510(k) Data Aggregation

    K Number
    K142825

    Validate with FDA (Live)

    Device Name
    Zmachine
    Date Cleared
    2015-01-20

    (112 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K101830, K120450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CRI Zmachine is a single-channel, EEG acquisition and analysis system, designed for use in the home or clinical environments. This device is intended to be used by qualified healthcare practitioners to monitor the wake and sleep states of adult patients and as an adjunct to their diagnosis of sleep disorders.

    Device Description

    The CRI Zmachine is a battery-operated, single-channel, EEG acquisition and analysis system. The Zmachine system includes the Zmachine device, disposable EEG sensors, sensor cable, and a wall charger. The device operates on data from the differential-mastoid EEG channel to determine the wake and sleep states (wake, light sleep, deep sleep and REM) of the patient every 30 seconds.

    AI/ML Overview

    Here's the breakdown of the acceptance criteria and study information for the Zmachine DT-200, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Zmachine DT-200 are implicitly established by demonstrating substantial equivalence to predicate devices, particularly the Sleep Profiler (K120450). The performance metrics used for comparison are P1 (sensitivity) and P2 (positive predictive value) for each sleep stage.

    Sleep StagePerformance MetricZmachine DT-200 (Reported)Sleep Profiler (Predicate)Acceptance Criteria (Implicit - "Similar to Predicate")Meets Criteria?
    WakeP10.9080.789≥ 0.789Yes
    P20.8430.831≥ 0.831Yes
    Light SleepP10.8350.855Similar to 0.855 (0.835 is slightly lower but deemed similar)Yes
    P20.8480.812≥ 0.812Yes
    Deep SleepP10.7380.757Similar to 0.757 (0.738 is slightly lower but deemed similar)Yes
    P20.7820.787Similar to 0.787 (0.782 is slightly lower but deemed similar)Yes
    REMP10.7210.719≥ 0.719Yes
    P20.7320.782Similar to 0.782 (0.732 is slightly lower but deemed similar)Yes
    OverallKappa Agreement0.716Not reported for Sleep ProfilerNot explicitly defined as a numerical threshold, but implied to be good.Yes

    Note: "Similar to Predicate" is the implicit acceptance criterion. The document states, "The P1 and P2 values obtained during clinical validation of the Zmachine Algorithm are similar to those obtained by the predicate Sleep Profiler algorithm."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 99 subjects (52 female / 47 male, 18-60 years, median age 32.7 years).
    • Data Provenance: The document does not explicitly state the country of origin. It indicates the study subjects included those reporting normal sleep and those with complaints consistent with various sleep disorders. The study is described as "clinical," suggesting it's prospective, but this is not explicitly stated. It is a clinical study in which overnight laboratory polysomnographic (PSG) data was acquired.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Two to four (2-4) certified polysomnographic technologists.
    • Qualifications of Experts: Certified polysomnographic technologists.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Majority voting rule. The individual score files from the 2-4 technologists were combined, on a 30-second epoch basis, using a majority voting rule to generate a single score file per subject ("Human Scores").

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • No MRMC comparative effectiveness study was done involving human readers with and without AI assistance. This study focused on the standalone performance of the Zmachine algorithm compared to human consensus.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone study was done. The Zmachine Algorithm's performance was evaluated by directly comparing its output to the "Human Scores" (ground truth). The objective was to assess the algorithm's ability to determine wake and sleep states independently.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus. The ground truth, referred to as "Human Scores," was established by comparing the Zmachine Algorithm's output against the scores generated by multiple certified polysomnographic technologists using a majority voting rule, based on standard PSG channels. The scoring guidelines used were Rechtschaffen and Kales (R&K) visual sleep staging guidelines.

    8. The Sample Size for the Training Set

    • The document does not provide information regarding the sample size used for the training set for the Zmachine DT-200. It focuses on the clinical performance testing sample.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not provide information on how the ground truth for any training set was established for the Zmachine DT-200.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120447

    Validate with FDA (Live)

    Device Name
    X4
    Date Cleared
    2012-06-20

    (127 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K120450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The X4 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit and display physiological signals from adult patients. The X4 System acquires, records, transmits and displays electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), accelerometer, acoustical and photoplethesmographic signals. The X4 system only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.

    Device Description

    The X4 system is used for configurable acquisition of physiological signals. Model X4-E provides for acquisition of three channels of electroencephalography (EEG) and one photoplethesmographic (PPG) signal from a head strip, with an optional channel connected to two sensors via a dual-lead connector with twice the gain. Model X4-M provides four channels of EEG with the dual-lead connector providing the input for reference sensors. Both models measure sound via an acoustic microphone, and movement and position measured via a 3-D accelerometer. The device is designed so it can be affixed by the patient and to record data. Alternatively, a technician can affix the device and display the signals via a wireless connection during acquisition. The X4 system firmware monitors signal quality to ensure that the sensors are properly applied and that high quality signals are being acquired.

    The X4 software provides a means to: a) initiate a study and track patient information, b) acquire and save signals to the memory of the device, c) acquire and wirelessly transmit signals from the device, d) upload data saved in the memory of the device to a PC, and e) visually inspect the signal quality.

    The acquired signals are saved in a universal data format (European Data Format – EDF) that is intended to be analyzed with third party software, including Advanced Brain Monitoring's Sleep Profile Software (K120450).

    AI/ML Overview

    The provided text describes the X4 System, a device for acquiring physiological signals. The 510(k) summary focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing and a limited clinical study.

    Here's an analysis of the acceptance criteria and study information provided:

    Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for the X4 System, as articulated in the clinical tests, is based on the interpretability of the acquired signals and the ease of device application.

    Acceptance CriteriaReported Device Performance
    Non-Clinical:
    Compliance with system-level requirementsAll features of the X4 System were compliant with the system level requirements.
    Electrical, biological safety, performance, and software testsConfirmation of conformity to FDA recognized consensus standards and voluntary standards (IEC 60601-1-1, IEC 60601-1-2, ISO 10993-1, IEC 60601-1-11, IEC 60601-2-26).
    Signals provide similar information to predicate device for physician interpretationSignals acquired with the X4 System provide similar information as compared to the predicate device that would allow a physician to interpret the signals. (Specific metrics or comparison data are not provided in this summary.)
    Clinical:
    Over 90% of overnight studies recorded with the X4 are interpretable and behave as expectedResult: Over 90% of studies recorded overnight with the X4 are interpretable and behave as expected.
    Device instructions can be applied without difficulty by patients in the homeResult: The X4 instructions were applied without difficulty, allowing all subjects to properly apply the device so that it remained in the proper position and allowing any problems that triggered an audio alarm to be properly resolved. (No specific quantitative metric for "difficulty" or "proper application" is given beyond the qualitative statement.)

    Study Details for Acceptance Criteria

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test set sample size: While a specific number for the sample size for the clinical test is not explicitly stated, it refers to "all subjects" and "over 90% of studies recorded overnight." This suggests a prospective clinical study involving multiple subjects.
    • Data provenance: Not explicitly stated, but the context of a 510(k) submission to the FDA for a US company suggests it would likely involve data from the US, potentially a prospective clinical trial.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • The summary states that "signals acquired with the X4 System provide similar information as compared to the predicate device that would allow a physician to interpret the signals." and that "over 90% of studies recorded overnight with the X4 are interpretable."
    • This implies that physicians are the experts who would interpret the signals to determine interpretability. However, the number of experts and their specific qualifications (e.g., board-certified sleep physicians, neurologists, years of experience) are not specified in the provided text.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • The text does not specify an adjudication method. It mentions that "physicians" would interpret the signals, but it does not detail how potential disagreements among interpreters would be resolved or if multiple interpreters were used for each case.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done.
    • The X4 System is described as a device for acquiring, recording, transmitting, and displaying physiological signals, with "no claims being made for analysis of the acquired signals with respect to the accuracy, precision and reliability." This indicates that the device itself does not perform AI-assisted analysis and therefore, a study on human reader improvement with AI assistance would not be applicable to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, a standalone (algorithm only) performance study was not done.
    • The device's intended use is to acquire, record, transmit, and display physiological signals. It explicitly states "no claims are being made for analysis of the acquired signals." Thus, there is no standalone algorithm being evaluated for diagnostic or analytical performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The ground truth for the clinical study appears to be expert interpretation (physician interpretability) of the acquired signals. The study aimed to demonstrate that the signals are "interpretable" and "behave as expected," which would rely on a physician's assessment.

    8. The sample size for the training set

    • Not applicable / Not provided. The X4 System is a signal acquisition and display device, not an analytical or AI-driven system. Therefore, it does not involve a "training set" in the context of machine learning algorithms.

    9. How the ground truth for the training set was established

    • Not applicable. As a signal acquisition device without analytical claims, there is no training set or associated ground truth establishment process for algorithm training.