Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161011

    Validate with FDA (Live)

    Date Cleared
    2016-11-08

    (211 days)

    Product Code
    Regulation Number
    878.4010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K123936, K130329, K140517, K130474, K133963, K141215

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SURGISEAL Topical Skin Adhesive is intended for topical applications only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, simple, thoroughly cleansed, trauma induced lacerations. SURGISEAL may be used in conjunction with, but not in place of, deep dermal sutures

    Device Description

    SURGISEAL® Topical Skin Adhesive is a sterile, professional liquid skin adhesive containing a monomeric (2-octyl cyanoacrylate) formulation and the colorant D&C Violet #2. It incorporates the Teardrop applicator that consists of a thermoformed blister tray with a heat sealed lid with an attached applicator sponge tip. The Stylus and Twist applicators contain the identical adhesive formula and incorporate a plastic ampoule, which houses the adhesive, contained within the double longer plastic sleeves with an attached applicator tip. When SURGISEAL is applied to the skin with these applicators, it polymerizes in minutes.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for topical skin adhesives (Surgiseal Topical Skin Adhesive, Surgiseal Stylus Topical Skin Adhesive, Surgiseal Twist Topical Skin Adhesive). It primarily focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than presenting a study proving the device meets specific acceptance criteria in the context of an AI/ML algorithm or performance study.

    Therefore, the information required to populate most of the sections of your request (e.g., sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details) is not present in this document. This document is related to a medical device approval and not a software or AI algorithm approval.

    However, I can provide information on acceptance criteria that are implicitly or explicitly mentioned for the biocompatibility and sterilization aspects of the device, which are part of its overall safety and effectiveness.

    Here's a breakdown based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
    BiocompatibilityNon-toxic, non-irritating, non-sensitizing, and biocompatible as per ISO-10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing".Previously conducted testing on predicate devices (SURGISEAL Teardrop (K082993), SURGISEAL Stylus (K130474), and SURGISEAL Twist (K130474)) deemed supportive. Both proposed labeled devices are considered to be non-toxic, non-irritating, non-sensitizing, and biocompatible. No additional biocompatibility testing was necessary as the subject devices are identical to those identified in K082993 and K130474.
    SterilizationSterilization by electron beam or gamma irradiation in accordance with ISO 11137-2:2006 (for Surgiseal Topical Skin Adhesive). Sterilization by gamma irradiation for the filled ampoule and ethylene oxide for the finished bulk applicator in accordance with ISO 11137-2:2006, ISO 11135-1:2008, and ISO 11135-2:2008 (for Surgiseal Stylus and Surgiseal Twist).Surgiseal Topical Skin Adhesive is terminally sterilized by electron beam irradiation as well as by gamma irradiation in accordance with ISO 11137-2:2006. Surgiseal Stylus and Surgiseal Twist Topical Skin Adhesives are sterilized with the filled ampoule by gamma irradiation (ISO 11137-2:2006) and the finished bulk applicator by ethylene oxide (ISO 11135-1:2008 and ISO 11135-2:2008).
    Shelf-LifeNot explicitly stated as a numerical criterion in this document, but a typical acceptance criterion would be to maintain product integrity and sterility over the declared shelf life.SURGISEAL has a 2-year shelf life.

    Missing Information (Not provided in the document):

    1. Sample sizes used for the test set and the data provenance: Not applicable in the context of an AI/ML algorithm evaluation; for biocompatibility, sample sizes for in vitro or in vivo tests are part of the ISO standards but not detailed here. The studies were previously conducted on predicate devices. Data provenance would be related to the lab conducting the biocompatibility tests.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for biocompatibility and sterilization is based on established scientific methods and standards.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/ML device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI/ML device.
    6. The type of ground truth used:
      • For Biocompatibility: The "ground truth" is defined by the biological response of cells/tissues to the material, as assessed by standardized tests (e.g., cytotoxicity, irritation, sensitization assays) according to ISO-10993.
      • For Sterilization: The "ground truth" is the sterility assurance level (SAL) achieved, verified by validated sterilization cycles and appropriate microbiological testing, according to ISO 11137 and ISO 11135.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of the Study:

    The document describes a 510(k) submission for new topical skin adhesive products (Surgiseal Topical Skin Adhesive, Surgiseal Stylus Topical Skin Adhesive, Surgiseal Twist Topical Skin Adhesive). The "study" or rather, the evidence presented to ensure safety and effectiveness and demonstrate substantial equivalence, relies primarily on:

    • Identity of Technological Characteristics: The device's formulation (monomeric 2-octyl cyanoacrylate) and mode of action are identical to previously cleared predicate devices.
    • Biocompatibility Testing: Prior biocompatibility testing conducted on the predicate devices (SURGISEAL Teardrop (K082993), SURGISEAL Stylus (K130474), and SURGISEAL Twist (K130474)) in accordance with ISO-10993 was leveraged. This testing confirmed the materials were non-toxic, non-irritating, non-sensitizing, and biocompatible. No new biocompatibility testing was required because the subject devices are identical to the predicates in terms of materials.
    • Sterilization Validation: Validation of sterilization methods (electron beam and gamma irradiation for Surgiseal, and gamma followed by ethylene oxide for Stylus and Twist) was conducted according to international standards (ISO 11137-2:2006, ISO 11135-1:2008, ISO 11135-2:2008). This ensures the devices are supplied sterile.
    • Shelf-Life Determination: A 2-year shelf life was established, implying stability testing was performed, though details of this testing are not provided in this specific document snippet.
    • Microbial Barrier Properties (In vitro): In vitro studies showed SURGISEAL acts as a physical barrier to microbial penetration as long as the adhesive film remains intact. However, the document explicitly states that clinical studies were not conducted to demonstrate microbial barrier properties, and a correlation between microbial barrier properties and a reduction in infection has not been established.

    In essence, the "study" for this 510(k) submission is a demonstration that the new devices are substantially equivalent to already approved predicate devices regarding their materials, design, intended use, and performance characteristics, supported by prior testing data and adherence to recognized standards for biocompatibility and sterilization.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1