Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241158

    Validate with FDA (Live)

    Date Cleared
    2024-06-25

    (60 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K031992, K023681, K120917, K141981

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This microcatheter is a medical device intended for angiography and/or infusion of various substances including diagnosis, embolization and treatment in the peripheral vasculature using an appropriate guide wire. The microcatheter is not intended for use in neurovasculature, coronary arteries and carotid arteries.

    Device Description

    The ASAHI Veloute, ASAHI Veloute C3, ASAHI Tellus and ASAHI Tellus C3 Microcatheters are sterile single use devices designed for use in the peripheral vasculature. The ASAHI Veloute, ASAHI Veloute C3, ASAHI Tellus and ASAHI Tellus C3 consist of a catheter shaft that is reinforced with braid wires to enhance pushability and maintain patency of the inner lumen. A radiopaque marker is fixed on the distal end of the catheter shaft to facilitate location of the catheter during angiography. The distal portion of the catheter shaft is flexible and available in two shapes, straight and angled, to provide improved trackability in tortuous vessels. A hydrophilic coating is applied on the outer surface of the catheter to provide a smooth transition. In addition, accessories, including either a stylet, syringe and RHV (rotating hemostasis valve), or a stylet, syringe, inserter and hemostasis valve are available for use with the ASAHI Veloute, ASAHI Veloute C3, ASAHI Tellus and ASAHI Tellus C3.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for various microcatheters (ASAHI Veloute, ASAHI Veloute C3, ASAHI Tellus, ASAHI Tellus C3). This document serves as a regulatory submission to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than a study proving the device meets acceptance criteria for an AI/ML medical device.

    Therefore, the requested information regarding acceptance criteria, study design, expert involvement, and ground truth for an AI/ML-based device cannot be extracted from this document. The document primarily focuses on the mechanical, material, and biocompatibility performance of a physical medical device (microcatheter) and its substantial equivalence to existing devices.

    However, I can extract the information related to the performance testing and acceptance criteria for the physical microcatheter device as detailed in the document.


    Acceptance Criteria and Device Performance (for the physical microcatheter device)

    The document details non-clinical laboratory testing and biocompatibility testing conducted to demonstrate the substantial equivalence of the ASAHI microcatheters.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implicit from 'Result' column for non-clinical) / Explicit for BiocompatibilityReported Device Performance
    Non-Clinical Laboratory Testing
    Appearance/Dimensions/Tip ShapeMet specified requirementsPass
    Corrosion resistanceMet specified requirementsPass
    Peak tensile StrengthMet specified requirementsPass
    Tip FlexibilityMet specified requirementsPass
    Liquid leakageMet specified requirementsPass
    Radio-detectabilityMet specified requirementsPass
    Air LeakageMet specified requirementsPass
    Burst PressureMet specified requirementsPass
    Flow RateMet specified requirementsPass
    Power InjectionMet specified requirementsPass
    Kink ResistanceMet specified requirementsPass
    SlidabilityMet specified requirementsPass
    ConnectorMet specified requirementsPass
    Coat integrity / Particulate Evaluation in a clinically relevant modelFor characterization only (not a pass/fail acceptance)This testing is for characterization only.
    Torque StrengthMet specified requirementsPass
    Biocompatibility Testing (for ASAHI Tellus, representative model)
    Cytotoxicity (MEM Elution Test)No signs of cellular reactivity (Grade 0) for both negative control and medium controlNon-cytotoxic
    Sensitization (KLIGMAN Maximization Test)No evidence of causing delayed dermal contact sensitization in guinea pigNon-Sensitizing
    Irritation (Intracutaneous Injection Test)Test extract and negative control must exhibit similar edema and erythema scoresNon-Irritant
    Systemic Toxicity (Acute Systemic Toxicity Test)Must not show significantly greater biological activity than the controlNo Systemic Toxicity
    Systemic Toxicity (Rabbit Pyrogen Test - material mediated)Not increase rectal temperature of any animals by more than 0.5 degrees CelsiusNon-pyrogenic
    Hemocompatibility (Rabbit Blood Hemolysis Test)Non-hemolyticNon-hemolytic
    Hemocompatibility (Unactivated Partial Thromboplastin Time Test)UPTT of plasma exposed to test article extract should not significantly decreased compared to untreated and negative controlsMinimal activator
    Hemocompatibility (Complement Activation Assay - SC5b-9)No significant increase in SC5b-9 when compared to activated NHS and negative control after 60 minutes exposureNot an Activator
    Hemocompatibility (Thrombogenicity Study in Dogs)Compare results of test article to predicate control for Thrombogenic response. Determine acceptability as part of risk management.Thromboresistant

    Regarding the AI/ML-specific details (Items 2-9 from your prompt):

    The provided document is a 510(k) summary for a physical medical device (microcatheter), not an AI/ML-driven software device. Therefore, information related to:

    1. Sample size for test set and data provenance: Not applicable. Tests were bench/in-vitro and animal studies for device performance and biocompatibility.
    2. Number of experts and qualifications for ground truth: Not applicable. Performance data is from physical tests. Biocompatibility standards are specific laboratory tests.
    3. Adjudication method: Not applicable.
    4. Multi Reader Multi Case (MRMC) comparative effectiveness study: Not applicable. This is for AI-assisted human reading.
    5. Standalone (algorithm only) performance: Not applicable. This is a physical device.
    6. Type of ground truth used: For non-clinical tests, the ground truth is the physical measurement and functional assessment against specified engineering criteria. For biocompatibility, it's defined by the specific ISO standards and their pass/fail criteria.
    7. Sample size for training set: Not applicable (no AI/ML model training).
    8. How ground truth for training set was established: Not applicable.

    In conclusion, this document demonstrates the safety and effectiveness of a microcatheter through standard predicate comparison, non-clinical bench testing, and biocompatibility studies, which are typical requirements for such devices. It does not contain any information about an AI/ML component or its associated validation studies.