Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K242401

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-12

    (30 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K110404, K121037, K130673, K071071, K951987

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EXULT Knee Replacement System is intended for the treatment of diseases as follows:

    • Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis.
    • Post-traumatic loss of knee joint configuration and function.
    • . Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
    • . Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure

    EXULT Knee Replacement System is intended for cemented application only.

    Device Description

    The purpose of the submission is a line extension of the EXULT Tibial Insert (CPS-type) to the EXULT Knee Replacement System (K192507). The CPS-type Tibial Inserts are designed based on the design of the other previously cleared Tibial Inserts of the EXULT Knee Replacement System (PS-type, CR-type and UC-type). The CPS-type, which is constrained posterior stabilized, has a thicker and wider post which provides additional constraint as compared to the PS-type insert. The CPS-type Tibial Insert matches with Femoral Components of one-size up and down. The CPS-type Tibial Insert is symmetrical and used in both the right and left knee. It is only used in combination with the Femoral component/Tibial baseplate of the predicate device (K192507).

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study that proves a device meets those criteria for software or AI/ML-driven medical device performance.

    The document is a 510(k) premarket notification for the EXULT Knee Replacement System, a physical orthopedic device. It discusses the device's indications for use, technological characteristics, and non-clinical testing performed to establish substantial equivalence to predicate devices (other knee replacement systems).

    The "Non-Clinical Testing" section lists various engineering tests (wear testing, fatigue testing, constraint testing, range of motion, surface roughness, disassembly) and references ISO and ASTM standards. These are traditional mechanical engineering tests for orthopedic implants, not performance studies for AI/ML algorithms.

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance (for AI/ML).
    • Sample size for the test set and data provenance.
    • Number and qualifications of experts for ground truth.
    • Adjudication method.
    • MRMC comparative effectiveness study.
    • Standalone performance study.
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K242046

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-01

    (20 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K192507, K110404, K121037, K130673

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EXULT Knee Replacement System is indicated for the treatment of diseases as follows:

    · Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis traumatic arthritis or avascular necrosis) or rheumatoid arthritis.

    · Post-traumatic loss of knee joint configuration and function.

    • Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

    · Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.

    EXULT Knee Replacement System is intended for cemented application only.

    Device Description

    The purpose of the submission is a line extension of the EXULT Tibial Insert (UC-type) to the EXULT Knee Replacement System (K192507). The UC-type Tibial Inserts are designed based on the design of the other Tibial Inserts of the EXULT Knee Replacement System (PS-type and CR-type). The UC-type, which is ultra-congruent, has high conformity and increased anterior/posterior edge height, which limits anterior/posterior motion instead of a post structure, like the PS-type. The UC-type Tibial Insert matches Femoral Components of one-size up and down. The UC-type Tibial Insert is symmetrical and used in both the right and left knee. It is only used in combination with the Femoral component/Tibial baseplate of the predicate device (K192507).

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Wear Testing as per ISO 14242-1&2, ASTM F2003-02Evaluated to demonstrate substantial equivalence to predicate devices.
    Constraint test as per ASTM F1223Evaluated to demonstrate substantial equivalence to predicate devices.
    Contact pressure as per ASTM F2083, ASTM F1672Evaluated to demonstrate substantial equivalence to predicate devices.
    Range of Motion as per ISO 21536Evaluated to demonstrate substantial equivalence to predicate devices.

    Important Note: The document does not provide specific numerical acceptance criteria or numerical reported device performance values for any of these tests. It only states that the device was "evaluated to demonstrate substantial equivalence to the predicate devices" based on these tests.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. The text mentions "non-clinical testing" but does not specify sample sizes for materials tested, nor does it refer to patient data (retrospective or prospective) or countries of origin.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and therefore not provided in the document. The device is a knee replacement system (implant), not an AI/software device that requires expert-established ground truth for image interpretation or diagnosis. The tests described are mechanical and material performance tests.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided in the document. As noted above, this is a mechanical device, not an AI/software device involving human reader interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable and therefore not provided in the document. This is a medical device (knee implant), not an AI-assisted diagnostic or treatment planning software.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable and therefore not provided in the document. This is a medical device (knee implant), not an AI algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical testing of this knee implant system would be the physical and mechanical properties of the device and its predicate, measured according to established international and ASTM standards. For instance, in wear testing, the "ground truth" is the actual material loss measured. In contact pressure testing, it's the measured pressure distribution.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided in the document. This is a medical device (knee implant), not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided in the document. As this is not an AI/ML device, there is no training set or ground truth in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222278

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-08-26

    (28 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K192507, K110404, K130673, K200395, K201851, K212034

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LOSPA TKR System is intended for use in total knee arthroplasty surgery for the following indications:
    · Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
    · Post-traumatic loss of knee joint configuration and function;
    • Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability:
    · Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.
    LOSPA TKR System is intended for cemented application only.

    EXULT Knee System is intended for the treatment of diseases as follows:
    · Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
    · Post-traumatic loss of knee joint configuration and function;
    • Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability:
    · Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.
    EXULT Knee System is intended for cemented application only.

    Device Description

    The subject devices in this special 510(k) are the surgical instruments for the LOSPA Modular Knee System and EXULT Knee System, and submitted as a line extension of optional instruments for Corentec's FDA-cleared reusable orthopedic knee joint replacement instruments. The subject instruments are part of Corenter' 510(k)-cleared knee product lines and are required to facilitate total knee arthroplasty procedures.

    LOSPA Modular Knee System Instruments: Implant specific trials - the design of the connection part of the trials to the instrument has been revised due to customer's needs. Guide, alignment instruments - the design of the cutting guides and alignments instruments have been revised due to customer's needs.

    EXULT Knee System Instruments: AP sizer external rotation instrument and cutting block instruments have been developed to facilitate to check in posterior access during the operation based on US surgeon's request.

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria or a study proving that a device meets acceptance criteria. The document is an FDA 510(k) clearance letter for Corentec Co., Ltd.'s LOSPA TKR System Instrumentation and EXULT TKR System Instrumentation. It primarily discusses the administrative details of the clearance, the indications for use of the devices, and a comparison to legally marketed predicate devices to establish substantial equivalence.

    Specifically, the document states:

    • No Clinical or Animal Testing: "No clinical or animal testing were required." (Page 6)
    • Non-Clinical Testing: "The following non-clinical laboratory testing and/or engineering analysis were performed to determine substantial equivalence: ・ Dimensions measurement testing ・ Assembly testing" (Page 6)

    Therefore, I cannot provide the requested information, which requires details about acceptance criteria, device performance studies, sample sizes, expert involvement, ground truth establishment, or multi-reader multi-case studies. The clearance was based on dimensional measurements and assembly testing, not on clinical performance metrics typically associated with acceptance criteria for device efficacy or diagnostic accuracy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212034

    Validate with FDA (Live)

    Device Name
    LOSPA TKR System
    Manufacturer
    Date Cleared
    2021-12-09

    (163 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K110404,K130673,K160157,K192507,K200395

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LOSPA TKR System is intended for use in total knee arthroplasty surgery for the following indications:

    · Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;

    · Post-traumatic loss of knee joint configuration and function;

    · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;

    · Correction of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.

    LOSPA TRK System is intended for cemented application only.

    Device Description

    The Additional components being added to the LOSPA TKR System are:

    Instrumentation (Tibial Insert Trial) -

    The subject LOSPA TKR System components specification inclusions are a line extension of Instrumentation (Tibial Insert trial). The following is the additional components.

    A) Instrumentation (Tibial Insert Trial)

    A copy of a final tibial prosthesis designed to be used during total knee arthroplasty to determine the correct alignment, size, and fit of the final prosthesis. It is one of a set of trial knee prostheses that match the different anatomical structures of the knee joint, and may be used in conjunction with a knee femoral component trial prosthesis and a tibial baseplate trial prosthesis. This instrument is a reusable device that must be sterilized prior to use.

    Additional Tibial Insert Trials are designed based on Tibial Insert Trials of predicate device cleared LOSPA TKR System Instrumentation under K110404 & K130673 & K160157 & K192507 & K200395. There is one eliminated size, 26T of each design (CR and PS), each for left and right sides. For both CR and PS designs, the ranges of dimensions are the same. The components have slightly modified height. Except for the differences of height, lengths are identical to the previously cleared Tibial insert trials. The subject LOSPA TKR System is available in nine proportional sizes.

    This Tibial Insert Trials are made of Propylux conforming to ASTM D4101: Standard Classification System and Basic for Specification for Polypropylene Injection and Extrusion Materials.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the LOSPA TKR System, specifically focusing on the inclusion of new instrumentation (Tibial Insert Trial). It does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the context of performance metrics like sensitivity, specificity, accuracy, or reader improvement.

    The document is a regulatory submission for a medical device (knee joint prosthesis instrumentation) and details:

    • Device Name: LOSPA TKR System
    • Manufacturer: Corentec Co., Ltd.
    • Regulatory Classification: Class II, Product Codes JWH, MBH
    • Indications for Use: Total knee arthroplasty for specific conditions (osteoarthritis, rheumatoid arthritis, post-traumatic loss of function, moderate varus/valgus/flexion deformity, revision of unsuccessful procedures). Intended for cemented application only.
    • Device Description: The submission is for additional instrumentation (Tibial Insert Trial) for the existing LOSPA TKR System. These trials are sized for correct alignment, size, and fit of the final prosthesis, are reusable, and made of Propylux.
    • Substantial Equivalence: The device is being submitted as substantially equivalent to previously cleared LOSPA Total Knee System products (K110404, K130673, K160157, K192507, K200395).

    The information requested in the prompt (acceptance criteria, study details, sample sizes, ground truth, expert qualifications, adjudication, MRMC studies, standalone performance, training set details) is typically associated with studies proving the clinical or analytical performance of diagnostic or AI-enabled devices. This document pertains to the modification of an existing mechanical prosthetic system, specifically the inclusion of new instrumentation (a trial component), not a diagnostic algorithm or a device requiring such performance studies.

    Therefore, I cannot extract the requested information from the provided text because it does not contain data on:

    1. Acceptance criteria and reported device performance: No specific performance metrics (e.g., sensitivity, accuracy) are mentioned for the Tibial Insert Trial.
    2. Sample size, data provenance: Not applicable for this type of submission which is for mechanical instrumentation.
    3. Number/qualifications of experts, adjudication method: Not relevant for this type of device.
    4. MRMC study: Not relevant.
    5. Standalone performance: Not relevant.
    6. Type of ground truth: Not relevant.
    7. Training set sample size/ground truth establishment: Not relevant, as this is not an AI/algorithmic device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182370

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-09-27

    (27 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K110404, K130673

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Acculoc Total Knee System is indicated in knee arthroplasty for relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, primary and secondary traumatic arthritis, avascular necrosis of the femoral condyle, posttraumatic loss of joint configuration. Moderate valgus, varus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. This device may also be indicated in correction of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. The Acculoc Total Knee System is designed for cemented use only.

    Device Description

    The Acculoc Tibial tray is a component of "Acculoc Total Knee System" cleared under K170753 which consist of Femoral component, Tibial tray, Tibial Insert, Patella components and Instrumentation- Acculoc Total Knee Instrumentation System for use with the implants components. The current premarket notification is related to the addition of the Acculoc Tibial tray manufactured from CoCrMo alloy (Cobalt-chromium-molybdenum alloy), this is same size and same design with the predicate device tibial tray (K170753) and only the material is different. The Tibial tray AP and ML size ranges are exact same as the predicate device Tibial tray (K170753).

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device called the "Acculoc Total Knee System." It specifically focuses on an update to this system, adding a new material (CoCrMo alloy) for the Tibial tray component.

    This document does not contain information about acceptance criteria for device performance in the context of an AI/ML algorithm or a clinical study proving its performance against such criteria. Instead, it details:

    • Device Name: Acculoc Total Knee System
    • Regulatory Information: Product code, regulatory class, regulation number.
    • Intended Use/Indications for Use: What the device is designed to do and for whom.
    • Predicate Devices: Previously cleared devices that the current one is being compared against for substantial equivalence.
    • Device Description: Information about the components and materials.
    • Performance Testing: Refers to mechanical and pyrogenicity testing of the physical implant components against relevant ASTM/ISO standards, not clinical performance or performance of a diagnostic algorithm. This is typical for orthopedic implants to demonstrate safety and durability.

    Therefore, I cannot answer the questions about acceptance criteria for an AI/ML device, sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance because the provided text does not describe such a study or requirements. The "performance testing" mentioned is for the physical knee implant itself, not an AI or diagnostic device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1