Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230156

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-06-30

    (162 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K132673, K103386

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medtronic guide catheter is designed to provide a pathway through which therapeutic devices are introduced. The guide catheter is intended to be used in the coronary or peripheral vascular system.

    Device Description

    Launcher Guide Catheters are comprised of a cylindrical catheter having a proximal and distal end. A single lumen extends from the proximal to the distal end of the catheter. The distal end is the section of the catheter towards the tip and the proximal end is towards the hub and strain relief. The guide catheter is comprised of a luer hub, strain relief, shaft, segments, tungsten marker band, sleeve, and soft tip. The catheter shaft is made of three distinct layers. The guide catheter is available in various lengths ranging from 45 cm to 130 cm and four French sizes [5F, 6F, 7F, and 8F]. Launcher Guide Catheters have a lubricious inner lumen that runs from the proximal to the distal end of the catheter via a lubricious liner material or coating on the inner lumen.

    AI/ML Overview

    The provided document is a 510(k) summary for the Medtronic Launcher Guide Catheter. It outlines the product's description, intended use, and substantial equivalence to a predicate device based on non-clinical performance and biocompatibility testing.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    | Acceptance Criteria Category | Specific Tests Performed | Reported Device Performance/Conclusion |
    |------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
    | Biocompatibility Assessment | 1) Cytotoxicity2) Sensitization3) Intracutaneous Reactivity4) Material-Mediated Pyrogenicity5) Hemocompatibility6) Acute Systemic Toxicity | Performed in accordance with FDA Guidance "Use of International Standard ISO 10993-1" and demonstrated no safety or effectiveness concerns. |
    | Product Performance Evaluation (Dimensional & Performance) | 1) Inner Diameters (shaft, tip, & hub)2) Outer Diameters (shaft, segment, soft tip sleeve, & segment/shaft overlap)3) Total Segment Length4) Effective Length5) Exposed Tip Length6) Luer Connector Performance*7) Leak Resistance*8) Draw-through stiffness of primary and secondary curve9) Shaft bending stiffness & bending kink resistance (Body temperature and room temperature)10) Arch bending stiffness & bending kink resistance (Body temperature and room temperature)11) Shaft & distal segment crush resistance12) Torsional Stiffness, Rotational Kink Angle, and Rotations to Separation13) Curve retention after simulated seating14) Peak shaft & segment tensile load15) Peak soft tip tensile load16) Soft tip stiffness17) Peak hub/shaft tensile load*18) Internal lubricity | Necessary tests performed met the acceptance criteria and demonstrated that there are no safety or effectiveness concerns. The performance testing, along with biocompatibility testing, demonstrated substantial equivalence to the predicate device. Specific standards used for some tests are noted: ISO 594-1, ISO 594-2, EN ISO 80369-7 for Luer Connector Performance; ISO 10555-1 for Leak Resistance and Peak Hub/Shaft Tensile Load. |

    Note: Tests marked with an asterisk () were specifically highlighted as being performed on the subject device to demonstrate substantial equivalence for modifications in this premarket notification.*

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the non-clinical and biocompatibility tests. It broadly mentions "Summary of Nonclinical Data/Performance Testing" and "Product Performance Evaluation". Given the nature of these tests for medical devices, they would typically involve a statistically relevant number of samples to ensure robust results, but the exact numbers are not provided in this summary.

    The data provenance is from non-clinical (laboratory) testing conducted by the manufacturer, Medtronic Vascular. No information regarding country of origin of the data or whether it was retrospective or prospective is mentioned (as these terms usually apply to clinical studies).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this submission. The "ground truth" concept usually applies to clinical studies where expert consensus or pathological findings are used to establish a definitive diagnosis or outcome. For non-clinical performance and biocompatibility testing, the "ground truth" is established by adherence to recognized international standards and engineering principles, with results evaluated against predefined acceptance criteria derived from these standards or predicate device performance. Experts involved would be engineers, scientists, and toxicologists specializing in medical device testing and regulatory compliance, but their specific numbers or detailed qualifications are not typically included in a 510(k) summary (beyond general statements about adherence to guidance documents like ISO 10993-1).

    4. Adjudication Method for the Test Set

    This information is not applicable to this submission. Adjudication methods (like 2+1, 3+1) are typically used in clinical trials where multiple experts assess cases and their disagreements need to be resolved to establish a ground truth. For the non-clinical tests described, the "adjudication" is inherent in the test methodology and adherence to international standards and acceptance criteria. Any deviations or failures would be investigated and resolved according to quality system procedures.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is a physical medical device (catheter), not an AI-powered diagnostic or imaging tool that would involve human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for the non-clinical performance and biocompatibility testing is established by:

    • Adherence to International/Recognized Standards: Such as ISO 594-1, ISO 594-2, EN ISO 80369-7, ISO 10555-1, and FDA Guidance "Use of International Standard ISO 10993-1".
    • Predicate Device Performance: The subject device's performance is compared to the predicate device, aiming to demonstrate substantial equivalence by showing that any differences in characteristics do not raise different questions of safety and effectiveness.
    • Engineering Specifications and Design Requirements: The tests verify that the device meets its own engineering specifications for parameters like dimensions, strength, flexibility, and resistance to various stresses.

    8. The Sample Size for the Training Set

    Not Applicable. There is no "training set" in the context of this device. Training sets are used in machine learning and AI model development.

    9. How the Ground Truth for the Training Set was Established

    Not Applicable. As there is no training set for this physical medical device, there is no ground truth to be established for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1