Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K232977

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-11-21

    (61 days)

    Product Code
    Regulation Number
    878.5400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K131113

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BIOPHOTAS Celluma CONTOUR is indicated for use as a non-invasive dermatological aesthetic treatment for the reduction of circumference of hips, waist, and thighs.

    The BIOPHOTAS Celluma CONTOUR is intended to deliver heat in the near infra-red spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis, and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

    The BioPhotas Celluma CONTOUR is intended to emit energy in the visible and infrared region of the spectrum for use in the treatment of full-face wrinkles.

    Device Description

    The BioPhotas Celluma CONTOUR is a Light emitting diode (LED) system that applies 640nm light for the purpose of reducing the circumference of the hips, waist, and thighs for aesthetic benefit, 640nm and 880nm light for the purpose of temporarily elevating skin temperature and 640nm and 880nm light for the purpose of treating full face wrinkles.

    The BioPhotas Celluma CONTOUR comprises of a flexible, shape-taking frame upon which is mounted an array of LEDs, this allows the device to be contoured to the treatment area. The LEDs are embedded within a biocompatible foam covering that holds a transparent polycarbonate cover recessed within it. The biocompatibility nature of allows the device to be placed in contact with the skin.

    The flexible LED panel is permanently connected by a three-foot long cable attached to a control panel that contains the circuitry and software that controls the device. The control panel contains several push buttons beneath a sealed cover. A power button that switches the device ON/OFF, a mode button that allows the user to select from 3 preprogrammed treatment modes "Contour" "wrinkles" and "Aches and Pains", and a Start button that activates the desired treatment mode. The control panel receives its power from a separate cable that connects via an AC adaptor for 110-220 Volts to a standard U.S. electrical power outlet. The control panel contains an automatic shut-off safety feature.

    AI/ML Overview

    The provided FDA 510(k) summary for the Biophotas Celluma CONTOUR (K232977) does not contain comprehensive details on acceptance criteria and specific study results in a format that would typically be described for a diagnostic AI device. This document is a premarket notification for a Class II medical device (Low Level Laser System For Aesthetic Use), and the information presented focuses on demonstrating substantial equivalence to predicate devices, rather than establishing de novo performance metrics through large-scale clinical trials with defined acceptance criteria and statistical proofs.

    However, based on the information provided, I can extract and infer some points regarding performance testing and acceptance, particularly concerning safety and usability for an Over-The-Counter (OTC) device.

    Here's a breakdown of the available information structured to address your questions, with the understanding that this device is not an AI diagnostic tool and thus many of your questions related to AI-specific metrics (e.g., MRMC, standalone algorithm performance, training set ground truth) are not applicable or explicitly stated for this type of device submission.


    Description of Acceptance Criteria and Studies

    1. Table of Acceptance Criteria and Reported Device Performance

    For this device, "acceptance criteria" are mainly focused on demonstrating compliance with safety standards, usability for OTC use, and equivalence of technological characteristics to predicates. Performance is assessed through compliance with these standards and comparison to predicates.

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance (as demonstrated)
    Electrical SafetyIEC 60601-1: ed 3.1Successfully passed testing.
    EMC TestingIEC 60601-1-2: ed 4Successfully passed testing.
    Software ValidationIEC 62304:2015Successfully passed testing; Software safety classification: Class A; FDA level of concern: Minor.
    Usability/Label ComprehensionIEC 62366:2015 (Medical devices Application of usability engineering to medical devices)No new use errors, hazardous situations, or hazard-related use scenarios were discovered. Improvement of the user interface design related to safety deemed unnecessary. Device is successfully used by laypersons (inferred from equivalence with OTC predicate and label comprehension study).
    Biocompatibility(Not explicitly stated, but implies compliance with relevant standards)Hardware is identical to previously cleared versions (K122237, K131113, K152280, K171323, K211038) in terms of material, manufacturing, tissue contact type, and duration. No change in biocompatibility.
    Technological EquivalenceEquivalent properties to predicates (e.g., energy type, wavelengths, intensity, treatment protocol, control, electrical power)Proposed device is within the bandwidth of predicate wavelengths, has equivalent output intensities, identical treatment times, and equivalent electrical power to predicates.

    2. Sample Size Used for the Test Set and Data Provenance

    • Usability/Label Comprehension Study:

      • Sample Size: 24 test subjects.
      • Data Provenance: Not explicitly stated, but typically these studies are conducted prospectively with recruited participants in a controlled environment. The country is not specified.
    • For other non-clinical performance tests (electrical safety, EMC, software validation, biocompatibility), these are typically engineering tests performed on the device itself or its components (e.g., materials for biocompatibility) rather than human subjects. Thus, the concept of "sample size" for a test set (like for a diagnostic device) doesn't directly apply in the same way.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is largely not applicable as this device is not an AI diagnostic tool that requires expert ground truth for performance evaluation in the context of disease detection or measurement.

    For the usability study, "ground truth" refers to whether lay users can correctly understand and operate the device safely based on labeling. This is assessed through questions and observation, not expert consensus on medical images or patient outcomes. The experts involved would be human factors engineers or usability specialists who design and analyze the study. The document does not specify the number or qualifications of such experts.

    4. Adjudication Method for the Test Set

    This question is not applicable for the type of studies described in this submission (safety, EMC, software validation, usability). Adjudication implies resolving disagreements among experts for ground truth establishment, which is relevant for diagnostic performance studies.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not Applicable. This device is a light-emitting diode (LED) system for aesthetic and pain relief uses. It does not involve AI for diagnostic interpretation or human reader assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not Applicable. This device does not feature a standalone diagnostic algorithm. The software mentioned is for controlling the device's operational modes and safety features, not for standalone diagnostic performance.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • For safety and performance characteristics (electrical, EMC, software): The "ground truth" is adherence to established international and national engineering standards (IEC 60601-1, IEC 60601-1-2, IEC 62304). Third-party testing labs or in-house engineering teams confirm compliance.
    • For usability/label comprehension: The "ground truth" is the correct understanding and safe operation of the device by lay users, as measured by their responses to questions and observed interactions, compared against the intended use and safety protocols.
    • For biocompatibility: The "ground truth" is the material's previously established biocompatibility or demonstration of compliance with relevant biocompatibility standards (e.g., ISO 10993 series), confirmed by testing (which was not repeated here due to material identity to cleared devices).

    8. The Sample Size for the Training Set

    Not Applicable. This device is not an AI/ML system that utilizes a training set for model development in the context of diagnostic performance. The software is noted as a Class A medical device software (lowest risk) and its validation focuses on life cycle processes and functional correctness, not machine learning training.

    9. How the Ground Truth for the Training Set was Established

    Not Applicable. As there is no "training set" in the context of an AI/ML algorithm for this device, this question does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211038

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-06-29

    (83 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K122237, K131113, K152280, K171323

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioPhotas Celluma RESTORE system is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

    Device Description

    The BioPhotas Celluma RESTORE system uses visible red light (640nm) and is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

    In addition, the Biophotas Celluma RESTORE is a portable, therapeutic device whose purpose is to provide even, cool, narrow-band wavelengths of polychromatic light (blue, 465nm red, 640nm and Near infra-red, 880nm) produced by super-luminous LEDs (light emitting diodes) to treat a variety of skin and musculoskeletal conditions as previously cleared under K122237, K131113, K152280 and K171323.

    The system comprises of a flexible, shape-taking frame upon which is mounted an array of LEDs, this allows the device to be contoured to the treatment area. The LEDs are embedded within a biocompatible foam covering that holds a transparent polycarbonate cover recessed within it. The biocompatibility nature of allows the device to be placed in contact with the skin. Nevertheless, the design of the device provides for maintaining a small distance between the surface of the skin and the surface of the device.

    The flexible LED panel is permanently connected by a three-foot long cable attached to a control panel that contains the circuitry and software that controls the device. The control panel contains several push buttons beneath a sealed cover. A power button that switches the device ON/OFF, a mode button that allows the user to select from 3 preprogrammed treatment modes "Hair", "Wrinkles", and "Aches and Pains", and a Start button that activates the desired treatment mode. The control panel receives its power from a separate cable that connects via an AC adaptor for 110-220 Volts to a standard U.S. electrical power outlet. The control panel contains an automatic shut-off safety feature.

    AI/ML Overview

    The provided text is a 510(k) summary for the Biophotas Celluma RESTORE device. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and the results of a standalone performance study.

    Specifically, the document does not contain the acceptance criteria or a study that directly proves the device meets specific performance metrics for hair growth in the same way a clinical trial might for a new drug or a direct performance study for a diagnostic AI. Instead, it relies on demonstrating that the device is "substantially equivalent" to previously cleared devices (predicates) that have already been deemed safe and effective for the indicated use.

    Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not applicable or not directly addressed in this type of submission. This document highlights non-clinical testing for safety and electrical performance, and a usability study, but not a study proving the efficacy of hair growth directly.

    Here's an analysis based on the information provided:

    1. A table of acceptance criteria and the reported device performance

    • No explicit performance acceptance criteria (e.g., % increase in hair count, specific hair density) for efficacy are listed or tested.
    • The document primarily focuses on safety and technological equivalence.
    • Safety and Electrical Performance:
      • Acceptance Criteria (Implied): Compliance with IEC 60601-1 (Electrical safety) and IEC 60601-1-2 (EMC testing).
      • Reported Performance: "The Biophotas Celluma RESTORE successfully passed testing per internal verification/validation requirements and national/international standards... The Biophotas Celluma RESTORE, and the predicate device have satisfied product safety testing to the IEC 60601-1 standard, and the electromagnetic safety testing to the IEC 60601-1-2 standard."
    • Software Validation:
      • Acceptance Criteria (Implied): Compliance with IEC 62304 and FDA Guidance document; software safety classification of Class A (low risk) and "Minor" level of concern.
      • Reported Performance: Software has been classified and validated accordingly.
    • Usability:
      • Acceptance Criteria (Implied): No new use errors, hazards, or hazardous situations discovered; no suggested revisions to the user manual; comprehension and understanding of the User Manual.
      • Reported Performance: "No new use errors, hazards, hazardous situations, or hazard-related use scenarios were discovered. Further improvement of the user interface design as it relates to safety was deemed unnecessary and there were no suggested revisions to the version of the user manual tested."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • For Efficacy: Not applicable, no direct efficacy test set in this submission.
    • For Usability: 25 test subjects.
    • Data Provenance: Not specified for the usability study. The document describes the device itself and its testing to standards, not a clinical trial with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable for efficacy data.
    • For the usability study, "experts" in establishing ground truth (i.e., correct understanding of the manual) are not explicitly mentioned in terms of their qualifications or number. The "ground truth" for usability here would be the intended correct understanding of the manual.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable for efficacy data.
    • Not specified for the usability study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC or comparative effectiveness study was done or reported in this 510(k) submission as it is not an AI-assisted diagnostic device, but rather a therapeutic device. The submission focuses on technological equivalence and safety, not on human interpretation or an AI component improving human performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This device is a therapeutic light device, not an AI algorithm. Therefore, "standalone algorithm performance" is not applicable. The "performance" is its ability to emit light within specified parameters and its safety.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For Efficacy: No direct efficacy ground truth is discussed as no clinical efficacy study on hair growth is presented here. The "ground truth" for substantial equivalence is the previously cleared predicate devices and their established safety and efficacy.
    • For Usability: The ground truth would be the intended correct interpretation and safe use of the device and its manual, as determined by the manufacturer's design and instructions.

    8. The sample size for the training set

    • This document does not describe the development of an AI model, so there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    • As there is no AI training set, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1