Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K222792

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-06-08

    (265 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SNUCONE Tissue Level Implant System is inticated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations and terminal or internediate Abutment support for fixed bridge work. Snucone implant system is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.

    Device Description

    SNUCONE Implant System Fixture, also known as an endosseous implant, is surgical component that interfaces with the bone of the jaw or skull to support a dental prosthesis such as a crown, bridge, denture. Snucone's abutment and prosthetic components and tools are compatible with the Snucone's fixture only. The system includes Fixtures and Cover Screw, and Abutment and Component (Solid Abutment, InOcta Closing Screw, InOcta Healing Abutment, InOcta Abutment, InOcta Angled Abutment, InOcta Temporary Abutment, Platform Switching Abutment, and Abutment Screw).

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device (SNUCONE Tissue Level Implant System). It contains details about the device, its intended use, comparison to predicate devices, and a list of non-clinical tests performed. However, it does not contain information about acceptance criteria or a study that proves the device meets those criteria, especially in the context of an FDA clearance for an AI/ML-driven device.

    The document explicitly states: "Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device." It details mechanical performance, sterilization, shelf-life, packaging, MR environment conditions, bacterial endotoxin, and biocompatibility.

    THERE IS NO MENTION OF AN AI/ML COMPONENT IN THIS DEVICE.

    Therefore, I cannot fulfill your request for information regarding:

    1. A table of acceptance criteria and the reported device performance for an AI/ML component. The document describes acceptance criteria for mechanical, material, and sterilization properties, but not AI/ML performance metrics (e.g., sensitivity, specificity, AUC).
    2. Sample size for test set and data provenance specific to an AI/ML model. The tests mentioned are laboratory bench tests, not data-driven performance studies on patient cohorts for an AI model.
    3. Number of experts and their qualifications for ground truth establishment. This is not relevant to the non-clinical tests described.
    4. Adjudication method for the test set. Not relevant.
    5. MRMC comparative effectiveness study or human reader improvement. Not relevant.
    6. Standalone (algorithm only) performance. Not relevant.
    7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.). Not relevant to the tests performed.
    8. Sample size for the training set. Not relevant, as no AI model is described.
    9. How ground truth for the training set was established. Not relevant, as no AI model is described.

    In summary, the provided document is a 510(k) submission for a physical dental implant system and does not involve any AI/ML components or related performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173141

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-09-19

    (355 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K143022, K120043, K081575, K112787

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CSM Submerged3-L Implant System is intended for use in support of single or multiple-unit restorations and partial or fully edentulous mandibles and maxilla. This system is intended for delayed loading.

    Device Description

    The CSM Submerged3-L Implant System is composed of dental fixtures (Sub3 Fixture) and various abutments such as Healing Abutment, Cementation Abutment (Hex, Non-Hex), Angled Abutment (Hex, Non-Hex), Abutment Screw, Submerged Cover Screw, Retainer Abutment, Retainer Cap, Retainer Retention Male, and Temporary Abutment (Hex, Non-Hex). There is no vertical anti-rotation slot in our fixtures. When it comes to anti-rotational feature, it is 11 °, which is internal hexagonal feature. Hence. 4.0. 4.8. 5.2. 5.6. 6.0 size fixtures have 2.5 hexagonal features, different from fixture with 3.6 diameter, which has 2.1 double hexagonal feature. The fin of submerged 3 fixtures is V-shape. In addition, a straight flat axial surface is on our fixture, functioning as tapering part. In case of abutments, 2.5 and 2.1 hexagonal features are usually used. The Sub3 fixtures and screws are made of Ti 6A1 4V ELI (Conforming to ASTM Standard F-136) and dental abutment is made of Ti-6A1-4V ELI (ASTM F136), POLYAMIDE 6.6. The surface of the fixtures is treated with RBM (Resorbable Blast Media) and Laser. The diameters of the CSM Submerged3-L Implants are Ø 3.62mm, 4.35mm, 4.75mm, 5.15mm, 5.55mm, 5.95mm and the lengths of the CSM Submerged3-L Implants are 7.3mm, 8.3mm, 9.3mm, 10.3mm, 11.3mm, 12.3mm, 13.3mm, 14.3mm. The dimension of each abutment ranges as below: (The tolerance of all products is ±0.03) - Healing Abutment: Ø 4.02mm, 4.5mm, 6.5mm (D) X 8.25mm, 8.75mm, 9.75mm, ● 10.35mm, 10.75mm, 11.75mm, 12.35mm, 12.75mm (L) - Cementation Abutment: Ø 4.5mm, 5.5mm, 6.5mm (D) X 8.0mm, 8.5mm, 9.0mm, 9.5mm, ● 10.5mm, 11.5mm, 12.5mm, 13.5mm (L) - Angled Abutment: Ø 4.0mm, 4.5mm, 5.5mm (D) X 9.0mm, 9.5mm, 10.0mm, 10.3mm, ● 10.4mm, 10.5mm, 11.3mm, 11.5mm, 12.3mm, 12.3mm, 13.3mm, 13.5mm (L) with 15° and 20° - . Abutment Screw: Ø 2.1mm, 2.33mm (D) X 8.3mm, 10.0mm (L) - Submerged Cover Screw: Ø 2.83mm, 3.33mm (D) X 6.0mm, 6.5mm (L) ● - Retainer Abutment: Ø 3.9mm (D) X 7.15mm, 7.65mm, 7.9mm, 8.65mm, 8.9mm, ● 9.65mm, 9.9mm, 10.65mm, 10.9mm, 11.65mm, 11.9mm, 12.9mm, 13.65mm, 13.9mm, 14.65mm, 14.9mm, 15.65mm, 15.9mm, 16.65mm, 16.9mm (L) - Retainer Cap: Ø 5.45mm (D) X 2.4mm (L) ● - Retainer Retention Male: Ø 4.7mm (D) X 1.85mm, 2.06mm (L) ● - Temporary Abutment: Ø 4.5mm, 5.5mm (D) X 12.8mm, 13.8mm, 13.8mm, 14.8mm, 15.8mm, ● 16.8mm (L) The implant-abutment connection is internal hex and Morse taper level. Implant-fixture and submerged cover screw are packed together and provided sterile. The abutments are provided non-sterile and must be sterilized before use.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the CSM Submerged3-L Implant System:

    This document is a 510(k) Premarket Notification summary for a dental implant system. The primary goal of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This process typically relies heavily on bench testing and comparison of technological characteristics rather than extensive clinical efficacy studies in the way you might see for novel pharmaceuticals or advanced AI diagnostics.

    Therefore, many of the questions you've asked (e.g., sample size for training/test sets, number of experts for ground truth, MRMC studies, standalone algorithm performance, data provenance) are not applicable to this type of medical device submission. These questions are typically relevant for AI/ML-based diagnostic devices or devices where human interpretation of data is a primary function.

    This submission focuses on demonstrating safety and effectiveness through:

    1. Comparison to predicate devices: Showing that the new device has similar technological characteristics and indications for use as devices already on the market.
    2. Non-clinical (bench) testing: Performing specific tests to ensure the device meets established performance standards.

    Acceptance Criteria and Reported Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list "acceptance criteria" in a quantitative, pass/fail manner for the device performance in the way an AI/ML study would (e.g., "accuracy > 90%"). Instead, the acceptance criteria are implicit in the compliance with recognized standards and the demonstration of substantial equivalence to predicate devices. The "reported device performance" is essentially the statement that the device met these standards and is substantially equivalent.

    Acceptance Criteria (Implicit)Reported Device Performance (as stated in submission)
    Material Composition: Conformance to ASTM Standard F-136 (Ti-6Al-4V ELI) and use of POLYAMIDE 6.6."The Sub3 fixtures and screws are made of Ti 6A1 4V ELI (Conforming to ASTM Standard F-136) and dental abutment is made of Ti-6A1-4V ELI (ASTM F136), POLYAMIDE 6.6." (Page 4&5) "Ti-6Al-4V ELI (ASTM-F136)" and "POLYAMIDE 6.6" are listed for individual components, aligning with predicates. (Pages 5-11)
    Sterilization Efficacy (Non-sterile components): Ability to be sterilized by end-user according to ISO 17665-1:2006, -2:2009 and ANSI/AAMI ST79."The end user sterilization test was performed for the subject abutments." "The results of the above tests have met the criteria of the standards..." (Page 12)
    Biocompatibility: Conformance to ISO 10993 series (cytotoxicity, irritation, sensitization, acute systemic, genotoxicity)."Biocompatibility Test such as cytotoxicity according to ISO 10993-5:2009, irritation according to ISO 10993-10:2010, sensitization according to ISO 10993-10:2010, acute systemic according to ISO 10993-11:2006, and genotoxicity according to ISO 10993-3:2014." "The biocompatibility tests were performed on the subject device and it demonstrated the subject device is biocompatible." "The results of the above tests have met the criteria of the standards..." (Page 12)
    Sterilization Efficacy (Pre-sterilized components): Conformance to ISO 11137-1,2,3 for sterilization validation."The sterilization validation test was performed for the predicate, K102635 and leveraged for the subject product because the product material, sterilization site, sterilization method, SAL, sterilization parameters are exactly identical to the predicate, K102635." (Page 12)
    Fatigue Strength: Conformance to ISO 14801."The fatigue testing was performed for the predicate device. K102635 and leveraged for the subject device because we compared the worst-case implants and the test sample from K102635 was the worst. Therefore, it supports mechanical properties." "The results of the above tests have met the criteria of the standards..." (Page 12)
    Shelf Life/Aging: Conformance to ASTM F1980-07."The accelerated aging testing was performed for predicate device, K102635 and leveraged for the subject device because the product category, material, manufacturing process, facility, packaging material and packaging procedure are the exactly the same as the predicate, K102635." "The results of the above tests have met the criteria of the standards..." (Page 12)
    Substantial Equivalence: Similar intended use, fundamental scientific technology, and performance parameters to predicate devices.The entire submission argues for this, concluding: "The CSM Submerged3-L Implant System... constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, CSM Submerged3-L Implant System, and its predicates are substantially equivalent." (Page 13)

    Study Details

    As mentioned, many of the requested fields are not applicable to a 510(k) submission for a non-AI/ML dental implant system.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This device is an implant and its components, not an AI/ML diagnostic system. The "testing" involves bench testing to international standards (e.g., ISO, ASTM) and comparison of physical characteristics to predicate devices. There is no "test set" in the context of imaging or patient data. Data provenance (country of origin, retrospective/prospective) is also not relevant here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. There is no "ground truth" established by experts in this context. Device performance is assessed against engineering standards and material specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No expert adjudication is involved.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI/ML-assisted device, nor is it a diagnostic tool that involves human "readers."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. There is no algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not Applicable. The "ground truth" here is the adherence to engineering specifications, material standards (e.g., ASTM F-136), and performance criteria outlined in international standards (e.g., ISO 17665, ISO 10993, ISO 14801, ASTM F1980-07).

    8. The sample size for the training set

    • Not Applicable. There is no training set for this type of device.

    9. How the ground truth for the training set was established

    • Not Applicable. There is no training set or associated ground truth.

    Summary of the "Study" that Proves the Device Meets Acceptance Criteria:

    The "study" consists of a combination of direct non-clinical testing on the subject device and leveraging data from legally marketed predicate devices.

    • Direct Testing on Subject Device:

      • End User Steam Sterilization Test: Performed on the abutments according to ISO 17665-1:2006, -2:2009 and ANSI/AAMI ST79.
      • Biocompatibility Tests: Performed on the subject device according to various parts of ISO 10993 (cytotoxicity, irritation, sensitization, acute systemic, genotoxicity).
      • Results: All these tests demonstrated that the subject device met the criteria of the respective standards.
    • Leveraged Testing from Predicate Device (K102635):

      • Sterilization Validation Test (ISO 11137-1,2,3): Data from the predicate K102635 was leveraged because the subject product's material, sterilization site, method, SAL, and parameters are identical.
      • Fatigue Test (ISO 14801): Data from the predicate K102635 was leveraged. The applicant specifically states that the worst-case implants from the predicate's testing encompassed the subject device's mechanical properties.
      • Accelerated Aging Test (Shelf Life, ASTM F1980-07): Data from the predicate K102635 was leveraged due to identical product category, material, manufacturing process, facility, packaging material, and packaging procedure.
      • Results: The leveraged predicate data, combined with the direct testing, collectively demonstrate that the CSM Submerged3-L Implant System meets the required performance and safety standards, thus establishing its substantial equivalence.

    In essence, the "study" is a comprehensive engineering and material science evaluation, rather than a clinical trial or an AI performance study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1