510(k) Data Aggregation

    K Number
    K192069

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-05-06

    (279 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K130643

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aerodentis System is indicated for use in movement and alignment of teeth during orthodontic treatment of malocclusion.

    Device Description

    The Aerodentis System is an orthodontic system intended to adjust the patient's teeth through movement from their initial position to the desired position as a result of pulsatile mechanical force. The device is designed to achieve orthodontic tooth movement over a period of approximately 10 hours per day (typically at night). The device is suitable for patients with full complement of permanent teeth in need of upper, lower or combined jaw treatments. The system is composed of two primary components: A custom-made, latex-free, plastic dental mouthpiece is inserted and worn by the patient according to a customized treatment prescription specified by the dental practitioner. The Personal Treatment Controller (PTC), which provides compressed air to the mouthpiece via a polyurethane tube.

    AI/ML Overview

    The provided document describes the Aerodentis System, an orthodontic device for teeth movement. However, it does not contain specific acceptance criteria or an explicit study that proves the device meets such criteria as typically found in a standalone performance study with quantifiable metrics.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices (Invisalign System and Acceledent Aura) for FDA 510(k) clearance. This means the device is compared to legally marketed devices based on technological characteristics and performance, rather than against predefined, quantitative acceptance criteria for a novel device.

    The "Performance Data" section describes a clinical trial for safety and efficacy, but the results are presented in a comparative and qualitative manner, not against pre-specified numerical thresholds.

    Given the information in the provided text, I will construct an answer that reinterprets "acceptance criteria" as the outcomes and findings from the clinical trial and performance testing described, as these are the "criteria" used to support substantial equivalence.


    Acceptance Criteria and Device Performance for the Aerodentis System

    The provided FDA 510(k) submission for the Aerodentis System does not explicitly state precise, quantifiable acceptance criteria in the format typically used for performance studies establishing specific statistical thresholds. Instead, the submission relies on demonstrating substantial equivalence to predicate devices by showing that the Aerodentis System is as safe and effective. The "acceptance criteria" can therefore be inferred from the outcomes presented in the clinical trial and performance data, which are compared to the performance characteristics expected of similar legally marketed devices.

    The clinical trial aimed to demonstrate safety (absence of adverse events and significant root resorption) and efficacy (achieving an average Little's Irregularity Index (LII) of 1.5 or less), as compared to the Invisalign control.

    1. Table of "Acceptance Criteria" (Inferred from Study Outcomes) and Reported Device Performance

    "Acceptance Criteria" (Inferred from Clinical Study Goals)Reported Device Performance (Aerodentis System)
    Safety:
    - Absence of adverse eventsNo adverse events reported.
    - No clinically significant root resorptionNo clinically significant root resorption observed.
    Efficacy:
    - Achieve average Little's Irregularity Index (LII) of 1.5 or lessAll subjects who completed treatment achieved an average LII score of 1.5 or less.
    Comparative Performance (vs. Invisalign control):
    - Safe as the predicate deviceExcellent safety profile, similar to predicate.
    - Effective in achieving tooth alignmentEffective in achieving an LII of 1.5 or less.
    - Potentially higher rates of tooth movement (when accounting for actual use hours)Statistically significantly higher rates of tooth movement when accounting for actual hours of device use.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Investigational Arm (Aerodentis System): 28 subjects enrolled, 5 dropped = 23 subjects completed.
      • Control Arm (Invisalign): 15 subjects enrolled, 2 dropped = 13 subjects completed.
      • Total completed subjects: 36 subjects.
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but the submitter (Dror Orthodesign, Inc.) is located in Jerusalem, Israel, suggesting the study may have been conducted there or internationally.
      • Retrospective or Prospective: Prospective. It was a "single center, controlled, clinical trial" where subjects were "followed until" specific endpoints, indicating a prospective study design.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts involved in establishing the "ground truth" (e.g., LII scoring). The LII is an objective measurement index, likely applied by trained personnel. However, the document does not detail this specific aspect of the study.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing ground truth or evaluating outcomes. The Little's Irregularity Index is a quantitative measure, and adverse events/root resorption evaluation would typically follow standard clinical assessment protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • No background AI assistance/AI-based interpretation. This device is an orthodontic treatment system, not an AI-assisted diagnostic or interpretation tool for human readers. Therefore, an MRMC comparative effectiveness study regarding human reader improvement with AI assistance is not applicable to this device. The clinical trial compared the device itself to a predicate device.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not applicable. The Aerodentis System is a physical medical device (mouthpiece and controller) used for orthodontic treatment involving a patient and clinical oversight, not a standalone algorithm. The "software validation" mentioned refers to the software controlling the device and developing patient-specific plans, not a standalone diagnostic algorithm for assessment.

    7. The Type of Ground Truth Used

    • Clinical Measurements and Observations:
      • Efficacy: The primary efficacy "ground truth" was the average Little's Irregularity Index (LII) score of 1.5 or less. LII is an objective measurement of incisor alignment.
      • Safety: Adverse event reporting and radiographic assessment of root resorption served as the safety ground truth.

    8. The Sample Size for the Training Set

    • Not applicable. This device is an orthodontic treatment system, and the provided document describes a clinical trial for its safety and effectiveness, not the development or training of an AI algorithm from a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set" or "ground truth for a training set" discussed in the context of AI model development for this device. The study described is a clinical trial to evaluate the device itself.
    Ask a Question

    Ask a specific question about this device

    K Number
    K153048

    Validate with FDA (Live)

    Device Name
    AcceleDent Aura
    Date Cleared
    2016-07-08

    (263 days)

    Product Code
    Regulation Number
    872.5470
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K130643,K143120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AcceleDent® Aura is intended for use during orthodontic treatment. It is used in conjunction with brackets and wires or aligners and helps facilitate minor anterior tooth movement.

    Device Description

    The AcceleDent® Aura is an orthodontic accessory for the treatment of tooth malocclusion. AcceleDent® Aura should be used by patients for twenty minutes per day in conjunction with standard orthodontic treatment.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modified medical device, the AcceleDent® Aura, which is an orthodontic accessory. The submission focuses on expanding the indications for use to include patients undergoing orthodontic treatment with aligners.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" in the traditional sense of a pass/fail threshold for a specific performance metric. Instead, it relies on demonstrating substantial equivalence to predicate devices, particularly regarding clinical outcomes with aligners. The key "performance" aspect is the demonstration of facilitation of tooth movement.

    Acceptance Criteria (Implied)Reported Device Performance (AcceleDent® Aura with aligners)
    Demonstrate substantial equivalence by facilitating minor anterior tooth movement during orthodontic treatment with aligners.In a retrospective cohort study, AcceleDent® Aura demonstrated a statistically significantly faster rate of tooth movement (p=0.0129) and shorter overall treatment time (p<0.0001) as compared to Invisalign® patients only (without AcceleDent® Aura).
    Not adversely impact performance or raise new safety/effectiveness concerns compared to predicate devices (K143120, OrthoPulse™).The comparative clinical study outcomes with aligners (Table 5-1) demonstrate that the difference in technological characteristics compared to the predicate (K143120) does not adversely impact performance and does not raise different questions of safety and effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: 97 subjects (for the AcceleDent Aura aligner study).
    • Data Provenance: Retrospective AcceleDent Aligner Study Report (Source mentioned in Table 5-1). The country of origin is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The document does not provide information on the number of experts or their qualifications used to establish ground truth for the clinical study. The "Outcome Measures" for the AcceleDent Aura study state "Treatment Time (Weeks) Rate of Tooth Movement (ABO DI per Week) Total time in orthodontic aligner treatment," which would likely be objectively measured by the treating orthodontist or study investigators.

    4. Adjudication Method for the Test Set:

    The document does not specify any adjudication method (e.g., 2+1, 3+1) for the clinical study data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not reported. The study described is a clinical cohort study comparing AcceleDent Aura users with non-users, and a comparison against a predicate device's reported outcomes.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This question is not applicable as the device, AcceleDent® Aura, is a physical medical device (an orthodontic accessory) and not an algorithm or AI system. Its function is to provide mechanical vibration to facilitate tooth movement, used by the patient.

    7. The Type of Ground Truth Used:

    The ground truth for the clinical study was based on clinical outcome measures such as:

    • Treatment Time (Weeks)
    • Rate of Tooth Movement (ABO DI per Week)
    • Total time in orthodontic aligner treatment

    These are objective measurements based on patient clinical records and assessments by treating orthodontists, rather than expert consensus on image interpretation or pathology results.

    8. The Sample Size for the Training Set:

    This information is not applicable. The device is a physical medical device, not an AI/ML algorithm that requires a "training set" in the computational sense. The clinical study described in the document is an evaluation of the device's performance, not for training an algorithm.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no "training set" in the context of an AI/ML algorithm being evaluated.