Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192633

    Validate with FDA (Live)

    Date Cleared
    2020-05-01

    (221 days)

    Product Code
    Regulation Number
    868.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K121969

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    GUS831 compressor nebulizer is designed for the production of compressed air to operate a nebulizer kit for the production of medical aerosol for respiratory disorders. The nebulizer kit is intended for multiple use by single patient for single medication. GUS831 compressor nebulizer is intended for domestic use with children from 5 years old, adolescent and adult patients and requires the order of a physician for medical use.

    Device Description

    GUS831 Compressor Nebulizer is a portable aerosol nebulizer electronically powered by an external power adapter and by an internal rechargeable lithium polymer battery. The compressed air, created by the membrane pump of the device, flows into the nebulizer container kit. The air pressure forced the liquid through the nozzle where it is atomized against a plate within the container. The device is equipped with a mouthpiece to easily delivery the medical aerosol.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the GUS831 Compressor Nebulizer, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (U-RIGHT Compressor Nebulizer, K121969) rather than explicitly stating pre-defined acceptance criteria with pass/fail thresholds against a specific standard for all performance parameters. However, the aerosol characterization section presents quantitative data for both the subject and predicate devices, implying those values serve as comparative criteria.

    Note: The "acceptance criteria" here are implicitly defined by the predicate device's performance, with the goal being "statistically identical" or "substantially equivalent" performance.

    Performance ParameterTest ConditionAcceptance Criteria (Predicate Device Performance)Reported Device Performance (GUS831)Conclusion on Meeting Acceptance Criteria
    Aerosol Characterization (Adult Conditions @ 28 LPM)
    Particle Size (MMAD) Albuterol Sulfate (2.5 mg/3 ml)Adult @ 28 LPM3.87±0.31 µm2.64±0.35 µmSubstantially Equivalent (implicitly)
    Particle Size (MMAD) Ipratropium Bromide (0.5 mg/2.5 ml)Adult @ 28 LPM3.37±0.46 µm2.51±0.36 µmSubstantially Equivalent (implicitly)
    Particle Size (MMAD) Cromolyn Sodium (20 mg/2.0 ml)Adult @ 28 LPM2.87±0.15 µm2.64±0.34 µmSubstantially Equivalent (implicitly)
    Geometric Standard Deviation Albuterol Sulfate (2.5 mg/3 ml)Adult @ 28 LPM3.77±0.483.32±0.57Substantially Equivalent (implicitly)
    Geometric Standard Deviation Ipratropium Bromide (0.5 mg/2.5 ml)Adult @ 28 LPM4.14±0.253.59±0.48Substantially Equivalent (implicitly)
    Geometric Standard Deviation Cromolyn Sodium (20 mg/2.0 ml)Adult @ 28 LPM3.70±0.553.37±0.70Substantially Equivalent (implicitly)
    Total Dose Delivered Albuterol Sulfate (2.5 mg/3 ml)Adult @ 28 LPM514±55 µg471±45 µgSubstantially Equivalent (implicitly)
    Total Dose Delivered Ipratropium Bromide (0.5 mg/2.5 ml)Adult @ 28 LPM89±12 µg91±9 µgSubstantially Equivalent (implicitly)
    Total Dose Delivered Cromolyn Sodium (20 mg/2.0 ml)Adult @ 28 LPM2772±441 µg3286±383 µgSubstantially Equivalent (implicitly)
    Total Respirable Dose (0.5-5 um) Albuterol SulfateAdult @ 28 LPM259±37 µg263±38 µgSubstantially Equivalent (implicitly)
    Total Respirable Dose (0.5-5 um) Ipratropium BromideAdult @ 28 LPM42±5 µg50±8 µgSubstantially Equivalent (implicitly)
    Total Respirable Dose (0.5-5 um) Cromolyn SodiumAdult @ 28 LPM1448±312 µg1913±328 µgSubstantially Equivalent (implicitly)
    Coarse Particle Dose (>4.7 um) Albuterol SulfateAdult @ 28 LPM253±38 µg178±23 µgSubstantially Equivalent (implicitly)
    Coarse Particle Dose (>4.7 um) Ipratropium BromideAdult @ 28 LPM41±8 µg35±6 µgSubstantially Equivalent (implicitly)
    Coarse Particle Dose (>4.7 um) Cromolyn SodiumAdult @ 28 LPM1143±106 µg1268±316 µgSubstantially Equivalent (implicitly)
    Fine Particle Dose (<4.7 um) Albuterol SulfateAdult @ 28 LPM261±17 µg293±45 µgSubstantially Equivalent (implicitly)
    Fine Particle Dose (<4.7 um) Ipratropium BromideAdult @ 28 LPM48±4 µg56±7 µgSubstantially Equivalent (implicitly)
    Fine Particle Dose (<4.7 um) Cromolyn SodiumAdult @ 28 LPM1628±338 µg2018±430 µgSubstantially Equivalent (implicitly)
    Ultra-Fine Particle Dose (<1.0 um) Albuterol SulfateAdult @ 28 LPM56±33 µg87±23 µgSubstantially Equivalent (implicitly)
    Ultra-Fine Particle Dose (<1.0 um) Ipratropium BromideAdult @ 28 LPM17±3 µg20±3 µgSubstantially Equivalent (implicitly)
    Ultra-Fine Particle Dose (<1.0 um) Cromolyn SodiumAdult @ 28 LPM540±80 µg687±133 µgSubstantially Equivalent (implicitly)
    Aerosol Characterization (Pediatric Conditions @ 12 LPM)
    Particle Size (MMAD) Albuterol Sulfate (2.5 mg/3 ml)Pediatric @ 12 LPM5.57±0.45 µm5.00±0.95 µmSubstantially Equivalent (implicitly)
    Particle Size (MMAD) Ipratropium Bromide (0.5 mg/2.5 ml)Pediatric @ 12 LPM6.80±0.61 µm4.57±1.12 µmSubstantially Equivalent (implicitly)
    Particle Size (MMAD) Cromolyn Sodium (20 mg/2.0 ml)Pediatric @ 12 LPM7.23±0.49 µm4.93±0.47 µmSubstantially Equivalent (implicitly)
    Geometric Standard Deviation Albuterol Sulfate (2.5 mg/3 ml)Pediatric @ 12 LPM3.90±0.293.00±0.07Substantially Equivalent (implicitly)
    Geometric Standard Deviation Ipratropium Bromide (0.5 mg/2.5 ml)Pediatric @ 12 LPM3.74±0.243.42±0.09Substantially Equivalent (implicitly)
    Geometric Standard Deviation Cromolyn Sodium (20 mg/2.0 ml)Pediatric @ 12 LPM3.63±0.794.35±0.62Substantially Equivalent (implicitly)
    Total Dose Delivered Albuterol Sulfate (2.5 mg/3 ml)Pediatric @ 12 LPM1148±69 µg1106±53 µgSubstantially Equivalent (implicitly)
    Total Dose Delivered Ipratropium Bromide (0.5 mg/2.5 ml)Pediatric @ 12 LPM232±11 µg221±26 µgSubstantially Equivalent (implicitly)
    Total Dose Delivered Cromolyn Sodium (20 mg/2.0 ml)Pediatric @ 12 LPM6633±595 µg5523±154 µgSubstantially Equivalent (implicitly)
    Total Respirable Dose (0.5-5 um) Albuterol SulfatePediatric @ 12 LPM423±15 µg541±93 µgSubstantially Equivalent (implicitly)
    Total Respirable Dose (0.5-5 um) Ipratropium BromidePediatric @ 12 LPM79±5 µg103±19 µgSubstantially Equivalent (implicitly)
    Total Respirable Dose (0.5-5 um) Cromolyn SodiumPediatric @ 12 LPM2263±207 µg2370±132 µgSubstantially Equivalent (implicitly)
    Coarse Particle Dose (>4.7 um) Albuterol SulfatePediatric @ 12 LPM701±107 µg576±36 µgSubstantially Equivalent (implicitly)
    Coarse Particle Dose (>4.7 um) Ipratropium BromidePediatric @ 12 LPM152±8 µg113±27 µgSubstantially Equivalent (implicitly)
    Coarse Particle Dose (>4.7 um) Cromolyn SodiumPediatric @ 12 LPM4343±426 µg2920±113 µgSubstantially Equivalent (implicitly)
    Fine Particle Dose (<4.7 um) Albuterol SulfatePediatric @ 12 LPM447±39 µg531±88 µgSubstantially Equivalent (implicitly)
    Fine Particle Dose (<4.7 um) Ipratropium BromidePediatric @ 12 LPM80±2 µg108±32 µgSubstantially Equivalent (implicitly)
    Fine Particle Dose (<4.7 um) Cromolyn SodiumPediatric @ 12 LPM2290±478 µg2603±111 µgSubstantially Equivalent (implicitly)
    Ultra-Fine Particle Dose (<1.0 um) Albuterol SulfatePediatric @ 12 LPM51±4 µg80±26 µgSubstantially Equivalent (implicitly)
    Ultra-Fine Particle Dose (<1.0 um) Ipratropium BromidePediatric @ 12 LPM17±3 µg17±5 µgSubstantially Equivalent (implicitly)
    Ultra-Fine Particle Dose (<1.0 um) Cromolyn SodiumPediatric @ 12 LPM571±161 µg737±51 µgSubstantially Equivalent (implicitly)

    Overall Conclusion: The document states that "almost all the performance parameters of the two different nebulizers were statistically identical for adult conditions and substantially equivalent in the therapeutic amount of medication delivered for pediatric conditions. The measured differences raise no new safety and efficacy issues; therefore, device equivalence is proved."


    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: The document does not specify the exact numerical sample size for the aerosol characterization testing (e.g., number of nebulizers tested, number of runs per condition). It reports mean and standard deviation, implying replicate measurements were performed.
    • Data Provenance: The testing was conducted by an "independent source" according to the relevant CDRH Guidance Document "Reviewer Guidance for Nebulizer, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH – 1993). The country of origin of the data is not explicitly stated. The study appears to be prospective for the device testing itself, as it was performed to support the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable to this type of device and study. The testing for a compressor nebulizer involves objective physical and chemical measurements (e.g., particle size, dose delivery) rather than expert interpretation of images or patient data. The ground truth is established by the standardized measurement methods themselves.


    4. Adjudication Method for the Test Set

    This information is not applicable for this type of device and study, as it did not involve expert review or consensus for establishing ground truth.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices where human readers interpret medical cases, sometimes with and without AI assistance. The GUS831 Compressor Nebulizer is a therapeutic device (drug delivery), not a diagnostic one.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Yes, a standalone performance evaluation was done. The aerosol characterization and biocompatibility testing were conducted directly on the device (GUS831 Compressor Nebulizer) and compared to the predicate device, without involving human users in a clinical setting for performance assessment.


    7. The Type of Ground Truth Used

    The ground truth for the aerosol characterization was established through objective physical and chemical measurements obtained using a cascade impactor, following the methodology outlined in the CDRH Guidance Document "Reviewer Guidance for Nebulizer, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH – 1993). This is a form of objective measurement data, not expert consensus, pathology, or outcomes data typically associated with diagnostic AI.

    The ground truth for biocompatibility was established by testing against recognized international standards (ISO 10993-1, ISO 18562 family).


    8. The Sample Size for the Training Set

    This information is not applicable. The GUS831 Compressor Nebulizer is a physical medical device; it is not an AI/ML algorithm that requires a training set. The "testing" referred to in the document is for the physical performance and biocompatibility of the device compared to a predicate, not for training a model.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as the device is not an AI/ML algorithm requiring a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1