Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200986

    Validate with FDA (Live)

    Date Cleared
    2021-08-02

    (474 days)

    Product Code
    Regulation Number
    862.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092686, K131126

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OPTI® B-Lac cassette is intended to be used for the in vitro measurement of pH, PO2, total hemoglobin (tHb), and % Saturated O2 in sodium heparinized venous blood samples on the OPTI CCA-TS and OPTI CCA-TS2 platform in a clinical laboratory location.
    · Measurements of blood gases (pCO2, pO2) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.

    • · Total hemoglobin (tHb) measurement is used to determine the hemoglobin content of human blood.
      · Oxygen saturation (SO2) measurement is used to determine the oxygen capacity of the hemoglobin.
    Device Description

    The OPTI CCA-TS/TS2 are portable devices, microprocessor-based instrument using optical fluorescence for the measurement blood gases, electrolytes and enzymes. The OPTI CCA-TS/TS2 utilize a color, graphical touch screen user interface. A disposable, single-use cassette contains all of the elements needed for calibration, sample measurement, and waste containment. Specific calibration from the cassette is scanned into the analyzer by holding the cassette package in front of the bar code scanner. The cassette is then placed into the measurement chamber. The analyzer warms the cassette to 37.0±0.1°C and performs a calibration verification. When calibration is verified, the analyzer aspirates the blood sample into the cassette and across the optode sensors. Fluorescence emission is then measured after equilibrating with the blood sample. After a single measurement, the cassette containing the blood sample is removed from the analyzer and discarded. The analyzer contains no reagents, blood, or waste. The B-Lac cassette is a disposable, single use cassette that contains four (4) sensors for in vitro quantitative measurements of PO2, PCO2, pH. There is an additional laser based measurement of total hemoglobin (tHb) and SO2. The B-Lac cassette is sealed in a foil pouch along with a desiccant and is marked with a barcode label that includes a lot identification number, calibration information, and expiration date.

    AI/ML Overview

    The provided document describes the OPTI® B-Lac Cassette for in vitro measurement of blood gases and related parameters. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the performance of the redesigned B-Lac cassette was determined to meet the performance claims made in the original B-Lac cassette submission (K093280) for all analytes. However, the specific quantitative acceptance criteria from K093280 are not detailed in this document. The reported device performance is indicated by the statement that the device "meets the performance claims" or "was demonstrated to meet the performance claims."

    Here's a table based on the information provided, noting where specific numerical criteria are not available in this document:

    ParameterAcceptance Criteria (from K093280 - not detailed in this document)Reported Device Performance
    pH Precision(Not detailed, but based on CLIA 1988 specifications)Meets performance claims from K093280
    PCO2 Precision(Not detailed, but based on CLIA 1988 specifications)Meets performance claims from K093280
    PO2 Precision(Not detailed, but based on CLIA 1988 specifications)Meets performance claims from K093280
    tHb Precision(Not detailed, but based on CLIA 1988 specifications)Meets performance claims from K093280
    SO2 Precision(Not detailed, but based on CLIA 1988 specifications)Meets performance claims from K093280
    pH Method Comparison(Not detailed, but comparison with ABL90 Flex)Meets performance claims from K093280
    PCO2 Method Comparison(Not detailed, but comparison with gravimetric target/ABL90 Flex)Meets performance claims from K093280
    PO2 Method Comparison(Not detailed, but comparison with gravimetric target/ABL90 Flex)Meets performance claims from K093280
    tHb Method Comparison(Not detailed, but comparison with ABL90 Flex)Meets performance claims from K093280
    SO2 Method Comparison(Not detailed, but comparison with E series cassettes)Meets performance claims from K093280
    Interferents (PCO2)(Not detailed)No interferents found
    Interferents (PO2)(Not detailed)Only one interferent found
    Interferents (pH)(Not detailed)Only one interferent found
    Interferents (tHb)(Not detailed)Several interferents found
    Interferents (SO2)(Not detailed)Several interferents found
    Altitude Performance(Not detailed, but assessed for linearity and bias at critical levels)Demonstrated to meet performance claims from K093280
    Shelf Life (B-Lac cassette)(Not detailed, but expected to be 12 months)Demonstrated to meet performance claims for up to 6 months (with real-time testing continuing for 13 months)

    2. Sample Size Used for the Test Set and Data Provenance

    • 20-Day Precision:

      • Sample Size: Paired samples run twice daily over 20 days. Three lots of B-Lac cassettes were used, and three levels of aqueous quality control solution.
      • Data Provenance: In-house (presumably US-based, as the company is in Georgia, USA). Retrospective data analysis of prospective testing.
    • Within-Run Precision:

      • Sample Size: Multiple repeats using three lots of B-Lac cassettes, three levels of aqueous quality controls, and whole blood manipulated to 3 different levels.
      • Data Provenance: In-house. Retrospective data analysis of prospective testing.
    • Method Comparison (in-house):

      • Sample Size: Whole blood samples tonometered to different levels using different O2/CO2 gas mixtures to generate test levels for pH, PCO2, PO2, and SO2. Samples manipulated for tHb. The exact numerical count of samples or measurements is not specified.
      • Data Provenance: In-house. Retrospective data analysis of prospective testing.
    • Method Comparison (Altitude):

      • Sample Size: Whole blood samples were tonometered to obtain samples that span the range for PCO2, PO2, and pH, and spiked or diluted for tHb. Aqueous solutions were measured. Number of samples/measurements not specified, but done at 4 distinct altitude sites (75 ft, 1080 ft, 5560 ft, 10151 ft).
      • Data Provenance: Conducted in the USA (Maine, Georgia, North Carolina, Colorado). Retrospective data analysis of prospective testing.
    • Interference Testing:

      • Sample Size: 16 interferents tested for each analyte (PCO2, PO2, pH, tHb, SO2). The number of samples per interferent is not specified.
      • Data Provenance: In-house. Retrospective data analysis of prospective testing.
    • Stability Testing:

      • Sample Size: Three lots of B-Lac cassettes were tested. One lot was subjected to two cycles of elevated and frozen temperatures.
      • Data Provenance: In-house. Retrospective data analysis of prospective testing for the initial 6 months, with real-time testing ongoing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document describes a medical device for in vitro diagnostic measurements. The "ground truth" for the test set is established by:

    • Reference Methods/Predicate Devices:

      • Gravimetric target for PCO2 and PO2 (based on gas concentration).
      • Predicate device ABL90 Flex for pH, PCO2, PO2, and tHb.
      • E series cassettes on the OPTI CCA-TS/TS2 for SO2.
    • No human "experts" (e.g., radiologists) were involved in establishing the ground truth in the way
      this question implies for imaging or subjective interpretation devices. The ground truth is
      based on established analytical methods and reference instruments.

    4. Adjudication Method for the Test Set

    Not applicable. This device provides quantitative measurements, and ground truth is established by reference methods/instruments, not through expert consensus requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This document does not mention any MRMC comparative effectiveness study where human readers improve with or without AI assistance. This device is an in vitro diagnostic instrument, not an AI-assisted diagnostic tool for human interpretation.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, the studies described are of standalone performance of the device (OPTI® B-Lac Cassette on OPTI CCA-TS/TS2 platforms). The measurements are performed automatically by the instrument and its embedded algorithms. There is no human-in-the-loop performance described or implied for the measurement process itself, although clinical interpretation of the results by healthcare professionals would follow.

    7. Type of Ground Truth Used

    The ground truth used for performance evaluation includes:

    • Gravimetric targets: For PCO2 and PO2 (based on gas concentration for tonometered samples).
    • Predicate device measurements: Measurements from the Radiometer ABL90 Flex for pH, PCO2, PO2, tHb, and from the OPTI CCA TS2 E-Series Cassettes for SO2.
    • Aqueous quality control solutions and manipulated whole blood samples: Used for precision and linearity studies, where the expected values are known or derived from previous characterization.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning. The device utilizes "new algorithms utilized to calculate concentrations for these sensors" (specifically for PCO2) and has updated software. However, the data sets described are for performance verification and validation, not for training a new algorithm from scratch in the classical AI sense. If there was an internal dataset used for algorithm development or "training" (e.g., to derive calibration curves or correction factors), that information is not provided. The described studies are primarily for demonstrating post-development performance.

    9. How the Ground Truth for the Training Set was Established

    As no specific "training set" is described for algorithm development, the method of establishing ground truth for such a set is not provided. The document focuses on the verification and validation of the device's performance against established clinical and analytical standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1