Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162539

    Validate with FDA (Live)

    Date Cleared
    2016-11-10

    (59 days)

    Product Code
    Regulation Number
    882.5600
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K113455, K123378, K141491

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The Solitaire™ 2 Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.
    2. The Solitaire™ Revascularization Device is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.
    Device Description

    The Solitaire™ 2 Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The Solitaire™ 2 Revascularization Device is designed for use in the neurovasculature such as the Internal Carotid Artery (ICA). M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ 2 Revascularization Device facilitates clot retrieval and has Iridium radiopaque markers on the proximal and distal ends. The devices are supplied sterile and are intended for single-use only.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the Solitaire™ 2 Revascularization Device meets those criteria. Here's a structured breakdown of the requested information:

    Acceptance Criteria and Reported Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document presents two main categories of performance data: biocompatibility and bench testing. Clinical data is used for comparative effectiveness rather than direct acceptance criteria for the device itself in this specific section.

    Biocompatibility Testing:

    Test CategoryTest DescriptionMethodAcceptance CriteriaConclusion
    CytotoxicityL929 MTT CytotoxicityISO 10993-5Viability is ≥70%.Acceptance criteria met
    SensitizationGuinea Pig Maximization SensitizationISO 10993-10Test article does not elicit a sensitization response.Acceptance criteria met
    IrritationIntracutaneous Irritation TestISO 10993-10Differences in the mean test and control scores of the extract dermal observations are < 1.0.Acceptance criteria met
    Acute Systemic ToxicityAcute Systemic Injection TestISO 10993-11No abnormal clinical signs and weight loss in excess of 10%.Acceptance criteria met
    Materials Mediated Rabbit Pyrogen-Temperature rise ≥0.5°C (Note: This looks like a rejection criterion, implies "no temperature rise ≥0.5°C" would be the acceptance)Acceptance criteria met
    Hemo-compatibilityHemolysisISO 10993-4No significant differences between the test article extract and negative control article results. The test article is considered non-hemolytic.Acceptance criteria met
    Partial Thromboplastin Time-Clotting times are similar to the negative control and the reference material (HDPE) indicating the device materials are not an activator of the intrinsic coagulation pathway.Acceptance criteria met
    Platelet and Leukocyte Count-Test article does not adversely affect the platelet and leukocyte components of the blood compared to the reference material.Acceptance criteria met
    Complement Activation C3a and SC5b-9 Assay-Levels of C3a and SC5b-9 are comparable and less than the positive control.Acceptance criteria met
    Thrombosis-Thrombo-resistance properties are acceptable in clinical use.Acceptance criteria met
    GenotoxicityBacterial Mutagenicity TestISO 10993-3Test article is considered non-mutagenic.Acceptance criteria met
    In-vitro Mouse Lymphoma Assay-Test article is considered non-mutagenic.Acceptance criteria met
    In-vivo Mouse Micronucleus Assay-Test article is considered non-mutagenic.Acceptance criteria met

    Performance Bench Testing:

    Test DescriptionMethodAcceptance CriteriaConclusion
    Delivery ForcePeak delivery force was measured through a representative tortuous anatomical model.Stent must be below delivery force specification.Acceptance criteria met
    Re-sheathing ForceRetrieval force was measured through a representative tortuous anatomical model.Stent must be below re-sheathing force specification.Acceptance criteria met
    Total System LengthTotal system length was measured from the proximal tip of the push-wire to the distal-most tip of the finger marker coils.System length must meet product specification.Acceptance criteria met
    Fluorosafe Marker LengthThe total length of the fluorosafe marker was measured.Fluorosafe marker length must meet product specification.Acceptance criteria met
    Multiple Re-sheathing DurabilitySamples were evaluated on their ability to withstand delivery and withdrawal forces in a representative tortuous model beyond the recommended number of passes and re-sheathings allowed per the Instructions for Use (IFU).Device must reliably deploy and resheath up to four times.Acceptance criteria met
    Body Marker TensileBody Marker tensile strength testing is performed to verify the strength of the laser weld of the Platinum/Iridium marker coil to the Nitinol distal finger of the device.Body marker should be greater than or equal to existing tensile strength specification.Acceptance criteria met
    Body Marker RadiopacityVerification analysis of body markers.The radiopaque body markers must be visible using standard catheter laboratory equipment.Acceptance criteria met
    Proximal Marker to Distal MarkerThe length of the laser cut and electro-polished stents are measured 100% in process.Length of stent must meet all inspection criteria.Acceptance criteria met
    Torque ResponseSamples were evaluated to determine the number of turns required to produce 1 rotation of the distal tip of the device.Device turns must be less than or equal to torque response criteria.Acceptance criteria met
    Torque StrengthSamples were evaluated to determine the number of turns required to break the device in a representative tortuous model.Device turns must be greater than torque strength criteria.Acceptance criteria met
    System TensileSamples were evaluated to determine the tensile strength of the full system.Devices must be greater than or equal to the system tensile criteria.Acceptance criteria met
    Af TemperatureIn-process 100% tracking of heat set parameters used to set final Af.Temperature should be less than or equal to existing Af temperature specification.Acceptance criteria met
    Radial ForceThe radial force was measured 100% in-process.Stent must be within existing radial force specification.Acceptance criteria met
    Kink ResistanceKink resistance testing verified the device flexibility across various levels of tortuosity.Device must be able to maintain vessel wall apposition at a minimum radius bend.Acceptance criteria met
    Clot Removal in a Simulated Neurovascular ModelBoth hard and soft clot retrieval success was evaluated in an in vitro tortuous anatomical model.Overall success rate of the device (usability and effectiveness) must be equal to or better than the predicate* device.Acceptance criteria met
    Particulate Under Simulated Use ConditionsDevice was evaluated for particulate generation under simulated use in a representative tortuous anatomical model per USP<788>.Device was evaluated for particulate generation under simulated use in a representative tortuous anatomical model per USP<788> (Same as method, implying conforming to USP<788> limits).Acceptance criteria met

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Clinical Study (SWIFT PRIME):
      • Total Randomized Subjects: 196 (98 in each group: IV t-PA alone vs. IV t-PA + Solitaire)
      • Analysis Cohort (after exclusions): 161 subjects (84 in the IV t-PA + Solitaire™ group and 77 with IV t-PA only). Further refined to 144 subjects for primary and secondary efficacy endpoints after additional exclusions.
      • Data Provenance: Global, multicenter. The study was a prospective, randomized, open, blinded endpoint (PROBE) clinical study (IDE G120142).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not explicitly state the exact number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the clinical study's endpoints. However, it does mention:

    • Blinded evaluation of modified Rankin Scale (mRS) for neurological disability outcomes. This implies that the mRS scores, which serve as a critical component of the ground truth for effectiveness, were assessed by experts who were blinded to the treatment arm.
    • Clinical Events Committee adjudication for adverse events. This indicates a panel of experts reviewed and categorized adverse events.
    • Core Laboratory assessed data for symptomatic ICH, infarct volume, and reperfusion ratio. This suggests specialized facilities with expert staff were responsible for these assessments.

    While the specific count and detailed qualifications are not provided, the involvement of blinded evaluators, a Clinical Events Committee, and a Core Laboratory indicates that ground truth was established through expert assessment according to established protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document mentions "Clinical Events Committee adjudication" for adverse events. While it doesn't specify a 2+1 or 3+1 method, "adjudication" implies a process where a committee of experts reviews and resolves discrepancies in event classification or assessment. It suggests a structured review by multiple parties, but the exact number of reviewers per case or tie-breaking mechanism is not detailed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done, or any mention of AI assistance. This study compared a medical device (Solitaire™ 2 Revascularization Device) with or without IV t-PA, not an AI system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done, as this document is about a mechanical thrombectomy device, not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the clinical study was established using a combination of:

    • Clinical Outcomes Data: Primarily, the 90-day global disability assessed via the blinded evaluation of Modified Rankin Scale (mRS). This is a widely accepted functional outcome scale often based on trained interviewer assessment.
    • Imaging-based Assessments: Volume of cerebral infarction, reperfusion ratio, and arterial revascularization (TICI 2b or 3) assessed by a Core Laboratory. This implies expert interpretation of medical images.
    • Adjudicated Adverse Events: Reviewed and categorized by a Clinical Events Committee.

    8. The sample size for the training set

    The document does not describe the development or training of an algorithm or AI. Therefore, there is no training set sample size mentioned. The clinical study (SWIFT PRIME) is a comparative effectiveness study for a medical device in patients.

    9. How the ground truth for the training set was established

    As there is no training set for an algorithm or AI described in the document, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1