Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180461

    Validate with FDA (Live)

    Date Cleared
    2018-05-09

    (78 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K131690, K140254, K161999

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in gynecology, obstetric, abdominal, pediatric, small parts (breast, testes, thyroid, etc.), neonatal cephalic, transcranial, cardiac, transvaginal, peripheral vascular, urology, orthopedic, and musculoskeletal (conventional and superficial) exams.

    Device Description

    The Navi e/ Navi s/Navi X Diagnostic Ultrasound System is a touch screen controlled ultrasonic system. Its function is to acquire and display ultrasound data in B-Mode, M-Mode, Color-Mode, Power (Dirpower)-Mode, PW-Mode, CW-mode, and the combined mode. The system can also measure anatomical structures and offer software analysis packages performance to provide information based on which the competent health care professionals can make the diagnosis. The Navi e/ Navi x/Navi X Diagnostic Ultrasound System consists of the main unit named Navi series, ultrasound probes, power adapter, connecting cable, probe extender, needle-guided bracket, batteries, mobile trolley, ECG module and batteries. Three models for the main units are included in this submission, that is Navi s, Navi e and Navi X. Eight different models of probes are available for the Navi series.

    AI/ML Overview

    The provided text describes the Navi e/Navi s/Navi X Diagnostic Ultrasound System and its substantial equivalence to predicate devices, but it does not contain detailed acceptance criteria or a specific study proving the device meets those criteria, especially not in the context of AI/ML performance.

    Instead, the document outlines various non-clinical tests conducted to ensure the device's safety and performance in a general sense, largely by comparing it to predicate ultrasound systems.

    Here's an analysis of what information is available and what is missing based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide a specific table of acceptance criteria with corresponding performance metrics for the device. It generally states that "all of the tested parameters met the predefined acceptance criteria" for performance testing, but the criteria themselves are not enumerated.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable as no specific test set (e.g., imaging dataset for AI/ML evaluation) is mentioned. The performance testing refers to general ultrasound system performance.
    • Data Provenance: Not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as no ground truth establishment for a test set (in the context of AI/ML or specific clinical performance evaluation) is mentioned.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable, as no clinical test set with ground truth requiring adjudication is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No, an MRMC comparative effectiveness study was not done.
    • Effect Size: Not applicable, as there's no mention of AI assistance for human readers or a study evaluating it.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No, a standalone algorithm performance study was not done. The document does not describe any specific AI/ML algorithms that would require such a study. The software verification and validation are for the overall system software, deemed "moderate" level of concern.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Not applicable. The document discusses regulatory compliance, biocompatibility, electrical safety, and general performance of the ultrasound system, not the performance of an AI/ML component against a clinical ground truth.

    8. The sample size for the training set:

    Not applicable, as no AI/ML training set is mentioned or described.

    9. How the ground truth for the training set was established:

    Not applicable, as no AI/ML training set or its ground truth establishment is mentioned.


    Summary of what the document does include regarding performance:

    The document primarily focuses on demonstrating substantial equivalence to predicate ultrasound devices through:

    • Biocompatibility testing: According to ISO 10993-1, ISO 10993-5, and ISO 10993-10, for cytotoxicity, sensitization, and skin irritation of probes and glue.
    • Electrical safety and electromagnetic compatibility (EMC) testing: Compliance with IEC 60601-1:2012 and IEC 60601-1-2:2007.
    • Performance testing: According to IEC 60601-2-37:2007 for ultrasonic medical diagnostic and monitoring equipment. It also mentions evaluation of "clinic measurement accuracy and system sensitivity" where "all of the tested parameters met the predefined acceptance criteria."
    • Acoustic output measurement and real-time display: Compliance with NEMA UD 2:2004 and NEMA UD 3:2004.
    • Software Verification and Validation Testing: Conducted per FDA guidance for "moderate" level of concern software.

    Conclusion:

    This submission for the Navi e/Navi s/Navi X Diagnostic Ultrasound System is for a conventional ultrasound imaging system. It does not describe any AI/ML components or their performance characteristics. Therefore, the specific information requested about AI acceptance criteria, clinical study designs (MRMC, standalone), ground truth, and training data is not present in the provided text. The "performance testing" mentioned refers to the general technical and functional performance of the ultrasound system itself, not the diagnostic performance of an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1