510(k) Data Aggregation

    K Number
    K163475

    Validate with FDA (Live)

    Device Name
    iVue
    Manufacturer
    Date Cleared
    2017-06-09

    (179 days)

    Product Code
    Regulation Number
    886.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K121739

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iVue is a non-contact, high resolution tomographic imaging device. It is intended for in-vivo imaging, axial crosssectional, and three-dimensional imaging and measurement of anterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, corneal epithelia, corneal stroma and anterior chamber of the eye. With the integrated normative database, the iVue is a quantitative tool for the comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects. The iVue is indicated for use as a device to aid in the diagnosis, documentation, and management of ocular health and diseases in the adult population.

    Device Description

    The iVue is used to capture, store, display and print spectral domain-optical coherence tomography (SD-OCT) images of the posterior and anterior structure of the eye. The device software includes a Normative Database (NDB), consisting of OCT data from a range of known normal subjects that can be used to compare a new patient's measurements in relation to the normal distribution.

    iVue is a computer-controlled ophthalmic imaging system using either a laptop computer or "All-in-One" computer. For laptop systems there are two control box options of 120 or 230 volts. The control box interfaces between the motorized table column and the medical-grade power supply for the computer.

    iVue System Key Functional Components:

    • Scanner Head
    • Computer
    • Control Box
    • Joystick and Chinrest Assembly
    • Footswitch (optional)
    • Motorized Table (optional)
    • Cornea Adapter Module
    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the Optovue iVue device, seeking clearance for a software modification that allows automated segmentation and measurement of corneal epithelial and stromal layer thickness. The document describes two clinical studies conducted to support the substantial equivalence claim.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria in a table format with pass/fail thresholds. Instead, it presents the results of repeatability, reproducibility, and agreement studies, and then concludes whether these results demonstrate "good repeatability and reproducibility" and "good agreement."

    Based on the summaries provided, the implicit acceptance criteria seem to be related to achieving clinically acceptable levels of repeatability, reproducibility, and agreement with manual measurements for corneal pachymetry, epithelial thickness, and stromal thickness, across various patient groups.

    Table of Implicit Acceptance Criteria and Reported Device Performance:

    MetricAcceptance Criteria (Implicit, based on conclusion)Reported Device Performance (Summary)
    Repeatability (SD)Demonstrated "good repeatability"Pachymetry:- Normal Group: 1.7 to 7.6 µm (across zones)- Corneal Patients (pooled): 3.3 to 8.6 µm- Subgroups (Contact Lens, Dry Eye, Post-LRS, KCN): Ranged from 1.5 to 13.4 µm (max in Dry Eye S_5_6_Pachy).Epithelial Thickness:- Normal Group: 0.8 to 1.4 µm- Corneal Patients (pooled): 1.2 to 1.9 µm- Subgroups: Ranged from 0.6 to 2.9 µm (max in Dry Eye I_5_6_Epi and IT_5_6_Epi, and Post-LRS C2 Epi)Stromal Thickness:- Normal Group: 1.6 to 7.6 µm- Corneal Patients (pooled): 2.8 to 8.8 µm- Subgroups: Ranged from 1.4 to 13.4 µm (max in Dry Eye S_5_6_Stroma)
    Reproducibility (SD)Demonstrated "good reproducibility"Pachymetry:- Normal Group: 2.2 to 7.7 µm- Corneal Patients (pooled): 3.6 to 8.8 µm- Subgroups: Ranged from 1.8 to 13.7 µm (max in Dry Eye S_5_6_Pachy).Epithelial Thickness:- Normal Group: 0.9 to 1.5 µm- Corneal Patients (pooled): 1.2 to 1.9 µm- Subgroups: Ranged from 0.7 to 2.9 µm (max in Dry Eye I_5_6_Epi and IT_5_6_Epi, and Post-LRS C2 Epi)Stromal Thickness:- Normal Group: 2.0 to 7.6 µm- Corneal Patients (pooled): 3.1 to 9.0 µm- Subgroups: Ranged from 1.8 to 13.7 µm (max in Dry Eye S_5_6_Stroma).The report notes: "Repeatability standard deviation was similar to the Reproducibility standard deviation for all study parameters and therefore is not detailed separately in the summary tables below." The tables for each parameter show Reproducibility SD values.
    Agreement (Mean of Differences)Demonstrated "good agreement"Mean of differences between software output and manual measurements for all zonal parameters:- Corneal Epithelial Mapping: < 0.7 µm (<1.3% of 54 µm average thickness)- Corneal Stroma Mapping: < 1.9 µm (<0.4% of 490 µm average thickness)- Pachymetry Mapping: < 2.3 µm (<0.4% of 543 µm average thickness)
    Agreement (LOA)Demonstrated "good agreement" (within acceptable limits)Limits of Agreement (LOA) between software output and manual measurements for all zonal parameters:- Corneal Epithelial Mapping: (-6.2 to 5.9) µm- Corneal Stroma Mapping: (-9.7 to 10.4) µm- Pachymetry Mapping: (-7.9 to 9.6) µm

    2. Sample Sizes and Data Provenance

    Study 1: Repeatability and Reproducibility (R&R) of Corneal Epithelial Thickness Mapping

    • Sample Size (Test Set): 59 subjects (from 598 total acquired scans, 527 qualified for final analysis).
      • Normal Subjects: 12 subjects
      • Corneal Patients: 47 subjects (further stratified into 4 subgroups, each with 11-12 subjects: Contact Lens, Dry Eye, Post-Refractive Surgery, Keratoconus).
    • Data Provenance: Not explicitly stated, but clinical studies are generally prospective. The multi-device, multi-operator protocol in Study 1 suggests it was a controlled, prospective collection. Given that Optovue is based in Fremont, CA, and the submission is to the FDA, it is highly probable the data was collected in the USA, although not explicitly stated.

    Study 2: Agreement of Corneal Epithelial Thickness Mapping with iVue SD-OCT to Manual Measurement

    • Sample Size (Test Set): 87 study eyes from 87 subjects.
      • Normal Subjects: 17
      • Contact Lens (normal): 16
      • Dry Eye: 18
      • Post-Laser Refractive Surgery: 20
      • Keratoconus: 16
    • Data Provenance: Not explicitly stated, but the mention of "3 study sites" collecting OCT data suggests a prospective clinical study. Again, given the FDA submission, most likely USA.

    3. Number of Experts and Qualifications for Ground Truth

    • Study 1 (R&R): Ground truth was based on the iVue software's automated segmentation, with manual edits performed by operators when needed. The document states: "For qualified scans, the operators reviewed the thickness maps for obvious segmentation error and reviewed the individual corneal meridian images to verify segmentation for erroneous maps. Segmentation edit tools were used to perform manual correction and then the epithelial map was reprocessed. Noticeable segmentation errors were manually corrected by the operator and marked for "Manual Correction" in the case report form."

      • The qualifications of these "operators" are not explicitly detailed. It implies they are trained users of the device capable of identifying and correcting segmentation errors, but their specific certifications (e.g., ophthalmologist, optometrist, ophthalmic technician) or years of experience are not mentioned.
      • Number of operators: The study used "device-specific designated operator" for each of the three iVue devices, so a minimum of 3 operators.
    • Study 2 (Agreement): Ground truth was established by manual measurements performed by "3 qualified graders."

      • The qualifications of these "qualified graders" are not specified (e.g., ophthalmologist, ophthalmic technician, years of experience). They used a "2-section caliper tool that was available in the previously cleared iVue software."

    4. Adjudication Method for the Test Set

    • Study 1 (R&R): There's no explicit mention of a formal multi-reader adjudication process (e.g., 2+1, 3+1). It states "manual edits were not required in Normal eyes scans, and ranged from 2.8% in Contact Lens sub-group to 17.9% in KCN sub-group...Noticeable segmentation errors were manually corrected by the operator and marked for "Manual Correction"". This suggests correction was done by a single operator per scan if needed, not a consensus or adjudicated process among multiple experts.
    • Study 2 (Agreement): The "agreement to manual measurement of the software was evaluated based on the agreement between the software output and the manual measurements by 3 qualified graders." It appears each grader provided their measurement independently, and then the agreement between the software and their measurements (or a combined representation of their measurements, though "manual measurements" is singular in "agreement to manual measurement") was assessed. It doesn't describe an adjudication process amongst the 3 graders to establish a single ground truth for each case from the graders. The comparison is between the software and the individual manual measurements, then reported collectively.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done comparing human readers with AI vs. without AI assistance. The study focuses purely on the device's technical measurement performance (repeatability, reproducibility, and agreement with existing manual measurement methods). The iVue is a measurement device, not an AI diagnostic aid in the sense of 'improving human reader performance.'

    6. Standalone (Algorithm Only) Performance

    • Yes, standalone performance was evaluated. The core of both clinical studies is the performance of the iVue device's software (algorithm) in segmenting and measuring the corneal layers.
      • In Study 1, it assessed the repeatability and reproducibility of the "modified iVue software" directly.
      • In Study 2, it assessed the "agreement between the software output and the manual measurements." This is a direct evaluation of the algorithm's performance against a human-derived ground truth.

    7. Type of Ground Truth Used

    • Study 1 (R&R): The ground truth was the automated segmentation by the iVue software, potentially refined by "operator correction". This is a form of expert-corrected algorithmic output.
    • Study 2 (Agreement): The ground truth was manual measurements performed by "qualified graders" using a digital caliper tool available in the predicate iVue software. This is a form of expert measurement. It's not pathology or outcomes data.

    8. Sample Size for the Training Set

    • Not specified in the provided document. The document describes clinical studies for validation and verification of a modified software, not the development or training of the original algorithm. Device approvals like 510(k) often focus on testing the finished product.

    9. How the Ground Truth for the Training Set Was Established

    • Not specified in the provided document. As the training set size is not mentioned, neither is the method for establishing its ground truth. This information would typically be found in the algorithm's development and internal validation documentation, not necessarily in a 510(k) summary focused on post-modification performance.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130656

    Validate with FDA (Live)

    Device Name
    IFUSION
    Manufacturer
    Date Cleared
    2013-07-03

    (113 days)

    Product Code
    Regulation Number
    886.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K121739,K122572

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The iFusion connects the iCam (K122572) and iVue (K121739) devices via a sliding bracket mechanism (iShuttle), to facilitate switching between the two devices. The iShuttle provides position adjustment ability of the iCam or iVue device during use.

    The iFusion interfaces with the iCam and iVue devices to enable the operation of the iCam and iVue devices from one computer unit.

    Device Description

    iFusion is a combination of the cleared iVue with NDB (K121739)("iVue") and the cleared iCam Fundus Camera (K122572)("iCam") on a shuttle platform, called the iShuttle. The iShuttle mounts to a joystick assembly using a central nut and bolt and four adjustable set screws. The iShuttle consists of a sliding metal plate supporting the iVue and the iCam. The iShuttle travels on tracks fitted to the X-Y-Z joystick assembly which helps position the iVue and the iCam devices. A release button is located at the front of the iShuttle to lock the iShuttle in position for the iVue and the iCam device. The iCam can be switched in position by pressing on the release button to unlock the iShuttle platform, then the iVue or the iCam can be pushed to slide to an opposite end until it is locked into position. In addition, the software of the iFusion system is created to allow the user to switch between the iVue and the iCam programs with the same computer. The iVue and the iCam programs use the same data communication and programming storage with a single shared database.

    AI/ML Overview

    The provided document is a 510(k) summary for the Optovue iFusion device. This submission is for a combination device that integrates two previously cleared predicate devices (iVue and iCam) onto a new sliding bracket (iShuttle) and includes minor software changes. The primary goal of this type of submission is to demonstrate substantial equivalence to existing legally marketed devices, rather than to prove clinical effectiveness and safety through de novo clinical studies with specific acceptance criteria related to disease diagnosis or treatment.

    Therefore, the information typically found in acceptance criteria tables and detailed study reports for showing diagnostic/predictive performance (like sensitivity, specificity, AUC) or comparative effectiveness (like MRMC studies) is not present in this document. This submission focuses on demonstrating that the changes (the iShuttle and minor software updates) do not negatively impact the safety and effectiveness of the already cleared component devices.

    Here's an breakdown of the available information based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not contain a table of acceptance criteria and reported device performance in the typical sense for diagnostic or prognostic claims (e.g., sensitivity, specificity, accuracy for a particular disease). Instead, the "acceptance criteria" for this 510(k) were focused on demonstrating that the new iFusion system, with its integrated iShuttle and minor software changes, maintained the safety and effectiveness of the predicate devices.

    The "performance data" section states:

    "Verifications and validations were performed for mechanical and functional testing according to IEC 60601-1 and 60601-1-2 for the new iShuttle bracket and the software changes."
    "The results demonstrated that the changes in iFusion did not affect the intended use or alter the fundamental technological characteristics of the predicate devices and that the iFusion is substantially equivalent to its predicate devices."

    This implies that the acceptance criteria for these tests would have been along the lines of "meet IEC 60601-1 and 60601-1-2 standards" and "no degradation in function or safety of iVue and iCam components." However, the specific quantitative results or acceptance thresholds for these engineering tests are not provided in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable in the context of this 510(k) submission. This is not a study assessing diagnostic performance on a patient dataset. The "test set" would have been the iFusion device itself subjected to mechanical and software integrity tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. There's no clinical ground truth established for this submission as it's not a study on diagnostic accuracy.

    4. Adjudication Method for the Test Set

    Not applicable for this type of submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study was not described or performed for this 510(k) submission. The iFusion device is essentially a physical integration and software interface for two existing devices, not a new diagnostic algorithm requiring human-in-the-loop performance measurement.

    6. If a Standalone Study Was Done

    Not applicable in the sense of an algorithm-only diagnostic performance study. The "standalone" testing here refers to the verification and validation of the iShuttle's mechanical and functional aspects, and the software changes, to ensure they didn't compromise the existing functionalities of the iVue and iCam.

    7. The Type of Ground Truth Used

    Not applicable in the context of a clinical ground truth. The "ground truth" for this engineering/integration submission would be adherence to safety standards (IEC 60601-1, 60601-1-2) and confirmation that the iVue and iCam components function as intended when integrated into iFusion.

    8. The Sample Size for the Training Set

    Not applicable. There is no mention of an algorithm being "trained" in this submission. The software changes are "minor" and related to user interface and data communication/storage, not a new AI/ML algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set for an AI/ML algorithm.