510(k) Data Aggregation

    K Number
    K133973

    Validate with FDA (Live)

    Date Cleared
    2014-08-14

    (231 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K130773, K123074, K101619, K131929, K120366

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for: 3 & 5 Lead ECG monitoring; 12 Lead ECG recording; real-time arrhythmia detection / alarming; QT measurement / alarming and ST measurements / alarming: impedance pneumography: noninvasive blood pressure (NIBP); end-tidal CO2 (ETCO2) and respiration rate; pulse oximetry (SpO2); contact temperature; and invasive pressure.

    The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).

    The device is indicated for adults, paediatrics and neonates.

    Device Description

    The Tempus Pro is a multi-parameter vital signs monitor designed for use in pre-hospital care and remote clinical locations by trained healthcare professionals. It provides 3&5 lead ECG monitoring, 12 lead ECG recording, real-time arrhythmia detection / alarming, QT measurement / alarming and ST measurements / alarming pulse oximetry, non-invasive blood pressure, sidestream capnometry, 2 channels of contact temperature, impedance respiration, up to 4 channels of invasive pressure and user configurable alarms.

    In addition, it provides the ability to transmit all vital signs data via wired or wireless connections to a software system (called i2i) expected to be based in a facility far from the user e.g. a response centre facility. In addition to sending all vital signs, the system can also send pictures or video via an integrated camera, geographic position by an integrated GPS receiver and voice via a wired or wireless headset.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Tempus Pro Patient Monitor with Enhanced Features, specifically the addition of arrhythmia detection/alarming, QT measurement/alarming, and ST measurements/alarming.

    Here's an analysis of the acceptance criteria and study information, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document mentions that the Arrhythmia detection, ST elevation, and QT interval measurements were evaluated in accordance with FDA guidance on arrhythmia detection and alarm, using the FDA recognized consensus standard EC57 (FDA Rec No: 3-118). However, it does not provide a specific table of acceptance criteria or quantitative performance metrics for these features. It only states: "A summary report demonstrating compliance with the guidance and associated standard is presented as part of this 510(k) submission." and "In each case the results of this testing confirmed that acceptance criteria defined by the relevant standard, or other appropriate reference document had been met."

    Without access to the "summary report" or the full 510(k) submission, we cannot extract specific numerical acceptance criteria or reported device performance for the new features. The table below reflects the information that is available.

    Feature / MetricAcceptance CriteriaReported Device Performance
    Arrhythmia Detection / AlarmingCompliance with FDA guidance on arrhythmia detection and alarm, and consensus standard EC57 (FDA Rec No: 3-118)."Results of this testing confirmed that acceptance criteria defined by the relevant standard... had been met." (No specific numerical performance provided in this document).
    ST Measurement / AlarmingCompliance with FDA guidance on arrhythmia detection and alarm, and consensus standard EC57 (FDA Rec No: 3-118)."Results of this testing confirmed that acceptance criteria defined by the relevant standard... had been met." (No specific numerical performance provided in this document).
    QT Measurement / AlarmingCompliance with FDA guidance on arrhythmia detection and alarm, and consensus standard EC57 (FDA Rec No: 3-118)."Results of this testing confirmed that acceptance criteria defined by the relevant standard... had been met." (No specific numerical performance provided in this document).
    General Device ParametersOperate to specification across stated performance and temperature range. Ensure all data is transmitted reliably and accurately.Confirmed by bench testing.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states that the evaluation was performed in accordance with "FDA guidance on arrhythmia detection and alarm" and "the FDA recognized consensus standard EC57 (FDA Rec No: 3-118), which defines methods of presenting results when testing against specified ECG databases."

    However, the provided text does not specify the sample size (e.g., number of patients, number of ECGs) used for the test set. It also does not specify the data provenance (e.g., country of origin, retrospective or prospective) of the ECG databases used. EC57 typically refers to standard databases like MIT-BIH Arrhythmia Database, but this is not explicitly stated in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on the number of experts used to establish ground truth or their specific qualifications for the test set. For standard ECG databases (like those referenced by EC57), ground truth is typically established by multiple expert cardiologists.

    4. Adjudication Method for the Test Set

    The document does not specify the adjudication method used for the test set. If standard ECG databases were used, the ground truth annotations would have been established prior to the device under test being evaluated.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to assess how much human readers improve with AI vs without AI assistance. The focus of this 510(k) appears to be on the device's standalone performance in arrhythmia detection and measurement against established standards.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    Yes, a standalone study was done. The document states that "Arrhythmia detection and alarms, ST elevation and QT interval measurements... were evaluated in accordance with FDA guidance on arrhythmia detection and alarm... This guidance calls for use of the FDA recognized consensus standard EC57 (FDA Rec No: 3-118), which defines methods of presenting results when testing against specified ECG databases." This indicates the algorithm's performance was evaluated independently against pre-annotated databases, which is a standalone assessment.

    7. Type of Ground Truth Used

    The ground truth used would be expert consensus annotations as provided in the "specified ECG databases" referenced by the EC57 standard. These databases contain ECG recordings with validated annotations of arrhythmias and other cardiac events, established by a panel of cardiologists.

    8. Sample Size for the Training Set

    The document does not provide any information regarding the sample size used for the training set of the algorithms. It is possible the algorithms were pre-trained and integrated, or that general ECG databases were used for training, but this is not detailed.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide any information on how the ground truth for the training set was established. If standard ECG databases were used for training, the ground truth would have been established by expert consensus annotations within those databases.