510(k) Data Aggregation

    K Number
    K131068

    Validate with FDA (Live)

    Date Cleared
    2013-08-27

    (132 days)

    Product Code
    Regulation Number
    872.3250
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K122421

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TEETHMATE DESENSITIZER is indicated for reduction of tooth hypersensitivity by the following treatments:

    [1] Treatment of dentin exposed by toothbrush abrasion, gingival recession, periodontal disease and/or acid erosion

    [2] Treatment of dentin after mechanical tooth cleaning, scaling and/or root planing

    [3] Treatment of tooth surface before and/ or after bleaching

    [4] Treatment of prepared dentin for fillings and/or prosthetic restorations

    Device Description

    The subject device consists of POWDER which contains Calcium Phosphate as a major component and LIQUID which mainly contains water. POWDER and LIQUID are mixed to obtain paste. The paste applied to the affected area transforms to a hardened material of hydroxyapatite and it suppresses hypersensitivity by sealing the dentin tubules and microcracks in the enamel.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called TEETHMATE DESENSITIZER. This submission is for a partial change to a predicate device, specifically altering the wording of one "Indications for use" statement.

    The core of the submission argues for substantial equivalence to the predicate device, K122421, rather than presenting a new study with explicit acceptance criteria and device performance metrics in the format requested. The document states that the effectiveness and performance are "substantially equivalent" because the chemical ingredients, hypersensitivity-suppressing mechanism, and technological characteristics are the same as the predicate device.

    Therefore, many of the requested fields cannot be directly extracted from the provided text as it focuses on demonstrating equivalence rather than reporting on a new performance study with specific acceptance criteria and detailed study methodology for the modified device.

    However, based on the information provided, here's what can be inferred and explicitly stated:

    1. Table of Acceptance Criteria and Reported Device Performance

    • Acceptance Criteria: The primary acceptance criterion for this 510(k) submission is to demonstrate substantial equivalence to the predicate device (K122421) under the modified indication for use. This means demonstrating that the device is as safe and effective as a legally marketed device. The specific criteria for demonstrating this substantial equivalence are outlined by the FDA's 510(k) regulations, which in this case, focus on:
      • Same intended uses (with the minor modification in wording)
      • Same chemical ingredients/safety
      • Same technological characteristics/effectiveness and performance
      • Same biocompatibility
    • Reported Device Performance: The document doesn't provide quantitative performance metrics for "TEETHMATE DESENSITIZER" against a new set of acceptance criteria. Instead, it asserts its performance is equivalent to the predicate.
      • It does mention two specific investigations regarding the effect of the modification:
        • "it was confirmed that there was no difference in sealing performance between before and after bleaching teeth when investigated the ability of the subject device to seal microcracks in the enamel before and after bleaching."
        • "it was confirmed that the subject device used before bleaching did not affect the bleaching performance when investigated the effect of the subject device on the performance of the bleaching material."
    Acceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance (as stated in the document)
    Intended Uses: Maintain substantial equivalence even with modified wording for indication [3] ("Treatment of tooth surface before and/or after bleaching").Intended Uses: Substantially equivalent to the predicate; the modified wording [3] does not substantially affect effectiveness and performance.
    Chemical Ingredients/Safety: Maintain identical chemical composition and safety profile as the predicate device.Chemical Ingredients/Safety: The chemical composition is the same as the predicate device, thus equivalent in terms of safety.
    Technological Characteristics/Effectiveness and Performance: Maintain identical technological characteristics, effectiveness, and performance.Technological Characteristics/Effectiveness and Performance: Contains same chemical ingredients and uses same hypersensitivity-suppressing mechanism as predicate; technological characteristics/effectiveness and performance are the same.
    Sealing Performance (before vs. after bleaching): No difference in sealing microcracks in enamel whether applied before or after bleaching.No difference in sealing performance between before and after bleaching teeth (when sealing microcracks in the enamel).
    Effect on Bleaching Performance: The device used before bleaching should not adversely affect the bleaching performance.The subject device used before bleaching did not affect the bleaching performance.
    Biocompatibility: Maintain identical biocompatibility as the predicate device.Biocompatibility: All chemical ingredients are the same as the predicate; biocompatibility already shown for predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified. The document refers to "investigated the ability of the subject device to seal microcrecks" and "investigated the effect of the subject device on the performance of the bleaching material," but does not provide details on the sample sizes used for these investigations.
    • Data Provenance: Not explicitly stated. Given the context of a 510(k) submission from "Kuraray Noritake Dental Inc." which is based in Japan, it's possible the testing was conducted in Japan, but this is not confirmed. The study type appears to be internal verification.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Not applicable/Not specified. The document describes internal investigations rather than clinical studies involving expert consensus on ground truth.

    4. Adjudication Method for the Test Set

    • Not applicable/Not specified.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI/software device, and no MRMC study was mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a dental desensitizing material, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the limited investigations mentioned (sealing performance and effect on bleaching) would likely be in-vitro laboratory measurements or potentially small-scale in-vivo observations. The document states, "it was confirmed that there was no difference in sealing performance..." and "...it was confirmed that the subject device used before bleaching did not affect the bleaching performance...". This suggests direct measurement of physical properties (e.g., microcrack sealing) and functional outcomes (e.g., bleaching efficacy).

    8. The sample size for the training set

    • Not applicable. This is a partial change application for a physical dental device. There is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    • Not applicable. As above, there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K051084

    Validate with FDA (Live)

    Date Cleared
    2005-06-13

    (46 days)

    Product Code
    Regulation Number
    870.5050
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K122421, K990924

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Denver Pleurx Catheter Kit and the Denver Pleurx Drainage Kit/Vacuum Bottle are indicated for intermittent long term drainage of symptomatic, recurrent pleural effusion, including malignant pleural effusion and other recurrent effusions that do not respond to medical management of the underlying disease. The devices are indicated for 1) the palliation of dyspnea due to pleural effusion and 2) providing pleurodesis (resolution of the pleural effusion).

    Device Description

    The Pleurx Drainage Kit includes a vacuum bottle with drainage line that is connected to the Pleurx catheter for removing fluid that has accumulated in the chest. It also includes a rooedure pack that includes all the supplies needed to perform the drainage procedure and to replace the dressing over the catheter.

    AI/ML Overview

    The provided text describes a special 510(k) for a modification to the Denver Pleurx Drainage Kit, focusing on changes to the vacuum bottle/drainage line assembly. The acceptance criteria and the study to prove the device meets these criteria are related to demonstrating substantial equivalence to the currently marketed product, specifically regarding flow rate and other design specifications, and applying risk management techniques.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Key safety-critical performance parameter: Flow rate - (Implicit: within the specification for the currently marketed device).Not explicitly stated what the specific flow rate was, but the sponsor "verified that the key safety-critical performance parameter, flow rate, was within the specification for the currently marketed device." This indicates successful a meeting of the acceptance criteria.
    Other aspects of the revised bottle design - (Implicit: met the specifications set for them).Not explicitly stated what these "other aspects" or their specifications were, but the sponsor "verified that other aspects of the revised bottle design met the specifications set for them." This indicates successful a meeting of the acceptance criteria.
    Risk management techniques applied to changes - (Implicit: potential impact on device safety assessed, appropriate risk control measures adopted).The sponsor applied "risk management techniques to assess the potential impact of the changes on device safety, and adopting appropriate risk control measures." This indicates successful a meeting of the acceptance criteria.

    2. Sample size used for the test set and the data provenance

    The document does not specify a sample size for any test set or the data provenance (e.g., country of origin, retrospective/prospective). The studies mentioned appear to be internal verification processes rather than clinical trials with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not mention the use of experts to establish ground truth for a test set. The verification process seems to be based on engineering specifications and risk management, not expert consensus on patient outcomes or interpretations.

    4. Adjudication method for the test set

    The document does not describe any adjudication method, as there's no mention of a test set requiring adjudication by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document does not describe an MRMC comparative effectiveness study. The device is a medical drainage kit, not an AI-assisted diagnostic tool, so such a study would not be applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not describe a standalone algorithm performance study. The device is a physical medical device, not an algorithm.

    7. The type of ground truth used

    The "ground truth" in this context is the engineering specifications and safety parameters of the device and its predicate. This is a technical ground truth related to mechanical performance and safety, rather than clinical ground truth (like pathology or outcomes data).

    8. The sample size for the training set

    The document does not mention a training set, as the described studies are not for machine learning or AI models.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set mentioned.

    Ask a Question

    Ask a specific question about this device