Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173266

    Validate with FDA (Live)

    Date Cleared
    2017-12-01

    (51 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K151313, K113757

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Octane aspiration system is intended for the removal/aspiration of embolic material (thrombus/debris) from vessels of the arterial and deep venous system, and to infuse/deliver diagnostic or therapeutic agents.

    Device Description

    The Octane aspiration system consists of two components: the Octane catheter and the Octane hand pump. The Octane aspiration catheter has a working length of 115cm, is compatible with introducer sheaths ≥ 8F, guidewires ≤ 0.018" in diameter, and it is intended for use in vessels ≥ 3 mm in diameter. The catheter has a rapid exchange design and includes sliding, co-axial inner and outer lumens. The catheter inner lumen tip has three side wall slot ports and the catheter's outer lumen has a beveled radiopaque tip. To alleviate plugging of the catheter tip caused by thrombus, the actuator on the proximal end of the catheter slides the catheter's outer lumen back and forth over the inner lumen distal tip. The actuator on the proximal end of the catheter is also used to control the directionality of the torque-able catheter. Directionality is further achieved through the gradual curve in the shaft materials at the distal end. The Octane hand pump provides suction to the inner lumen of the Octane aspiration catheter and is manually operated using a spring-loaded handle. It consists of a pump handle fitted with a 30ml piston syringe, tubing, a series of one-way valves, and a 500ml collection bag. The hand pump tubing has a standard luer fitting for connection to the Octane aspiration catheter hub. The Octane aspiration system has been sterilized with ethylene oxide.

    AI/ML Overview

    This document is a 510(k) Summary for the Octane Aspiration System, a medical device intended for the removal of embolic material from blood vessels. The summary describes the device, its intended use, and the studies conducted to demonstrate its substantial equivalence to a predicate device (Pronto XL). However, it does not contain information about acceptance criteria and device performance in a format that directly addresses the user's request for AI/algorithm-related performance metrics or human reader studies.

    The document specifies bench testing and biocompatibility testing to demonstrate substantial equivalence, focusing on the physical and material properties of the device, rather than the diagnostic or interpretative performance of an AI system.

    Therefore, many of the requested categories related to AI performance, ground truth, expert opinions, and sample sizes for AI training/testing sets cannot be directly answered from the provided text.

    Here's a breakdown of what can be extracted and what cannot:

    1. Table of acceptance criteria and the reported device performance:

    The document broadly states: "The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues." However, it does not provide a specific table detailing the numerical acceptance criteria for each test and the corresponding reported device performance values. It only lists the tests performed.

    Acceptance Criteria CategoryReported Device Performance
    Bench Testing
    Track forceMet specified criteria
    RadiopacityMet specified criteria
    Collection bag pressureMet specified criteria
    Syringe seal attachmentMet specified criteria
    Check valve crack pressureMet specified criteria
    Synthetic clot extractionMet specified criteria
    Proximal seal leakMet specified criteria
    AspirationMet specified criteria
    Liquid leakMet specified criteria
    Bond/joint tensileMet specified criteria
    Torque capacityMet specified criteria
    Corrosion resistanceMet specified criteria
    Therapeutic agent conditioningMet specified criteria
    Static pressureMet specified criteria
    Package integrityMet specified criteria
    Biocompatibility Testing
    CytotoxicityPassed
    SensitizationPassed
    IrritationPassed
    Acute systemic toxicityPassed
    Material mediated pyrogenicityPassed
    HemolysisPassed
    Complement activationPassed
    ThrombogenicityPassed

    2. Sample size used for the test set and the data provenance:

    • The document does not specify sample sizes for each bench or biocompatibility test. These tests typically involve a certain number of device units.
    • Data provenance (e.g., country of origin, retrospective/prospective) is not applicable or provided as these are bench and lab tests, not clinical or image-based studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable. The studies described are physical and material tests of a medical device, not diagnostic or AI-driven interpretations requiring expert consensus for ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable. These are objective engineering and laboratory tests, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC or comparative effectiveness study involving human readers or AI assistance was performed or reported in this document. The device is an aspiration catheter, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • No standalone algorithm performance study was done as the device is a physical medical instrument, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the bench tests would be established by engineering specifications, physical measurements, and standardized laboratory procedures. For biocompatibility, it's defined by the ISO 10993 standards. It's not based on expert consensus, pathology, or outcomes data in the way an AI diagnostic would be.

    8. The sample size for the training set:

    • Not applicable, as this device does not involve an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable, as this device does not involve an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1