510(k) Data Aggregation

    K Number
    K170326

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-07-20

    (168 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K960385, K120165, K140148, K133520, K070598

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The dynaMX™ Intramedullary Fixation Device is indicated to stabilize and aid in the fixation of fractures, fusions, and osteotomies of the phalanges.

    Device Description

    The dynaMX™ Intramedullary Implant is intended to stabilize and aid in the fixation of fractures, fusions, and osteotomies of the phalanges. The implant is fully cannulated with threads on the distal side and two barbs on the proximal side. The implant is offered in multiple combinations of diameter and length to accommodate various anatomies. The dynaMX™ Intramedullary Implant is made of biocompatible Nitinol. The barbs of the implant are designed to exhibit superelastic properties at room and body temperature. As manufactured, one end of the implant has a screw thread, and the other end a pair of barbs. The screw thread is inserted into a drilled hole in the middle phalange. The barbed end is implanted into a broached canal in the proximal phalange. While inserting the barbed end of the implant, the barbs deflect inward to allow the implant to pass through the broached canal. Once inside the canal, the barbs superelastically return to their flared outward configuration to securely hold the implant in the bone. The dynaMX™ Intramedullary Implant is sold as part of a single use disposable kit. The kit includes an implant, pre-loaded onto a delivery device, a drill bit, a broach and a quide wire.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "dynaMX™ Intramedullary Implant." This type of document is submitted to the FDA to demonstrate that a device is substantially equivalent to a legally marketed predicate device, and it typically does not contain detailed acceptance criteria and study results in the same way a clinical trial report would.

    However, based on the information provided, I can infer what constitutes the "acceptance criteria" through the bench and cadaver testing listed and how the device aims to meet these through comparison with predicate devices.

    Here's an analysis based on your request, structured as much as possible with the available information:

    Key Takeaway: The document outlines a substantial equivalence claim based on laboratory studies (bench and cadaver tests) rather than human clinical trials involving complex statistical criteria for AI performance. Therefore, many of your specific questions related to AI-specific metrics, human reader studies, and large-scale data provenance are not applicable to this type of device and submission.


    Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

    The acceptance criteria for the dynaMX™ Intramedullary Implant are demonstrated through a series of laboratory studies (bench tests and cadaver tests) designed to verify the suitability of the device for its intended use, establish substantial equivalence with predicate devices, and confirm reproducibility of packaging. The successful completion of these tests, showing comparable performance to predicate devices, serves as the primary evidence for meeting acceptance criteria for this Class II medical device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Measure/TestReported Device Performance (as stated in the document)
    Mechanical Performance / Structural IntegrityElastic Static Bending TestingConducted to verify suitability and establish Substantial Equivalence with predicate devices. (Specific quantitative results or pass/fail thresholds are not detailed in this summary.)
    Bending Fatigue TestingConducted to verify suitability and establish Substantial Equivalence with predicate devices. (Specific quantitative results or pass/fail thresholds are not detailed in this summary.)
    Implant Pull-Out ForceConducted to verify suitability and establish Substantial Equivalence with predicate devices. The device's design, with a screw thread on one end and superelastic barbs on the other, is described to "securely hold the implant in the bone."
    Rotational StabilityConducted to verify suitability and establish Substantial Equivalence with predicate devices. (Specific quantitative results or pass/fail thresholds are not detailed in this summary.)
    BiocompatibilityCorrosion TestingConducted. (Specific results not detailed, but generally, acceptance would be low/negligible corrosion appropriate for implants.)
    Transformation Temperature Determination (for Nitinol)Conducted. (Specific results not detailed, but acceptance would be within a range demonstrating appropriate superelastic properties at body temperature.)
    PyrogenicityEndotoxin testing conducted to confirm the implant is non-pyrogenic. (Implies successful completion.)
    Packaging & SterilizationPackage Seal Strength VerificationConfirmed reproducibility of the packaging. (Implies successful completion.)
    Shelf Life / StabilityA shelf-life/stability protocol has been designed to support expiry dating of up to 36 months, with an initial proposed shelf-life of 6 months following completion of the initial study.
    Material EquivalenceMaterial CompositionFabricated from Nitinol per ASTM F2063-12. This is explicitly compared to predicate devices, some also using Nitinol,others using stainless steel or titanium.
    Functional EquivalenceStabilize and aid in fixation of phalangeal fractures, fusions, and osteotomiesDemonstrated through laboratory studies and substantial equivalence comparison to predicate devices with similar indications for use.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample size for the bench and cadaver tests. It generally states a "series of laboratory studies." For cadaver tests, typically a limited number of cadaveric specimens would be used.

    The data provenance is laboratory-based (bench tests) and cadaveric. There is no information regarding country of origin, but it is assumed the tests were conducted within the facilities of Arthrex, Inc. (or MX Orthopedics, Corp.) or their designated testing laboratories. These are effectively retrospective in the sense that they are conducted on inert materials or cadaveric tissue in a controlled lab setting, not in live patients.

    3. Number of Experts Used to Establish the Ground Truth For The Test Set and Qualifications of Those Experts

    This information is not applicable as the device is an implantable mechanical device, not an AI or diagnostic device that relies on expert ground truth for interpretation of output. The "ground truth" for this device's performance would be the physical properties and mechanical behavior observed in the specified laboratory tests.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication typically refers to expert consensus in interpreting diagnostic results or clinical outcomes. The performance of this device is assessed directly through engineering measurements and established material science principles in laboratory settings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was The Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    This information is not applicable. This device is a surgical implant, not a diagnostic or AI-powered system, and therefore, human reader studies with AI assistance are irrelevant to its evaluation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. This device is a physical implant; there is no algorithm involved in its direct function for standalone performance evaluation in this context.

    7. The Type of Ground Truth Used

    The ground truth used for this device's evaluation is primarily engineering specifications, material science properties, and established biomechanical performance standards for bone fixation devices. This is demonstrated through direct physical measurements and observations during bench and cadaver tests. The comparison to predicate devices also establishes a benchmark for "ground truth" in terms of acceptable clinical performance and safety.

    8. The Sample Size for the Training Set

    This information is not applicable. This device is a physical implant, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. There is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133636

    Validate with FDA (Live)

    Date Cleared
    2014-03-28

    (122 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K120645, K120165, K073674, K061164, K071498, K120449

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HammerFiX device is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe and mallet toe.

    Device Description

    The HammerFiX device is a bone fixation device consisting of a sterile, threaded PEEK implant and a set of instruments used for implant site preparation and delivery. The device is offered in small, medium and large implant sizes to allow for use in the proximal interphalangeal (PIP) joints of the lesser toes of the foot.

    AI/ML Overview

    The provided text describes a 510(k) summary for the HammerFiX device, a bone fixation device. However, it explicitly states that no clinical testing was performed. Therefore, the information requested regarding acceptance criteria and a study proving the device meets them cannot be fully provided from the given document as it pertains to clinical performance. The 510(k) relies on non-clinical bench testing to establish substantial equivalence to predicate devices, rather than clinical efficacy or specific performance criteria from a clinical study.

    Here's a breakdown of the available information based on the provided text, acknowledging the lack of clinical study data:


    Acceptance Criteria and Study Details for HammerFiX Device (Based on Non-Clinical Testing)

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since no clinical study was performed to evaluate performance against clinical acceptance criteria, the table below reflects the non-clinical testing performed and the general statement of equivalence based on those tests. Specific quantitative acceptance criteria or detailed numerical performance metrics from the bench testing are not provided in this summary.

    Acceptance Criteria Category (Non-Clinical)Reported Device Performance
    Mechanical Performance
    Pull-out StrengthPerformed and compared to predicate devices. (Details not specified)
    Torque StrengthPerformed and compared to predicate devices. (Details not specified)
    Static BendingPerformed and compared to predicate devices. (Details not specified)
    Dynamic BendingPerformed and compared to predicate devices. (Details not specified)
    Overall EquivalenceThe HammerFiX device is substantially equivalent to its predicate device based on indications for use, principles of operation, design, and mechanical test data.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: Not specified in the provided summary. This refers to the number of devices or constructs tested during non-clinical bench testing.
    • Data Provenance: The testing was "bench testing," meaning it was conducted in a laboratory setting, likely at the manufacturer's facility or a contracted testing lab. No country of origin for data is stated, and it is inherently prospective for the specific bench tests conducted to support the 510(k).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Not Applicable. For non-clinical bench testing, "ground truth" as it relates to expert clinical opinion is not established. The "ground truth" for mechanical testing is typically defined by standardized test methods and engineering specifications. The tests compare the device's measured mechanical properties to those of predicate devices or established engineering standards.

    4. Adjudication Method for the Test Set:

    • Not Applicable. Adjudication methods (like 2+1, 3+1) are relevant for clinical studies, especially when multiple readers or experts assess subjective outcomes. For non-clinical mechanical testing, results are typically objective measurements, not subject to adjudication in this manner.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. The document explicitly states: "No clinical testing was performed." Therefore, no MRMC study was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. The HammerFiX device is a physical bone fixation implant, not a software algorithm or AI-based device. No standalone algorithm performance evaluation would be relevant.

    7. The Type of Ground Truth Used:

    • Non-Clinical Mechanical Standards and Predicate Device Performance: For the non-clinical testing, the "ground truth" was established by comparing the HammerFiX device's mechanical properties (pull-out, torque, static, and dynamic bending) against:

    8. The Sample Size for the Training Set:

    • Not Applicable. This question is relevant for AI/ML algorithm development. Since the HammerFiX device is a physical implant and no AI/ML component is mentioned, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. See explanation above for training set.

    Summary: The 510(k) clearance for the HammerFiX device was based on non-clinical bench testing demonstrating substantial equivalence to predicate devices, rather than a clinical study evaluating human performance or clinical outcomes. The document clearly states that "No clinical testing was performed."