Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K152151

    Validate with FDA (Live)

    Device Name
    ATAL 9
    Manufacturer
    Date Cleared
    2015-11-10

    (99 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K113812

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATAL 9 is indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, angiographic, and mammographic applications.

    Device Description

    The ATAL 9 is a digital radiography system, featuring an integrated flat panel digital detector (FPD). ATAL 9 is designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the modified panel are the same as our previous panel with a Wi-Fi wireless feature added.

    AI/ML Overview

    The provided document is a 510(k) summary for the ATAL 9 digital radiography system. It focuses on demonstrating substantial equivalence to a predicate device (ATAL 8, ATAL 8c, K113812), primarily due to the addition of wireless functionality. As such, the study design and acceptance criteria are geared towards proving this equivalence rather than establishing de novo performance for a new type of device or AI algorithm.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit)Reported Device Performance (ATAL 9 vs. ATAL 8/8c)
    Technological Characteristics
    Pixel Pitch: Same as predicateSAME (139 μm)
    Limiting Resolution: Same as predicateSAME (Over 3 lp/mm)
    DQE (CSI) at 2 lp/mm: Comparable to predicate26% (Predicate: 27%)
    MTF (CSI) at 2 lp/mm: Comparable to predicate42% (Predicate: 43%)
    A/D Conversion: 14 bits or better16 bits (Predicate: 14 bits)
    Active Area: Same as predicateSAME (17 x 17 inch)
    Pixels: Same as predicateSAME (3,072 x 3072)
    Software: Same functionalitySAME as K113812
    DICOM output: YesYES
    Scintillator: Same optionsUNCHANGED (Csl/GOS)
    Interface: Gigabit Ethernet and/or wireless functionalityWired: Gigabit Ethernet (1000BASE-T); Wireless: IEEE802.11ac, backward compatible
    Power source: AC line and/or rechargeable batteryAC Line and/or Rechargeable Lithium Battery (10 hr run time)
    Dimensions/Weights: Changed (due to wireless/battery)460(W)×461(L)×15(H)/ 2.9kg (2.4kg w/o Battery) (Predicate: 500(W)×500(L)×25(H)/ 7.8kg)
    Safety and Effectiveness
    Electrical Safety: Compliance with IEC 60601-1Complies with IEC 60601-1
    EMC: Compliance with IEC 60601-1-2Complies with IEC 60601-1-2
    Wireless operation: Compliance with IEEE 802.11ac and FCCComplies with IEEE 802.11ac, Meets FCC requirements
    Image Quality (Clinical): As good as or better than predicateConcluded "as good as or better than" by board-certified radiologist.
    Risk Analysis: ConductedConducted (as per bench testing summary)
    Software Verification: ConductedConducted (as per bench testing summary)
    Wireless connectivity: VerifiedVerified

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Clinical images were acquired and evaluated by a board certified radiologist..." However, it does not specify the number of clinical images (sample size) used for the test set.

    • Sample Size: Not explicitly stated for the clinical image evaluation.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). While "clinical images were acquired," it doesn't specify if these were newly acquired for the study (prospective) or selected from an existing archive (retrospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: One.
    • Qualifications of Experts: A "board certified radiologist." No further details on years of experience or sub-specialty are provided.

    4. Adjudication Method for the Test Set

    The document indicates that "Clinical images were acquired and evaluated by a board certified radiologist who concluded the images from the new panel are as good as or better than the images acquired with the predicate panel." This implies a single-reader evaluation against the predicate device, not a multi-reader adjudication process. Therefore, the adjudication method was essentially none in the sense of comparing multiple experts' opinions. The single expert provided the judgment.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • MRMC Study: No, an MRMC study was not done.
    • AI Improvement Effect Size: Not applicable. This submission is for a digital radiography panel, not an AI software/algorithm intended to assist human readers or provide automated diagnoses. The device being cleared (ATAL 9) is a hardware device for image acquisition.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: No, a standalone performance study as described (algorithm only without human-in-the-loop performance) was not done. The ATAL 9 is a digital X-ray panel, where the "performance" relates to image acquisition and quality, not an algorithm's diagnostic output. The clinical evaluation involved a human radiologist examining the acquired images.

    7. The Type of Ground Truth Used

    For the clinical evaluation, the "ground truth" was established by the expert opinion/conclusion of a single board-certified radiologist. The radiologist compared images from the new panel against those from the predicate panel. This is a form of expert consensus, albeit from a single expert, comparing image quality.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable/not provided. The ATAL 9 is a hardware device (digital X-ray panel). It is not an AI algorithm that undergoes a "training" phase with a large dataset. The "software" component remains the same as its predicate (K113812), implying no new algorithm training was performed for the ATAL 9.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable. As explained above, this device does not involve an AI algorithm with a training phase requiring a "ground truth" to be established for training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141743

    Validate with FDA (Live)

    Date Cleared
    2014-10-03

    (95 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K113812, K112527

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This flat panel digital imaging system is intended for use in generating radiographic images of human anatomy. This device is intended to replace film/screen systems in all general purpose diagnostic procedures. This device is intended for use by qualified medical personnel. (Not for mammography)

    Device Description

    This is a digital x-ray panel coupled with image acquisition software. The acquisition software is installed on a Windows compatible workstation. No routine user calibration or generator connection is required. AED and ACC are standard features. AED is automatic exposure detection. No synchronization connection to the generator is required. ACC is Automatic Calibration Control. Calibration values are automatically loaded at power-up. The TFT is inherently linear, and all that is necessary is to set a Start (Dark) and Stop (Bright) value to define the dynamic range you want to operate in. It is completely independent of the x-ray source. That is why this can be done at the factory. MECHANICAL: FITS IN THE BUCKY TRAY WITHOUT MODIFICATION... Ultra-Light Cassette Sized Detector easily moves from table top to wall to stretcher with ease. The system is compatible with CPI CMP 200 and Sedecal SHF series generators.

    AI/ML Overview

    The provided document is a 510(k) summary for the Fusion Digital Diagnostic X-Ray Upgrade Kit (K141743). It explicitly states that "No new testing was done because no modifications have been made other than deleting the generator and tube stand" for both non-clinical and clinical tests (Sections 7 and 8 on page 4).

    Therefore, this document does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria for the K141743 submission.

    The document refers to a previous clearance (K132294) for the "IDENTICAL digital x-ray panel/software/workstation." To obtain the requested information, one would need to review the 510(k) submission for K132294.

    Based on the provided document K141743, the following questions cannot be answered:

    1. A table of acceptance criteria and the reported device performance
    2. Sample sized used for the test set and the data provenance
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    4. Adjudication method for the test set
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    7. The type of ground truth used
    8. The sample size for the training set
    9. How the ground truth for the training set was established
    Ask a Question

    Ask a specific question about this device

    K Number
    K132921

    Validate with FDA (Live)

    Device Name
    GALAXY PLUS
    Date Cleared
    2014-01-17

    (122 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K130883, K093816, K113812

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Galaxy Plus Digital Radiography System is indicated for use in generating radiographic images of human anatomy. This device is not intended for mammographic applications.

    Device Description

    The digital diagnostic x-ray system consists of a generator, a tube, an x-ray console, a beam limiting device, an image receptor, a u-arm stand and software. The digital panels supplied are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications. Product features: This system is designed to maintain the alignment between a tube and the image receptor, regardless of the angle on the image receptor or image tilt positions. This typical configuration permits a qualified/trained doctor or technologist to take a range of head-to-toe diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on patients. Its flexibility makes the system ideal for all patient examinations including the standing, sitting and lying patient positions.

    AI/ML Overview

    The provided 510(k) summary for K132921 describes the Galaxy Plus Digital Radiography System.

    Here's an analysis based on the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly state quantitative "acceptance criteria" in terms of performance metrics (e.g., sensitivity, specificity, resolution targets) for the device. Instead, the primary acceptance criterion appears to be demonstrated substantial equivalence to the predicate device (DRGEM Corporation's DIAMOND-5A, 6A, 8A, K102408) through non-clinical and clinical testing, and the ability to produce images of "good diagnostic quality."

    Given this, the table would look like this:

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance
    Electrical SafetyCompliance with EN/IEC 60601-1: 2006 (for Stand & System)Test reports submitted; unit successfully underwent electrical safety testing.
    EMCCompliance with IEC 60601-1-2:2007, EN/IEC 60601-1-2:2007, CISPR 11:2009/A1:2010, IEC 61000-3-2:2005/A1:2008/A2:2009, IEC 61000-3-3:2008Test reports submitted; unit successfully underwent electromagnetic compatibility testing.
    Radiation ProtectionCompliance with EN/IEC 60601-1-3; EN/IEC 60601-2-28; EN/IEC 60601-2-54Test reports submitted; compliance demonstrated.
    Software ValidationSoftware validation conducted.Software validation performed; updated software (previously cleared in K093816) employed.
    Risk AnalysisRisk analysis conducted.Risk analysis performed.
    BiocompatibilityPerformed in accordance with applicable ISO standards.Biocompatibility testing performed.
    Image Quality (Clinical)Images for all four panels to be of "good diagnostic quality" as evaluated by board-certified radiologists, and obtained in accordance with FDA Guidance Document on Solid State Imaging Devices. The device is "as safe, as effective, and performs as well as or better than" the predicate device.Clinical images for all four panels obtained, evaluated by board-certified radiologists, and found to be of "good diagnostic quality." This implicitly means the device performs at least as well as the predicate in producing diagnostically acceptable images for its intended use (generating radiographic images of human anatomy, excluding mammography).
    Non-clinical PerformanceDemonstrated by respective panel manufacturers.Non-clinical performance tests provided by the panel manufacturers.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a numerical sample size for the clinical test set (i.e., the number of patients or images). It vaguely states "Clinical images for all four panels were obtained."
    • Data Provenance: Not explicitly stated (e.g., country of origin). It states the images were obtained "in accordance with the FDA Guidance Document on Solid State Imaging Devices," implying they met U.S. regulatory standards for data collection, but no further details on location or retrospective/prospective nature are provided. It is likely retrospective or a limited prospective case series given the summary nature.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not specified. The document states "They were evaluated by board certified radiologist." This could imply one or more radiologists.
    • Qualifications: "board certified radiologist." The experience level in years is not provided.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (e.g., 2+1, 3+1, none). It simply states the images "were evaluated by board certified radiologist." This suggests either independent evaluations or a consensus approach, but the specific process is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention an MRMC study. The comparison is primarily against the predicate device's technological characteristics and the general standard of "good diagnostic quality" for radiographic images, rather than a comparative effectiveness study with human readers.

    6. Standalone Performance

    Yes, a standalone evaluation was performed. The clinical images generated by the algorithm/system only (i.e., the Galaxy Plus Digital Radiography System itself) were evaluated by radiologists for their diagnostic quality. This evaluation was independent of human interpretation improvements; it assessed the quality of the raw output from the device.

    7. Type of Ground Truth Used

    The ground truth used for the clinical evaluation was expert consensus/evaluation by "board certified radiologist[s]" who determined the images were of "good diagnostic quality." There is no mention of pathology, outcomes data, or other objective measures of truth. For a diagnostic imaging device like an X-ray system, "good diagnostic quality" determined by an expert is a common form of ground truth for substantial equivalence.

    8. Sample Size for the Training Set

    The document does not mention a "training set" or provide any sample size for training. This device is a digital radiography system, not an AI/ML algorithm that typically requires a large training set in the way a diagnostic AI would. The software components are an update of previously cleared software (K093816), implying its core functionality was already established. The emphasis is on the image acquisition hardware and its ability to produce images.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned in the context of this 510(k) summary (because it's not a de novo AI/ML solution), the method for establishing ground truth for a non-existent training set is not applicable here. The software is an update of a previously cleared version, suggesting its underlying principles and performance were evaluated during its initial clearance (K093816).

    Ask a Question

    Ask a specific question about this device

    K Number
    K132294

    Validate with FDA (Live)

    Date Cleared
    2013-11-12

    (112 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K113812, K112527

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)

    Device Description

    This is a complete stationary diagnostic x-ray system employing a digital x-ray panel coupled with image acquisition software. The acquisition software is installed on a Windows compatible workstation. The high frequency generator is available in four different power ratings. All components are either 510(k) exempt or previously cleared. The only difference between the two models is the configuration of the tube stand. Galaxy has a straight arm and the Galaxy+ Plus is in a U-Arm configuration. Generators are available in 40-50-65-or 80 kW (High Frequency) models.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Galaxy and Galaxy+ Plus Digital Diagnostic X-Ray System. It does not contain information about specific acceptance criteria or a detailed study proving the device meets them in the way typically found for AI/algorithm-driven devices.

    The document focuses on demonstrating substantial equivalence to a predicate device (Sedecal X-Plus LP Plus Digital Diagnostic X-Ray Systems, K090238) rather than meeting pre-defined performance acceptance criteria for an AI algorithm. The core of the submission is to show that the new device has "essentially the same technological characteristics" and similar clinical indications for use as the predicate.

    Here's an analysis based on the information provided, highlighting what is present and what is not:


    Acceptance Criteria and Device Performance:

    The document doesn't explicitly state "acceptance criteria" in terms of specific performance metrics (e.g., sensitivity, specificity, AUC) for an AI component, nor does it report such metrics for the device. Instead, the "acceptance criteria" for this 510(k) submission appear to be implicitly about demonstrating that the new device is as safe and effective as the predicate device.

    Table of Acceptance Criteria and Reported Device Performance:

    Criterion Category (Implicit)Description / Device Performance (from text)
    Indications for Use EquivalenceIdentical indications for use as the predicate device: "Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)"
    Technological Characteristics Equivalence"Essentially the same technological characteristics" as the predicate—differences are in manufacturer for digital panels, x-ray generator, and tube stands, but with "functionally identical capabilities." Key technological comparison points include: - Digital Receptor Panel: Atlaim ATAL 8 and ATAL 8C (Cleared in K113812), compared to Canon panels in predicate. - Panel Communication: Tethered Ethernet, one panel (Same as predicate). - Panel Resolution: Pixel size 139 x 139 μm, Image matrix size 3072 x 3072 pixels, Approx. 9.4 million pixels (Predicate: 160 x 160 μm, 2208 x 2688 pixels, Approx. 5.9 million pixels). The new device has higher resolution. - DICOM: Yes (Same software as cleared in K112527) (Same as predicate). - Tube Stand: One model is a straight arm, the other is similar to U-Arm (Predicate: U-Arm). - Generator: CPI, 40-50-65-or 80 kW (High Frequency) (Predicate: Sedecal 65 kW, optional 80 kW).
    Electrical Safety & EMC"The unit has undergone electrical safety and electromagnetic compatibility testing... The unit meets IEC safety and EMC standards." (Implicitly, this is an acceptance criterion).
    Software Validation & Risk Analysis"software validation and risk analysis" were performed. (Implicitly, this is an acceptance criterion).
    Clinical Image Quality"Clinical images were obtained in accordance with the FDA Guidance Document on Solid State Imaging Devices. They were evaluated by professional radiologist and found to be of good diagnostic quality." (This is the primary "performance" metric related to clinical utility).

    Detailed Study Information:

    The document describes non-clinical and clinical tests, but these are general descriptions rather than a detailed study protocol for an AI/algorithm.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not specified. The clinical tests mention "Clinical images were obtained," but no quantity is given.
      • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not specified, but referred to as "professional radiologist" (plural).
      • Qualifications of Experts: "professional radiologist." No years of experience or specific subspecialty mentioned.
    3. Adjudication method for the test set:

      • Not specified. The text simply states images "were evaluated by professional radiologists and found to be of good diagnostic quality." No mention of consensus, voting, or specific adjudication rules.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was performed or described. This submission is for a conventional digital X-ray system, not an AI-driven diagnostic aid. The "clinical tests" assessed the diagnostic quality of the images produced by the system itself, not the performance of human readers with or without AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No standalone algorithm performance study was done. This device is a diagnostic imaging system, not an algorithm. The "performance" being evaluated is the image quality produced by the system for human interpretation.
    6. The type of ground truth used:

      • The "ground truth" for the clinical images appears to be the expert opinion/consensus of professional radiologists regarding the "good diagnostic quality" of the images generated by the device. It's not pathology, outcomes data, or a pre-defined objective reference standard for specific病灶 detection, but rather an assessment of general image utility.
    7. The sample size for the training set:

      • Not applicable. This device is a digital X-ray system, and the submission does not mention any machine learning or AI components that would require a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, as no training set for an AI algorithm is mentioned or implied.

    Summary of the Device and Submission Context:

    This 510(k) pertains to a new digital X-ray system. The "study" described is a testing and evaluation process for a medical device's physical and functional characteristics, not a clinical trial or performance study for an AI algorithm. The device produces images, and the "clinical tests" verified that these images are of "good diagnostic quality" as assessed by radiologists, aligning with the performance expected from similar legally marketed predicate devices. The core argument for clearance is substantial equivalence based on comparable indications for use, technological characteristics (with the new device having higher resolution panels), and meeting relevant safety and EMC standards.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1