510(k) Data Aggregation

    K Number
    K170777

    Validate with FDA (Live)

    Date Cleared
    2017-05-01

    (47 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K140378

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bovie J-Plasma Precise Flex Handpiece is used for the delivery of radiofrequency energy and/or a helium gas plasma for electrosurgical cutting, coagulation of soft tissue. It is intended for use with a grasper during minimally invasive surgical procedures.

    The Bovie J-Plasma Precise Flex Handpiece is compatible with Bovie Ultimate Electrosurgical Generators BVX-200H, and BVX-200P.

    Device Description

    The Bovie J-Plasma Precise Flex handpiece, is an electrosurgical accessory that is connected to a Bovie Ultimate Electrosurgical Generator and used to deliver helium gas plasma (J-Plasma) and standard electrosurgical monopolar coagulation during surgical procedures. The accessory has a flexible shaft that can be advanced through a cannula during minimally invasive surgical procedures. The handpiece is activated by the generator footswitch.

    The primary difference between the Bovie J-Plasma Precise Flex handpiece and currently available Bovie J-Plasma handpieces is that the Bovie J-Plasma Precise Flex handpiece has a flexible shaft that can be controlled with graspers, forceps, or grasping instruments during minimally invasive surgical procedures.

    The Bovie J-Plasma Precise Flex handpiece uses the same basic technology and delivers the same J-Plasma helium plasma energy as currently available J-Plasma handpieces (K151325, K152570, K170188) and are compatible only with Bovie Ultimate electrosurgical generators.

    AI/ML Overview

    The provided text describes the Bovie J-Plasma Precise Flex Handpiece, an electrosurgical accessory. However, it does not contain the specific details required to answer all parts of your request regarding acceptance criteria and a study that proves the device meets those criteria. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed performance study results with acceptance criteria.

    Here's an analysis based on the information available:

    1. A table of acceptance criteria and the reported device performance

    The document provides a table of tests performed but does not explicitly state specific acceptance criteria in measurable terms. It describes the objective of each test and the protocols used, but not the pass/fail thresholds.

    TestObjectiveReported Performance (from summary description)
    Electrical VerificationVerify the electrical functionality and safety of the handpieceMeets objective by complying with IEC Electrical Safety Standards
    Bench EquivalencyMeasure effect of generator settings on the plasma stream, plasma stream characterizationDemonstrated equivalent tissue effect as the predicate device J-Plasma handpiece
    Mechanical PerformanceVerify the mechanical functionality of the handpieceDemonstrated ability of the J-Plasma Precise Flex device to be grasped and maneuvered
    Electromagnetic CompatibilityEMC testing for emissions and immunityImplied compliance, as no issues are reported and it's a standard regulatory test for electrosurgical devices

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size for any of the performance tests. It also does not mention the data provenance (e.g., country of origin, retrospective/prospective). The tests listed are primarily bench and electrical verification, which typically do not involve human patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The tests described (electrical, bench, mechanical, EMC) are primarily engineering and technical verifications, not typically requiring expert interpretation of clinical data for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. Given the nature of the tests (bench, electrical, mechanical), an adjudication method like 2+1 or 3+1, which is common in clinical imaging studies for ground truth establishment, would not be applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned in the document. This type of study is typically associated with AI-powered diagnostic devices, not electrosurgical handpieces. The device described is a surgical instrument, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is an electrosurgical handpiece, not an AI algorithm. It's a hand-held tool used by a surgeon.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the bench equivalency testing, the "ground truth" for tissue effect was established by comparison to the predicate J-Plasma handpiece. This would likely involve observable and measurable tissue effects (e.g., coagulation depth, cutting efficiency) under controlled bench conditions. For electrical, mechanical, and EMC tests, the ground truth is simply meeting predefined engineering specifications and regulatory standards.

    8. The sample size for the training set

    This information is not provided. The device is not an AI/ML product, so the concept of a "training set" in the context of machine learning does not apply.

    9. How the ground truth for the training set was established

    This information is not applicable as the device is not an AI/ML product.

    In summary, the provided document is a 510(k) summary for an electrosurgical handpiece, focusing on substantial equivalence. It outlines technical performance tests but lacks the detailed clinical study data, acceptance criteria, ground truth establishment methods, and sample sizes that would be found in a clinical trial report or for AI/ML device submissions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151229

    Validate with FDA (Live)

    Device Name
    ABC D-FLEX PROBE
    Manufacturer
    Date Cleared
    2016-06-02

    (392 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K050161, K140378

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CONMED ABC D-Flex Probes are intended for use with a CONMED ABC Generator for delivery of argon gas and electrosurgical current for non-contact ablation and hemostasis at the operative site. The ABC D-Flex Probes are indicated for use during general, urological, and gynecological laparoscopic surgical procedures when manipulated by CONMED manual laparoscopic graspers or similar manual instruments.

    Device Description

    CONMED ABC® D-Flex Probes™ are monopolar, electrosurgical devices indicated for non-contact ablation and hemostasis of the operative site in laparoscopic surgery. The CONMED ABC® D-Flex Probes™ are composed of a catheter tube, ceramic tip, the D-Flex attachment, insufflation seal, and a shrink tube. The CONMED ABC® D-Flex Probes™ are provided sterile for single use only and are controlled by the appropriate ABC® footswitch used in conjunction with compatible CONMED ABC generators.

    The ABC® D-Flex Probes™ are intended for use with a CONMED ABC® Generator for delivery of argon gas and electrosurgical current for non-contact ablation and hemostasis at the operative site. The probes should be used with a maximum output of 6.5kV at a power setting of 80W.

    AI/ML Overview

    The provided document is a 510(k) summary for the CONMED ABC D-Flex Probes, an electrosurgical device. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain information related to specific acceptance criteria for diagnostic performance (such as sensitivity, specificity, accuracy), nor does it describe a study involving human readers, ground truth establishment by experts, or sample sizes related to AI performance evaluation.

    The "Performance Data" section (Section I) refers to:

    • Compliance with standards: AAMI / ANSI ES60601-1:2005/(R)2012, IEC 60601-2-2:2009, ANSI/AAMI ST67:2011, and ANSI/AAMI/ISO 10993-1:2009 (Biological Evaluation). These are general safety, essential performance, and biocompatibility standards, not diagnostic performance metrics.
    • Benchtop verification testing: Grasping, release, manipulation, axial pull force, and argon flow.
    • Functionality testing: Beam initiation, non-contact ablation, and reliability.
    • Ex-vivo comparison testing: Assessment of thermal spread depth at comparable power settings.

    These tests are designed to demonstrate the physical and functional aspects of the electrosurgical device itself, not an AI or diagnostic algorithm. Therefore, I cannot extract the requested information in the format provided, as it is not present in the document.

    Specifically, the document does not include:

    1. A table of acceptance criteria and reported device performance related to diagnostic accuracy (e.g., sensitivity, specificity).
    2. Sample sizes for test sets or data provenance for an AI study.
    3. Number or qualifications of experts used for establishing ground truth.
    4. Adjudication methods.
    5. Information about a multi-reader multi-case (MRMC) comparative effectiveness study or human reader improvement with AI.
    6. Results from a standalone (algorithm only) performance study.
    7. Type of ground truth used for AI-related evaluation.
    8. Sample size for a training set.
    9. How ground truth for a training set was established.

    This document pertains to the regulatory clearance of a physical medical device (electrosurgical probes) based on established electrical safety, biocompatibility, and functional performance, rather than the evaluation of a diagnostic AI system.