Found 3 results

510(k) Data Aggregation

    K Number
    K170624

    Validate with FDA (Live)

    Device Name
    TalarLift STS
    Date Cleared
    2017-08-11

    (162 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K140053, K142290, K161449

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TalarLift STS arthroereisis implant is designed for use in the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is designed to block the posterior and inferior displacement of the talus, allowing normal subtalar joint motion while blocking excessive pronation and the resulting sequelae. The TalarLift STS is intended for the following pathological conditions resulting from disease, injury, or other trauma:
    • Hypermobile pes valgus;
    • Posterior tibial tendon dysfunction;
    • Severe pronation;
    • Subtalar instability;
    • Hypermobile flexible congenital flat foot.

    Device Description

    The TalarLift STS consists of a threaded implant, designed to be inserted between the posterior and middle facets of the subtalar joint, and corresponding instrumentation to facilitate insertion. The system consists of three configurations, conical, cylindrical, and anatomic. All three devices are manufactured from titanium alloy (Ti-6Al-4V) complying with ASTM F136.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the TalarLift STS device, which is a subtalar spacer. It describes the device, its intended use, and claims substantial equivalence to predicate devices. This document does not describe an AI/ML powered device, nor does it contain information about acceptance criteria or a study proving that an AI/ML device meets those criteria.

    Therefore, I cannot provide the requested information about acceptance criteria or a study that proves an AI/ML device meets those criteria based on the provided text. The document primarily focuses on non-clinical testing for mechanical properties and pyrogenicity of a physical implant. It explicitly states, "No clinical studies were performed."

    Ask a Question

    Ask a specific question about this device

    K Number
    K170038

    Validate with FDA (Live)

    Date Cleared
    2017-08-07

    (214 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Predicate For
    Why did this record match?
    Reference Devices :

    K140053, K142290

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FuzeFix Mini System is intended for use in the stabilization of bone fractures, osteotomies, non-unions, and reconstruction, tendon reattachment, and arthrodesis of the hand, foot, wrist, and ankle. The FuzeFix Mini System is indicated for: scaphoid fractures, metacarpal fractures, phalangeal fractures, ulnar styloid fractures, small ioint fusions, humeral head fractures, intercarpal fractures, tarsal fusions, patellar fractures, interfragmentary ulnar fractures, small hand and wrist bone fractures, forefoot interfragmentary fractures, trapezial fractures, metatarsal fractures, radial head fractures, osteo-chondral fractures, interphalangeal fractures, malleolar fractures, metaphyseal fractures, interfragmentary radius fractures, distal metatarsal osteotomies, midfoot interfragmentary fractures. Not intended for use in the spine.

    The FuzeFix Large System is intended for use in the stabilization and fixation of bone fractures, osteotomies, non-unions, and reconstruction, tendon reattachment, and arthrodesis of the hand, foot, wrist, and ankle. The FuzeFix Large System is indicated for: joint reconstructions, joint fusions, multiple fragment joint fractures, simple metaphyseal fractures of the wrist, ankle, elbow, and shoulder, metatarsal fractures and other fractures of the foot, ligament fixation of the proximal humerus, malleolar fractures, navicular fractures of the calcaneus, fractures of the talus, arthrodesis of the ankle joint, and fractures of the tarsal region. The devices and implants are not intended for use in the spine.

    The FuzeFix Headless FT System is intended for use in the stabilization and fixation of bone fractures, osteotomies, non-unions, and reconstruction, tendon reattachment, and arthrodesis of the hand, foot, wrist, and ankle. The FuzeFix Headless FT System is indicated for: Osteochondral fragments (talar vault, femoral condyle), apical fragments (radial head, patellar rim, navicular, metacarpal/metatarsal), cancellous fragments (talus), Carpal, meta-carpal, and small hand bone, tarsal and metatarsals, phalanges, Intra-articular fractures, ankle, proximal and distal humerus, proximal and distal radius, proximal and distal ulna, osteochondral fixation and fractures, Osteochondritis Dissecans, Fixation of fractures and osteotomies about the knee, Oblique fractures of the fibula, Reconstructive surgeries of the foot, and malleolar fixation. Not intended for use in the spine.

    The FuzeFix Twist Off System is intended for use in the stabilization of bone fractures. osteotomies, and for bone reconstruction of the hand and foot. The FuzeFix Twist Off System is indicated for: fixation of small bone fragments, Weil osteotomy, mono-cortical fixation, osteotomies and fractures fixation in the foot and hand.

    Device Description

    The FuzeFix Screw System consists of headed and headless, partial and full threaded cannulated, self-compressive screws and snap off screws and washers for the management of orthopedic osteotomies and trauma. The screws are self-drilling, self-tapping. The system consists of multiple screw lengths and diameters, and the necessary instruments to facilitate the placement of these implants.

    AI/ML Overview

    The provided document describes the Fusion Orthopedics, LLC FuzeFix Screw System and its substantial equivalence to predicate devices, based on non-clinical testing. It explicitly states that "No clinical studies were performed." Therefore, the information requested about acceptance criteria and studies proving the device meets them, in the context of clinical performance or AI/software, is not largely applicable or available in this document.

    However, I can extract information related to the device's non-clinical performance acceptance criteria and the study that proves it meets them, as documented for regulatory submission.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria and reported device performance in terms of specific numerical values for each test. Instead, it lists the types of non-clinical tests performed and states that the "testing demonstrated that the subject device meets the recommended maximum endotoxin level of 20 EU per device" and "The results of these evaluations indicate that the FuzeFix Screw System is equivalent to predicate devices." This implies that the device met the acceptance criteria for each test by showing equivalence or compliance with established standards.

    Acceptance Criteria (Implied by Test Type)Reported Device Performance
    Torque to Failure (ASTM F543-07 Annex A1)Device performance was found to be "equivalent to predicate devices."
    Axial Pullout Strength (ASTM F543-07 Annex A2)Device performance was found to be "equivalent to predicate devices."
    Driving/Insertion and Removal Torque (ASTM F543-07 Annex A3)Device performance was found to be "equivalent to predicate devices."
    Engineering analysesDevice performance was found to be "equivalent to predicate devices."
    Pyrogenicity (Limulus amebocyte lysate (LAL) assay)Meets the recommended maximum endotoxin level of 20 EU per device.

    The underlying acceptance criteria for "equivalence" would be that the device's mechanical properties (torque, pullout, driving/insertion/removal torque) perform within a comparable range or demonstrate similar failure modes and strengths to the established predicate devices, as typically determined through statistical comparison or by falling within a predefined performance envelope during mechanical testing.

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for each non-clinical test (e.g., number of screws tested for torque to failure). It also does not explicitly mention data provenance in terms of country of origin for these non-clinical tests, and non-clinical tests are inherently "prospective" in the context of generating new data for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The "ground truth" for non-clinical mechanical and biocompatibility testing is established by conforming to recognized industry standards (e.g., ASTM F543-07 for mechanical tests, and general biocompatibility standards for pyrogenicity). It does not involve expert readers or clinical ground truth determination.

    4. Adjudication method for the test set

    This question is not applicable as the tests are non-clinical, objective measurements against established standards, not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was performed, as stated in the document: "No clinical studies were performed." This is not an AI/software device, but a physical medical implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This is a physical medical implant, not an AI algorithm.

    7. The type of ground truth used

    For the non-clinical tests, the "ground truth" is based on:

    • Established Industry Standards: ASTM F543-07 Annex A1, A2, and A3 for mechanical properties (torque to failure, axial pullout, driving/insertion/removal torque).
    • Biocompatibility Standards: Implicit standards leading to the "recommended maximum endotoxin level of 20 EU per device" for pyrogenicity.
    • Predicate Device Performance: The primary comparative "ground truth" for equivalence claims is the performance of the legally marketed predicate devices.

    8. The sample size for the training set

    This question is not applicable. There is no artificial intelligence or machine learning component mentioned that would require a training set.

    9. How the ground truth for the training set was established

    This question is not applicable. There is no artificial intelligence or machine learning component mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142290

    Validate with FDA (Live)

    Date Cleared
    2014-11-24

    (98 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K140053

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    KangLi® pedicle screw spinal system is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-Sl vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 – S1.

    Device Description

    The spinal system consists of pedicle screws, nuts, rods and crosslink etc. It is made of titanium alloy (Ti6Al4V ELI), which meets ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well-known biocompatibility. The proposed devices are provided non-sterile. It is required to be sterilized via autoclave method to reach a SAL of 106 by the hospital prior to surgery. The recommended sterilization method was validated per ISO 17665-1 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the KangLi® Pedicle Screw Spinal System. This is a medical device for spinal fixation, and the document focuses on demonstrating its substantial equivalence to a predicate device, rather than proving a new or novel performance against specific acceptance criteria for a diagnostic or AI-driven medical device.

    Therefore, the information requested in your prompt (e.g., acceptance criteria for device performance like sensitivity/specificity, sample size for test sets, expert qualifications for ground truth, MRMC studies, standalone performance, training set details) is not applicable to this document because it pertains to a mechanical spinal implant, not a diagnostic or AI-assisted device assessed by such metrics.

    The "Non-Clinical Testing" section (item 6) refers to bench tests to ensure the device meets design specifications and is substantially equivalent to the predicate. These tests are mechanical in nature, verifying physical properties and performance under stress, not diagnostic accuracy.

    Here's a breakdown of what is available from the text relevant to mechanical equivalent performance:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Standard)Reported Device Performance (Compliance)
    ASTM F1717-13: Static compression bending testComplies with standard
    ASTM F1717-13: Dynamic compression bending testComplies with standard
    ASTM F1717-13: Static torsion testComplies with standard
    ISO 17665-1 (Sterilization validation)Sterilization method validated per standard (SAL of 10-6)
    ASTM F136 (Material biocompatibility)Made of Ti6Al4V ELI meeting standard

    2. Sample sized used for the test set and the data provenance: Not applicable. These were bench tests, not clinical studies involving patient data or test sets for diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for mechanical testing is established by engineering standards and measurements, not expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This refers to diagnostic study design, not mechanical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For mechanical properties, the "ground truth" is defined by the objective physical measurements and material properties specified in the ASTM and ISO standards (e.g., strength parameters, biocompatibility, sterilization efficacy).

    8. The sample size for the training set: Not applicable. No training set for an AI or diagnostic algorithm was used.

    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document focuses on demonstrating substantial equivalence of a physical medical implant through adherence to established material and mechanical performance standards, not on diagnostic or AI performance metrics.