Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K220348

    Validate with FDA (Live)

    Date Cleared
    2022-07-14

    (157 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K162309 StealthstationS8 System Platforms, K172548 Wiltrom Spinal Fixation System, K132884 PATHFINDER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anatase Spine Surgery Navigation System is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery when reference to a rigid anatomical structure, such as the vertebra, can be identified relative to a patient's fluoroscopic or CT imagery. It is intended as a planning and intraoperative guidance system to enable open or percutaneous image guided surgery by means of registering intraoperative 2D fluoroscopic projections to pre-operative 3D CT imagery.

    Example procedures include but are not limited to:

    Posterior-approach spinal implant procedures, such as pedicle screw placement, within the lumbar region.

    Device Description

    The Anatase Spine Surgery Navigation System, also known as an Image Guided System, is comprised of a platform, clinical software, surgical instruments, and a referencing system. The system uses optical tracking technology to track the position of instruments in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient. The system helps guide surgeons during spine procedures such as spinal fusion. The software functionality in terms of its feature sets is categorized as imaging modalities, registration, planning, interfaces with medical devices, and views.

    The modified Anatase Spine Surgery Navigation System, the subject of these 510(k) applications, introduces software, hardware and instruments modifications to the original Surgery Navigation System cleared in 510(k) K180523.

    AI/ML Overview

    The Anatase Spine Surgery Navigation System, Model number: SNS-Spine2-S and SNS-Spine2-V, is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery.

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria/StandardReported Device Performance
    SterilizationISO 17665-1:2006Moist heat sterilization of reusable accessories validated.
    Repeated ReprocessingISO 11737-2: 2019Reliability of reusable instruments validated.
    BiocompatibilityFDA guidance for ISO 10993-1 (June 16, 2016), ISO 10993-1:2009Accessories in contact with patient evaluated.
    SoftwareFDA guidance for software in medical devices (May 11, 2005)Software verified and validated.
    Electrical SafetyANSI/AAMI ES60601-1:2005/(R)2012, A1:2012, C1:2009/(R)2012, A2:2010/(R)2012Complied with requirements.
    Electromagnetic CompatibilityIEC 60601-1-2:2014Complied with requirements.
    UsabilityANSI/AAMI HE75:2009/(R)2013, IEC 62366-1:2015, IEC 60601-1-6:2010 + A1:2013System usability validated.
    AccuracyASTM F2554-18Positional accuracy evaluated. (Specific results not given in summary)
    Risk AssessmentISO 14971:2007Effectiveness of risk control measures verified.
    Design VerificationNot explicitly stated, but "all design input requirements"Design output fulfills all design input requirements.

    2. Sample size used for the test set and the data provenance:

    The provided document does not specify sample sizes for any test sets nor the data provenance (e.g., country of origin, retrospective/prospective). The studies are non-clinical, meaning they did not involve patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document as these were non-clinical tests.

    4. Adjudication method for the test set:

    This information is not provided in the document as these were non-clinical tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No clinical testing, including MRMC studies, was conducted. The document explicitly states: "No clinical testing has been conducted."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The provided information focuses on the entire system's performance, which is an image-guided navigation system that would inherently involve human interaction (a surgeon). While software verification and validation were performed, the document does not distinguish between human-in-the-loop and algorithm-only performance for a standalone assessment in a manner that would typically be seen for an AI diagnostic device. The "Accuracy" test implies an assessment of the system's ability to track and display positions, which is a standalone performance metric for the navigation component, but it's not described as an AI-specific algorithm performance.

    7. The type of ground truth used:

    For the accuracy testing, the ground truth would likely be established through precise physical measurements to determine the true positional accuracy of the system against a known standard. However, the document does not specify the exact methodology for establishing the ground truth beyond referencing ASTM F2554-18. For other tests like electrical safety, EMC, and sterilization, the "ground truth" is defined by compliance with the referenced standards.

    8. The sample size for the training set:

    As this is a navigation system and not explicitly an AI diagnostic device in the context of machine learning model training, the concept of a "training set" in that sense is not directly applicable or discussed in the document. Software verification and validation were performed, but details on data used for these processes are not provided.

    9. How the ground truth for the training set was established:

    Same as above, the concept of a "training set" with established ground truth as typically understood in AI/machine learning is not applicable here. Software verification and validation would use various testing methods to ensure the software performs as designed and meets requirements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192133

    Validate with FDA (Live)

    Date Cleared
    2019-10-29

    (83 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K133444, K132884, K151974, K183550

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zimmer Biomet Universal Navigation System is indicated for use during the preparation and insertion of Zimmer Biomet screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The universal adaptors are specifically designed for use with the Zimmer Biomet ROSA One System, which is indicated for providing spatial positioning and orientation of instrument holders or tool guides based upon an intraoperative plan developed with three dimensional imaging software provided that the required fiducial markers and rigid patient anatomy can be identified on 3D CT scans. The ROSA One Spine System is intended for the placement of pedicle screws in vertebrae with a posterior approach in the thoracolumbar region.

    Device Description

    The Zimmer Biomet Universal Navigation System includes Universal Adaptors that are intended to be used with instrumentation from Vitality Spinal Fixation System (including Vitality MIS), Pathfinder NXT Minimally Invasive Pedicle Screw System, and Polaris Spinal Fixation System (including Cypher MIS Screw System) to facilitate preparation and insertion of Zimmer Biomet screws using navigation.

    AI/ML Overview

    The provided text describes the Zimmer Biomet Universal Navigation System and its 510(k) clearance. However, it does not contain specific details about acceptance criteria, a detailed study proving performance against those criteria, or the methodology (sample size, expert qualifications, ground truth, MRMC study, standalone performance) typically associated with a robust clinical or technical validation study for an AI/ML medical device.

    The closest information provided is:

    • "Accuracy testing of the Zimmer Biomet Universal Navigation System Instruments was performed in accordance ASTM F2554-18 Standard Practice For Measurement of Positional Accuracy of Computer Assisted Surgical Systems."
    • "Clinical data was not needed for the Zimmer Biomet Universal Navigation System."

    This indicates that the validation was focused on the system's positional accuracy and relied on a standardized test method (ASTM F2554-18), rather than a clinical study evaluating diagnostic or treatment outcomes with human readers.

    Therefore, many of the requested details cannot be extracted from this document.

    Here's a breakdown of what can and cannot be answered from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified directly in the text. It states that "Validation activities are used to confirm the instruments meet performance requirements," but the specific numerical or qualitative targets are not detailed."Accuracy testing of the Zimmer Biomet Universal Navigation System Instruments was performed in accordance ASTM F2554-18 Standard Practice For Measurement of Positional Accuracy of Computer Assisted Surgical Systems." Specific performance metrics (e.g., target registration error, root mean square error) are not reported in this document.
    Conclusion: The document states that performance requirements were met through accuracy testing following a standard, but it does not provide the specific acceptance criteria or the numerical results of that testing.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not specified. The document mentions "accuracy testing" but does not detail the number of trials, cases, or specific measurements performed.
    • Data provenance: Not applicable in the context of clinical data, as it states "Clinical data was not needed." For the accuracy testing, the "data" would be the measurements taken during the experimental procedure according to ASTM F2554-18. The origin of this experimental setup is not stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. The validation was a technical accuracy test, not a clinical study requiring expert interpretation of medical images or outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. There was no clinical test set requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states, "Clinical data was not needed for the Zimmer Biomet Universal Navigation System." Therefore, an MRMC study and AI assistance comparison were not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, implicitly. The "Accuracy testing of the Zimmer Biomet Universal Navigation System Instruments" performed per ASTM F2554-18 is a standalone technical performance test of the device's ability to measure spatial positioning accurately, independent of human interpretation of clinical outcomes. However, the specific metrics of this standalone performance are not reported.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the accuracy testing, the ground truth would be based on precisely established physical reference points and measurements in a controlled environment, as defined by the ASTM F2554-18 standard for measuring the positional accuracy of computer-assisted surgical systems. It's a "physical ground truth" related to spatial measurement, not a biological or clinical ground truth.

    8. The sample size for the training set

    • Not applicable. This device is a navigation system, not an AI/ML algorithm that requires a training set in the conventional sense (e.g., for image classification or risk prediction). Its "training" would be its engineering design and calibration.

    9. How the ground truth for the training set was established

    • Not applicable, as there isn't a "training set" for an AI/ML algorithm in this context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190556

    Validate with FDA (Live)

    Date Cleared
    2019-10-24

    (233 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K132884, K151974, K171907, K172275

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zimmer Biomet Universal Navigation System is indicated for use during the preparation and insertion of Zimmer Biomet screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These adaptors are specifically designed for use with the Medtronic StealthStation System which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone or vertebra can be identified relative to an acquired image: CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

    Device Description

    The Zimmer Biomet Universal Navigation System includes Universal Adaptors that are intended to be used with instrumentation from Vitality Spinal Fixation System (including Vital Power Instrument System), Pathfinder NXT Minimally Invasive Pedicle Screw System, and Polaris Spinal Fixation System (including Cypher MIS Screw System) to facilitate preparation and insertion of Zimmer Biomet screws using navigation.

    AI/ML Overview

    Due to the nature of the provided document, which is a 510(k) summary for a medical device and not a study detailing specific acceptance criteria and performance data in the typical sense of AI/ML device testing, many of the requested fields cannot be directly extracted or are not applicable.

    The document describes the Zimmer Biomet Universal Navigation System, a stereotaxic instrument for spinal surgery. The performance data provided focuses on accuracy testing of the instruments in accordance with a particular ASTM standard, rather than clinical performance metrics often associated with AI/ML systems (e.g., sensitivity, specificity, AUC).

    Here's a breakdown based on the available information:

    Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for performance in the format of a table with numerical targets and reported values. Instead, it refers to validation activities and accuracy testing. The primary performance claim is related to accuracy as measured by a standard practice.

    Implied Acceptance Criterion: The device should meet the performance requirements for accuracy as defined by ASTM F24554-18.

    Here's a table based on the summary of performance data presented in the document:

    Acceptance Criteria CategorySpecific Metric / StudyReported Device Performance
    AccuracyPositional Accuracy according to ASTM F24554-18 Standard Practice For Measurement of Positional Accuracy of Computer Assisted Surgical Systems."Validation activities are used to confirm the instruments meet performance requirements, providing assurance of device performance for their intended use, similar to predicate systems." Specific numerical results are NOT provided in this 510(k) summary.

    Missing Information: Crucially, the specific numerical results of the accuracy testing are not included in this 510(k) summary. The document states that "validation activities ... confirm the instruments meet performance requirements," implying that the acceptance criteria were met, but the criteria themselves (e.g., "accuracy must be within X mm") and the measured performance (e.g., "accuracy achieved Y mm") are not detailed here.


    Additional Requested Information (Based on Document Content)

    • Sample sized used for the test set and the data provenance: Not explicitly stated. The testing mentioned in the document is "Accuracy testing of the Zimmer Biomet Universal Navigation System Instruments" in accordance with an ASTM standard. This likely refers to in-vitro/bench testing, not a clinical test set with patient data. Therefore, concepts like "country of origin of the data" or "retrospective/prospective" are not applicable here.

    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. This device is not an AI/ML diagnostic or predictive tool that relies on expert interpretation to establish ground truth in a test set. The validation is for the accuracy of navigation instruments, which is typically measured against physical standards or known positions.

    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for the reasons mentioned above.

    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic or interpretation device that would involve human readers or an MRMC study. It's a navigation system designed to aid surgeons in precisely locating anatomical structures.

    • If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The "accuracy testing" mentioned is likely a standalone performance assessment of the navigation system's instruments, independent of a human surgeon's interaction during a live procedure, but it's not an "AI algorithm" in the typical sense. It is testing the instrument's ability to accurately track or guide.

    • The type of ground truth used (expert consensus, pathology, outcomes data, etc): For positional accuracy testing of a navigation system, the ground truth would typically be established by highly precise measurement devices or known physical reference points in a controlled laboratory setting, not expert consensus or pathology.

    • The sample size for the training set: Not applicable. This device is not described as an AI/ML device that undergoes a "training" phase.

    • How the ground truth for the training set was established: Not applicable for the reasons mentioned above.

    Conclusion based on provided text:

    The provided document is a 510(k) summary for a navigation system, not an AI/ML diagnostic or predictive device. As such, many of the questions related to AI/ML device testing methodologies (like training sets, expert adjudication of ground truth, MRMC studies) are not applicable to the information presented. The key performance aspect is the accuracy of the navigation instruments, which was validated according to a recognized ASTM standard, though specific numerical results of this validation are not included in the summary. The FDA deemed this sufficient for substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1