Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K142299

    Validate with FDA (Live)

    Date Cleared
    2014-11-25

    (99 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K092521, K132894

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CoRoent® Small Interlock System is a stand-alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with cervical disc disease (DDD) at one level from C2-T1. The CoRoent Small Interlock System is intended for use with autogenous and/or allogeneic bone graft comprised of cancellous, cortical, and/or corticocancellous bone graft to faciliate fusion. The cervical devices are to be used in patients who have had at least six weeks of non-operative treatment.

    Device Description

    The NuVasive CoRoent Small Interlock System is a standalone anterior cervical interbody device consisting of a PEEK (polyetheretherkevtone) implant cage with titanium alloy and tantalum radiographic markers, titanium alloy washers, and three (3) titanium alloy bone fixation screws. The devices are manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026, titanium alloy conforming to ASTM F136, and tantalum conforming to ASTM F560. The implants are available in a variety of sizes to accommodate anatomical conditions. The CoRoent Small Interlock System is a standalone system intended to be used with the bone screws provided, and when used as such requires no additional supplementary fixation systems.

    AI/ML Overview

    I am sorry, but the provided text does not contain the information required to answer your request. The document is an FDA 510(k) clearance letter and summary for a medical device (NuVasive® CoRoent® Small Interlock™ System). It focuses on demonstrating substantial equivalence to predicate devices rather than reporting on a clinical study with detailed acceptance criteria and performance data as you've outlined.

    Specifically, the document states: "A systemic literature analysis of published clinical data for cervical interbody fusion devices similar to the CoRoent Small Interlock System was provided as performance data to support the expanded Indications for Use." and "Additionally, Finite Element Analysis and comparative methods were used to demonstrate that the subject CoRoent Small Interlock System is substantially equivalent to CoRoent Small Interlock System (K102547)."

    This indicates that new clinical studies with defined acceptance criteria and performance metrics (as typically seen in AI/software device submissions) were not performed for this submission. The "performance data" referred to is a summary of existing literature and engineering analysis, not a de novo study with a test set, ground truth experts, or MRMC studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141500

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-09-23

    (109 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K132894

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When Optio-C Anterior Cervical Plate is used with structural allograft it is intended for one-level anterior screw fixation of the cervical spine (C2-T1). The implant has been designed for use with structural allograft to provide stabilization as an adjunct to cervical fusion. Indications for use of the Optio-C Anterior Plate with structural allograff/ autograft include degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion. The Optio-C Anterior Cervical Plate is intended to be used with a structural allograft and with three Optio-C bone screws.

    When Optio-C Anterior Cervical Plate is used with an Optio-C PEEK IBF Spacer it becomes an Optio-C Anterior Cervical Intervertebral Body Fusion Device (IBFD), indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2 to T1. Degenerative disc disease (DDD) is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Optio-C IBFD is comprised of one Optio-C PEEK IBF Spacer, one Optio-C Anterior Cervical Plate and three Optio-C bone screws.

    The Optio-C IBFD is to be used with autograft and implanted via an open, anterior approach in patients who have had six weeks of non-operative treatment.

    Device Description

    The Optio-C Anterior Cervical System consists of two different configurations, an Optio-C Anterior Cervical PEEK Intervertebral Body Fusion Device (IBFD) (K132894, SE on January 16, 2014) and the subject of this submission, the Optio-C Anterior Cervical Plate System. Both configurations share the same cervical plate, bone screws and instrumentation and are used only in anterior surgical procedures and at the same spinal level of C2 to T1.

    The subject cervical plate and bone screws of this submission are designed for use with structural allograft or autograft and is supplied sterile to the end user. The bone screws and instrumentation are supplied non-sterile and are intended to be sterilized by the end user.

    The cervical plate with an Optio-C PEEK IBF Spacer or an allograft/autograft is placed in the cervical disc space, flush with the adjacent vertebral bodies. Bone screws pass through the screw holes of the plate and affix to bone to help prevent implant migration. The implant construct can be implanted in two orientations: Standard orientation, two screws cephalic and one screw caudal or Inverted orientation, one screw cephalic and two screws caudal.

    The Optio-C Anterior Cervical Plate System is a stand-alone system designed to provide structural stability in skeletally mature individuals following discectomy. The implants are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. Screws are inserted through the anterior titanium portion of the implant into adjacent vertebral bodies for bony fixation. The Optio-C Anterior Cervical Plate System is used with structural bone graft material and is manufactured from Titanium with Titanium bone screws that allow intradiscal fixation to the vertebral body. The superior and inferior surfaces of the implant have a pattern of ridges that provide increased stability and help prevent migration of the device.

    The Optio-C Anterior Cervical System is provided for single use only.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Optio-C Anterior Cervical System, which is a medical device. It focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a clinical study with detailed performance metrics. Therefore, many of the requested elements for describing acceptance criteria and a study proving device performance are not applicable or cannot be extracted from this type of document.

    However, I can extract information related to the device description, indications for use, and a summary of performance testing that aims to establish substantial equivalence.

    Here's a breakdown of the information that can be extracted, and where the requested details are not applicable:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria in terms of specific performance targets (e.g., accuracy, sensitivity, specificity, or clinical outcomes) for the Optio-C Anterior Cervical System as would be found for a diagnostic or AI-based device. Instead, it describes performance testing conducted to demonstrate substantial equivalence to predicate devices. The reported device performance is qualitative, indicating acceptability for intended use and substantial equivalence to predicates.

    Acceptance Criteria (General Goal)Reported Device Performance (as demonstrated by testing)
    Acceptable performance for intended use (mechanical stability)Bench Testing (ASTM 1717 for Static Torsion, Static and Dynamic Compression Bending) confirmed performance is acceptable for intended use, similar to predicate devices.
    Acceptable performance for intended use (clinical application)Design Validation / Cadaver Testing confirmed performance is acceptable for intended use and substantial equivalence to predicate(s). Biomechanical cadaveric testing conducted.
    Sterility of sterile implant componentsGamma Sterilization conducted per ISO 11137 and ISO 11737.
    Maintenance of sterile barrier by packaging materialsPackaging Sterility Testing conducted per ISO 11607.
    Sterilization for end-user facility (non-sterile components)Sterilization conducted per ISO 17665 and AAMI TIR12 to ensure equivalence to predicate devices. Dry time and cleaning instructions assessed to be similar.
    Biocompatibility of materialsBiocompatibility Assessment per ISO 10993-1 conducted, confirming materials are biocompatible and similar to predicate devices.
    Technological characteristics similar to predicate devicesDesign, dimensions, intended use, materials, and performance characteristics are similar to predicate devices.
    Implant construct design does not change stabilization fixationThe implant construct design does not change the stabilization fixation of the cervical vertebra found in the predicate devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. The document describes bench testing (mechanical tests) and cadaveric testing (biomechanical tests), but not a patient-based "test set" in the context of clinical data for an AI/diagnostic device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The ground truth for this type of device (spinal implant) is related to mechanical and biocompatibility standards, and biomechanical performance, not expert assessment of clinical images or data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept is relevant for studies involving human interpretation or clinical endpoints, not for the engineering and biological tests described for this spinal implant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a spinal implant, not an AI-assisted diagnostic device or a system for human readers to interpret.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (spinal implant), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance testing of this device is based on:

    • Engineering standards: ASTM 1717 for mechanical properties.
    • Biomechanical principles: Used in cadaveric testing.
    • Biological safety standards: ISO 11137, ISO 11737, ISO 11607, ISO 17665, AAMI TIR12 for sterilization and packaging; ISO 10993-1 for biocompatibility.
    • Comparison to predicate devices: The ultimate "standard" for substantial equivalence.

    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" as it is not an AI/machine learning model.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set."

    Summary of the Study that Proves the Device Meets Acceptance Criteria (as interpreted for this device type):

    The "study" described in the document is a series of non-clinical performance tests designed to demonstrate that the Optio-C Anterior Cervical System is substantially equivalent to legally marketed predicate devices and is safe and effective for its intended use.

    Components of the Study:

    • Bench Testing: Performed according to ASTM 1717 for Static Torsion, Static and Dynamic Compression Bending. This tested the mechanical properties of the cervical plate. The results concluded that its performance is "acceptable for its intended use" and comparable to the predicate devices (Synthes CSLP and Trinica devices).
    • Design Validation / Cadaver Testing: This type of testing assesses the biomechanical performance and confirms the device's acceptability for its intended use, ensuring substantial equivalence to the predicate(s).
    • Biomechanical Cadaveric Testing: Conducted to evaluate the device's performance in a more realistic anatomical setting.
    • Sterilization Validation:
      • Gamma Sterilization: Performed for sterile implant components under international standards ISO 11137 and ISO 11737.
      • Packaging Sterility Testing: Conducted per ISO 11607 to ensure the packaging maintains a sterile barrier.
      • End-user Sterilization: Validated for non-sterile components (instrumentation and bone screws) under ISO 17665 and AAMI TIR12 to ensure equivalence to predicate devices, including dry time and cleaning instructions.
    • Biocompatibility Assessment: Conducted per ISO 10993-1 to ensure the materials used in the Optio-C System are biocompatible and similar to those of the predicate devices.

    Conclusion of the Study:

    The document explicitly states: "The test results conclude the Optio-C™ Anterior Cervical Plate System to be substantially equivalent to the predicate devices listed above." It further notes that "The Optio-C Anterior Cervical Plate System performance, intended use and fundamental scientific technology remain unchanged from the predicate devices. The implant construct design does not change the stabilization fixation of the cervical vertebra found in the predicate devices. The Optio-C Anterior Cervical Plate is substantially equivalent to the predicate devices, a traditional anterior cervical plate."

    In essence, the "acceptance criteria" for this 510(k) submission are met by demonstrating that the device performs mechanically, biologically, and functionally in a manner that is substantially equivalent to already approved predicate devices for the same indications for use.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1