    K Number
    K192617

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-20

    (150 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K121286

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Dental Soft Tissue Indications

    Incision, excision, vaporization, ablation of oral soft tissues including marginal and inter-dental gingival and epithelial lining of free gingiva and the following specific indications:

    • · Excisional and incisional biopsies
    • · Exposure of unerupted teeth
    • Fibroma removal
    • · Frenectomy
    • · Frenotomy
    • · Gingival troughing for crown impressions
    • · Gingivectomy
    • · Gingivoplasty
    • Gingival incision and excision
    • Hemostasis and coagulation
    • · Implant recovery
    • Incision and drainage of abscess
    • · Leukoplakia
    • · Operculectomy
    • · Oral papillectomies
    • · Pulpotomy
    • Pulpotomy as an adjunct to root canal therapy
    • · Reduction of gingival hypertrophy
    • · Soft tissue crown lengthening
    • · Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
    • · Treatment of aphthous ulcers.
    • · Vestibuloplasty
    • · Tissue retraction for impression
    • · Lesion (tumor) removal.

    Laser Periodontal Procedures.

    • · Laser soft tissue curettage.
    • · Laser removal of diseased, Inflamed and necrosed soft tissue within the periodontal pocket.

    · Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium.

    · Sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)

    • · Reduction of bacterial level (decontamination) and inflammation

    Pain therapy

    · Topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, minor sprains and minor muscular back pain, the temporary increase in local blood circulation; the temporary relaxation of muscle.

    Device Description

    Gemini is an 810nm + 980nm soft tissue laser is intended for the incision, ablation, vaporization, hemostasis and treatment of oral soft tissue and for pain relief using photobiomodulation. The Gemini laser operates at a wavelength of 810nm +/- 10nm or at 980 +/- 10nm or a combination of both 810nm + 980nm wavelengths, with a maximum average power of 2 watts +/- 20% and maximum peak pulse power of 20 watts +/- 20% when in dual wavelength mode. The dual wavelength diode laser radiation is delivered through a flexible optical fiber and the Photobiomodulation (PBM) Adapter. Infrared laser energy is emitted from the fiber tips and the PBM Adapter when the wireless footswitch is depressed. The laser diodes are directly coupled to the flexible fiber optic able that connects the laser unit to the surgical hand piece and to the disposable fiber tip and PBM Adapter that emits the energy to the target area. The laser diodes can simultaneously emit energy when Gemini is set to dual wavelength mode.

    The Gemini unit comprises of six main assemblies: a Laser unit, a flexible fiber optic delivery svstem tethered to an anodized aluminum hand piece, disposable single-use fiber tips. PBM Adapter with spacers, a wireless footswitch, and an auxiliary power supply. The system also contains a 5mW 650nm laser diode coupled to the same fiber optic cable to produce the red aiming light. The laser system is contained within a compact lightweight impact resistant molded plastic housing that contains a laser diode assembly, a laser power controller PCB, an Electroluminescent Display connected to an interface PCB, a touch sensitive selection panel with status lights, and a rechargeable Lithium Battery. The hand piece used by the practitioner consists of an anodized aluminum cylindrical body which encloses the optics used to transfer the laser energy to the single-use fiber tips and the PBM Adapter. The LED illumination system for lighting the work area will be off for PBM Adapter.

    The laser's visible light is designed to aid the clinician in aiming the tip of the delivery fiber into the target tissue. Additionally, a bright white light from the LEDS in the hand piece illuminates target work area during procedures through the translucent disposable laser tips.

    The PBM Adapter is an accessory attachment to the Gemini laser system to increase the spot size of the laser beam, allowing the Gemini Laser to provide near-infrared laser energy to a tissue surface for the purpose of photobiomodulation. Affected muscles and/or joints must be exposed to an adequate level of laser energy over a period of time to provide effective results. The spot size of the PBM Adapter is 25 mm. The PBM Adapter operates at 810nm ± 10nm wavelength with a maximum average power of 1 watt ± 20%.

    A PBM spacer is attached to the PBM Adapter. The spacer is a single-use disposable piece designed to ensure proper working distance to target tissues and limit the risk of cross contamination between patients. The PBM Adapter is designed to only work with the Gemini Laser System.

    Gemini 810 + 980 Diode Laser utilizes non-volatile, preprogrammed firmware that cannot be modified by the user. During the development process, requirements are met, hazards are evaluated and mitiqated for the safety of the patient and /or the operator and the firmware. It does not capture any data related to patients. The Gemini software focuses on the functional and technical design for the Gemini laser. This includes the graphical user interface (GUI), the control of the laser drivers, sound, tip illumination and aiming light.

    AI/ML Overview

    The provided text is a 510(k) summary for the Gemini 810+980 Diode Laser. It primarily focuses on demonstrating substantial equivalence to predicate devices based on indications for use, technological characteristics, and non-clinical performance data.

    Crucially, this document does not describe a study involving a "test set" with "ground truth established by experts" or a "multi-reader multi-case (MRMC) comparative effectiveness study" for an AI or imaging device. It is for a laser surgical instrument and pain therapy device.

    Therefore, many of the requested items related to AI/imaging device validation (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, ground truth type for test set and training set) are not applicable to this submission.

    The document states: "No clinical data was submitted for this Traditional 510(k)." This further confirms that there were no human subject studies to evaluate the device's performance in a clinical setting in the way an AI/imaging device would be evaluated.

    However, I can extract information related to the acceptance criteria and the studies that were performed to support the substantial equivalence claim for this specific type of device.

    Here's a breakdown of the available information based on your request:

    1. A table of acceptance criteria and the reported device performance:

    Since this is a laser device and not an AI/imaging diagnostic tool, the "acceptance criteria" are compliance with various national and international standards, and "reported device performance" refers to the results of bench testing confirming these compliances and functional equivalence to predicate devices.

    Acceptance Criteria (Compliance with Standards/Functional Equivalence)Reported Device Performance (Study Results)
    Laser Safety & Performance Standards:- 21 CFR 1040.10 & 1040.11: Found in compliance (except for deviations pursuant to laser notice 50 dated June 24, 2007).
    - IEC 60601-2-22 Edition 3.1 2012-10- Conformed to IEC 60601-2-22 (Particular Requirements for basic safety and essential performance of surgical, cosmetic, therapeutic, and diagnostic laser equipment).
    - IEC 60825-1 Edition 2.0 2007-03- Conformed to IEC 60825-1 (Safety of laser products – Part 1: Equipment classification and requirements).
    Electrical Safety & EMC Standards:- AAMI/ANSI ES60601-1:2005/ (R) 2012 and A1:2012: Conformed to IEC 60601-1 (General requirements for basic safety and essential performance).
    - IEC 60601-1-2 Edition 4.0 2014-02- Conformed to IEC 60601-1 -2 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance. Collateral Standard: Electromagnetic compatibility).
    Cleaning & Sterilization Standards:- AAMI/ANSI ST81:2004/(R)2016
    - AAMI/ANSI ST79:2017- Cleaning Validation (Anodized Aluminum Surgical Hand Piece with Fiber Connector): Achieved 78.5% protein recovery and 77.6% hemoglobin recovery. Cleaned devices contained < 5.4 uq/cm² protein and < 1.6 µq/cm² hemoglobin. Cleaning procedure validated.
    - AAMI/ANSI/ISO 17665-1:2006/(R)2013- Sterilization Validation (Anodized Aluminum Hand Piece Shell): Fifteen-minute gravity steam sterilization cycle at 135°C validated. No growth of biological indicators; verified half cycle indicates 12 log reduction capability and 10^-6 sterility assurance level for a worst-case population. 15-minute gravity cycle at 135°C validated.
    - AAMI/ANSI/ISO 17665-2:2009-01-15
    Biocompatibility Standards:- AAMI/ANSI/ISO 10993-5:2009/(R)2014 (The document lists this standard but does not explicitly state the results of such testing, only that the device was tested in accordance and found in compliance with the listed standards.)
    Software Validation:- Conformed to IEC 62304 (Medical device software - Software lifecycle processes).
    Functional Equivalence for Pain Management Indication:- Comparative bench testing with Biolase Epic 10 (pain therapy handpiece): Resulted in similar power density and spot size. Concluded Gemini with PBM Adapter is substantially equivalent to Epic 10 for pain management laser energy output.
    Therapeutic Heating Mechanism (Pain Therapy):- Skin temperature testing: Demonstrated that Gemini 810+980 Diode Laser with PBM Adapter increases topical heat at the tissues level to 40 degrees Celsius and can maintain this temperature for a minimum of 10 minutes during treatment, which aligns with the mechanism of action for ILY product code (therapeutic heating to 40-45°C for ≥10 minutes).
    General Functional Compliance (21 CFR 1040.10):- Bench testing: Found that Gemini meets the features and functions as identified in 21 CFR 1040.10.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable in the context of an "AI/imaging test set": As stated, no clinical data was submitted. The testing involved bench testing of the device and its components, not a "test set" of patient data for diagnostic evaluation. Sample sizes for bench tests (e.g., number of units tested for electrical safety, cleaning/sterilization) are standard engineering and quality assurance practices and are not detailed in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable: No "ground truth" was established by medical experts for a test set of patient data, as this device is a laser surgical and pain therapy instrument, not an AI diagnostic tool.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable: No clinical test set requiring adjudication of findings was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is not an AI-assisted diagnostic device. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is a physical medical device, not an algorithm. Bench testing confirmed its standalone performance against engineering specifications and comparison with predicate devices.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable in the typical AI/imaging sense. For this device, "ground truth" for its performance is derived from:
      • Compliance with recognized industry standards: The standards themselves define acceptable parameters and measurement methodologies.
      • Bench test results: Physical measurements of power output, wavelength, temperature achieved, cleaning efficacy, sterilization efficacy, etc., verified by instruments and documented procedures.
      • Comparison to predicate device specifications/performance: Demonstrating that its physical and energy output characteristics are equivalent or superior to already cleared devices.

    8. The sample size for the training set

    • Not Applicable: This device does not use a "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    • Not Applicable: No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K130465

    Validate with FDA (Live)

    Device Name
    EPIC 10S
    Date Cleared
    2013-04-11

    (45 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K121286

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Epic 10S is intended for use as a laser surgical instrument in general and plastic surgery and in dermatology. It is indicated for the following expanded Indications for Use: Ear, Nose and Throat and Oral Surgery, Arthroscopy, Gastroenterology, Orthopedics, General Surgery, Dermatology & Plastic Surgery, and Podiatry, GI/GU, Gynecology, Neurosurgery, Ophthalmology, Pulmonary Surgery, Cardiac Surgery, Thoracic Surgery, Urology, Dermatology/Aesthetics, Vascular Surgery.

    Device Description

    The EPIC 10S system uses an Indium Arsenide Phosphorous (InGaAsP) solid state laser diode to emit infrared laser energy which is transmitted via a flexible fiber optic cable to a handpiece that emits the energy to the target site. A visible light is emitted at the same time to visually identify the treatment location. The EPIC 10S laser is comprised of a base console, a detachable delivery system, tips, and a wireless footswitch. Various types of the single use tips are included for different applications and the device is activated by means of a wireless footswitch. The Epic 10S delivery system consists of the following: Fiber Optic Assembly, Surgical Handpiece, and Single-use tips.

    AI/ML Overview

    The provided document is a 510(k) summary for the Biolase Epic 10S laser device. It explicitly states that no non-clinical or clinical test data was performed on this device for the purpose of this 510(k) submission.

    Instead, the submission claims substantial equivalence to a legally marketed predicate device, the Diolase 10S (K121327), based on a comparison of technological characteristics and expanded Indications for Use. Therefore, the information requested regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert involvement, and comparative effectiveness studies is not present in this document.

    The document indicates that the Epic 10S is substantially equivalent to the Diolase 10S for the stated indications.

    Here's a breakdown of why the requested information cannot be provided based on the input:

    1. A table of acceptance criteria and the reported device performance, Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective), Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience), Adjudication method (e.g. 2+1, 3+1, none) for the test set, If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance, If a standalone (i.e. algorithm only without human-in-the-loop performance) was done, The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The document states: "Non-clinical testing was not performed on this device since the purpose of this 510(k) is to only expand the Indications for Use from the original 510(k) clearance under 510(k) No.: K121286. All indications included in this 510(k) have already been cleared by the FDA for an equivalent medical device manufactured by Biolase Technology, Inc." (Page 8).
      • It also states: "Clinical testing was not conducted on this device." (Page 8).
      • Therefore, there is no performance data, test set, ground truth, or details on reader studies for this submission. The device is being cleared based on its similarity to a previously cleared device.
    2. The sample size for the training set, How the ground truth for the training set was established:

      • Since no new clinical or non-clinical studies were conducted for this submission (as stated above), there is no training set mentioned or described within this document. The submission relies on the prior clearance of a predicate device.

    In summary, the provided submission (K130465) relies on a claim of substantial equivalence to a predicate device (Diolase 10S, K121327) and did not involve new testing, clinical or non-clinical, to establish acceptance criteria or device performance.