Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K143195

    Validate with FDA (Live)

    Date Cleared
    2014-12-22

    (46 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K132026, K123074, K132037, K092449

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Passport 17M and Passport 12M patient monitors are intended for monitoring. displaying, reviewing, and transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-mvasive blood pressure (NIBP), invasive blood pressure(IBP), pulmonary artery wedge pressure (PAWP), cardiac output (C.O.), continuous cardiac output (CCO), mixed/central venous oxygen saturation (Sy()//Scy(), carbon dioxide (CO2), Oxygen (O2), anesthetic gas (AG), impedance cardiograph (ICG), bispectral index (BIS), and respiration mechanics (RM). The equipment also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

    • . The arrhythmia detection. ST Segment analysis of Mortara algorithm. BIS, RM, CCO. SvOz/Scv02, and PAWP monitoring are intended for adult and pediatric patients only:
    • ST Segment analysis of Mindray algorithm is intended for adult patients only; ●
    • C.O. monitoring is restricted to adult patients only: ●
    • ICG monitoring is only for use on adult patients who meet the following ● requirements: height: 122 to 229cm, weight: 30 to 155kg.

    The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

    The T1 Patient Monitor is intended for monitoring, displaying, reviewing, storing, and transferring of multiple physiological parameters including ECG (3-lead, or 12-lead selectable, arrhythmia detection, ST Segment analysis, and heart rate (HR)), respiration (Resp), temperature (Temp), pulse oxygen saturation (SpOz), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), pulmonary artery wedge pressure (PAWP), and carbon dioxide (CO2). The equipment also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

    The arrhythmia detection, ST Segment analysis of Mortara algorithm, and PAWP monitoring are intended for adult and pediatric patients;

    ST Segment analysis of Mindray ECG algorithm is intended for adult patients only;

    The monitor is to be used in healthcare facilities by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for helicopter transport, hospital ambulance, or home use.

    Device Description

    The subject Passport Series Patient Monitors includes three monitors:

    • . Passport 12M Patient Monitor
    • Passport 17M Patient Monitor ●
    • T1 Patient Monitor

    All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters. The Passport 12M and 17M Patient monitors provide patient monitoring capabilities based on the user selected modules that are plugged into the main monitor. The T1 patient monitor is one of the available modules that can be plugged into the Passport 17M or 12M monitor. The T1 can also be used as a standalone monitor and when used as a standalone monitor, it provides a subset of the functions that are provided by Passport 17M and 12M.

    Each of the three monitors that are the subject of this 510(k) are modified versions of previously cleared Mindray devices. The Passport 17M and Passport 12M were previously cleared in K132075, on April 18, 2014 and the T1 was previously cleared in K123074 on July 3, 2013.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Mindray Passport Series Patient Monitors (K143195). This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a single study proving that the device meets those criteria.

    The core of the submission is that the subject devices are modified versions of previously cleared Mindray devices (K132075 and K123074). The performance data cited refers to functional and system-level testing to evaluate the performance of the modifications and ensure continued adherence to specifications and relevant consensus standards. It does not contain information about a specific single study with acceptance criteria and results.

    Given the nature of the K143195 submission as an amendment/modification to existing devices, the information provided focuses on the changes and how these changes do not alter the substantial equivalence to the predicate devices. Therefore, the requested information about acceptance criteria and a detailed study proving the device meets those criteria for the entire device as if it were a novel submission is not explicitly present in the provided text.

    However, I can extract information related to the performance characteristics, which serve as implicit acceptance criteria based on the predicate device's performance and general recognized standards for patient monitors.

    Here's an attempt to structure the answer based on the available information, noting where specific details are not provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal table of acceptance criteria with corresponding performance results in a standalone "study summary" format. Instead, it compares the specifications of the subject devices to the predicate device. The "acceptance criteria" are implied to be the specifications of the predicate device and relevant industry standards. The "reported device performance" is generally stated as "Same" or that the new (modified) features "meet specifications" or "are equivalent to the predicate device."

    Here's an illustrative table based on the provided "Device Comparison Table," highlighting areas that could be inferred as "acceptance criteria" and "performance":

    Feature CategoryParameterImplicit Acceptance Criteria (from Predicate Device K132075)Reported Subject Device Performance (K143195)
    GeneralIndications for UseMonitoring, displaying, reviewing, alarming, and transferring multiple physiological parameters for adult, pediatric, and neonatal patients (with specific exceptions).For Passport 12M/17M, "Same" as predicate device. For T1, Indications are a "subset" of the predicate. New feature: 12-lead ECG interpretation, noted as present in cleared Mindray V Series (K132026), implying equivalence.
    DisplayIntegrated display and touchscreen (17M)17", 1280*1024 pixelsSame
    Integrated display and touchscreen (12M)12", 800*600 pixelsSame
    Integrated display and touchscreen (T1)(Predicate K123074: 5", 480*272 pixels)5", 480*272 pixels (same as K123074)
    Secondary display (Passport 12M/17M)Independent control and display (17M), Linked to integrated display (12M)Same
    Secondary display (T1)N/A in predicate K132075Independent display and control via a VGA port in the T1 docking station (new feature for T1)
    ConnectivityWireless module (Passport 17M)ASUS module for network connectionSame
    Wireless module (Passport 12M)ASUS module for network connectionAdded support for Silex wireless module. "Mindray conducted Wireless functionality testing to ensure the performance of the new wireless modules meet specifications and are equivalent to the predicate device."
    Wireless module (T1)N/A in predicate K132075Added wireless function using the Cyberlink wireless module. "Mindray conducted Wireless functionality testing to ensure the performance of the new wireless modules meet specifications and are equivalent to the predicate device."
    Physiological ParamsECG (features)3-lead, 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, heart rateSame
    12-lead ECG interpretationNot supported (Predicate K132075)Supported (All subject devices). Justified by equivalence to K132026. Performance not detailed here.
    ECG J-point auto detectionSupportedSame (17M, 12M). Added (T1). Performance not detailed.
    Respiration rate (Resp) Measurement & AccuracyAdult: 0-120 rpm, Pediatric/Neonate: 0-150 rpm. Accuracy: 7-150 rpm: ±2 rpm or ±2%, whichever is greater; 0-6 rpm: Not specified.Same
    Temperature (Temp) Measurement & Accuracy0-50°C (32-122°F). Accuracy: ±0.1°C or ±0.2°F (without probe).Same
    SpO2 Measurement & Accuracy (Mindray SpO2)Range: 0-100%. Accuracy: 70-100%: ±2% (adult/pediatric), ±3% (neonate); 0-69%: Not specified.Same
    SpO2 Measurement & Accuracy (Masimo SpO2)Range: 1-100%. Accuracy: 70-100%: ±2% (no motion, adult/pediatric), ±3% (no motion, neonate), ±3% (with motion); 1-69%: Not specified.Same
    SpO2 Measurement & Accuracy (Nellcor SpO2)Range: 0-100%. Accuracy: 70-100%: ±2% (adult/pediatric), ±3% (neonate); 0-69%: Not specified.Same
    Pulse rate (PR) from SpO2 ModuleMindray SpO2: 20-254 bpm, ±3 bpm. Masimo SpO2: 25-240 bpm, ±3 bpm (no motion), ±5 bpm (with motion). Nellcor SpO2: 20-300 bpm, 20-250 bpm: ±3 bpm.Same
    Pulse rate (PR) from IBP Module25-350 bpm, ±1 bpm or ±1%.Same
    NIBP Measurement & AccuracyAdult Systolic: 40-270 mmHg, Diastolic: 10-210 mmHg, Mean: 20-230 mmHg. Pediatric Systolic: 40-200 mmHg, Diastolic: 10-150 mmHg, Mean: 20-165 mmHg. Neonate Systolic: 40-135 mmHg, Diastolic: 10-100 mmHg, Mean: 20-110 mmHg. Accuracy: Max mean error: ±5 mmHg, Max standard deviation: 8 mmHg.Same
    IBP Measurement & AccuracyRange: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor).Same
    Pulse Pressure Variation (PPV)Supported feature of IBP (Passport 12M/17M)Same (17M, 12M). Added (T1).
    CO2 Measurement & AccuracySidestream: 0-99 mmHg; Accuracy: 0-40 mmHg: ±2 mmHg, 41-76 mmHg: ±5%, 77-99 mmHg: ±10%. Microstream: 0-99 mmHg; Accuracy: 0-38 mmHg: ±2 mmHg, 39-99 mmHg: ±5% +0.08% of (reading-38). Mainstream: 0-150 mmHg; Accuracy: 0-40 mmHg: ±2 mmHg, 41-70 mmHg: ±5%, 71-100 mmHg: ±8%, 101-150 mmHg: ±10%.Specifications unchanged. Modules repackaged to occupy a single slot. (All subject devices).
    Other3 Additional SpO2 SensorsNone (Predicate K132075)Added LNCS NeoPt, LNCS Neo, LNCS Inf. Cleared in K132037 (implies their performance against specific standards is already established).
    4 New NIBP cuffsNone (Predicate K132075)Added CM1306, CM1307, CM1506, CM1507. Substantial equivalent to NIBP cuffs cleared in K132075 and K092449 (implies their performance against specific standards is already established).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not detail specific sample sizes for test sets for individual parameters or features. It states:

    • "Mindray conducted functional and system level testing on the subject devices. The testing provided an evaluation of the performance of the device relevant to each of the modifications to the subject devices since their previous clearance."
    • "Mindray has conducted testing to ensure the subject devices meet relevant consensus standards."
    • "Mindray has also followed the FDA Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm issues on October 28, 2003."
    • "Mindray conducted Wireless functionality testing to ensure the performance of the new wireless modules meet specifications and are equivalent to the predicate device."

    There is no mention of data provenance (e.g., country of origin, retrospective/prospective). The studies appear to be internal V&V testing.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    There is no information provided regarding the number or qualifications of experts used to establish ground truth for any test sets. The testing referenced appears to be engineering/technical validation against specifications and standards rather than clinical studies requiring expert adjudication of data. For interpretations like 12-lead ECG, it is stated that the feature is equivalent to a previously cleared device (K132026), implying that its validation methods would have been established there.


    4. Adjudication Method for the Test Set

    No adjudication method is described.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No MRMC comparative effectiveness study is mentioned. This type of study is typically done for diagnostic imaging devices or algorithms where human interpretation is a key component and the AI is intended to assist, altering reader performance. For a patient monitor, the performance validation is typically against direct physiological measurements and established standards.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The entire device (the patient monitor) is a standalone system in the sense that its algorithms process physiological signals and present data and alarms. The performance data described ("functional and system level testing") would inherently be evaluating the algorithms and hardware in isolation from a human interpretation loop. For example, SpO2 accuracy is measured against a reference, not against a human's interpretation of SpO2. The 12-lead ECG interpretation, being an automated algorithm output, would be assessed for its accuracy as a standalone function.


    7. The Type of Ground Truth Used

    Based on the nature of the device (physiological monitor) and the parameters listed, the ground truth would typically be established by:

    • Reference standards/equipment: For parameters like NIBP, IBP, SpO2, Temperature, CO2, etc., the device's measurements are compared against highly accurate reference measurement equipment or simulators.
    • Known input signals: For features like arrhythmia detection, standardized ECG waveforms with known arrhythmias are often used.
    • Clinical data: For capabilities like 12-lead ECG interpretation, ground truth is usually established by highly experienced cardiologists' interpretations, though the document only refers to equivalence to a previously cleared device for this feature.
    • Specifications: The "acceptance criteria" are derived from the performance specifications of the predicate device and relevant international standards.

    8. The Sample Size for the Training Set

    The document does not describe the development or training of AI/ML algorithms. It is a 510(k) for hardware and software modifications to an existing patient monitor, rather than a novel AI/ML device. Therefore, no information on training set sample size is provided. The functions described (e.g., arrhythmia detection, ST segment analysis) typically rely on established signal processing algorithms, not necessarily machine learning requiring explicit "training sets" in the modern sense.


    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned for AI/ML algorithms, this information is not applicable and not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133973

    Validate with FDA (Live)

    Date Cleared
    2014-08-14

    (231 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K130773, K123074, K101619, K131929, K120366

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for: 3 & 5 Lead ECG monitoring; 12 Lead ECG recording; real-time arrhythmia detection / alarming; QT measurement / alarming and ST measurements / alarming: impedance pneumography: noninvasive blood pressure (NIBP); end-tidal CO2 (ETCO2) and respiration rate; pulse oximetry (SpO2); contact temperature; and invasive pressure.

    The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).

    The device is indicated for adults, paediatrics and neonates.

    Device Description

    The Tempus Pro is a multi-parameter vital signs monitor designed for use in pre-hospital care and remote clinical locations by trained healthcare professionals. It provides 3&5 lead ECG monitoring, 12 lead ECG recording, real-time arrhythmia detection / alarming, QT measurement / alarming and ST measurements / alarming pulse oximetry, non-invasive blood pressure, sidestream capnometry, 2 channels of contact temperature, impedance respiration, up to 4 channels of invasive pressure and user configurable alarms.

    In addition, it provides the ability to transmit all vital signs data via wired or wireless connections to a software system (called i2i) expected to be based in a facility far from the user e.g. a response centre facility. In addition to sending all vital signs, the system can also send pictures or video via an integrated camera, geographic position by an integrated GPS receiver and voice via a wired or wireless headset.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Tempus Pro Patient Monitor with Enhanced Features, specifically the addition of arrhythmia detection/alarming, QT measurement/alarming, and ST measurements/alarming.

    Here's an analysis of the acceptance criteria and study information, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document mentions that the Arrhythmia detection, ST elevation, and QT interval measurements were evaluated in accordance with FDA guidance on arrhythmia detection and alarm, using the FDA recognized consensus standard EC57 (FDA Rec No: 3-118). However, it does not provide a specific table of acceptance criteria or quantitative performance metrics for these features. It only states: "A summary report demonstrating compliance with the guidance and associated standard is presented as part of this 510(k) submission." and "In each case the results of this testing confirmed that acceptance criteria defined by the relevant standard, or other appropriate reference document had been met."

    Without access to the "summary report" or the full 510(k) submission, we cannot extract specific numerical acceptance criteria or reported device performance for the new features. The table below reflects the information that is available.

    Feature / MetricAcceptance CriteriaReported Device Performance
    Arrhythmia Detection / AlarmingCompliance with FDA guidance on arrhythmia detection and alarm, and consensus standard EC57 (FDA Rec No: 3-118)."Results of this testing confirmed that acceptance criteria defined by the relevant standard... had been met." (No specific numerical performance provided in this document).
    ST Measurement / AlarmingCompliance with FDA guidance on arrhythmia detection and alarm, and consensus standard EC57 (FDA Rec No: 3-118)."Results of this testing confirmed that acceptance criteria defined by the relevant standard... had been met." (No specific numerical performance provided in this document).
    QT Measurement / AlarmingCompliance with FDA guidance on arrhythmia detection and alarm, and consensus standard EC57 (FDA Rec No: 3-118)."Results of this testing confirmed that acceptance criteria defined by the relevant standard... had been met." (No specific numerical performance provided in this document).
    General Device ParametersOperate to specification across stated performance and temperature range. Ensure all data is transmitted reliably and accurately.Confirmed by bench testing.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states that the evaluation was performed in accordance with "FDA guidance on arrhythmia detection and alarm" and "the FDA recognized consensus standard EC57 (FDA Rec No: 3-118), which defines methods of presenting results when testing against specified ECG databases."

    However, the provided text does not specify the sample size (e.g., number of patients, number of ECGs) used for the test set. It also does not specify the data provenance (e.g., country of origin, retrospective or prospective) of the ECG databases used. EC57 typically refers to standard databases like MIT-BIH Arrhythmia Database, but this is not explicitly stated in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on the number of experts used to establish ground truth or their specific qualifications for the test set. For standard ECG databases (like those referenced by EC57), ground truth is typically established by multiple expert cardiologists.

    4. Adjudication Method for the Test Set

    The document does not specify the adjudication method used for the test set. If standard ECG databases were used, the ground truth annotations would have been established prior to the device under test being evaluated.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to assess how much human readers improve with AI vs without AI assistance. The focus of this 510(k) appears to be on the device's standalone performance in arrhythmia detection and measurement against established standards.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    Yes, a standalone study was done. The document states that "Arrhythmia detection and alarms, ST elevation and QT interval measurements... were evaluated in accordance with FDA guidance on arrhythmia detection and alarm... This guidance calls for use of the FDA recognized consensus standard EC57 (FDA Rec No: 3-118), which defines methods of presenting results when testing against specified ECG databases." This indicates the algorithm's performance was evaluated independently against pre-annotated databases, which is a standalone assessment.

    7. Type of Ground Truth Used

    The ground truth used would be expert consensus annotations as provided in the "specified ECG databases" referenced by the EC57 standard. These databases contain ECG recordings with validated annotations of arrhythmias and other cardiac events, established by a panel of cardiologists.

    8. Sample Size for the Training Set

    The document does not provide any information regarding the sample size used for the training set of the algorithms. It is possible the algorithms were pre-trained and integrated, or that general ECG databases were used for training, but this is not detailed.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide any information on how the ground truth for the training set was established. If standard ECG databases were used for training, the ground truth would have been established by expert consensus annotations within those databases.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1