Found 2 results

510(k) Data Aggregation

    K Number
    K203378

    Validate with FDA (Live)

    Device Name
    Pulsehaler
    Manufacturer
    Date Cleared
    2021-03-31

    (134 days)

    Product Code
    Regulation Number
    868.5690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K123400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pulsehaler™ is indicated for use as a Positive Expiratory Pressure (PEP) Device.

    • The use of Pulsehaler™ improves clearance of secretions .
    • The use of Pulsehaler™ may reduce the need for postural drainage ●
    • Pulsehaler™ facilitates opening of airways in patients ●
    • Pulsehaler™ may be used to prevent or reverse atelectasis ●
    • Pulsehaler™ mav also be useful in the removal of mucus from the lungs ●

    Pulsehaler™ is intended for single patient, adult users in a home or hospital environment.

    Device Description

    Pulsehaler™ is a non-invasive handheld treatment device used daily by the patient to promote airway opening and secretion clearance by vibrating the airways at a variety of different frequencies. It consists of Base Unit with an LCD touch screen that provides a pressure source, and a Hand Unit with a rotating disc that periodically interrupts the pressure source to deliver pressure pulsations to the patient. These components are connected by an air hose and an electrical power and data cable.

    While sitting in a relaxed position, the patient holds the Hand Unit and breathes normally through its mouthpiece. A steady flow of pressurized air is delivered to the Hand Unit component by the Base Unit and enters the air inlet of the Hand Unit. The pressurized air passes along the internal airflow duct of the Hand Unit through an air filter until it reaches the rotating disc. When it rotates the disc interrupts the continuous flow of air from the Base Unit and then releases the flow in pulses at a rate corresponding to the rotation speed. The rotation speed of the disc is determined by the software using a pre-set protocol.

    The exhalation port is restricted in size, which together with the positive pressure generated by the Base Unit, induces PEP.

    The patient is treated with the device for 20 minutes at a time, up to three times a day according to the prescribing physician's instructions. During each 20-minute treatment, Pulsehaler™ delivers pulses varying from 5 to 50 pulses per second. The patient may use the User Interface to pause the treatment at any time or resume after a pause.

    Pulsehaler™ is intended for single adult patient, multiple reuse, at home or in a healthcare facility and is non-sterile. The instructions for use contain cleaning and high-level disinfection instructions that involve the use of Cidex.

    Expected Use Life of the device is 5 years and Shelf Life is 6 Months.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Pulsehaler™ device, which is a Positive Expiratory Pressure (PEP) Device. The document details the device, its indications for use, and a comparison to predicate and reference devices to establish substantial equivalence.

    Based on the provided text, the Acceptance Criteria and the Study that proves the device meets the Acceptance Criteria are primarily focused on non-clinical testing and comparison to predicate devices, rather than a clinical study with human subjects, especially not a multi-reader, multi-case (MRMC) comparative effectiveness study of an AI-powered device. The device is a mechanical PEP device, not an AI-powered diagnostic or therapeutic device that would typically involve AI-specific performance metrics like sensitivity, specificity, or AUC, or human reader improvement studies.

    Therefore, the requested information regarding AI device acceptance criteria, MRMC studies, ground truth establishment for AI models, and training/test set sizes for AI data are not applicable to this specific submission. The submission relies on demonstrating substantial equivalence to a legally marketed predicate device (K140772) through technical and functional comparisons, and non-clinical testing.

    Here's an attempt to answer the questions based only on the provided document, noting the limitations of the document for an AI-focused query:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a quantitative, pass/fail manner for performance in the way one might see for an AI device (e.g., "Sensitivity must be >X%"). Instead, the device's "performance" is demonstrated by its functional characteristics being similar or equivalent to its predicate device. The document summarizes non-clinical testing and then provides a comparison table.

    CharacteristicAcceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (Pulsehaler™)
    Indications for UseMust be same as or similar to predicate (Aerosure, K140772): Improves clearance of secretions, reduces need for postural drainage, facilitates opening of airways, prevents/reverses atelectasis, aids mucus removal. Intended for single patient, adult users in home or hospital.Same: Improves clearance of secretions, may reduce need for postural drainage, facilitates opening of airways, may be used to prevent or reverse atelectasis, may also be useful in the removal of mucus from the lungs. Intended for single patient, adult users in home or hospital environment.
    Intended Use EnvironmentMust be Hospital and Home Use.Hospital and Home Use.
    Patient PopulationSimilar to predicate (patients age 21 and above with various lung conditions susceptible to secretion problems/atelectasis).Adult patients. (Note in comparison table: "Similar to predicate, no specific disease conditions are claimed.")
    Oscillation MechanismSimilar to predicate (Rotating Disk).Rotating Disc.
    PEP MechanismOutput PEP pressures, vibration frequencies, and amplitudes are equivalent to predicate and reference device.Flow vs resistance from oscillation mechanism, restricted orifice, and blower. (Note in comparison table: "Subject device generates PEP with a blower, however, output PEP pressures, vibration frequencies and amplitudes are equivalent to those produced by predicate and reference device.")
    Operating ModesOutput PEP pressures, vibration frequencies and amplitudes produced at each speed are equivalent to predicate.Single mode with multiple speeds. (Note in comparison table: "Subject device has multiple speeds. Output PEP pressures, vibration frequencies and amplitudes produced at each speed are equivalent.")
    Software ControlSimilar to predicate (Yes).Yes.
    Mean FrequencyWithin range of predicate (30-52 Hz) and reference (7-22 Hz).6 Hz to 65 Hz. (Note in comparison table: "Mean frequencies generated by subject device are within the range generated by predicate and reference device.")
    Pressure AmplitudeWithin range of predicate (2-32 cmH2O) and reference (1-21 cmH2O).9-23 cmH2O. (Note in comparison table: "Pressure amplitude generated by subject device is within the range generated by predicate and reference device.")
    Mean PressureWithin range of predicate (1-11 cmH2O).7-10 cmH2O. (Note in comparison table: "Mean pressure generated by subject device is within the range generated by predicate.")
    Pressure LimitationOutput pressure during single and multiple faults must be <20 cmH2O.Based on hardware blower limitations. (Note in comparison table: "Output pressure of blower limited during single and multiple faults to <20cmH2O.")
    Power SupplySimilar to predicate (AC Mains connected DC power supply).AC Mains connected DC power supply, 110~240V 50/60Hz. (Note in comparison table: "Wall power to Base Unit - same as predicate; however, Hand Unit is connected and powered by Base Unit, instead of rechargeable batteries.")
    SterilityUsed non-sterile.Used non-sterile.
    Standards ComplianceCompliant with relevant medical device standards shown by predicate (e.g., IEC 60601-1-2, ISO 10993).ISO 10993; IEC 60601-1-2; IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, ISO18562. (Note: "Additional applicable standards met.")
    Cleaning & DisinfectionValidated cleaning and disinfection methods.Validation done, involving Cidex.
    Shelf/Use LifeValidated shelf and use life.Shelf Life (6 Months) and Use Life (5 years) Testing completed.
    BiocompatibilityCompliant with relevant biocompatibility standards.Testing including VOC and Particulate Matter 2.5 Micron Testing - External Communicating, tissue / bone / dentin and Surface contact, Mucosal Membrane, permanent contact duration, per ISO 10993 suite and ISO 18562-3:2002.
    Functional/Software TestingFunctional and Software validation to Design Input Requirements; Software V&V per FDA guidance.Functional Testing and Software Validation to Design Input Requirements. Software verification and validation testing conducted, documentation provided as recommended by FDA's Guidance. Software level of concern: moderate.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document describes non-clinical testing and comparison to predicate devices, not typically "test sets" of data for algorithmic performance. The testing involved physical devices. There is no mention of "data provenance" in terms of patient data or clinical images from specific regions or whether it was retrospective/prospective, as this is not a data-driven AI device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. There is no "ground truth" establishment in the context of expert review of images or clinical data, as this is a mechanical device submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no adjudication required for this type of non-clinical device test.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an AI-assisted diagnostic or therapeutic device. No MRMC study was performed or required.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-only device. Its performance is inherent to its mechanical and software design, tested physically.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable in the context of clinical "ground truth" for diagnostic accuracy. The "ground truth" for this device's performance is its physical measurements (e.g., pressure, frequency) and safety characteristics, confirmed through engineering tests and adherence to recognized standards. For the purpose of substantial equivalence, the "ground truth" is also the performance and characteristics of the legally marketed predicate device.

    8. The sample size for the training set

    Not applicable. There is no "training set" as this is not a machine learning or AI-based device.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160636

    Validate with FDA (Live)

    Date Cleared
    2016-08-03

    (149 days)

    Product Code
    Regulation Number
    868.5690
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K123400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The D R Burton iPEP Therapy System or vPEP is intended for use as a Positive Expiratory Pressure (PEP) by patients capable of generating an exhalation flow of 10 lpm for 3-4 seconds and an Incentive Spirometer as an inspiratory, deep breathing positive exerciser.

    Intended for single-patient, multi-use.

    iPEP System for hospital and clinical settings vPEP for hospital, clinical, and home care setting

    Device Description

    D R Burton Healthcare as indicated above proposes to offer 2 devices which we will refer to as the:

    • iPEP (integrated Incentive spirometer with PEP) system and ●
    • . vPEP which is a standalone PEP therapy device

    iPEP system

    • Volumetric Incentive Spirometer up to 4000 cc
    • Oscillatory PEP module (cartridge) ●

    The design of the iPEP system permits the PEP cartridge to be removed and used as an independent PEP device, which we refer to as the Pocket PEP.

    In order to use the Pocket PEP the iPEP system includes a separate mouthpiece, carrying / storage case and dust cover which are needed when one wants to use the PEP cartridge independent of the iPEP systems (Pocket PEP).

    When the PEP cartridge is removed from the iPEP system, one can insert a cover where the PEP cartridge has been removed to allow the incentive spirometer to be used one its own.

    vPEP

    • The vPEP is the standalone device that is pre-assembled and is not intended to be used in the iPEP ● System. The vPEP incorporates the identical internal components of the PEP cartridge of the iPEP system.
    AI/ML Overview

    The provided document is a 510(k) Summary for the D R Burton iPEP System and vPEP, which are devices intended for Positive Expiratory Pressure (PEP) therapy and as Incentive Spirometers. The study described is a non-clinical bench testing and usability study, focused on demonstrating substantial equivalence to predicate devices. There is no clinical study described.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents two comparison tables (Table 1 for PEP function, Table 2 for Incentive Spirometer function) comparing the proposed device against its predicates. The acceptance criteria for the proposed device are implicitly set by its performance falling within a certain range (e.g., 15%) of the predicate devices' performance.

    AttributeAcceptance Criteria (Implicit from Predicate Performance)Reported Device Performance (Proposed iPEP System and vPEP)
    PEP Function (Table 1)
    Mean Frequency (Hz) @ 5 lpm7.0 Hz (Predicate: DHD Acapella)8.0 Hz
    Mean Frequency (Hz) @ 25 lpm13.7 Hz (Predicate: DHD Acapella)13.0 Hz
    Mean Amplitude / Pressure (cmH2O) @ 5 lpm5.9 cmH2O (Predicate: DHD Acapella)5.5 cmH2O
    Mean Amplitude / Pressure (cmH2O) @ 25 lpm23.2 cmH2O (Predicate: DHD Acapella)22.7 cmH2O
    Min Pressure (cmH2O) @ 5 lpm0 cmH2O (Predicate: DHD Acapella)0 cmH2O
    Min Pressure (cmH2O) @ 25 lpm2.9 cmH2O (Predicate: DHD Acapella)2.5 cmH2O
    Max Pressure (cmH2O) @ 5 lpm5.9 cmH2O (Predicate: DHD Acapella)5.5 cmH2O
    Max Pressure (cmH2O) @ 25 lpm26.1 cmH2O (Predicate: DHD Acapella)25.1 cmH2O
    Incentive Spirometer Function (Table 2)
    Volume range2500 cc and 5000 cc (Predicate: DHD Cliniflo / Coach)4000 cc
    Volume accuracy @ 2500 cc14 – 21% (Predicate: DHD Cliniflo / Coach)Not directly comparable at 2500cc.
    Volume accuracy @ 5000 cc16 – 21% (Predicate: DHD Cliniflo / Coach)Not directly comparable at 5000cc.
    Volume accuracy @ 4000 ccImplies similar accuracy to predicate models at comparable volumes. (Predicate values are 14-21% at similar volumes.)1.6 to 2.2% (Significantly better than reported predicate range, but applied to a different volume. The summary states "The results demonstrated equivalent performance," implying this accuracy is considered acceptable and substantially equivalent to the predicate's performance range.)

    Note on "Criteria within 15%": This phrase appears in Table 1 for Frequency and Amplitude/Pressure, suggesting that the proposed device's performance metrics were expected to be within 15% of the predicate device's values. The reported values for the iPEP and vPEP appear to meet this implicit criterion.

    Additional Bench Testing:

    • Aging, Simulated Life (cleaning), Drop Testing, Cleaning: Performed with pre- and post-exposure evaluation.
    • Biocompatibility Materials: Cytotoxicity, sensitization, and irritation testing for materials with limited duration (<24 hours) patient contact.
    • Usability Studies: Confirmed appropriate design and instructions for use for intended user groups.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Description: The "test set" for the non-clinical, comparative bench testing is not explicitly defined in terms of a specific number of units or data points beyond the flow rates (5 lpm and 25 lpm) at which measurements were taken. However, for "Usability Studies," there were at least 15 users per identified group (lay users and Healthcare Providers).
    • Data Provenance: The data is generated from bench testing and usability studies conducted by D R Burton Healthcare LLC. This is prospective data from internal testing. There is no mention of country of origin for the data beyond the company's location in Farmville, North Carolina, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is largely not applicable as the described studies are non-clinical bench tests and usability tests.

    • For the bench testing, the "ground truth" is established by direct measurement against engineering specifications and comparison to the performance of predicate devices per existing literature or internal testing of those devices. No independent clinical experts are mentioned for establishing ground truth in this context.
    • For the usability studies, the "ground truth" (i.e., whether the device is usable and instructions are clear) is assessed by the performance and feedback of the users themselves. The document states "2 user groups, lay users and Healthcare Providers," but does not specify their individual qualifications beyond being "Healthcare Providers."

    4. Adjudication Method for the Test Set

    Not applicable as there is no mention of a formal adjudication process for establishing a "ground truth" from expert opinions for the non-clinical bench testing or usability studies. The bench tests rely on objective measurements, and usability studies on user feedback and performance.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical instrument (incentive spirometer, PEP device), not an AI-powered diagnostic or interpretive tool. Therefore, an MRMC study related to human reader improvement with AI assistance is not relevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device does not have an "algorithm" in the sense of AI or image analysis. Its performance is mechanical and measured directly in the bench tests described. The device's function is standalone in its mechanical operation, but it requires human interaction for its intended therapeutic purpose (patient breathing into it).

    7. The Type of Ground Truth Used

    • For PEP and Incentive Spirometer performance (frequency, amplitude, pressure, volume accuracy): The ground truth is based on direct physical measurements using calibrated equipment, compared to the established performance characteristics and specifications of the predicate devices.
    • For Biocompatibility: Ground truth is established by standardized laboratory testing against accepted biological safety norms (cytotoxicity, sensitization, irritation).
    • For Usability: Ground truth is established by observations of user interactions and user feedback against predefined usability metrics and risk analysis requirements.

    8. The Sample Size for the Training Set

    Not applicable. This device is a mechanical medical product. There is no mention or implication of a machine learning or AI model that would require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no ground truth to establish for it.