating Device)Reported Device Performance (Inferred/Stated)
    Reliable Heartbeat Detection: Ability to accurately detect and track the cardiac cycle."The heartbeat of the patient is detected by positioning an ultrasound sensor...""Clinical tests and performance tests show that the subject device smart-sync is safe and effective for the intended patient population.""The bench tests show that the subject device smart-sync is effective in providing gating signals for cardiac MR acquisition...""Clinical studies from clinicians at different hospitals across Europe and the US show that the subject device smart-sync is effective in improving the image quality for cardiac MRI examinations..."
    Accurate Gating Signal Generation: Ability to produce a timely and correct trigger signal for the MRI scanner."...a trigger signal is derived for every heartbeat. smart-sync sends the Doppler ultrasound signal and the trigger signal (synchronization signal) to the MRI scanner utilizing a wireless connection.""The bench tests show that the subject device smart-sync is effective in providing gating signals for cardiac MR acquisition, which can be used by the MRI to improve the image quality."
    Improved MR Image Quality: Reduction of motion artifacts in MR images due to cardiac motion."smart-sync is intended to improve the image quality by reducing movement artefacts for MR images of the whole body, where motion compensation of the heart movement is necessary.""Clinical studies from clinicians at different hospitals across Europe and the US show that the subject device smart-sync is effective in improving the image quality for cardiac MRI examinations..."
    Safety and Compatibility: MRI safety, electrical safety, electromagnetic compatibility."MRI Safety Classification: MRI Conditional""RF power levels: 4 W/kg""The bench tests also prove the safety of the subject device smart-sync and the compliance with applicable international safety and performance standards.""Usability tests show that the application of the subject device is safe to use for the intended user group radiographers and radiologists."
    Measurement Range (Heart Rate):Fetal: 60 - 240 BPMAdult: 30-120 BPM

    1. A table of acceptance criteria and the reported device performance:
    See table above. Quantitative acceptance thresholds (e.g., minimum percentage reduction in motion artifact, or specific ECG/gating signal accuracy) are not provided in the document. The performance is described qualitatively as "effective in improving image quality" and "safe."

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not explicitly stated in terms of a numerical count of patients or studies. The text mentions "Clinical studies from clinicians at different hospitals across Europe and the US." This implies a multi-site study, but without specific numbers.
    • Data Provenance: "Europe and the US" (country of origin). The studies appear to be "clinical studies," implying a prospective or at least a clinical trial setting rather than purely retrospective database analysis.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable in the context of AI/ML-driven diagnostic device evaluation.
    • For a device like smart-sync, the "ground truth" would be the actual cardiac rhythm and the resulting improvement (or lack thereof) in image quality as assessed by radiologists in a clinical setting. The document mentions "clinicians at different hospitals" and "radiographers and radiologists" as the intended user group who would assess image quality.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable (as this is not an AI diagnostic device and no AI-specific adjudication of ground truth for classification tasks is described).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. The device is a motion compensation tool, not an AI assistance tool for human readers. No MRMC study comparing human performance with/without AI assistance is mentioned because AI assistance is not its function.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. The device is described as generating a gating signal for the MR scanner, which is a standalone function in this context (i.g., it produces a signal based on its sensor input). However, this is not an "AI algorithm only" performance test in the typical sense of diagnostic AI. The performance is then assessed in a clinical context by human readers evaluating the resulting MR images.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the device's effectiveness appears to be primarily clinical assessment of image quality by clinicians/radiologists based on whether motion artifacts are reduced and images are improved. The accuracy of heartbeat detection would be the underlying "ground truth" for the device's direct function.

    8. The sample size for the training set:

    • Not Applicable. This is not an AI/ML device that requires a "training set" in the context of machine learning model development.

    9. How the ground truth for the training set was established:

    • Not Applicable. As per point 8.

    Summary of Key Findings from the Document regarding "smart-sync":

    • Device Type: MRI-compatible Doppler ultrasound system for cardiac gating.
    • Purpose: To generate a gating signal for MR scanners (1.5T and 3T) to compensate for heart movement, thereby improving image quality by reducing motion artifacts.
    • Patient Groups: Adults and pregnant women (fetus examined from 30th gestational week).
    • Technology: Uses Doppler ultrasound to detect myocardial wall and blood flow motion, deriving a trigger signal. Wireless transmission to MRI scanner.
    • Performance Evaluation:
      • Bench Tests: Demonstrated effectiveness in providing gating signals and compliance with safety standards.
      • Clinical Studies: Conducted at hospitals in Europe and the US, indicating effectiveness in improving image quality for cardiac MRI examinations in the intended patient population.
      • Usability Tests: Confirmed safe use by radiographers and radiologists.
    • Substantial Equivalence: Claimed based on clinical evidence, non-clinical bench testing, electrical safety, electromagnetic compatibility, product verification and validation, showing it does not raise different questions of safety and effectiveness compared to predicate devices (ECG-based gating cables) and a reference device (Doppler-based fetal monitors). The different acquisition method (Doppler ultrasound vs. ECG) was addressed through comparison with a Doppler reference device and analytical comparison showing similar cardiac information can be derived.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1