Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K210752

    Validate with FDA (Live)

    Date Cleared
    2021-04-14

    (30 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K130537, K161021

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroMARK Breast Biopsy Site Marker is intended to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.

    Device Description

    The HydroMARK Breast Biopsy Site Marker [subject device] is a two-component marker that provides permanent marking of a breast biopsy site; a resorbable hydrogel component and a metallic component, and is not intended to be removed unless the marked tissue is determined to require surgical removal. The marker is supplied pre-loaded in a sterile, disposable applicator that is compatible with specified commercially available biopsy devices.

    The HydroMARK Breast Biopsy Site Marker is a resorbable component that is a highly expandable solid cylinder of polymerized and desiccated hydrogel. The hydrogel has features that are unique and highly desirable for breast tissue marking. The hydrogels absorb fluid, they are readily visible by ultrasound imaging. During a breast biopsy procedure, the marker is deployed through a delivery tool into the cored-out space created by a breast biopsy device. Upon expansion, the hydrogel fills the space and conforms to the site of biopsy. Embedded in the hydrogel is a coiled metallic wire made of Titanium or Stainless Steel. The wire is coiled into loops to provide a unique identifier under ultrasound, x-ray, and MRI imaging. The embedded metallic wire coil is visible under ultrasound for up to 6 weeks and is permanently visible under X-ray and MRI.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the HydroMARK Breast Biopsy Site Marker. The notification is for modifications to the labeling of an already cleared device, not for a new device or significant changes to its fundamental scientific technology, materials, manufacturing, or intended use.

    Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are focused on demonstrating that the labeling changes do not raise new or different issues of safety and effectiveness compared to the previously cleared devices.

    Acceptance Criteria and Reported Device Performance (Table):

    Acceptance CriteriaReported Device Performance
    Device meets all system labeling requirementsPassed Document Analysis of Labeling Changes
    Device is as safe and effective as predicateDemonstrated through verification testing by document analysis of labeling changes and comparison of technological characteristics
    Device performs as well as predicateDemonstrated through verification testing by document analysis of labeling changes and comparison of technological characteristics
    No new or different issues of safety and effectiveness arise from labeling changesConclusion based on comparison of technological characteristics and verification testing

    Detailed Breakdown of the Study and Related Information:

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not applicable in the traditional sense of a clinical or performance study involving patient data. The "test set" here refers to the revised labeling documents.
      • Data Provenance: The study involved document analysis of labeling changes to the existing HydroMARK Breast Biopsy Site Marker and its predicate/reference devices (K121113, K130537, K161021). No patient-specific data or data from a specific country of origin is mentioned for this particular submission, as it focuses on regulatory labeling compliance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not explicitly stated. The "ground truth" for labeling changes would likely be established by the manufacturer's regulatory affairs and quality assurance teams, in consultation with relevant internal subject matter experts (e.g., medical, engineering) to ensure accuracy, clarity, and compliance with regulations. External regulatory consultants might also be involved. Specific numbers or qualifications of individual experts are not provided in this summary.
    3. Adjudication method for the test set:

      • Not applicable as this was a document analysis assessing compliance with regulatory and safety information, not a comparative review of clinical outcomes by multiple experts. The "adjudication" would be internal review and approval processes within the submitting company to ensure the labeling changes are justified and complete.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This submission is for a breast biopsy site marker (a physical device), not an AI-powered diagnostic tool. The concept of "human readers improve with AI" is not relevant to this device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No, a standalone algorithm performance study was not done. This device is a physical implantable marker, not a software algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this submission is regulatory compliance and demonstration that the modified labeling does not introduce new safety or effectiveness concerns, based on the established safety and performance of the physical device (which was previously cleared). This is primarily based on:
        • Comparison to the predicate device's cleared labeling and performance.
        • Internal design verification and risk assessment processes (implied by ISO 14971:2007 mention).
        • Applicable regulatory standards and guidelines for medical device labeling.
    7. The sample size for the training set:

      • Not applicable. This is not a machine learning or AI-based device, so there is no "training set" in the context of data used to train an algorithm.
    8. How the ground truth for the training set was established:

      • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203097

    Validate with FDA (Live)

    Date Cleared
    2020-12-16

    (63 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K121113, K130537

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroMARK® Breast Biopsy Site Marker is indicated to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.

    Device Description

    The HydroMARK® Breast Biopsy Site Marker [subject device] is a two-component marker that provides permanent marking of a breast biopsy site; a resorbable hydrogel component and a metallic component and is not intended to be removed unless the marked tissue is determined to require surgical removal. The marker is supplied pre-loaded in applicator devices designed to fit into the following:

    • 8G Flexible Applicator System fits into:
      • O Mammotome revolve® 8G Probe (9cm with and without Specimen Management System and 12cm without Specimen Management System)
      • EnCor 7G Directional Vacuum-Assisted Biopsy Devices o
    • . 10G Flexible Applicator System fits into:
      • Mammotome revolve® 10G Probe (9cm with and without Specimen Management O System and 12cm without Specimen Management System)
      • EnCor 10G Directional Vacuum-Assisted Biopsy Devices o
      • ATEC 9G Biopsy Handpiece Introducers O

    The HydroMARK® Breast Biopsy Site Marker is a resorbable component that is a highly expandable solid cylinder of polymerized and desiccated hydrogel. The hydrogel has features that are unique and highly desirable for breast tissue marking. The hydrogels absorb fluid, they are readily visible by ultrasound imaging. During a breast biopsy procedure, the marker is deployed through a delivery tool into the cored-out space created by a breast biopsy device. Upon expansion, the hydrogel fills the space and conforms to the site of biopsy. Embedded in the hydrogel is a coiled metallic wire made of Titanium. The wire is coiled into loops to provide a unique identifier under ultrasound, x-ray and MRI imaging. The embedded metallic wire coil is visible under ultrasound for up to 6 weeks and is permanently visible under X-ray and MRI. The wire coil is available in (3) three shapes:

    • Barrel Shape
    • Open Coil Shape
    • Butterfly Shape
    AI/ML Overview

    The provided document is a 510(k) summary for the HydroMARK® Breast Biopsy Site Marker. It details the device's substantial equivalence to predicate devices and the non-clinical performance data supporting this claim.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for each test in a table format. Instead, it describes general "In-House Testing Standards" and states that all tests "PASSED," concluding that the device is "as safe, as effective, and performs as well as, the legally marketed predicate device."

    However, based on the descriptions of the tests, we can infer the performance criteria. The device aims to demonstrate substantial equivalence to its predicate for various aspects, including safety, effectiveness, and performance characteristics.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Non-Clinical Performance Bench Testing
    Design VerificationMeet all design and functional requirements.PASSED
    FEA (Safety Factor of Cannula)Demonstrate comparable or improved safety factor compared to predicate/reference cannula to reduce tip shear.PASSED
    Hydrogel Verification (Hydration, Moisture)Meet established in-house standards for hydrogel hydration and moisture content, based on and more stringent than predicate.PASSED
    Compatibility Verification (Deployment/Removal)Successful deployment and removal of markers in each compatible probe, based on and more stringent than predicate.PASSED
    Mechanical Verification (Deployment, Self-Retraction, Bond Strength)Meet established in-house standards for marker deployment, pushrod self-retraction, and bond strength, based on and more stringent than predicate.PASSED
    Human Factors/Usability Testing
    Simulated Use TestingSafe and effective interaction for intended users, uses, and environments per FDA recognized standards (IEC 62366-1, IEC 60601-1-6, FDA Guidance).PASSED
    Formative StudiesFindings support safe and effective use.PASSED
    Biocompatibility Testing
    Cytotoxicity (MEM Elution)No unacceptable adverse biological response. Comply with ISO 10993-1, 10993-5.PASSED
    Chemical Characterization Testing
    Exhaustive ExtractablesNo unacceptable adverse biological response from leachable substances. Comply with ISO 10993-12, 10993-18.PASSED
    Toxicological Risk AssessmentNo unacceptable toxicological risks. Comply with ISO 10993-1, 10993-18, 10993-17.PASSED
    Sterilization & Shelf-Life Testing
    Bioburden TestingMeet established bioburden limits. Comply with ANSI/AAMI/ISO 11737-1.PASSED
    Sterilization TestingEffective sterilization. Comply with ANSI/AAMI/ISO 11737-1.PASSED
    Sterility Testing (Natural Product/BI)Demonstrate product sterility. Comply with ANSI/AAMI/ISO 11737-1.PASSED
    Endotoxin TestingMeet endotoxin limits. Comply with USP <85>, ANSI/AAMI ST72.PASSED
    Residual Analysis TestingMeet residual limits related to sterilization. Comply with USP <85>, ANSI/AAMI ST72.PASSED
    Shelf-Life / Transit Verification (12-month)Maintain integrity, sterility, and performance after 12 months simulated aging. Comply with ASTM F1980-16, F2095-11, F38/F88M-15.PASSED
    Shelf-Life / Transit Verification (18-month)Maintain integrity, sterility, and performance after 18 months simulated aging. Comply with ASTM F1980-16, F2095-11, F38/F88M-15.PASSED

    2. Sample sizes used for the test set and data provenance

    The document provides a summary of non-clinical bench testing. It does not specify sample sizes for individual tests. The data provenance is "In-House Testing Standards as established by Devicor Medical Products, Inc." which indicates retrospective testing performed by the manufacturer, relying on internal procedures and industry best practices. There is no mention of country of origin for the data beyond the manufacturer being in Cincinnati, OH, U.S.A.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document focuses on non-clinical (bench, human factors, biocompatibility, sterilization) tests, not clinical studies. Therefore, there is no mention of experts establishing ground truth for a test set in the context of clinical outcomes or imaging interpretation. The "ground truth" for these engineering and safety tests is defined by the established testing standards and internal design requirements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This document describes non-clinical testing for substantial equivalence, not a clinical study involving human readers or interpretation of medical images. Hence, there is no adjudication method described for the test set. The results are reported as "PASSED" against established testing standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or mentioned. This document is for a medical device (breast biopsy site marker), not an AI-powered diagnostic tool. Therefore, there's no discussion of human readers improving with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This point is not applicable as the device is a physical breast biopsy site marker, not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests described:

    • Design Verification, FEA, Hydrogel, Mechanical, Compatibility Testing: The "ground truth" is defined by the device's design requirements and the manufacturer's internal testing standards, often based on industry best practices for physical properties, functionality, and performance.
    • Human Factors/Usability Testing: The "ground truth" relates to compliance with recognized standards (IEC 62366-1, IEC 60601-1-6, FDA Guidance) for user safety and effectiveness through simulated use and formative studies.
    • Biocompatibility, Chemical Characterization, Bioburden, Sterilization, Endotoxin, Residual Analysis, Shelf-Life Testing: The "ground truth" is adherence to recognized international and national standards (e.g., ISO 10993 series, ANSI/AAMI, ASTM, USP) which define acceptable biological response, chemical safety, sterility, and material stability.

    8. The sample size for the training set

    This document describes the testing for a physical medical device (breast biopsy site marker), not an AI algorithm. Therefore, there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    As there is no training set mentioned, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170803

    Validate with FDA (Live)

    Date Cleared
    2017-06-15

    (90 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K121113, K130537

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HydroMARK® Breast Biopsy Site Marker is indicated to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.

    Device Description

    The HydroMARK® Breast Biopsy Site Marker contains a resorbable hydrogel component and a metallic component for permanent marking. The hydrogel has features that are unique and highly desirable for breast tissue marking. The HydroMARK® Breast Biopsy Site Marker is provided pre-loaded in a sterile, disposable applicator that is compatible with specified commercially available biopsy devices. The marker is deployed by the delivery system and is left in the tract created during the biopsy procedure. This Traditional 510(k) addresses the qualification and validation of a new source of raw material polymer of the same original formulation, produced in a new facility and in smaller batch sizes, to be used by the manufacturer of the hydrogel component of the HydroMARK® Breast Biopsy Site Marker. The hydrogel component is manufactured by Coldstream Laboratories, Inc. The new source of polymer for the hydrogel is Corden Pharma, replacing Genzyme polymer.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the HydroMARK® Breast Biopsy Site Marker. The notification aims to demonstrate substantial equivalence to previously cleared predicate devices, primarily due to a change in the raw material polymer supplier for the hydrogel component.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" alongside "reported device performance" in a quantitative manner for specific benchmarks (e.g., "Visibility under ultrasound: >95% detected"). Instead, it lists various tests performed and states that the device "met all original finished product specifications" or "performed as intended according to the specifications established for the original device."

    However, we can infer the acceptance criteria from the tests conducted and the statements of equivalence:

    Acceptance Criteria CategorySpecific Criteria / Test DescriptionReported Device Performance (New Device)
    Material EquivalencePolymer Raw Material:
    - Gel Permeation Chromatography (GPC): Equivalent molecular weight distribution (Number Average (Mn) and Weight Average (Mw)) and polydispersity between old and new polymer sources. No evidence of changed polymer composition."Results show the number (Mn) and weight averaged (Mw) molecular weights obtained from both sources are comparable. The polymer molecular weight distributions for both manufacturers are equivalent. There is no evidence of changed polymer composition..."
    - Nuclear Magnetic Resonance (NMR): Equivalent ratios of chemicals from first and second reaction steps, indicating no new chemical entities."Ratios of chemicals from the first reaction step demonstrate equivalency. NMR was also used to determine molar ratios of the secondary reaction step and also showed equivalency. ...No different or new chemical entities are identified."
    - Degradation Profiles: Hydrolysis products of both old and new polymers contain the same components in the same proportions."Finally, hydrolysis of both predicate and new polymers contained the same components, in the same proportions. Both raw material polymers are chemically equivalent."
    Finished Product QualityPhysical Characteristics:
    - Visual Inspection: Free of visual defects."met all original finished product specifications including visual inspection..."
    - Critical Dimensions: Outer diameter and length (0.070 inches and 0.20 inches for marker, 0.035-0.0625 inches for coil length, 0.039 inches for coil width)."met all original finished product specifications including... dimensions..." (Implied to match the predicate device dimensions listed in the comparison table: 0.070 inches OD, 0.20 inches length, 0.035-0.0625 inches coil length, 0.039 inches coil width).
    - % Moisture: Within specified range pre- and post-sterilization."met all original finished product specifications including... % moisture..."
    - Functional Deployment: Deploys as intended."met all original finished product specifications including... functional deployment..."
    - Hydration Rate: Within specified range."met all original finished product specifications including... hydration rate..."
    Imaging Visibility- Visibility under Ultrasound: Visible for at least 6 weeks."met all original finished product specifications including... visibility by ultrasound..." (The Indications for Use state "be visible under ultrasound for at least 6 weeks").
    - Visibility under X-ray: Permanently visible."met all original finished product specifications including... visibility by... x-ray..." (The Indications for Use state "be permanently visible by x-ray").
    - Visibility under MRI: Permanently visible and safe in an MRI environment (no significant radio frequency induced heating, image artifacts, magnetically induced torque, or displacement force)."met all original finished product specifications including... visibility by... MRI..." (The Indications for Use state "be permanently visible by... MRI." This implies compliance with ASTM F2182-11a, ASTM F2119-07, ASTM F2213-06, ASTM F2503-13, ASTM F2052-15). The document states the "Finished devices performed as intended according to the specifications established for the original device." concerning these tests.
    Biocompatibility- Sterility: Sterile, meeting ISO 11135 requirements."met all original finished product specifications including... sterility..."
    - Bioburden: Within acceptable limits."met all original finished product specifications including... bioburden..."
    - Pyrogens: Apyrogenic."met all original finished product specifications including... pyrogens..."
    - EO Residuals: Within acceptable limits (if Ethylene Oxide sterilized)."met all original finished product specifications including... ethylene oxide residuals."
    - Biological Evaluation: Meeting ISO 10993 requirements (Cytotoxicity, Systemic Toxicity, Chemical Characterization).The document lists the ISO 10993 standards and states that the "Finished devices performed as intended according to the specifications established for the original device," implying that these criteria were met.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: For the polymer equivalence testing, "Three samples each of the predicate polymer and the new polymer" were tested by GPC. The same number seems implied for NMR and degradation profiles. For finished product testing, "The three manufacturing batches [of the new polymer] were combined and used to prepare HydroMARK® Breast Biopsy Site Markers which met all original finished product specifications". The exact number of finished devices tested is not specified, but it was enough to represent these combined batches.
    • Data Provenance:
      • Polymer Raw Material: The new polymer batches were from the Corden Pharma Liestal, Switzerland facility. The predicate polymer was from Genzyme (original supplier/process).
      • Finished Product: The finished devices were manufactured using the new Corden Pharma polymer. The testing protocols were "identical to the tests used in the original 510(k)". The study is retrospective in the sense that it aims to demonstrate equivalence to previously established specifications and predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information (number of experts, their qualifications, and their role in establishing ground truth) is not applicable to this submission. This 510(k) is not for a diagnostic algorithm or a device requiring expert interpretation for performance evaluation. It's a submission for a physical medical device (breast biopsy site marker) that primarily relies on objective chemical, physical, and imaging characteristic tests, not human interpretive performance.

    4. Adjudication method for the test set

    Not applicable for the same reasons as #3. There was no "ground truth" established by expert consensus or adjudication in the context of human interpretation. The "ground truth" here is adherence to objective engineering, material science, and biological specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This submission is for a physical medical device (a biopsy site marker), not an AI/CAD system or a device that directly assists human readers in interpreting images. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. As stated above, this is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" in this context refers to the established specifications and performance characteristics of the predicate device and relevant industry standards (e.g., ISO, ASTM). Specifically:

    • Chemical Equivalence: Established by comparing molecular weight distribution, NMR spectra, and degradation profiles to the known composition and characteristics of the original Genzyme polymer.
    • Physical/Functional Performance: Established by existing specifications for visual inspection, critical dimensions, moisture content, functional deployment, hydration rate, invisibility under various imaging modalities, and biological safety (sterility, biocompatibility).
    • Imaging Visibility/Safety: Established by adherence to specific ASTM standards for MRI compatibility.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires a training set. The "training" in this context refers to the qualification and validation of the manufacturing process for the new polymer source, which involved testing "3 consecutive test batches" from the new facility.

    9. How the ground truth for the training set was established

    Not applicable in the AI/ML sense. However, if interpreting "training set" as the batches used to qualify the new polymer manufacturing process:
    The "ground truth" for qualifying these batches was established by comparison to the "original specifications" designed for the predicate device's polymer. This means the acceptable parameters (e.g., molecular weight range, chemical ratios) were already defined based on the performance of the previously approved device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1