510(k) Data Aggregation

    K Number
    K162905

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-03-15

    (149 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K120757

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SleepWeaver 3D Nasal Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or Bi-Level therapy. This mask is intended for single-patient reuse in the home and single-patient reuse in the hospital / institutional environment. This mask is to be used on patients greater than 66 lbs (30 kg).

    Device Description

    The SleepWeaver 3D Nasal Mask consists of a fabric cushion, headgear, and a swivel connector. The device is to be used as an interface for Continuous Positive Airway Pressure (CPAP) or Bi-Level therapy. The mask consists of an elastic polyester fabric that provides a seal around the user's nose once positive airway pressure therapy is applied. The mask is held in place with a 4-point adjustable elastic fabric headgear that contains an integrated plastic component to direct the headgear away from the patient's ears. The headgear is attached through slots in the fabric mask to support the mask fit. The mask and headgear are designed to be hand washed in the home (single patient reuse) and the hospital / institutional environment (single patient reuse). Attached to the cushion is an L-shaped swivel connector that can rotate 360 degrees and is compatible with the industry standard 22mm air tubing.

    AI/ML Overview

    This document describes the premarket notification for the SleepWeaver 3D Nasal Mask (K162905). It focuses on demonstrating substantial equivalence to a predicate device, the SleepWeaver élan Nasal CPAP Mask (K120757), rather than primary efficacy or performance against new clinical endpoints. Therefore, the information regarding acceptance criteria and study design will reflect this context of substantial equivalence.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative table format alongside "reported device performance" for each feature. Instead, it lists performance tests and their outcomes, indicating that the device "met all predefined acceptance criteria" or "results were satisfactory to demonstrate substantial equivalence."

    For non-clinical tests, here's a summary:

    Acceptance Criteria (Implied by Standards/Predicate)Reported Device Performance (SleepWeaver 3D Nasal Mask)
    Fixed Leak Rate (Acceptable per ISO 17510:2015)Met
    CO2 Rebreathing (Acceptable per ISO 17510:2015)Met
    Pressure Drop (Resistance to Flow) at 50 LPM0.3 cmH2O
    Pressure Drop (Resistance to Flow) at 100 LPM0.5 cmH2O
    Sound Power Level (dBA)26 dBA
    Sound Pressure Level at 1m (dBA)18 dBA
    Reliability (Wash cycles without damage/loss of function)90 wash cycles without damage or loss of function
    Biocompatibility (for new skin-contacting materials based on ISO 10993 series)All tests successfully completed with acceptable results

    2. Sample Size Used for the Test Set and Data Provenance

    The document specifies no clinical tests were performed or required to demonstrate substantial equivalence. All performance assessments were based on non-clinical (laboratory/bench) tests. Therefore, there is no "test set" in the context of human subjects or patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. As no clinical studies were conducted and the assessment was based on engineering and manufacturing standards, there was no ground truth requiring expert consensus from medical professionals. The "ground truth" here would be the standards themselves (e.g., ISO 17510:2015).

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring adjudication was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. This type of study is not mentioned and would not typically be performed for a device seeking substantial equivalence based on non-clinical testing of a nasal mask.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The SleepWeaver 3D Nasal Mask is a physical medical device (nasal mask), not an algorithm or a software-based diagnostic tool. Performance relates to its physical properties and interaction with a CPAP system, not independent algorithmic output.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests was based on established international standards (e.g., ISO 17510:2015, ISO 14971:2007, ISO 10993 series). Device performance was compared against the requirements and acceptable limits specified within these standards.

    8. Sample Size for the Training Set

    Not applicable. There is no training set mentioned, as this is a physical medical device and not an AI/ML model that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151683

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-04-05

    (288 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K092362, K120757, K130481

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SleepWeaver Advance Pediatric Nasal Mask is intended to provide an interface for noninvasive Positive Airway Pressure (PAP) ventilation therapy. This mask is intended for single-patient use in the home and single-patient use in the hospital / institutional environment. This mask is to be used on patients between 2 and 7 years of age.

    Device Description

    The SleepWeaver Advance Pediatric Nasal CPAP Mask serves as a mechanism for connecting a pediatric patient between 2 and 7 years of age for whom PAP therapy has been prescribed to maintain an open airway. The nasal mask is placed over a patient's nose and held in place by use of an adjustable elastic headgear. A cloth cushion contacts the patient's face. The mask assembly has a tubing swivel connector which is compatible with the industry standard 22 mm air tubing. Air is supplied to the mask by a therapy device. The patient inhales air from the mask and exhales into the mask where continuous airflow from the therapy device purges the exhaled carbon dioxide from the mask through the mask exhalation holes. The SleepWeaver Advance Pediatric Nasal CPAP Mask can be used in the home environment (single-patient use) or the hospital/institutional environment (single-patient use). The mask features a cloth cushion made of a blue polyester fabric with a polyester / nylon / lycra / polyurethane interface material. The mask exhalation feature is incorporated into the cloth cushion. Attached to the cushion is a 22-mm tubing swivel connector that can rotate 360°. The tubing connector can be removed by unwrapping the white medical tape for cleaning and cushion replacement. The cloth headgear is connected to the mask through slots in the cushion wings. The headgear is removable for cleaning and replacement. The SleepWeaver Advance Pediatric Nasal CPAP Mask product also offers an optional accessory: the Feather Weight tube. The optional SleepWeaver FeatherWeight tube may be used with this mask, but it remains outside the scope of this submission as it has been previously cleared by its inclusion in K120757. The Feather Weight tube is an 18" plastic extension tube compliant to ISO 5356-1 and ISO 5367 that connects between the SleepWeaver Advance Pediatric Nasal CPAP Mask and standard CPAP tubing.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "SleepWeaver Advance Pediatric Nasal CPAP Mask." This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study where an AI-driven device's performance is compared against human readers or specific acceptance criteria for AI output.

    Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, stand-alone performance, training set sample size, how ground truth for training set was established) are not applicable to this type of submission. The device described is a physical medical device (a CPAP mask), not an AI/ML-driven software device.

    However, I can extract the acceptance criteria and reported device performance from the "Non-Clinical Testing Summary" and "Technological Characteristics" sections, as these effectively serve as the performance evaluation for this type of device.


    Acceptance Criteria and Reported Device Performance

    Device: SleepWeaver Advance Pediatric Nasal CPAP Mask (K151683)
    Predicate Device: ResMed Ltd., Pixi Pediatric Mask (K102224)

    Acceptance Criteria / Performance MetricPredicate Device (ResMed Pixi Pediatric Mask) PerformanceSleepWeaver Advance Pediatric Nasal CPAP Mask PerformanceEquivalence Narrative / Compliance
    Intended UseTo provide an interface for noninvasive Positive Airway Pressure (PAP) ventilation therapy for patients between 2 and 7 years of age. Single-patient re-use in home/hospital environment.To provide an interface for noninvasive Positive Airway Pressure (PAP) ventilation therapy for patients between 2 and 7 years of age. Single-patient use in home/hospital environment.Same.
    Patient PopulationPediatric patients between 2 and 7 years of age.Pediatric patients between 2 and 7 years of age.Same.
    Therapy Pressure Range3 to 20 cm H2O4 - 20 cm H2OProposed device range is within predicate range.
    Sizing1 size1 sizeBoth sized to accommodate intended population.
    Physical Dead Space101.7 mL130 mLHigher than predicate, but performance testing (CO2 rebreathing tests under ISO 17510-2) confirms no CO2 build-up due to exhalation holes.
    Resistance (Pressure Drop)0.8 cm H2O at 50 L/min; 3.1 cm H2O at 100 L/min0.4 cm H2O at 50 L/min; 1.0 cm H2O at 100 L/minLower than predicate, which is considered preferential.
    Carbon Dioxide Washout ProfileNot available (Predicate)Amount of CO2 being rebreathed < 20% max allowable increase under normal conditions (4, 5, and 10 cm H2O).The SleepWeaver Pediatric meets ISO 17510-2.
    Leak RatePressure (cm H2O) 3: 18 L/min 7: 28 L/min 12: 36 L/min 16: 43 L/min 20: 49 L/minPressure (cm H2O) 4: 14 L/min 5: 16 L/min 10: 25 L/min 15: 33 L/min 20: 40 L/minLower than predicate. Its leak rate is greater than patient exhalation per minute, and it passed CO2 rebreathing tests under ISO 17510-2.
    A-weighted Sound LevelsNot available (Predicate)Sound Power Level: 26 dBA; Sound Pressure Level at 1m: 18 dBANot applicable (no predicate data for comparison).
    BiocompatibilityISO 10993-1 compliant materialsMaterials used in legally marketed devices (SleepWeaver Advance Nasal CPAP Mask (K092362), SleepWeaver Elan Nasal CPAP Mask (K120757), and SleepWeaver Anew Full Face Mask CPAP (K130481)).Materials are biocompatible and in use in other cleared devices.

    Study Details (as applicable to a physical medical device submission)

    1. Sample size used for the test set and the data provenance:

      • Specific human subject sample sizes for each test are not explicitly stated. The "Fit evaluation and anthropometric analysis" implies human subjects were used, but the number is not provided.
      • The "CO2 rebreathing testing per ISO 17510-2" and "Fixed leak rate per ISO 17510-2" were likely conducted in a lab setting using test lungs or simulated breathing circuits rather than human subjects for the core performance metrics.
      • Data provenance: Primarily bench-top testing, not clinical trial data from specific countries.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. This is a physical device, and the "ground truth" is established by compliance with engineering standards (like ISO 17510-2) and direct measurements from bench-top tests, rather than expert interpretation.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are typically for subjective assessments or discrepancy resolution in clinical data, which is not the primary basis of this submission.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI-driven device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI-driven device. The "standalone" performance is the device's measured physical characteristics on a test bench.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this medical device consists of established engineering standards (e.g., ISO 17510-2 for CO2 rebreathing), physical measurements (e.g., dead space, resistance, leak rate), and biocompatibility testing results.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no training set.