Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K142983

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-03-04

    (139 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K122685, K133998

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA Scoring Balloon Catheter is intended for dilatation of lesions in the illiac, femoral, ilio-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt PTA Scoring Balloon Catheter with HydroCross™ Coating is a standard twolumen catheter with a scoring balloon near the distal end of the catheter has a conventional nylon-blend balloon with a scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure. As shown below, the catheter has a segment which is coated with a hydrophilic coating (HydroCross™ Coating).

    AI/ML Overview

    This section describes the acceptance criteria and study proving the device meets said criteria for the "AngioSculpt® PTA Scoring Balloon Catheter with HydroCross™ Coating" based on the provided text.

    Based on the provided document, the device described is a medical device, and the information presented is a 510(k) summary for premarket notification to the FDA, demonstrating substantial equivalence to a predicate device. This document does not describe the acceptance criteria and a study in the context of an AI/algorithm-driven device's performance metrics (e.g., sensitivity, specificity) against a ground truth established by experts.

    Instead, the "acceptance criteria" discussed are related to the safety and effectiveness of a physical medical device, primarily through mechanical testing and biocompatibility testing, to demonstrate its substantial equivalence to a previously cleared predicate device.

    Therefore, many of the requested categories for AI/algorithm performance studies are not applicable to this document. I will fill in the applicable information based on the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with specific numerical targets and corresponding device performance values for each criterion in the manner typical for AI/algorithm validation. Instead, it describes categories of tests performed to verify design outputs met design requirements and to confirm proper function and durability. The general outcome is that the device "demonstrated adequate device performance," and is "non-sensitizing, non-irritating, not systemically toxic, non-hemolytic, and non-mutagenic."

    Acceptance Criteria Category (Evaluated Aspect)Reported Device Performance (Summary)
    Mechanical Performance
    Catheter Diameter and Balloon ProfileDesign outputs met design requirements; proper function and durability confirmed.
    Device Preparation, Deployment, and RetractionDesign outputs met design requirements; proper function and durability confirmed.
    Balloon Inflation and Deflation TimeDesign outputs met design requirements; proper function and durability confirmed.
    Catheter Bond (Tensile) StrengthDesign outputs met design requirements; proper function and durability confirmed.
    Tip Pull StrengthDesign outputs met design requirements; proper function and durability confirmed.
    Torque StrengthDesign outputs met design requirements; proper function and durability confirmed.
    Pushability, Trackability, Secure EdgesDesign outputs met design requirements; proper function and durability confirmed.
    Guidewire CompatibilityDesign outputs met design requirements; proper function and durability confirmed.
    Catheter Effective LengthDesign outputs met design requirements; proper function and durability confirmed.
    Catheter Surface AppearanceDesign outputs met design requirements; proper function and durability confirmed.
    Luer CompatibilityDesign outputs met design requirements; proper function and durability confirmed.
    Coating LengthDesign outputs met design requirements; proper function and durability confirmed.
    Coating ThicknessDesign outputs met design requirements; proper function and durability confirmed.
    Coating LubricityDesign outputs met design requirements; proper function and durability confirmed.
    Coating IntegrityDesign outputs met design requirements; proper function and durability confirmed.
    Particulate EvaluationDesign outputs met design requirements; proper function and durability confirmed.
    Rated Burst PressureDesign outputs met design requirements; proper function and durability confirmed.
    ComplianceDesign outputs met design requirements; proper function and durability confirmed.
    Device FatigueDesign outputs met design requirements; proper function and durability confirmed.
    Flexibility and KinkDesign outputs met design requirements; proper function and durability confirmed.
    Biocompatibility
    CytotoxicityNon-cytotoxic (MEM Elution Test using L-929 Mouse Fibroblast cells).
    SensitizationNon-sensitizing (Guinea Pig Maximization Sensitization Test).
    IrritationNon-irritating (Intracutaneous Reactivity/Irritation Test).
    Systemic Toxicity (Acute & Material-Mediated)Not systemically toxic (Acute Systemic Injection Test, Material Mediated Pyrogens).
    HemocompatibilityNon-hemolytic (Partial Thromboplastin Time (PTT), Hemolysis Direct Contact/Extract Method, Direct Contact Complement Activation).
    Thrombogenicity (in vivo)No thrombo-embolism observed in GLP animal study under simulated use conditions.
    GenotoxicityNon-mutagenic (Reverse Bacterial Mutation, In vitro Mouse Lymphoma Assay-Extended Treatment, Rodent bone Marrow Micronucleus Assays).

    Note: The document states that "Design verification and validation testing demonstrated adequate device performance and confirmed that no new questions of safety or effectiveness for peripheral balloon angioplasty devices were raised."

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: The document refers to "Test articles consisted of finished sterilized catheters" for bench testing. It does not specify the exact number of devices tested for each mechanical or biocompatibility test. For biocompatibility, it indicates studies were conducted, including a GLP animal study for thrombogenicity.
    • Data Provenance: The studies were conducted by Spectranetics, Inc. (the submitter). The context implies these were internal tests conducted to support the 510(k) submission. No information on country of origin of data (beyond the applicant being US-based) or retrospective/prospective nature is explicitly stated, though a GLP animal study would be prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable to this type of device submission. The "ground truth" for a physical medical device's performance is established by standardized testing methods and regulatory requirements, not expert consensus on diagnostic interpretations.

    4. Adjudication method for the test set:

    • Not applicable for this type of device validation. Adjudication methods are typically for evaluating human or AI performance against a "truth" in diagnostic settings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a physical medical instrument (a balloon catheter), not an AI diagnostic tool that assists human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical medical instrument, not an AI algorithm.

    7. The type of ground truth used:

    • The "ground truth" for the performance of this physical medical device is based on engineering specifications, established standards, and regulatory requirements for mechanical integrity, functional performance, and biocompatibility. For example, for biocompatibility, the ground truth is "non-toxic," "non-irritating," etc., as defined by ISO 10993-1:2009 and FDA GLP regulations. For mechanical tests, the ground truth is that the device must meet predefined design requirements and exhibit proper function and durability.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device requiring a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K133998

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-04-18

    (113 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K112182/K122685, K013459

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioSculpt PTA.Scoring Balloon Catheter is intended for dilatation of lesions in the iliac, femoral, lito-femoral, popliteal, infra popliteal, and renal arteries, and for the treatment of obstructive lestons of native or synthelic anteriovenous dialysis fistulae. Not for use in the coronary or neuro-vasculature.

    Device Description

    The AngioSculpt PTA Scoring Balloon Catheter with HydroCross™ Coating is a standard twolumen catheter with a scoring balloon near the distal end of the catheter has a conventional nylon-blend balloon with a scoring element that wraps around the balloon. The scoring element creates focal concentrations of dilating force which minimizes balloon slippage and assists with luminal expansion of stenotic arteries. The balloon has radiopaque markers to aid in positioning the balloon in the stenosis, and is designed to provide an expandable segment of known diameter and length at a specific pressure. As shown below, the catheter has a segment which is coated with a hydrophilic coating (HydroCross™ Coating).

    AI/ML Overview

    The provided text describes a medical device, the AngioSculpt® PTA Scoring Balloon Catheter with HydroCross™ Coating, and its clearance process with the FDA. However, the document does NOT contain information about acceptance criteria or a study proving the device meets specific performance criteria in the way typically expected for an AI/ML medical device submission (e.g., sensitivity, specificity, F1-score, AUC, etc., derived from clinical data).

    Instead, this document is a 510(k) summary for a physical medical device (a balloon catheter) seeking substantial equivalence to already marketed devices. The "studies" described are primarily bench testing (mechanical, material properties) and biocompatibility testing, and an animal study to demonstrate safety and deliverability. These types of tests are standard for physical devices to ensure they perform as intended and are safe for use.

    Therefore, many of the requested fields regarding acceptance criteria related to a device's performance in diagnosing or predicting (like sensitivity/specificity) and specific details about ground truth, expert readers, MRMC studies, or training sets for AI algorithms are not applicable to this document.

    Here's a breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a physical medical device, the "acceptance criteria" are not reported as statistical performance metrics like sensitivity or specificity. Instead, they are defined by successful completion of various engineering and biological tests, demonstrating the device meets design requirements and safety standards. The "reported device performance" is the successful outcome of these tests.

    Acceptance Criteria CategorySpecific Tests ConductedReported Device Performance
    Mechanical Performance (Bench Testing)- Catheter Diameter and Balloon ProfileDesigned to meet design outputs and requirements, confirming proper function and durability.
    - Minimum Burst Strength (RBP)
    - Balloon Compliance (Diameter vs. Pressure)
    - Balloon Inflation and Deflation Time
    - Device Fatigue
    - Bond (Tensile) Strength
    - Tip Pull Strength
    - Catheter Diameter and Balloon Profile (with Scoring Element)
    - Flexibility and Kink
    - Torque Strength
    - Pushability, Trackability and Secure Edges
    - Balloon Preparation, Deployment and Retraction
    - Freedom from Stent Interference
    - Focal Force
    - Corrosion Resistance
    - Coating Length, Thickness, Lubricity, Integrity
    - Particulate Evaluation
    Biocompatibility- Cytotoxicity (MEM Elution Test)Non-cytotoxic, non-irritating, not systemically toxic, non-hemolytic, non-mutagenic. No thrombo-embolism observed in GLP animal study.
    - Sensitization (Guinea Pig Maximization Sensitization Test)
    - Irritation (Intracutaneous Reactivity Test)
    - Systemic Toxicity (Acute Systemic Injection Test, Material Mediated Pyrogens)
    - Hemocompatibility (Partial Thromboplastin Time, Hemolysis Direct Contact/Extract Method, Direct Contact Complement Activation Testing, Thrombosis (in vivo))
    - Genotoxicity (Reverse Bacterial Mutation, In vitro Mouse Lymphoma Assay, Rodent bone Marrow Micronucleus Assays)
    Animal Study (Safety & Deliverability)- Successful introduction with standard guidewires (6F sheath or 7F guide catheter).Successfully introduced and expanded in target tissue; No evidence of dissection, perforation, or embolization. Maintained integrity with no loss of components. Procedures performed with ease. No adverse events. Catheters found to be clinically acceptable in peripheral (femoral) arteries.
    - Expansion in targeted tissue.
    - No evidence of dissection, perforation, or embolization.
    - Device integrity maintained with no loss of components.
    - No adverse events observed.

    2. Sample Size for Test Set and Data Provenance

    • Bench Testing: "Test articles consisted of finished sterilized catheters." The exact number of catheters tested for each bench test is not specified in this summary.
    • Biocompatibility Testing: The number of biological samples or animals used for each specific biocompatibility test (e.g., L-929 Mouse Fibroblast cells for cytotoxicity, Guinea pigs for sensitization) is not specified.
    • Animal Study: The sample size (number of animals) for the acute GLP animal study is not explicitly stated.
    • Data Provenance: The document does not specify country of origin for the data; it refers to tests conducted in accordance with FDA GLP regulations (21 CFR Part 58) and ISO standards for biocompatibility. The animal study was conducted as a GLP study. Given the company is US-based (Fremont, CA), it's highly likely the testing was performed in the USA or by labs adhering to these international/US standards. All studies sound prospective as they were specifically designed to characterize this device.

    3. Number of Experts and Qualifications for Ground Truth

    • Not Applicable: This type of information is generally relevant for studies establishing clinical ground truth (e.g., presence of a disease from imaging) for interpretive devices. For this physical device, "ground truth" is established by direct physical, chemical, and biological measurements, and observed adverse events (or lack thereof) in an animal model. The "experts" are likely laboratory technicians and veterinarians performing and assessing the animal study, certified in their respective fields, but not defined in terms of their role as "ground truth adjudicators" in a clinical sense.

    4. Adjudication Method for Test Set

    • Not Applicable: Adjudication methods like 2+1 or 3+1 are used to resolve disagreements among multiple human readers when establishing ground truth, typically in the context of diagnostic performance studies. This is not descriptive of the testing performed for a physical device. Outcomes of bench tests are based on objective measurements against specifications. The animal study outcomes would be assessed by the study's veterinary and pathology team.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No: No MRMC study was performed or described. This type of study is relevant for comparing the diagnostic performance of an AI system, with and without human assistance, against human readers. The AngioSculpt catheter is a physical interventional device, not an AI diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    • Not Applicable: There is no algorithm-only performance study because this is a physical medical device, not a software algorithm.

    7. Type of Ground Truth Used

    • Bench Testing: The ground truth for bench tests is objective physical and chemical measurements against pre-defined engineering specifications and accepted industry standards.
    • Biocompatibility Testing: The ground truth is based on established biological responses (e.g., cell viability, immune response, genetic mutation) as measured by validated assays against accepted toxicity limits and ISO standards.
    • Animal Study: The ground truth is direct observation by trained personnel (veterinarians, histopathologists) of in-vivo events and tissue responses in an animal model, assessing for safety endpoints like dissection, perforation, embolization, and device integrity.

    8. Sample Size for Training Set

    • Not Applicable: No "training set" in the context of machine learning was used or described. The device's design and performance are based on engineering principles, material science, and iterative testing, not on training an AI model with data.

    9. How Ground Truth for Training Set Was Established

    • Not Applicable: As there is no training set, this question is not relevant.