Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K234142

    Validate with FDA (Live)

    Date Cleared
    2024-09-18

    (264 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TiGEN Abutment, PMMA Abutment and Scan Healing Abutment are intended for use on endosseous dental implants in the edentulous or partially edentulous maxilla or mandible, as an aid in prosthetic rehabilitation.

    The PMMA Abutment is indicated to be used pror to the final components to maintain, stabilize and shape the soft tissue during the healing plase. They must be placed out of occlusion and are for temporary use (3 months).

    For TiGEN Abutment and PMMA Abutment, all digitally designed abutments for use with PMMA Abutment and TiGEN Abutment are intended to be sent to a MegaGen-validated milling center for manufacture.

    Device Description

    The TiGEN Abutment is machined with the final prosthetic in accordance with the intraoral structure. It is machined by using dental CAD/CAM technology in accordance with customized patient's information in MegaGen-validated milling center. The TiGEN Abutment is made of Ti-6Al-4V ELI alloy. And It is provided with abutment screw. All TiGEN Abutment is provided non-sterile. The milled TiGEN Abutment must be sterilized by users prior to use.

    The PMMA Abutment is a temporary prosthesis used until the final prosthesis is placed for up to three months. The PMMA Post is machined with the temporary prosthetic in accordance with the intraoral structure by using dental CAD/CAM technology. The PMMA Cuff is made of Ti-6Al-4V ELI alloy and available in various gingival heights. The PMMA Post is made of Polymethyl methacrylate and available in various diameters and lengths so that it can be used according to individual patient conditions. All PMMA Abutment is provided non-sterile with abutment screw. The milled PMMA Abutment must be sterilized by users prior to use.

    The Scan Healing Abutment designed to aid in soft tissue contouring during the healing period after implant placement, creating an emergence profile for the final prosthesis. And they have the added design feature to be scannable an intraoral impression by digital scanner. The Scan Healing Abutment is provided with abutment screw and is provided gamma-sterile.

    AI/ML Overview

    This document pertains to a 510(k) premarket notification for dental implant abutments and does not contain information about an AI/ML medical device. Therefore, a table of acceptance criteria and a study proving the device meets the criteria, as requested by the prompt, cannot be extracted from the provided text for an AI/ML context.

    The document discusses the substantial equivalence of the "TiGEN Abutment, PMMA Abutment and Scan Healing Abutment" to already marketed predicate devices. The studies mentioned are primarily bench tests, biocompatibility evaluations, and sterilization/shelf-life validations, which are standard for physical medical devices. There is no mention of an AI/ML component, AI/ML device performance metrics, or related study methodologies like multi-reader multi-case (MRMC) studies.

    Without information on an AI/ML component, the requested details such as sample size for test sets (for AI), data provenance, expert ground truth establishment, MRMC studies, standalone AI performance, and training set details are not applicable and cannot be provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220751

    Validate with FDA (Live)

    Date Cleared
    2022-06-02

    (79 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K122192

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Straumann® temporary prosthetic components are indicated to be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase. They must not be placed in occlusion. Straumann® temporary abutments are indicated to be placed into Straumann® dental implants to provide a support structure for the functional and esthetic oral rehabilitation of partially edentulous patients with temporary crowns and bridges.

    Device Description

    Straumann® BLX Temporary Abutments PMMA are temporary abutments intended for placement on Straumann BLX Dental Implants with RB/WB and WB connection platforms. The temporary abutments are made of polymethyl methacrylate (PMMA), with a Titanium Alloy (TAN) inlay.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device (Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA). It outlines the device description, indications for use, technological characteristics, and a comparison to predicate devices, but it does not contain information about the acceptance criteria and the study that proves the device meets those criteria in the context of an AI/ML-enabled medical device performance study.

    Specifically, the document focuses on the physical and material properties, and mechanical performance testing (dynamic fatigue and static strength) of a dental implant abutment, rather than the performance of a software or AI/ML algorithm.

    Therefore, I cannot extract the requested information (acceptance criteria table, sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth types and establishment for training/test sets, training set sample size) from this document as it pertains to an AI/ML device study.

    The "Performance Testing" section (5.7) mentions:

    • "Dynamic fatigue and static strength tests were conducted according to the FDA guidance document 'Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments' and demonstrated the BLX Temporary Abutments PMMA performed as well as the reference devices."
    • This refers to physical testing of the device's durability and strength, not AI/ML performance.

    The "Conclusion" (5.8) states that the device is "substantially equivalent to the legally marketed primary predicate and reference devices" based on nonclinical tests (which would include the mechanical and material tests mentioned).

    In summary, the provided text does not contain the information needed to answer your request about acceptance criteria and study details for an AI/ML-enabled medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180477

    Validate with FDA (Live)

    Date Cleared
    2019-01-04

    (316 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Straumann PURE Ceramic Implant is indicated for the restoration of single-tooth gaps and in edentulous or partially edentulous jaws. The prosthetic restorations used are single crowns, fixed partial or full dentures, which are connected to the implants through the corresponding components.

    Closure and Healing caps are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Closure and Healing caps should be used only with suitable implant connections. Do not use healing components for longer than 6 months.

    The provisional components are intended to serve as a base for temporary crown or bridge restoration out of occlusion for the Straumann® PURE Ceramic Implant System. The Straumann® Temporary Abutment VITA CAD-Temp® for the Straumann® PURE Ceramic Implant is indicated for temporary usage of up to 180 days.

    CI RD Straumann PUREbase abutment is a titanium base placed onto Straumann ceramic dental implants to provide support for customized prosthetic restorations and is indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. All digitally designed copings and/or crowns for use with the Straumann® Variobase Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.

    Device Description

    The Straumann PURE Ceramic Implant is a dental implant made out of yttrium-stabilized zirconium dioxide (Y-TZP). The Straumann PURE Ceramic Implant is based on features of the Straumann PURE Ceramic implant Monotype. Straumann PURE Ceramic Implants can be placed using the existing instruments using the same osteotomy preparation protocol as for Bone Level implants. The subject implant is a two -piece implant and is available in Ø4.1 mm with lengths of 8, 10, 12, and 14 mm.

    The Closure and Healing caps are screws machined as one piece and come in three gingival heights to accommodate individual gingival thickness. The material of the devices is titanium grade 4. The Closure caps are screwed into the implant to protect the inner configuration and shoulder of the implant during the healing phase in cases of submerged (submucosal) healing protocols and do not support a prosthetic restoration. Healing caps are screwed into the implant to protect the inner configuration in cases of transmucosal healing protocols and are placed out of occlusion and do not support a prosthetic restoration.

    The Temporary Abutments are used to serve as a temporary crown or bridge restoration for the Straumann PURE Ceramic Implant System.

    The CI RD Straumann PUREbase Abutment for Ceramic implant is a premanufactured (stock) abutment (the first piece of the two-piece abutment), sometimes referred to as "Tibase", and is used as a base when fabricating a CAD/CAM customized restoration (the second piece of the two-piece abutment). The assembly of the two-pieces becomes a finished medical device after cementation of the CAD/CAM customized restoration on the PURE base abutment.

    AI/ML Overview

    The Straumann PURE Ceramic Implant System did not present specific, quantified acceptance criteria with corresponding performance data in the provided text. Instead, the document focuses on demonstrating substantial equivalence to predicate and reference devices through various tests.

    However, based on the provided "Performance Data 5.7" section and the comparison tables, we can infer the types of acceptance criteria and the nature of the study that proves the device meets those criteria. The acceptance criteria generally revolve around demonstrating equivalence to legally marketed predicate and reference devices in terms of safety and effectiveness, as per FDA guidelines for 510(k) submissions.

    Here's an attempt to structure the information based on the request, interpreting "acceptance criteria" as the performance standards expected for demonstrating substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since explicit numerical acceptance criteria were not stated, this table describes the type of performance evaluated and the outcome in relation to the predicate/reference devices.

    Acceptance Criterion (Inferred from regulatory requirements and comparative studies)Reported Device Performance
    Biocompatibility: Device materials are safe for biological contact and do not cause adverse reactions. (In accordance with ISO 10993-1:2009).The subject device materials are identical to the predicate and reference device materials. Therefore, no new issues regarding biocompatibility were raised.
    Sterilization Efficacy (for sterile components): Sterilization process reliably achieves a Sterility Assurance Level (SAL) of 10-6. (Per ISO 11135).Sterilization validation was performed using the Half Cycle Overkill Approach, demonstrating the process can reliably and consistently sterilize to a minimum SAL of 10-6.
    Pyrogenicity: Device does not contain pyrogens above acceptable limits. (Based on FDA Guidance, testing limit of 20 EU/device for blood contacting/implanted devices).The device meets pyrogen limit specifications as determined by LAL Endotoxin Analysis, with a testing limit of 20 EU/device.
    Reprocessing Validation (for non-sterile components to be reprocessed by end-user): Reprocessing methods (e.g., autoclaving) are effective. (Per ISO 17665-1:2006 and FDA guidance).The sterilization process for the Straumann PURE temporary abutment and CI RD Straumann PUREbase abutments (as recommended in labeling) was validated according to applicable recommendations in FDA guidance and ISO 17665-1:2006.
    Mechanical Strength/Fatigue: Device can withstand dynamic loading forces typical of oral function without failure. (Per ISO 14801:2016 and FDA guidance).Dynamic fatigue tests demonstrated the Straumann PURE Ceramic Implant, Straumann PURE temporary abutments, and CI RD Straumann PUREbase abutments are equivalent to the predicate and reference devices.
    Wear on Implant-to-Abutment Connection: Connection surfaces exhibit comparable wear behavior to predicate/reference devices.Post-fatigue bench tests concluded comparable behavior of the subject device to the reference devices in terms of wear on the implant-to-abutment connection.
    Screw Loosening: Risk of screw loosening is comparable to predicate/reference devices.Post-fatigue bench tests concluded comparable behavior of the subject device to the reference devices in terms of screw loosening. A literature review was also provided to evaluate the risk of screw loosening with ceramic abutments.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes bench testing for mechanical performance, biocompatibility, and sterilization. It does not refer to a "test set" in the context of clinical data or data from human subjects.

    • Sample Size: Not explicitly stated for each test, but typically, mechanical and sterilization validations involve a statistically representative number of units or replicates to demonstrate consistency and meet statistical confidence levels defined in the relevant ISO standards (e.g., ISO 11135, ISO 14801). Biocompatibility is based on material composition equivalency, not a specific sample size.
    • Data Provenance: The studies are described as bench tests (laboratory studies), biocompatibility assessments based on material equivalency, and sterilization/reprocessing validations. There is no mention of country of origin of data in the context of test subjects, as these are not human studies for this specific assessment. The data is generated from laboratory testing and analytical comparisons.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is typically relevant for studies involving human interpretation (e.g., image analysis by radiologists). For this device, which involves material properties and mechanical performance, the "ground truth" is established by adherence to international standards (ISO) and FDA guidance documents, and the expertise lies in the validation engineers and materials scientists performing and interpreting these tests. There is no mention of "experts" in the sense of clinical adjudicators for a "test set" in the provided text.

    4. Adjudication Method for the Test Set

    Not applicable as this is not a study involving human interpretation of medical data (e.g., images) requiring adjudication of findings. Performance is evaluated against objective engineering and scientific standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an endosseous dental implant system, not an AI-powered diagnostic or assistive technology.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm or software. It is a physical medical device.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluations (mechanical, sterilization, biocompatibility) is defined by the International Organization for Standardization (ISO) standards and FDA guidance documents to which these tests were conducted.

    • Biocompatibility: ISO 10993-1:2009.
    • Sterilization: ISO 11135 and FDA Guidance on "Submission and Review of Sterility Information...".
    • Reprocessing: ISO 17665-1:2006 and FDA guidance "Reprocessing Medical Devices in Health Care Settings...".
    • Mechanical Testing (Dynamic Fatigue): ISO 14801:2016 and FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
    • Wear and Screw Loosening: Bench tests demonstrating comparable behavior to reference devices, supported by a literature review for screw loosening risk.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This device is not an AI/machine learning model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1