510(k) Data Aggregation

    K Number
    K203735

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-23

    (123 days)

    Product Code
    Regulation Number
    882.5805
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K122288, K173540

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BrainsWay Deep TMS System is indicated for the treatment of depressive episodes in adult patients suffering from Major Depressive Disorder who failed to achieve satisfactory improvement from previous anti-depressant medication treatment in the current episode.

    Device Description

    The BrainsWay Deep TMS System enables direct non-invasive activation of deep brain structures. Transcranial magnetic stimulation (TMS) is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.
    The BrainsWay Deep TMS System is composed of the following main components:

      1. Cart
      • a) TMS Neurostimulator
      • b) Cooling System
      • c) Positioning Device
      1. Helmet
      • a) Aiming Apparatus (i.e., ruler/grid)
      • b) Electromagnetic Coil (H1-Coil)
      • c) Cap
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the BrainsWay Deep TMS System with iTBS Protocol, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this 510(k) submission are established through demonstrating non-inferiority of the new iTBS protocol compared to an already FDA-cleared High Frequency (HF) protocol for the same device. The primary endpoint for non-inferiority was the change in HDRS-21 score.

    Acceptance Criteria (Study Hypothesis - Null Hypothesis)Reported Device Performance (Summary of Results)
    The mean change in HDRS-21 score from baseline at 5 weeks in the iTBS Deep TMS group is inferior to the mean change in HDRS-21 score from baseline at 5 weeks in the HF Deep TMS group, by more than 3 points. (Non-inferiority margin: 3 points)The non-inferiority analysis of the change from baseline between the two treatment groups was assessed. Since the upper limit of the confidence interval was less than the non-inferiority margin of 3 points, the null hypothesis of inferiority was rejected, indicating non-inferiority between the iTBS Deep TMS and HF Deep TMS treatments, when adjusted for baseline HDRS-21 scores.
    Secondary Endpoints (No specific numerical acceptance criteria mentioned, but demonstrate similar performance)HDRS-21 Response Rate at 5 weeks, HDRS-21 Remission Rate at 5 weeks, Change in CGI-S from baseline to 6 weeks, CGI-S Response Rate at 6 weeks, CGI-S Remission Rate at 6 weeks. (The document states "the change from baseline was similar between the two treatment groups at 5 weeks," implying similar performance on these secondary endpoints as well.)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 146 subjects
      • 66 subjects received iTBS stimulation protocol.
      • 80 subjects received HF stimulation protocol.
    • Data Provenance: Not explicitly stated as retrospective or prospective, nor country of origin. However, given it's a clinical study for FDA clearance, it's highly likely to be a prospective multi-center study, though the specific locations are not provided in this summary. The mention of "multicenter MDD study" for age eligibility criteria might hint at a multi-site study.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This particular study does not involve establishing ground truth through expert review of specific data like images. Instead, the "ground truth" for the effectiveness of the treatment is based on patient-reported outcomes and clinician-administered scales (HDRS-21, CGI-S), which inherently involve expert clinicians in their administration and interpretation during the study.

    4. Adjudication Method

    Not applicable in the typical sense of radiological or pathological adjudication. The primary and secondary endpoints are based on standardized clinical assessment scales (HDRS-21, CGI-S) administered by qualified clinical personnel as part of the study protocol.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This study is a clinical trial comparing two different stimulation protocols (iTBS vs. HF) on human patients for Major Depressive Disorder (MDD), not a reader study involving interpretation of cases by multiple human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a medical device for treatment, not an AI/algorithm for diagnosis or analysis where standalone performance would typically be evaluated. The "performance" being assessed is the clinical effectiveness of the device (iTBS protocol) in treating MDD. Therefore, a standalone algorithm-only performance study is not applicable.

    7. The Type of Ground Truth Used

    The ground truth used for determining the effectiveness of the device in this clinical study is:

    • Clinical Outcomes Data: Specifically, changes in patient depression levels as measured by established clinical rating scales (HDRS-21 and CGI-S scores) from baseline to follow-up, along with derived response and remission rates.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a clinical study to evaluate the effectiveness of a medical device's protocol, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set mentioned or implied for a machine learning algorithm. The "training" for this device refers to the clinical application of the stimulation protocol in the context of treating patients.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN170078

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-08-17

    (322 days)

    Product Code
    Regulation Number
    882.5802
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K122288, K992911, K051864, K162935

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Brainsway Deep Transcranial Magnetic Stimulation System is intended to be used as an adjunct for the treatment of adult patients suffering from Obsessive-Compulsive Disorder.

    Device Description

    The Brainsway DTMS System enables direct non-invasive activation of brain structures. TMS is a non-invasive technique used to apply brief magnetic pulses to the brain. The pulses are administered by passing high currents through an electromagnetic coil placed adjacent to a patient's scalp. The pulses induce an electric field in the underlying brain tissue. When the induced field is above a certain threshold, and is directed in an appropriate orientation relative to the brain's neuronal pathways, localized axonal depolarizations are produced, thus activating neurons in the targeted brain structure.

    The Brainsway DTMS System is composed of the following five main components:

    • Electromagnetic Coil (HAC-Coil)
    • TMS Neurostimulator
    • Cooling System
    • Positioning Device
    • Personal Head Cap
    AI/ML Overview

    The provided text describes a clinical study to establish the safety and effectiveness of the Brainsway Deep Transcranial Magnetic Stimulation (DTMS) System for the treatment of Obsessive-Compulsive Disorder (OCD).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for effectiveness appear to be based on the primary effectiveness endpoint: a statistically significant reduction in Yale-Brown Obsessive Compulsive Scale (YBOCS) scores compared to sham treatment. Additionally, clinically meaningful reductions in OCD severity are highlighted.

    Acceptance Criterion (Effectiveness)Reported Device Performance
    Primary Endpoint: Statistically significant difference in change from baseline YBOCS scores to 6 weeks (post-randomization) between DTMS and Sham groups in the mITT analysis set.Met: In the mITT analysis set (N=94), the DTMS group showed an adjusted mean YBOCS score reduction of 6.0 points (95% CI: [4.0;8.1]), while the Sham group showed a reduction of 3.3 points (95% CI: [1.2;5.3]). The difference between the slopes of 2.8 points was statistically significant (p-value: 0.0127), indicating superior effectiveness of DTMS. The ITT analysis set (N=99) did not achieve statistical significance (p-value: 0.0988) due to a higher reduction in YBOCS in the Sham group (4.1 points) compared to the mITT Sham group (3.3 points), but a clinically meaningful reduction (6.0 points) was still observed in the DTMS group.
    Secondary Endpoints: - Response rate (≥30% YBOCS reduction from baseline) at 6 weeks. - Partial Response rate (≥20% YBOCS reduction from baseline) at 6 weeks. - Improvement in CGI-I ("Improved" category: moderately improved to very much improved) at 6 weeks. - Improvement in CGI-S ("Improved" category) at 6 weeks. - Sustained YBOCS reduction at 10 weeks.Met (with qualifications): - Response Rate: DTMS: 38.1% (16/42), Sham: 11.1% (5/45) in mITT. - Partial Response Rate: DTMS: 54.8% (23/42), Sham: 26.7% (12/45) in mITT. - CGI-I "Improved": DTMS: 49% (20/41), Sham: 21% (9/43) in mITT. - CGI-S "Improved": DTMS: 61% (25/41), Sham: 33% (14/43) in mITT. - YBOCS at 10 Weeks: Adjusted mean YBOCS reduction of 6.5 points (95% CI: [4.3;8.7]) in the DTMS group, indicating stability. (Note: Formal statistical testing was stopped at the first secondary endpoint, so p-values for other secondary endpoints were unadjusted for multiplicity, but results show favorable trends for DTMS).
    Safety: Device is safe and tolerable, with adverse events being minor, reversible, and typical for TMS devices, and no significant differences in vital signs, physical/neurological exams, or cognitive status between treatment groups. Absence of serious related adverse events.Met: - AEs reported by 73% (DTMS) vs. 69% (Sham). - Most frequent AE was headache (37.5% DTMS, 35.3% Sham). - Other pain/discomfort (administration site, jaw, facial, muscle, neck, dizziness) were mild/moderate and resolved. - No reports of hypoacusis (hearing loss). - No notable differences in vital signs, physical/neurological exams between groups. - One SAE reported, assessed as not device-related (suicidal thoughts preceding study).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Study Population):
      • Total enrolled: 100 subjects.
      • ITT Analysis Set (received at least one treatment): 99 subjects (48 DTMS, 51 Sham).
      • mITT Analysis Set (met eligibility criteria): 94 subjects (47 DTMS, 47 Sham). This was the principal data analysis set for efficacy.
      • Per-Protocol Analysis Set (without major protocol deviations): 93 subjects (46 DTMS, 47 Sham).
    • Data Provenance: Prospective, randomized, double-blinded, multi-site, sham-controlled clinical trial. The study was conducted at 11 study sites in the United States (9 sites), Israel (1 site), and Canada (1 site).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • The ground truth for the effectiveness of the treatment was established through clinical assessments using standardized scales like the YBOCS, Sheehan Disability Scale (SDS), and Clinical Global Impression - Severity (CGI-S) and Improvement (CGI-I).
    • These assessments were performed by independent raters. While the exact number of independent raters or their specific qualifications (e.g., years of experience as psychiatrists or psychologists) is not explicitly stated, the study design mentions that "the independent rater" was blinded to the treatment administered. This implies that trained clinical professionals, likely psychiatrists or psychologists, administered these critical rating scales.

    4. Adjudication Method for the Test Set

    • The study design involved a double-blinded approach where both study personnel (including the operator and independent rater) and study subjects were blinded to whether active or sham treatment was administered.
    • The primary and secondary outcome measures (YBOCS, SDS, CGI-S, CGI-I) are standardized clinical scales that are typically administered by trained personnel. There is no explicit mention of an "adjudication method" like 2+1 or 3+1 for discordance between multiple experts for individual case ground truth, as these scales are designed for direct clinical assessment. The "ground truth" here is the patient's symptom severity and improvement as measured by these scales.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This study directly evaluated the effectiveness of a medical device (DTMS system) for treating OCD in patients, not an AI system assisting human readers. Therefore, the concept of "how much human readers improve with AI vs. without AI assistance" is not applicable to this study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • No, a standalone (algorithm only) performance study was not done. The study evaluated a medical device that is administered to patients by trained operators, with outcomes assessed by independent clinical raters. This is a human-in-the-loop system in the context of patient treatment and assessment, not an AI algorithm acting autonomously.

    7. The Type of Ground Truth Used

    • The ground truth used for determining treatment effectiveness was based on expert clinical assessments and standardized patient-reported outcomes. Specifically:
      • Yale-Brown Obsessive Compulsive Scale (YBOCS): A clinician-administered scale that measures the severity of OCD symptoms.
      • Sheehan Disability Scale (SDS): A patient-rated scale that assesses functional impairment.
      • Clinical Global Impression - Severity (CGI-S) and Improvement (CGI-I): Clinician-rated scales that measure global severity and improvement of illness.
      • Response/Remission Definitions: Pre-defined thresholds of YBOCS score reduction or absolute score for clinical response and remission.
    • For safety, ground truth was assessed through adverse event reporting, vital signs, physical/neurological examinations, and cognitive tests.

    8. The Sample Size for the Training Set

    • This was a clinical trial evaluating a medical device, not a machine learning model. Therefore, there was no "training set" in the context of machine learning model development. The sample size used for the clinical study (94-99 subjects in the analysis sets) was for direct evaluation of the device's efficacy and safety.

    9. How the Ground Truth for the Training Set Was Established

    • As there was no machine learning "training set," this question is not applicable. The clinical trial directly provided the data (including outcome measures like YBOCS scores) used to assess the device's performance.